← Product Code [MEA](/productcode/MEA) · K162165

# ambIT PCA*PIB, ambIT PIB, ambIT PIB*PCA, ambIT PIEB, and ambIT Programmable Intermittent Pump (K162165)

_Summit Medical Products, Inc. · MEA · Aug 29, 2017 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K162165

## Device Facts

- **Applicant:** Summit Medical Products, Inc.
- **Product Code:** [MEA](/productcode/MEA.md)
- **Decision Date:** Aug 29, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5725
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The ambIT PCA*PIB Pump is used to infuse medicines and/or fluids into patients primarily for pain management. The routes of administration are generally intravenous, epidural and/or regional. The ambIT PCA*PIB Pump is not intended to supersede, augment, or replace any other medical device or drug indications for use or intended uses. The ambIT PCA*PIB Pump is intended to be used in the home and in healthcare facilities.

## Device Story

Electro-mechanically controlled, rotary-peristaltic infusion pump; consists of reusable pump driver and sterile, single-use cassette (tubing set). Microprocessor-controlled motor drives rollers in cassette at set frequency to achieve desired infusion rate. Powered by two 1.5V AA batteries. User interface includes two buttons (pause/infuse, bolus request) and LCD display for status, alarms, and history. Used in home and healthcare facilities by patients and healthcare providers. Provides PCA (patient-controlled analgesia), PIB (programmed intermittent bolus), and combined PIB+PCA therapy. Monitors downstream occlusions and pump malfunctions. Healthcare providers program infusion parameters; device delivers medication via catheter. Benefits include automated, precise pain management delivery with programmable lockout and dose intervals to optimize therapy.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, usability testing, and assurance case analysis. Bench testing confirmed volumetric accuracy, occlusion alarm performance, and compliance with electrical/safety standards (IEC 60601 series). Usability testing with lay users and healthcare providers confirmed no new safety concerns.

## Technological Characteristics

Rotary-peristaltic infusion pump. Materials: ABS, polycarbonate, nylon, glass-filled nylon. Power: 2x 1.5V AA batteries. Connectivity: None (standalone). Ingress Protection: IP22. Software: Microprocessor-controlled, designed per IEC 62304. Standards: ANSI/AAMI ES60601-1, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-8, IEC 60601-1-11, IEC 60601-2-24, IEC 62366-1.

## Regulatory Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

## Predicate Devices

- ambIT PCA Pump ([K002434](/device/K002434.md))

## Reference Devices

- CADD-Solis Ambulatory Infusion Pump ([K130394](/device/K130394.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image is a logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 29, 2017

Summit Medical Products, Inc Mr. LeVoy Haight CEO/President 504 West 8360 South Sandy, Utah 84070

Re: K162165

Trade/Device Name: ambIT PCA*PIB, ambIT PIB, ambIT PIB*PCA, ambIT PIEB, and ambIT Programmable Intermittent Pump Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion pump Regulatory Class: Class II Product Code: MEA, FRN Dated: July 27, 2016 Received: August 3, 2016

Dear Mr. Haight:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply 

{1}------------------------------------------------

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Image /page/1/Picture/9 description: The image shows the name "Tara A. Ryan -S" in a large, bold font. The name is written in black and is centered on the page. The background is white, and there is a faint watermark in the background.

for Lori Wiggins, MPT, CLT Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K162165

Device Name

ambIT PCA*PIB, ambIT PIB, ambIT PIB*PCA, ambIT PIEB, and ambIT Programmable Intermittent Pump

#### Indications for Use (Describe)

The ambIT PCA*PIB Pump is used to infuse medicines and/or fluids into patients primarily for pain management.

The routes of administration are generally intravenous, epidural and/or regional.

The ambIT PCA*PIB Pump is not intended to supersede, augment, or replace any other medical device or drug indications for use or intended uses.

