The C-Flex Ureteral Stent Sets are intended for temporary internal drainage from the ureteropelvic junction to the bladder. Ureteral stents have been used to relieve obstruction in a variety of benign, and post-traumatic conditions. The stents may be placed using endoscopic, percutaneous, or open surgical techniques. The 3.7 French C-Flex stents are indicated for pediatric patients.
Device Story
Single-lumen ureteral stent set; provides temporary internal drainage from ureteropelvic junction to bladder; used for benign, malignant, or post-traumatic obstructions. Stent constructed from radiopaque C-Flex material; ranges 3.7-5.0 French diameter; 6.0-32.0 cm length. Set includes stent, positioner, wire guide, and catheter. Placed via endoscopic, percutaneous, or open surgical techniques by clinicians. Indwelling duration up to 6 months. Sterile, single-use device. Radiopacity allows visualization during placement. MR conditional. Benefits include relief of ureteral obstruction.
Clinical Evidence
Bench testing only. Performance data includes: retention strength and break strength (post-30-day artificial urine soak and accelerated aging), gravity flow rate, radiopacity (fluoroscopic evaluation), biocompatibility, and MR conditional testing. All predetermined acceptance criteria met.
Technological Characteristics
Single-lumen ureteral stent; radiopaque C-Flex material; 3.7-5.0 French diameter; 6.0-32.0 cm length. Supplied sterile, single-use. MR conditional. No software or electronic components.
Indications for Use
Indicated for temporary internal drainage from ureteropelvic junction to bladder in patients with benign, malignant, or post-traumatic obstruction. 3.7 French stents indicated for pediatric patients.
Regulatory Classification
Identification
A ureteral stent is a tube-like implanted device that is inserted into the ureter to provide ureteral rigidity and allow the passage of urine. The device may have finger-like protrusions or hooked ends to keep the tube in place. It is used in the treatment of ureteral injuries and ureteral obstruction.
Predicate Devices
Universa Soft Ureteral Stents and Stent Sets (K151051)
Bioteq Double Pigtail Ureteral Stent Set (K033210)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of three stylized human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the largest and the bottom profile being the smallest.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 24, 2017
Cook Incorporated Minjin Choi Regulatory Affairs Specialist 750 Daniels Way Bloomington, IN 47404
Re: K162104 Trade/Device Name: C-Flex Ureteral Stent Sets Regulation Number: 21 CFR§ 876.4620 Regulation Name: Ureteral Stent Regulatory Class: II Product Code: FAD Dated: March 17, 2017 Received: March 20, 2017
Dear Minjin Choi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Charles Viviano -S
For Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
A-5-1
510(k) Number (if known) K162104
Device Name C-Flex Ureteral Stent Sets
Indications for Use (Describe)
The C-Flex Ureteral Stent Sets are intended for temporary internal drainage from the ureteropelvic junction to the bladder. Ureteral stents have been used to relieve obstruction in a variety of benign, and post-traumatic conditions. The stents may be placed using endoscopic, percutaneous, or open surgical techniques.
The 3.7 French C-Flex stents are indicated for pediatric patients.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda. hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in white, uppercase letters on a red background. Below the word "COOK" is the word "MEDICAL" in white, uppercase letters on a red background. The red background is shaped like a rectangle with a diagonal cut on the bottom right corner.
COOK INCORPORATED 750 DANIELS V
#### 2.0 510(k) Summary
# C-Flex® Ureteral Stent Set 21 CFR §807.92 Date Prepared: April 21, 2017
## Submitted By:
| Submission: | Traditional 510(k) Premarket Notification |
|--------------------|---------------------------------------------------------------|
| Applicant: | Cook Incorporated |
| Contact: | Minjin Choi |
| Applicant Address: | Cook Incorporated<br>750 Daniels Way<br>Bloomington, IN 47404 |
| Contact Phone: | (812) 339-2235 x104901 |
| Contact Fax: | (812) 332-0281 |
#### Device Information:
Trade Name: Common Name: Classification Name: Classification Regulation: Device Class/Classification Panel: C-Flex® Ureteral Stent Set Stent, Ureteral Ureteral Stent 21 CFR §876.4620, Product Code FAD Class II, Gastroenterology/Urology
## Predicate Device:
- Primary predicate device: Universa " Soft Ureteral Stents and Stent Sets (K151051). ■ Secondary predicate device:
Bioteq® Double Pigtail Ureteral Stent Set (K033210).
