← Product Code [MOS](/productcode/MOS) · K162079

# TRILLIUM Oval Head Coil 32 (K162079)

_Hitachi Medical Systems America, Inc. · MOS · Sep 16, 2016 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K162079

## Device Facts

- **Applicant:** Hitachi Medical Systems America, Inc.
- **Product Code:** [MOS](/productcode/MOS.md)
- **Decision Date:** Sep 16, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The TRILLIUM OVAL Head Coil 32 is a 32 channel recieve-only multiple array device used with the Hitachi TRILLIUM OVAL 3.0 Tesla systems for imaging of the head region that can be interpreted by a trained physician.

## Device Story

The TRILLIUM OVAL Head Coil 32 is a 32-channel, receive-only, rigid multiple-array coil designed for use with Hitachi TRILLIUM OVAL 3.0 Tesla MRI systems. It functions as a receiver to detect MR signals (RF emissions/echoes) from hydrogen nuclei in the head region during MRI procedures. The coil enhances reception sensitivity and signal-to-noise (S/N) ratio compared to standard configurations. The system uses these signals to reconstruct cross-sectional images (transverse, coronal, sagittal, oblique, curved) representing spatial proton distribution. These images are interpreted by a trained physician to assist in clinical diagnosis. The device is intended for use in clinical settings where MRI systems are operated by trained personnel.

## Clinical Evidence

No clinical trials were conducted. Effectiveness was supported by the submission of clinical head images and bench testing confirming conformance to NEMA MS 1, NEMA MS 3, AAMI/ANSI ES60601-1, IEC 60601-1-2, and IEC 60601-2-33 standards.

## Technological Characteristics

32-channel receive-only multiple array head coil; rigid housing; active and passive decoupling; 3.0 Tesla MRI system compatibility; NEMA MS 1/MS 3 performance standards; IEC 60601-1/60601-1-2/60601-2-33 safety standards; Class II device.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- TRILLIUM OVAL WIT Posterior Head/Neck coil and WIT Anterior Head attachment ([K142734](/device/K142734.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of the department's name arranged in a circular fashion around a symbol. The symbol is a stylized representation of three human profiles facing to the right, with flowing lines beneath them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 16, 2016

Hitachi Medical Systems America, Inc. % Mr. Doug Thistlethwaite Manager of Regulatory Affairs 1959 Summit Commerce Park TWINSBURG OH 44087

Re: K162079

Trade/Device Name: TRILLIUM OVAL Head Coil 32 Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: MOS Dated: July 28, 2016 Received: July 29, 2016

Dear Mr. Thistlethwaite:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael O'Hara

For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration |                                                                                                                                                                                                                           |
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                                         | <b>Indications for Use</b>                                                                                                                                                                                                |
| 510(k) Number (if known)                                                | K162079                                                                                                                                                                                                                   |
| Device Name                                                             | TRILLIUM OVAL Head Coil 32                                                                                                                                                                                                |
| Indications for Use (Describe)                                          | The TRILLIUM OVAL Head Coil 32 is a 32 channel recieve-only multiple array device used with the Hitachi<br>TRILLIUM OVAL 3.0 Tesla systems for imaging of the head region that can be interpreted by a trained physician. |

  

Type of Use (Select one or both, as applicable)  

| <span style="padding-right: 5px"><svg height="15px" width="15px"><rect fill="black" height="15px" stroke="black" width="15px" x="0" y="0"></rect></svg></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span><svg height="15px" width="15px"><rect fill="white" height="15px" stroke="black" width="15px" x="0" y="0"></rect></svg></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the  
time to review instructions, search existing data sources, gather and maintain the data needed and complete  
and review the collection of information. Send comments regarding this burden estimate or any other aspect  
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services  

Food and Drug Administration  

Office of Chief Information Officer  

Paperwork Reduction Act (PRA) Staff  

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of  
information unless it displays a currently valid OMB number."

| FORM FDA 3881 (8/14) | Page 1 of 1 |
|----------------------|-------------|
|----------------------|-------------|