The ambIT PCA*PIB Pump is intended to be used in the home and in healthcare facilities.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Summit Medical Products. The logo features three overlapping triangles, resembling mountains, in green and gray. Below the triangles, the words "SUMMIT MEDICAL" are stacked above the word "PRODUCTS" in a sans-serif font.

August 29, 2017

Summit Medical Products, Inc. 504 West 8360 South Sandy, UT 84070 Phone: 801-352-1888 FAX: 801-352-1818 Contact Person: LeVoy Haight

Trade Names: ambIT® PCA*PIB, ambIT® PIB, ambIT® PIB*PCA, ambIT® PIEB, and ambIT® Programmable Intermittent Pump Common Name: PCA Infusion Pump or Infusion Pump Classification Name: Infusion Pump Regulation Number: 21 CFR 80.5725 Product Codes: MEA, PCA Infusion Pump FRN, Infusion Pump

Predicate Device K002434—ambIT® PCA Pump

# Reference Device

K130394 – CADD®-Solis Ambulatory Infusion Pump

# Device Description

The new ambIT PCA*PIB is an electro-mechanically controlled, rotary-peristaltic, infusion pump that consists of two components: (1) the reusable pump driver; and (2) the sterile, single-use cassette (tubing set).

The ambIT PCA*PIB Pump has a microprocessor-controlled motor that turns the rollers in the cassette at a set frequency to provide the desired infusion rate/therapy. The amb/T PCA*PIB Pump is powered by two (2) standard 1.5-volt "AA" batteries, and cannot be powered by electrical mains. The batteries are replaceable but cannot be recharged inside the pump.

The controls consist of two buttons and an OFF/ON switch. One button toggles the pump between pause and infusing. The other button (BOLUS button) is used to request a bolus. An LCD display shows the pump status, alarms, and history. The pump monitors downstream occlusions, pump malfunctions, and other pump error conditions. The

{4}------------------------------------------------

materials of construction are all medical grade plastics. Specifically, the materials are ABS, polycarbonate, nylon, and glass-filled nylon.

The part of the pump that comes into indirect contact with the body is the cassette or tubing set. The medications flow through the tubing into a catheter and then into the body. The tubing set is classified to be used for less than 30 days. 1 The length of time the pump is used is based on the therapy requirements.

The ambIT PCA*PIB product has been validated and verified to meet healthcare providers', lay users', and patients' needs for pain management infusions. The pump meets all applicable national and international consensus standards.

# Indications for Use

See Table 1 for a comparison of the Indications for Use between the new and predicate pumps.

| Table 1: Indications for Use - The differences between the two indications for use are |  |  |  |  |  |
|----------------------------------------------------------------------------------------|--|--|--|--|--|
| in italics.                                                                            |  |  |  |  |  |

| ambIT PCA*PIB Pump                                                                                                                                       | ambIT PCA Pump                                                                                                  |
|----------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| The ambIT PCA*PIB Pump is used to infuse<br>medicines and/or fluids into patients<br>primarily for pain management.                                      | The ambIT PCA Pump is used to infuse<br>medicines and/or fluids into patients<br>primarily for pain management. |
| The routes of administration are generally<br>intravenous, epidural and/or regional.                                                                     | Modes of action are generally intravenous, epidural<br>and/or regional.                                         |
| The ambIT PCA*PIB Pump is not intended to<br>supersede, augment, or replace any other<br>medical device or drug indications for use or<br>intended uses. |                                                                                                                 |
| The ambIT PCA*PIB Pump is intended to be<br>used in the home and in healthcare<br>facilities.                                                            |                                                                                                                 |

The differences in the indications for use are as follows:

The phrase "modes of action" has been changed to "routes of administration."

The last two sentences are added to the indication for use statement for completeness.

None of these changes or additions raise new questions of safety or efficacy in relation to the new ambIT PCA*PIB Pump.

<sup>1</sup> The classification is external communicating device, blood path-indirect, prolonged use.

