← Product Code [LRK](/productcode/LRK) · K162192

# BluePro (K162192)

_Bluesom · LRK · Dec 15, 2016 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K162192

## Device Facts

- **Applicant:** Bluesom
- **Product Code:** [LRK](/productcode/LRK.md)
- **Decision Date:** Dec 15, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.5570
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic, Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K162192 · Dec 15, 2016 | BluePro | Bluesom | Post-market commercial sales data (outside the US) | The sponsor utilized real-world clinical experience from international commercial distribution (since October 2014) to support the device's performance, durability, and substantial equivalence to the predicate device. | Commercial experience; Real-time clinical experience; Post-market data |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| International commercial experience; Post-market observational experience; Follow-up/Duration: Since October 2014; Study Period: October 2014 to 2016 | Adult patients using the device for snoring and obstructive sleep apnea | Not applicable for this study | Device performance, cleaning efficacy, and mechanical durability |

## Indications for Use

The BluePro® is a mandibular repositioning device intended to reduce or alleviate night-time snoring and mild to moderate obstructive sleep apnea in adult patients 18 years or older.

## Device Story

BluePro is a mandibular repositioning device (MRD) for snoring and mild to moderate obstructive sleep apnea. Device consists of upper and lower pre-formed trays; customized via 'boil and bite' method by dentist. Operates by holding mandible in advanced position; integral interlocking mechanism allows for fixed advancement up to 10 mm. Used in home, dental/physician offices, or sleep labs. Patient wears device at night; device allows lateral and vertical jaw movement. Benefits include reduction of snoring and apnea symptoms. Cleaning via water rinse and toothbrush. No electronic or software components.

## Clinical Evidence

No clinical testing. Evidence based on non-clinical performance testing, mechanical durability testing (simulated 1-year use via Braem testing), and commercial experience outside the US since 2014.

## Technological Characteristics

Mandibular repositioning device; pre-formed thermoplastic trays; 'boil and bite' customization; integral interlocking advancement mechanism; 10 mm maximum adjustment; surface-contacting (mucosal membrane) materials; biocompatibility per ISO 10993 (cytotoxicity, sensitization, irritation).

## Regulatory Identification

Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.

## Special Controls

*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”

## Predicate Devices

- ApneaRx ([K113569](/device/K113569.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{1}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus, a symbol of medicine, with three figures representing health, hope, and well-being. The seal is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement. The image is in black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 15, 2016

Bluesom Paul Dryden Consultant BlueSom 24 Rue Leon Gaumont Orvault, 44700 FR

Re: K162192 Trade/Device Name: BluePro® Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices for Snoring and Intraoral Devices for Snoring and Obstructive Sleep Apnea Regulatory Class: Class II Product Code: LRK Dated: November 14, 2016 Received: November 15, 2016

Dear Mr. Paul Dryden:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{2}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Michael J. Ryan -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K162192

Device Name

### BluePro®

Indications for Use (Describe)

The BluePro® is a mandibular repositioning device intended to reduce or alleviate night-time snoring and mild to moderate obstructive sleep apnea in adult patients 18 years or older.

Type of Use (Select one or both, as applicable)