## Device Description:
The C-Flex Ureteral Stents are single-lumen, ureteral stents inserted using endoscopic, percutaneous, or open surgical techniques in order to provide temporary internal drainage from the ureteropelvic junction to the bladder. They are long-term indwelling devices not to exceed 6 months in the body.
The C-Flex® Ureteral Stent Set is comprised of a stent, positioner, wire guide, and catheter. The C-Flex stent is constructed from a radiopaque C-Flex material and ranges from 3.7 to 5.0 French in outer diameters with working lengths of 6.0 to 32.0 cm. The stents are available in either specified-length or multi-length stent configurations.
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in large, bold, white letters on a red background. Below the word "COOK" is the word "MEDICAL" in smaller, white letters on a red background. The red background is shaped like a rectangle with a diagonal cut on the bottom left corner.
The sets will be supplied sterile and are intended for one-time use. The sets are packaged in a peel-open pouch with a three-year shelf life.
## Indications for Use:
The C-Flex® Ureteral Stent Sets are used for temporary internal drainage from the ureteropelvic junction to the bladder. Ureteral stents have been used to relieve obstruction in a variety of benign, malignant, and post-traumatic conditions. The stents may be placed using endoscopic, percutaneous, or open surgical techniques. The 3.7 French C-Flex stents are indicated for pediatric patients.
## Comparison to Predicate Device:
The C-Flex® Ureteral Stent Set and its primary predicate device, the Universa Soft Ureteral Stents and Stent Sets (K151051), are substantially equivalent in that these devices have similar indications for use, methods of operation, designs, and fundamental technological characteristics. The modifications from the predicate device include:
- 트 Extended stent sizes
- 트 Stent material
- Additional accessories
The proposed C-Flex® Ureteral Stent Set is also similar in indications for use, dimensions, and methods of operation to the Bioteg Double Pigtail Ureteral Stent Set (K033210). Differences between the proposed device and the predicate device include:
- 트 Extended stent sizes
- I Stent material
- 트 Additional accessories
Differences between the characteristics of the proposed device set and the predicate devices are supported by testing.
## Performance Data:
The following testing was performed in order to demonstrate that the proposed C-Flex® Ureteral Stent Set met applicable design and performance requirements.
- 트 Retention Strength and Break Strength - Retention strength testing shows the curl retention during proper clinical use should retain the C-Flex Ureteral Stent Set within the intended anatomy. Testing also shows the curl retention strength must allow for removal of the C-Flex Ureteral Stent Set from the intended anatomy. Break strength testing shows the tensile force during proper clinical use should not fracture the C-Flex Ureteral Stent Set. Additional retention strength and break strength tests were conducted following a 30-day artificial urine soak after
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Cook Medical logo. The word "COOK" is in white, bold, sans-serif font on a red background. Below the word "COOK" is the word "MEDICAL" in white, sans-serif font on a red background. The red background is shaped like a banner.
accelerated aging to the equivalence of three years both with and without a 30-day urine soak. All predetermined acceptance criteria were met.
- 트 Gravity Flow Rate - Testing characterized the gravity flow rate of the C-Flex Ureteral Stent Set. The evaluation was conducted on the stents at time zero.
- . Radiopacity - Testing assessed the radiopacity of the C-Flex Ureteral Stent Set by subjecting it to a comparative fluoroscopic evaluation. Testing was conducted on the stents at time zero.
- . Biocompatibility – Testing shows that the proposed device sets conform with the biocompatibility requirements based on its intended use. All predetermined acceptance criteria were met.
- . Magnetic Resonance (MR) – Testing shows that the proposed device sets are MR conditional based on defined, tested conditions. All predetermined acceptance criteria were met.
# Conclusion:
The results of these tests support a conclusion that the C-Flex® Ureteral Stent Set will perform as intended. The proposed device sets do not raise new questions of safety or effectiveness as compared to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.