{5}------------------------------------------------

energy remaining in

batteries.3

# Comparison to Predicate Pump

boluses and doses]

The new ambIT PCA*PIB Pump and the predicate ambIT PCA Pump utilize the same technology (microprocessor-controlled, rotary peristaltic pump), and the same cassettes (tubing set). They are built using the same materials and mechanical design, with the same dimensions and validated processes. They also have the same volumetric accuracy (±6%), operating conditions, storage conditions, LCD display, alarms/alerts, and power source. The user needs and specifications are the same for the predicate and new pump except for the addition of a programmable intermittent bolus (PIB) mode and more specificity and range in selecting the bolus volumes, lockout times, and the total volume to be infused.

The new ambIT PCA*PIB Pump and the predicate ambIT PCA Pump cannot be connected to a network or a computer.

The indications for use for the new ambIT PCA*PIB Pump and predicate ambIT PCA Pump are discussed above.

As was mentioned above, the new ambIT PCA*PIB Pump has additional user needs that led to software changes. The software changes allow for more specificity in choosing infusion volumes, bolus volumes, and lockout times when compared to the predicate pump. In addition, the PIB mode has been added as an option to replace the basal infusion rate with automated boluses (i.e., doses). It should be noted that throughout the remainder of this document, the term dose is defined as an automated bolus whose frequency is programmed by the healthcare provider. 2 The changes to the ambIT PCA*PIB Pump have been verified, validated, and shown to raise no new questions with reqards to safety or efficacy. See Table 2 for details on the software differences.

| Parameter         | ambIT<br>PCA*PIB Pump | Predicate<br>ambIT PCA Pump | Comments/Safety-related<br>Discussion                      |
|-------------------|-----------------------|-----------------------------|------------------------------------------------------------|
| Delivery Rate [of | 125-210               | 100 ml/hr                   | The delivery rate is not<br>adjustable. It varies based on |

ml/hr

| Table 2. The differences between the predicate ambIT PCA Pump and new amblT |  |
|-----------------------------------------------------------------------------|--|
| PCA*PIB Pump.                                                               |  |

<sup>2</sup> To avoid confusing healthcare providers, the automated/Intermittent Bolus will be referred to as a dose and a patient-controlled analgesia (PCA) bolus will be called a bolus. The reason is that healthcare providers think of PCA when the term bolus is used and we are maintaining convention with regard to PCA boluses.

<sup>3</sup> The rate will be 210 ml/hr until a low-battery alarm occurs, and then the infusion-rate may drop as low as 125 ml/hr. The variance in delivery rate maximizes the battery life

{6}------------------------------------------------

| Parameter             | ambIT<br>PCA*PIB<br>Pump | Predicate<br>ambIT PCA<br>Pump | Comments/Safety-related<br>Discussion                                                                                                                                                                                                                                                                                                                                                                                                   |
|-----------------------|--------------------------|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Bolus Volume          | 0-50 ml                  | 0-20 ml                        | The initial or "loading" bolus is<br>generally larger than 20 ml<br>and can be as high as 50 ml.<br>This allows for longer lockouts<br>and the ability to "reload" the<br>site later.                                                                                                                                                                                                                                                   |
| Lockout Time4         | 1 minute to<br>24 hours  | 5 minutes to<br>12 hours       | The new pump allows the<br>lockout to be set in 1 minute<br>increments, while the<br>predicate device allows only a<br>limited number of specific<br>values (00:05, 00:10, 00:20,<br>00:30, 00:45, 01:00, 01:30,<br>02:00, 04:00, 08:00, 12:00).<br>The new pump also increases<br>the range for the lockout time.<br>Overall, the new pump allows<br>for greater flexibility in setting<br>lockout times to match the<br>bolus volume. |
| Dose Volume           | 0-50 ml                  | NA                             | NA                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Dose Interval5        | 1 minute to<br>24 hours  | NA                             | This setting is required for a<br>dose to occur at programmed<br>intervals.                                                                                                                                                                                                                                                                                                                                                             |
| Total Infusion Volume | 25 -1,000<br>mL          | 25-1,000 mL                    | The new pump allows infusion<br>volumes to be set in 1 ml<br>increments. The predicate<br>device allows only a limited<br>number of specific values (25,<br>50, 75, 100, 150, 200, 250,<br>300, 350, 400, 450, 500, 600,                                                                                                                                                                                                                |
| Parameter             | ambIT<br>PCA*PIB<br>Pump | Predicate<br>ambIT PCA<br>Pump | Comments/Safety-related<br>Discussion                                                                                                                                                                                                                                                                                                                                                                                                   |
|                       |                          |                                | 700, 800, 900, 1000). The<br>overall range is the same.                                                                                                                                                                                                                                                                                                                                                                                 |
| Ingress Protection    | IP22                     | IPX1                           | The new pump has a higher<br>ingress protection. The<br>increase is in accordance with<br>consensus home use<br>standard, IEC 60601-1-11:<br>2010. The predicate ambIT<br>PCA Pump currently has the<br>same ingress protection, but<br>when the 510(k) for the<br>predicate ambIT PCA Pump<br>was cleared, it was not tested<br>to see if it met IP22.                                                                                 |