XX Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

{4}------------------------------------------------

### BLUESOM

24 rue Léon Gaumont 44700 Orvault FRANCE Tel - +33 2 28 25 61 43

| Official Contact:                                           | Erwan Floch - General Manager                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                      |
|-------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary or Trade Name:                                  | BluePro®                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                      |
| Common/Usual Name:                                          | Intra-oral appliance                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                      |
| Classification Name:                                        | LRK - Device, anti-snoring, Intraoral devices for snoring<br>and intraoral devices for snoring and obstructive sleep apnea<br>21 CFR 872.5570, Class 2                                                                                                                                        |                                                                                                                                                                                                                                                                      |
| Predicate Device:                                           | K113569 – Apnea Sciences – ApneaRx                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                      |
|                                                             | Predicate<br>ApneaRx                                                                                                                                                                                                                                                                          | Subject Device<br>BluePro®                                                                                                                                                                                                                                           |
|                                                             | K113569                                                                                                                                                                                                                                                                                       | K162192                                                                                                                                                                                                                                                              |
| Indications for Use                                         | The Apnea Sciences Corporation<br>"ApneaRx" is intended for use on<br>adult patients 18 years of age or older<br>as an aid for the reduction of mild to<br>moderate obstructive sleep apnea<br>(OSA), and/or snoring.                                                                         | The BluePro® is a mandibular<br>repositioning device intended to<br>reduce or alleviate night-time snoring<br>and mild to moderate obstructive<br>sleep apnea in adult patients 18 years<br>or older.                                                                |
| Environments of use                                         | Home, dental and Physician offices,<br>Sleep laboratories                                                                                                                                                                                                                                     | Home, dental and Physician offices,<br>Sleep laboratories                                                                                                                                                                                                            |
| Patient Population                                          | Adult patients 18 years and older                                                                                                                                                                                                                                                             | Adult patients 18 years and older                                                                                                                                                                                                                                    |
| Contraindications                                           | • have central sleep apnea<br>• have severe respiratory disorders<br>• have loose teeth or advanced<br>periodontal disease<br>• are under 18 years of age                                                                                                                                     | • have central sleep apnea<br>• have severe respiratory disorders<br>• have loose teeth or advanced<br>periodontal disease<br>• are under 18 years of age                                                                                                            |
| Prescription                                                | Prescription use                                                                                                                                                                                                                                                                              | Prescription use                                                                                                                                                                                                                                                     |
| Single patient, multi-use                                   | Yes                                                                                                                                                                                                                                                                                           | Yes                                                                                                                                                                                                                                                                  |
| Limitation of duration of use                               | No limitation                                                                                                                                                                                                                                                                                 | No limitation                                                                                                                                                                                                                                                        |
| Principle of operation / means of<br>mandibular advancement | Adjustment of the relative position of<br>the trays by the use of fixed<br>advancement.                                                                                                                                                                                                       | Adjustment of the relative position of<br>the trays by the use of fixed<br>advancement.                                                                                                                                                                              |
| Fixed tray sizes                                            | Yes                                                                                                                                                                                                                                                                                           | Yes                                                                                                                                                                                                                                                                  |
| Method to advance the mandible                              | Integral interlocking advancement<br>method                                                                                                                                                                                                                                                   | Integral interlocking advancement<br>method                                                                                                                                                                                                                          |
| Allows lateral and vertical movement                        | Yes                                                                                                                                                                                                                                                                                           | Yes                                                                                                                                                                                                                                                                  |
| Employs boil and bite design                                | Yes                                                                                                                                                                                                                                                                                           | Yes                                                                                                                                                                                                                                                                  |
| Maximum adjustment by the user                              | 10 mm                                                                                                                                                                                                                                                                                         | 10 mm                                                                                                                                                                                                                                                                |
| Works by holding lower jaw forward                          | Yes                                                                                                                                                                                                                                                                                           | Yes                                                                                                                                                                                                                                                                  |
| Cleaned by simple rinsing with water                        | Yes                                                                                                                                                                                                                                                                                           | Yes                                                                                                                                                                                                                                                                  |
| Cleaning with toothbrush                                    | Yes                                                                                                                                                                                                                                                                                           | Yes                                                                                                                                                                                                                                                                  |
| Biocompatibility and Patient contact                        | Surface Contacting, Mucosal<br>membrane, with duration of use<br>prolonged > 24 hours < 30 days, per<br>FDA Guidance Class II Special<br>Controls Guidance Document:<br>Intraoral Devices for Snoring and/or<br>Obstructive Sleep Apnea; Guidance<br>for Industry and FDA dated<br>11/12/2002 | ISO 10993<br>Cytotoxicity<br>Sensitization<br>Irritation<br>Surface Contacting, Mucosal<br>membrane, with duration of use<br>prolonged > 24 hours < 30 days, per<br>FDA Guidance Intraoral Devices for<br>Snoring and/or Obstructive Sleep<br>Apnea dated 11/12/2002 |
| Age testing                                                 |                                                                                                                                                                                                                                                                                               | Yes                                                                                                                                                                                                                                                                  |
| Real-time clinical experience                               |                                                                                                                                                                                                                                                                                               | Yes                                                                                                                                                                                                                                                                  |

### Device Description

The BluePro® oral appliance design concept is based upon the use of a standard set of upper and lower trays, referred also as splints in documentation, that are then customized for fit by a dentist. The design is commonly referred to as a Mandibular Repositioning Device (MRD).

