K162110 · Corpak Medsystems, Inc. A Division of Haylard Health · KNT · Sep 15, 2016 · Gastroenterology, Urology
Device Facts
Record ID
K162110
Device Name
CORGRIP NG/NI Tube Retention System
Applicant
Corpak Medsystems, Inc. A Division of Haylard Health
Product Code
KNT · Gastroenterology, Urology
Decision Date
Sep 15, 2016
Decision
SESK
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Indications for Use
The CORGRIP Nasogastric/Nasointestinal (NGNI) Tube Retention System is indicated for use with enteral feeding tubes of 8 FR and greater and NG decompression, suction and drainage tubes up to 18 FR to prevent inadvertent removal or displacement of the tubes for adult patients.
Device Story
Device prevents inadvertent displacement/removal of nasogastric/nasojejunal tubes in adult patients; secures tube to nasal septum. Components: insertion catheter, retrieval catheter, tube clamp. Clinician inserts catheters into patient's nares; maneuvers to connect magnets behind nasal septum; pulls insertion catheter through nare using retrieval catheter; removes catheters until umbilical tape exits nares; secures tape to feeding tube using clamp. Device provides mechanical retention; benefits patient by maintaining tube placement for feeding/decompression. Used in clinical settings by healthcare providers.
Clinical Evidence
Bench testing only. Laboratory verification testing performed on 14, 16, and 18 Fr clamps using Covidien Salem Sump™ Dual Lumen Stomach Tubes. Tests included visual inspections, tether grip strength, tube grip strength, and tube occlusion. Results demonstrated that additional clamp sizes meet or exceed design requirements and are substantially equivalent to the predicate device.
Technological Characteristics
Mechanical tube retention system consisting of insertion/retrieval catheters with magnetic tips and umbilical tape with a plastic clamp. Dimensions: accommodates tubes 8 FR to 18 FR. Non-powered, manual operation. Biocompatible materials used consistent with predicate device.
Indications for Use
Indicated for adult patients requiring enteral feeding tubes (≥8 FR) or NG decompression, suction, and drainage tubes (up to 18 FR) to prevent inadvertent tube removal or displacement.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of a stylized eagle or bird-like figure, with three overlapping profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 15, 2016
CORPAK MedSystems, Inc. A Division of Halyard Health Stephanie Wasielewski VP, RA/QA 1001 Asbury Drive Buffalo Grove, IL 60089
Re: K162110
> Trade/Device Name: CORGRIP NG/NI Tube Retention System Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal Tube and Accessories Regulatory Class: II Product Code: KNT Dated: July 27, 2016 Received: July 29, 2016
Dear Stephanie Wasielewski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{1}------------------------------------------------
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162110
Device Name CORGRIP NG/NI Tube Retention System
Indications for Use (Describe)
The CORGRIP Nasogastric/Nasointestinal (NGNI) Tube Retention System is indicated for use with enteral feeding tubes of 8 FR and greater and NG decompression, suction and drainage tubes up to 18 FR to prevent inadvertent removal or displacement of the tubes for adult patients.
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|--------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a logo for "CORPAK MEDSYSTEMS". The logo is set against a blue background with rounded corners. The word "CORPAK" is in large, bold, white letters, and there is a yellow line underneath it. Below the line, the word "MEDSYSTEMS" is written in smaller, white letters.
## 510(k) Summary
- l. SUBMITTER
CORPAK MedSystems, Inc. 1001 Asbury Dr. Buffalo Grove, IL 60089
P: 847 403-3400 Contact: Stephanie Wasielewski Date prepared: July 27, 2016
#### II. DEVICE
Trade name: CORGRIP NG/NI Tube Retention System Common name: Tube retention Classification name: Gastrointestinal tube and accessories (21 CFR 876.5980) Regulatory Class: II Product Code: KNT
#### III. PREDICATE DEVICE
CORGRIP NG/NI Tube Retention System (K133599)
#### IV. DEVICE DESCRIPTION
The CORGRIP NG/NI Tube Retention System is an accessary to prevent unwanted displacement or removal of nasogastric/nasojejunal tubes for adult patients by retaining the tube to the patient's nasal septum. The device consists of three main components: insertion (delivery) catheter, retrieval catheter, and tube clamp.
The catheters are placed in either nare of the patient and the clinician maneuvers the catheters to connect the magnets at the back of the patient's nasal septum. Both catheters have a flexible tip to aid in magnet connection. Once connection is made the magnet on the insertion catheter is pulled through the nare by the retrieval catheter. Both catheters are then removed from each nare until the ends of the umbilical tape are exiting either nare. The umbilical tape is then secured to the feeding tube with the provided clamp.
The proposed device adds the option of a 14, 16 and 18 Fr clamp.
## V. INDICATIONS FOR USE
The CORGRIP Nasogastric/Nasointestinal (NG/NI) Tube Retention System is indicated for use with enteral feeding tubes of 8 FR and greater and NG decompression, suction and drainage tubes up to 18 FR to prevent inadvertent removal or displacement of the tubes for adult patients.
{4}------------------------------------------------
#### VI. COMPARISON
The predicate device is identical to the proposed device with the exception of the additional clamp sizes (14,16 and 18Fr).
#### VII. Performance Data
The 14, 16, and 18 Fr CORGRIP NG/NI Tube Retention System clamp testing was conducted in a laboratory setting to establish verification of design requirements. Testing found that all clamps and materials met or exceeded design requirements established for the components when used with Covidien Salem Sump™ Dual Lumen Stomach Tubes. The following tests showed that the additional clamp sizes are substantially equivalent to those of the predicate device:
- . Visual inspections
- Tether grip strength
- Tube grip strength ●
- . Tube occlusion
Based upon the clamp test results, the proposed device design remains substantially equivalent to the predicate device.
#### VIII.Conclusions
There are no material or design changes being made to any components of the CORGRIP. In order to accommodate the larger suction tubes CORPAK has added three clamp sizes 14, 16 and 18 Fr.
The use, safety and effectiveness, of the proposed device clamps are substantially equivalent to the predicate device in terms of intended use, mode of operation, patient population, design, acceptance criteria, and biocompatibility.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.