<sup>4</sup> The lockout time is the time during which a bolus can be requested but won't be delivered. It starts after a bolus or dose has been delivered.

<sup>్</sup> The dose interval is the length of time that must pass before another dose will be able to be delivered.

{7}------------------------------------------------

The changes to the programming either provide a greater range (maximum bolus volume of 50 ml), allow for more specificity in programming the pump (in 1 ml increments or 1minute increments), or allow for doses to replace the basal rate. The changes are based on feedback from healthcare providers. The programming changes were tested in a simulated environment with healthcare providers to see if the changes created any hazardous situations. The testing showed that no hazardous situations were created due to the software changes.

# Comparison to Reference Pump

To address the software differences between the new and predicate pumps, the amblT PCA*PIB Pump was compared to a FDA cleared reference product (CADD-Solis Pump – K130394). This comparison is given in Table 3 below. Comments following the table discuss how any differences between the two pumps do not call into question the safety and efficacy of the ambIT PCA*PIB Pump.

{8}------------------------------------------------

Table 3. The comparison between the reference CADD-Solis Pump and the new ambIT PCA*PIB Pump. Parameters in italics are the parameter names as found in the CADD-Solis specification literature (if different from the names used by Summit).

| Parameter                                                          | ambIT PCA*PIB<br>Pump                                                                      | CADD-Solis Pump6                                                                                                                                                                                  | Comment |
|--------------------------------------------------------------------|--------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Routes of<br>administration found<br>in the indications for<br>use | Intravenous                                                                                | Intravenous                                                                                                                                                                                       |         |
|                                                                    | Epidural                                                                                   | Epidural Space                                                                                                                                                                                    | 1       |
|                                                                    | Regional                                                                                   | NA                                                                                                                                                                                                |         |
|                                                                    | NA                                                                                         | Subcutaneous,<br>intraperitoneal, in close<br>proximity to nerves,<br>intraoperative site (soft<br>tissue, body<br>cavity/surgical wound<br>site)                                                 |         |
|                                                                    | N/A                                                                                        | Subarachnoid space                                                                                                                                                                                |         |
| Programming modes                                                  | PCA                                                                                        | PCA                                                                                                                                                                                               |         |
|                                                                    | PIB                                                                                        | PIB                                                                                                                                                                                               | NA      |
|                                                                    | PIB+PCA (P+P)                                                                              | PIB+PCA                                                                                                                                                                                           |         |
| Maximum Delivery<br>Rate [of boluses or<br>doses]                  | 210 ml/hr                                                                                  | 250 ml/hr7                                                                                                                                                                                        | 2       |