The principle of advancing a lower or upper tray so that it advances the mandible for the treatment of snoring and / or mild to moderate obstructive sleep apnea is well known and there are a number of predicate devices.

### Indications for Use

The BluePro® is a mandibular repositioning device intended to reduce or alleviate night-time snoring and mild to moderate obstructive sleep apnea in adult patients 18 years or older.

### Contraindications

The device is contraindicated for patients who:

- . have central sleep apnea
- . have severe respiratory disorders
- . have loose teeth or advanced periodontal disease
- . have loose dental work, dentures, or other oral conditions which would be adversely affected by wearing dental appliances
- . are under 18 years of age

#### Warnings

Use of this device may cause:

- tooth movement or changes in dental occlusion
- gingival or dental soreness
- . pain or soreness to the temporomandibular joint
- . obstruction of oral breathing
- . excessive salivation
- . excessive dry mouth
- . loosening of dental restorations
- . loosening of teeth

{5}------------------------------------------------

## Environment of Use

Home, Dental and Physician offices, and Sleep laboratories

# Predicate Device Comparison:

Table 1 compares the predicate and the subject device.

{6}------------------------------------------------

### K162192 510(k) Summary 7-Dec-16

The BluePro® is viewed as substantially equivalent to the predicate device because:

Indications - Similar to predicate - ApneaRx - K113569; indicated for treating snoring and mild to moderate obstructive sleep apnea (OSA).

Discussion - The indications for use between the subject device and predicate are similar and therefore they can be found as substantially equivalent.

Technology / Principle of Operation - Similar to predicate - ApneaRx - K113569. Both devices use separate pre-formed trays with a means to advance the mandible / lower jaw. Discussion - Both devices use separate pre-formed trays with a means to advance the mandible / lower jaw are similar and therefore they can be found as substantially equivalent.

Environment of Use - Similar to predicate - ApneaRx - K113569. They are used in Home, Dental offices, Physician offices, Sleep laboratories.

Discussion - Both devices have similar environments of use and therefore they can be found as substantially equivalent.

Patient Population - Similar to predicate - ApneaRx - K113569. 18 years and older Discussion - The patient population is similar and therefore they can be found as substantially equivalent.

### Non-clinical performance testing

#### Commercial experience

The BluePro® has been marketed outside the US since October 2014. The results of this experience support the substantial equivalence of the BluePro® for the intended use, as compared to the predicate device.

The observations support:

- Cleaning via rinsing with water and use of a toothbrush ●
- . Real-time clinical experience from sales outside the US
- . Mechanical and durability after real-time use plus simulated 1 year use via the Braem testing

### Biocompatibility of Materials

All the materials are considered per FDA G95-1 as Surface Contacting, Mucosal membrane, with duration of use prolonged > 24 hours < 30 days, per FDA Guidance Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA dated 11/12/2002

Based upon the Guidance the ISO 10993-1 testing would be:

- Cytotoxicity .
- Sensitization .
- . Irritation

Testing was performed on finished, final devices.

Clinical No clinical testing.

{7}------------------------------------------------

### K162192 510(k) Summary 7-Dec-16

## Discussion of Differences and Conclusion

The subject and predicate devices share equivalent intended use. that of treating snoring and mild to moderate obstructive sleep apnea (OSA). Both the subject and predicate devices share equivalent technology, in that they both use separate pre-formed trays with a means to advance the mandible / lower jaw. Both the subject and predicate devices employ a boil and bite design to provide a custom impression for each patient. Both devices have similar environments of use, patient populations and biocompatibility.

Any differences between the subject and predicate devices are not significant and does not affect the substantial equivalence of the proposed device to the predicate device.

### Substantial Equivalence Conclusion

Based upon the performance testing and comparison to the legally marketed predicate device for indications for use, technology, and performance we believe we have demonstrated that the BluePro® is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K162192](https://fda.innolitics.com/device/K162192)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com