{3}------------------------------------------------

Section 5

510(k) Summary

{4}------------------------------------------------

| Submitter:           | Hitachi Medical Systems America, Inc.<br>1959 Summit Commerce Park<br>Twinsburg, Ohio 44087-2371 |
|----------------------|--------------------------------------------------------------------------------------------------|
| Registration Number: | 1528028                                                                                          |
| Contact:             | Douglas J. Thistlethwaite                                                                        |
| Telephone number:    | 330-425-1313                                                                                     |
| Telephone number:    | 330-963-0749                                                                                     |
| E-mail:              | thistlethwaited@hitachimed.com                                                                   |
| Date:                | July 11, 2016                                                                                    |

### Submitter Information

### Manufacturer Information

| Manufacturer:        | HITACHI, LTD.<br>2-1, Shintoyofuta<br>Kashiwa-shi, Chiba-Ken, Japan 277-0804 |
|----------------------|------------------------------------------------------------------------------|
| Registration Number: | 8030405                                                                      |

# Device Information

| Regulation Number:      | 892.1000                                                                                 |
|-------------------------|------------------------------------------------------------------------------------------|
| Regulation Name:        | Coil, Magnetic Resonance, Specialty                                                      |
| Product Code            | MOS                                                                                      |
| Class                   | 2                                                                                        |
| Panel                   | Radiology                                                                                |
| Trade/Proprietary Name: | TRILLIUM OVAL Head Coil 32                                                               |
| Predicate Device(s):    | TRILLIUM OVAL WIT Posterior Head/Neck coil and WIT Anterior Head<br>attachment (K142734) |

## Device Intended Use

The TRILLIUM OVAL Head Coil 32 is a 32 channel recieve-only multiple array device used with the Hitachi TRILLIUM OVAL 3.0 Tesla systems for imaging of the head region that can be interpreted by a trained physician.

## Device Description

### Function

The TRILLIUM OVAL Head Coil 32 is a receive-only device that detects the MR signal used to produce transverse, coronal, sagittal, oblique, and/or curved cross-sectional images that display the internal structure of the head. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.

#### Scientific Concepts

Magnetic Resonance imaging (MRI) is based on the fact that certain atomic nuclei have electromagnetic properties that cause them to act as small spinning bar magnets. The most ubiquitous of these nuclei is hydrogen, which makes it the primary nuclei currently used in magnetic resonance imaging. When placed in a static maqnetic field. these nuclei assume a net orientation or alignment with the magnetic field, referred to as a net magnetization vector.

{5}------------------------------------------------

The introduction of a short burst of radiofrequency (RF) excitation of a wavelength specific to the magnetic field strength and to the atomic nuclei under consideration can cause a reorientation of the net magnetization vector. When the RF excitation is removed, the protons relax and return to their original vector. The rate of relaxation is exponential and varies with the character of the proton and its adiacent molecular environment. This re-orientation process is characterized by two exponential relaxation times, called T1 and T2. A RF emission or echo that can be measured accompanies these relaxation events.

The emissions are used to develop a representation of the relaxation events in a three dimensional matrix. Spatial localization is encoded into the echoes by varying the RF excitation. applying appropriate magnetic field gradients in the x, y, and z directions, and changing the direction and strength of these gradients. Images depicting the spatial distribution of the NMR characteristics can be reconstructed by using image processing techniques similar to those used in computed tomography.

#### Physical and Performance Characteristics

The Head Coil 32 is used with TRILLIUM OVAL MRI System. The coil consists of 32 elements multiple array coils used for obtaining diagnostic images of the human head region in MRI System.

The MRI system produces images by using the transmitter coil. Head Coil 32 is used as receiver in the system. The use of transmitter coil ensures uniform volume excitation, while the receiver coils enhance the reception sensitivity or S/N ratio for optimal image quality.

#### Performance Evaluation

Quality assurance measures were conducted on the subject device. Testing included:

- NEMA MS 1
.