| Bolus Volume or<br>PCA Dose                                        | 0-50 ml                                                                                    | 0-50 ml                                                                                                                                                                                           | NA      |
| Parameter                                                          | ambIT PCA*PIB<br>Pump                                                                      | CADD-Solis Pump6                                                                                                                                                                                  | Comment |
| Lockout Time8 or PCA<br>Dose Lockout                               | 1 minute to 24 hours<br>in 1-minute<br>increments                                          | 1 minute to 24 hours in<br>the following<br>increments: 1-minute<br>increments for values<br>between 1 and 20<br>minutes; 5-minute<br>increments for values<br>between 20 minutes and<br>24 hours | 3       |
| Dose Volume or<br>Intermittent Bolus                               | 0-50 ml                                                                                    | 0-50 ml                                                                                                                                                                                           | NA      |
| Dose Interval9 or<br>Intermittent Bolus<br>Interval                | 1 minute to 24 hours<br>in 1-minute<br>increments                                          | The amount of time<br>from the start of one<br>intermittent bolus to the<br>start of the next<br>intermittent bolus; 0 to<br>4 hours                                                              | 4       |
| Total Infusion Volume<br>or Reservoir Volume                       | 25 - 1,000 mL;<br>programmable in 1<br>mL increments,<br>displayed in 0.1 mL<br>increments | 0 to 9999 ml;<br>programmable in 1 mL<br>increments, displayed in<br>0.1 ml increments                                                                                                            | 5       |
| Infusion Accuracy                                                  | ±6 %                                                                                       | ±6 %                                                                                                                                                                                              | NA      |
| Pump Type                                                          | Peristaltic                                                                                | Peristaltic                                                                                                                                                                                       | NA      |

<sup>6</sup> Smiths Medical CADD-Solis Ambulatory Pain Management System with Programmed Intermittent Bolus (PIB) Brochure.

<sup>7</sup> This maximum delivery rate is related to the standard volume tubing, For high volume tubing, the maximum rate is 500 ml/hr. It is not 'route of administration'-specific.

{9}------------------------------------------------

Comment 1 – The CADD-Solis Pump includes additional routes of administration in the indications for use than the ambIT PCA*PIB Pump.

Comment 2 – While the 100 ml/hr rate is acceptable in the predicated device, higher delivery rates for boluses and doses are added in an attempt to more closely mimic an actual injection.

<sup>8</sup> The lockout time is the time after a bolus has been delivered during which a bolus can be requested but won't be delivered.

<sup>9</sup> The dose interval is the minimum time from the start of one dose until the start of the next dose.

{10}------------------------------------------------

The CADD-Solis Pump has a maximum delivery rate of 250 ml/hr. The maximum delivery rate of the ambIT PCA*PIB Pump is 210 ml/hr. Thus, a previously cleared device has been shown to have as high or higher a delivery rate and the same maximum bolus and dose amounts. Thus, a delivery rate of 210 ml/hr via any of the proposed routes of administration does not raise any new questions of safety or efficacy.

Comment 3 – The ambIT PCA*PIB Pump has the same range as the CADD-Solis Pump for the Lockout Time/PCA Dose Lockout. The only difference is in the intervals available; the ambIT PCA*PIB Pump allows more specificity in selecting a lockout time after the first 20 minutes. This does not raise any new questions of safety or efficacy.

Comment 4 – The definitions of amblT's 'dose interval range' and CADD's 'intermittent bolus interval' are the same – "The amount of time from the start of one intermittent bolus to the start of the next intermittent bolus" 10; the difference between them is that the ambIT PCA*PIB Pump has a ranqe of 1 minute to 24 hours, while the CADD-Solis Pump has a range of 0 to 4 hours. The availability of longer dose intervals allows the healthcare provider to provide a longer time for the drug to be cleared from the body. Thus, the difference between the ambIT PCA*PIB Pump's dose interval range and the CADD-Solis Pump's intermittent bolus interval does not raise any new questions of safety or efficacy.

Comment 5 – The total infusion volume intervals for the CADD-Solis Pump are the same as the ambIT PCA*PIB Pump - 1 ml intervals. Since the range of the ambIT PCA*PIB Pump's total infusion volume is the same as the predicate ambIT PCA Pump, having a narrower range than the CADD-Solis Pump does not raise any questions of safety or efficacy.

# Non-Clinical Performance Data

Both the predicate ambIT PCA Pump and the new ambIT PCA*PIB Pump meet the same non-clinical performance data. The bench testing on the ambIT PCA*PIB Pump demonstrates that the new pump meets the same volumetric accuracy requirements (when delivering high, low, and median flow rates and bolus volumes), the same usability requirements, and the same downstream occlusion alarm specifications. The volumetric accuracy and occlusion alarm were tested and the results show that the new pump meets the same specifications as the predicate pumps. The usability test results showed that no new concerns related to safety were found when used by lay users or healthcare providers.

Both the predicate ambIT PCA Pump and the new ambIT PCA*PIB Pump meet the following standards: ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012; IEC 60601-1: 2012 Edition 3.1; IEC 60601-1-2:

<sup>10</sup> Smiths Medical CADD-Solis Ambulatory Pain Management System with Programmed Intermittent Bolus (PIB) Brochure.

{11}------------------------------------------------

2014 (Fourth Edition_2014-02); IEC 60601-1-6: 2010; IEC 60601-1-8: 2012 Edition 2.1; IEC 60601-1-11: 2010; IEC 60601-2-24; and IEC 62366-1: 2015 Edition 1.0. The software for all ambIT Pumps were designed following IEC 62304: 2006.

All bench/performance testing was completed using the cassettes (tubing sets). The cassettes also meet the standards listed above and sterilization standards. There were no changes to the fluid contacting components from the predicate device.

Following FDA quidance for infusion pumps, an assurance case was used to demonstrate that the new ambIT PCA*PIB Pump is adequately safe, adequately reliable and adequately designed for its intended use. The processes/techniques explained in ISO 14971 were used in identifying hazards and risks, ensuring that the risks were appropriately mitigated, and that the mitigation was effective. The assurance case shows that the four infusion pump system hazards11 identified in the FDA infusion pump guidance were considered and mitigated. All eight sources of hazardous situations 12 specified in the FDA infusion pump guidance were evaluated by the risk management team. The risk management team determined what hazardous situations from each source could lead to the four infusion pump hazards. The risk management team consisted of clinical, engineering, marketing, and operational personnel.

Additional sources of information used to identify hazards and associated hazardous situations included the following: (1) MDR/Vigilance reports for predicate devices; (2) recall and MDR/Viqilance reports for other infusion pumps; (3) feedback from users via sales representatives and the customer support hotline; (4) non-conforming product reports for predicate products; (5) formative and summative human factors testing; (6) published and non-published articles; (7) specific consensus standards; (8) FDA guidance documents, and (9) nurses, healthcare providers, engineers (software, hardware, and mechanical), and RA/QA personnel.

# Clinical Performance Data

Clinical Performance Data was not relied on to show substantial equivalence.

# Summary

Based on the comparison of the new ambIT PCA*PIB Pump and the predicate ambIT PCA Pump, the new ambIT PCA*PIB Pump is substantially equivalent to the predicate amblT PCA Pump.

<sup>11</sup> The four infusion pump system hazards are (1) infusion delivery error, (2) incorrect therapy, (3) biological/chemical contamination, and (4) traumatic injury.

<sup>12</sup> The eight sources of hazardous situations are (1) operational, (2) environmental, (3) electrical, (4) hardware, (5) software, (6) mechanical, (7) biological and chemical, and (8) use.

---

**Source:** [https://fda.innolitics.com/device/K162165](https://fda.innolitics.com/device/K162165)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