- AAMI / ANSI ES60601-1 ●
- IEC 60601-1-2

- NEMA MS 3 . Heat testing
- . IEC 60601-2-33

{6}------------------------------------------------

# Device Technological Characteristics

The technological characteristics (intended use, hardware, and performance) of the subject device are very similar to the predicate device as indicated in Table 1.

|  |  | Table 1 Head Coil 32 Predicate Comparison Table |  |
|--|--|-------------------------------------------------|--|

| Parameter                        |                                        | Predicate Device: WIT Posterior<br>Head/Neck coil and WIT Anterior Head<br>attachment                                                                       | Subject Device: Head Coil 32                                                                                          |  |
|----------------------------------|----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|--|
| Intended Use                     | Indications for<br>Use Statement       | The WIT Posterior Head/Neck coil with<br>WIT Anterior Head attachment is recieve-<br>only multiple array device used for MRI<br>imaging of the head region. | The TRILLIUM OVAL Head Coil 32 is a<br>recieve-only multiple array device used for<br>MRI imaging of the head region. |  |
|                                  | Coil length                            | 575 mm                                                                                                                                                      | 520 mm                                                                                                                |  |
|                                  | Coil height                            | 359 mm                                                                                                                                                      | 380 mm                                                                                                                |  |
|                                  | Coil width                             | 550 mm                                                                                                                                                      | 550 mm                                                                                                                |  |
| Dimensions                       | Figure                                 | 359                                                                                                                                                         | 520                                                                                                                   |  |
|                                  | Coil type                              | Receive only, 15 channel                                                                                                                                    | Receive only, 32 channel                                                                                              |  |
| Coil<br>Architecture             | Housing type                           | Rigid                                                                                                                                                       | Rigid                                                                                                                 |  |
|                                  | Coil Elements                          | 19 Element                                                                                                                                                  | 32 Element                                                                                                            |  |
|                                  | Primary<br>Decoupling                  | Active and passive                                                                                                                                          | Active and passive                                                                                                    |  |
| Performance                      | NEMA MS 1<br>Method 4                  | 145 +/- 20%                                                                                                                                                 | 245 +/- 20%                                                                                                           |  |
| Electrical<br>Safety             | Electrical safety<br>standards applied | UL-60601-1                                                                                                                                                  | UL-60601-1                                                                                                            |  |
|                                  | Safety Class (IEC<br>60601-1)          | Class II                                                                                                                                                    | Class II                                                                                                              |  |
|                                  | Waterproof                             | No                                                                                                                                                          | No                                                                                                                    |  |
|                                  | Explosion proof                        | No                                                                                                                                                          | No                                                                                                                    |  |
| Operator's<br>Labeling<br>Manual |                                        | WIT Posterior Head/Neck coil Instruction<br>Manual                                                                                                          | Head Coil 32 Instruction Manual                                                                                       |  |

{7}------------------------------------------------

 The technological differences from the TRILLIUM OVAL Head Coil 32 to the predicate device are analyzed in Table 2.

| FDA<br>Requirements               | Analyze why any differences between the subject device and predicate(s) do not render the device NSE (e.g., does not<br>constitute a new intended use; and any differences in technological characteristics are accompanied by information that<br>demonstrates the device is as safe and effective as the predicate and do not raise different questions of safety and effectiveness<br>than the predicate ), affect safety or effectiveness, or raise different questions of safety and effectiveness (see section<br>513(i)(1)(A) of the FD&C Act and 21 CFR 807.87(f)).<br>21 CFR 807.87(f)). |             |              |                                  |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|--------------|----------------------------------|
| Device<br>Modification<br>Summary | Coil was designed with 32 channels.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |             |              |                                  |
| Significant<br>Changes            | □ Manufacturing Process                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | □ Labeling  | □ Technology | □ Performance                    |
|                                   | □ Engineering                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | □ Materials | □ Others     | ☑ None (See rationale statement) |
| HITACHI<br>Rationale<br>Statement | The performance and technological characteristics of coil are similar to the predicate device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |             |              |                                  |

#### Table 2 Differences Analysis

There are no functionality differences from the TRILLIUM OVAL Head Coil 32 to the predicate device as analyzed in Table 3.

| Table 3 Functionality Comparison Analysis |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |             |              |                                  |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|--------------|----------------------------------|
| FDA Requirements                          | Analyze why any differences between the subject device and predicate(s) do not render the device NSE (e.g., does not constitute a new intended use; and any differences in technological characteristics are accompanied by information that demonstrates the device is as safe and effective as the predicate and do not raise different questions of safety and effectiveness than the predicate ), affect safety or effectiveness, or raise different questions of safety and effectiveness (see section 513(i)(1)(A) of the FD&C Act and 21 CFR 807.87(f)). |             |              |                                  |
| Device Modification Summary               | None                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |             |              |                                  |
| Significant Changes                       | □ Manufacturing Process                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | □ Labeling  | □ Technology | □ Performance                    |
|                                           | □ Engineering                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | □ Materials | □ Others     | ☑ None (See rationale statement) |
| HITACHI Rationale Statement               | There is no new intended use.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |             |              |                                  |

{8}------------------------------------------------

# Substantial Equivalence

A summary decision was based on analysis of Table 8.

| Table 8 Rationale Analysis: TRILLIUM OVAL V5.1 MRI vs. Predicate |  |  |  |
|------------------------------------------------------------------|--|--|--|
|------------------------------------------------------------------|--|--|--|

| ITEM          | Overall Rationale Analysis                                                                                                                                     |
|---------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Hardware      | Different specifications do not constitute a new intended use. There are no significant changes in technological<br>characteristics, safety and effectiveness. |
| Functionality | Enhanced features do not constitute a new intended use. There are no significant changes in technological characteristics,<br>safety and effectiveness.        |

Therefore, based on a thorough analysis and comparison of the functions, scientific concepts, physical and performance characteristics, performance comparison and technological characteristics, the proposed TRILLIUM OVAL Head Coil 32 is considered substantially equivalent to the currently marketed predicate device.

# Summary of Non-Clinical Testing

The TRILLIUM OVAL Head Coil 32 is in conformance with the applicable parts of the following standards:

- . NEMA MS 1-2008, Determination of Signal-to-noise Ratio (SNR) in Diagnostic Magnetic Resonance Images
- . NEMA MS 3-2008, Determination of Image Uniformity in Diagnostic Magnetic Resonance Images
- AAMI / ANSI ES60601-1:2005/(R) 2012 and A1:2012, c1:2009/(r) 2012 and a2:2010/(r ) . 2012 (consolidated text) medical electrical equipment - part 1: general requirements for basic safety and essential performance (iec 60601-1:2005, mod).
- . IEC 60601-1-2 Edition 3:2007-03. medical electrical equipment - part 1-2: general requirements for basic safety and essential performance - collateral standard: electromagnetic compatibility - requirements and tests.
- IEC 60601-2-33 Edition 3.1 2013-04, medical electrical equipment part 2-33: particular . requirements for the basic safety and essential performance of magnetic resonance equipment for medical diagnostic.

In addition, heat testing was conducted on the new TRILLIUM OVAL Head Coil 32.

# Summary of Clinical Testing

No clinical tests were conducted to support the TRILLIUM OVAL Head Coil 32 and the substantial equivalence conclusion. However, clinical images of the head were provided to support the effectiveness of the subject device.

### Conclusions

It is the opinion of Hitachi Medical Systems America. Inc. the TRILLIUM OVAL Head Coil 32 has the same intended use and the same basic technological characteristics as the predicate device. While there are some technical features that vary with respect to the predicate device, the conclusions from the non-clinical data suggest that the subject device bears an equivalent safety and performance profile as that of the predicate device.

Therefore, the TRILLIUM OVAL Head Coil 32 is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K162079](https://fda.innolitics.com/device/K162079)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
