← Product Code [KCT](/productcode/KCT) · K162063

# Steri-Cassette and Steri-Cage Sterilization Packaging System (K162063)

_Sybron Dental Specialties · KCT · Apr 21, 2017 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K162063

## Device Facts

- **Applicant:** Sybron Dental Specialties
- **Product Code:** [KCT](/productcode/KCT.md)
- **Decision Date:** Apr 21, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6850
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The Steri-Cassettes and Steri-Cages are intended to contain dental instruments for cleaning, sterilization, organization, storage and handling. The Steri-Cassettes and Steri-Cages are suitable for pre-vacuum and gravity displacement steam sterilization methods. The cassettes and cages are not intended to maintain sterility; they are intended to be used in conjunction with validated, FDA-cleared sterilization wrap in order to maintain sterility of the enclosed devices.

## Device Story

Steri-Cassettes and Steri-Cages are perforated, stackable containers used to organize and secure dental instruments during cleaning, steam sterilization, and storage. Available in various sizes and colors, they feature hinged lids with latching mechanisms and removable instrument holder inserts. The devices are manufactured via injection molding using high-heat resin (Ultem 1000) or polypropylene. They are used in dental clinics; operated by dental staff. The devices are placed in autoclaves for steam sterilization; they do not maintain sterility independently and must be used with FDA-cleared sterilization wrap. The perforations allow for sterilant penetration and drying. The system facilitates efficient instrument handling and organization at the point of use.

## Clinical Evidence

Bench testing only. No clinical data. Validation included steam sterilization efficacy (pre-vacuum and gravity displacement cycles), cleaning validation, material compatibility, and packaging integrity testing in accordance with ANSI/AAMI ST77, ISO 17665-1, and AAMI TIR30/TIR12 standards.

## Technological Characteristics

Materials: High-heat resin (Ultem 1000) or polypropylene. Design: Perforated (slotted) base and lid, hinged with latching mechanism. Form factor: Stackable cassettes/cages in various sizes. Sterilization: Steam (pre-vacuum and gravity displacement). Standards: ANSI/AAMI ST77, ISO 10993-1, ISO 17664, ISO 14937, ANSI/AAMI ST79.

## Regulatory Identification

A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.

## Predicate Devices

- PolyVac Surgical Instrument Delivery System's Instrument Cassettes ([K012105](/device/K012105.md))

## Reference Devices

- PeelVue+ sterilization pouches ([K894437](/device/K894437.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, layered on top of each other, creating a sense of depth and unity.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 21, 2017

Sybron Dental Specialties Jennifer Dzidrums Regulatory Affairs Associate Ii 1717 W. Collins Ave. Orange, California 92867

Re: K162063

Trade/Device Name: Steri-Cassette and Steri-Cage Sterilization Packaging System Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: March 15, 2017 Received: March 20, 2017

Dear Jennifer Dzidrums:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

# Michael J. Ryan -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

### 510(k) Number (if known) K162063

### Device Name

Steri-Cassette and Steri-Cage Sterilization Packaging Systems

### Indications for Use (Describe)

The Steri-Cassettes and Steri-Cages are intended to contain dental instruments for cleaning, sterilization, organization, storage and handling. The Steri-Cassettes and Steri-Cages are suitable for pre-vacuum and gravity displacement steam sterilization methods. The cassettes and cages are not intended to maintain sterility; they are intended to be used in conjunction with validated, FDA-cleared sterilization wrap in order to maintain sterility of the enclosed devices.

The validated steam sterilization cycle parameters are as follows:

| Steam Sterilization<br>Cycle                                | Minimum Exposure<br>Temperature | Minimum<br>Exposure Time | Minimum Dry Time |
|-------------------------------------------------------------|---------------------------------|--------------------------|------------------|
| Gravity<br>Displacement<br>(wrapped)                        | 250°F (121°C)                   | 30 Minutes               | 15 Minutes       |
| Gravity<br>Displacement<br>(wrapped)                        | 270°F (132°C)                   | 15 Minutes               | 15 Minutes       |
| Gravity<br>Displacement<br>(wrapped)                        | 275°F (135°C)                   | 10 Minutes               | 30 Minutes       |
| Gravity<br>Displacement<br>(unwrapped for<br>immediate use) | 132°C (270°F)                   | 3 Minutes                | None             |
| Pre-Vacuum<br>(wrapped)                                     | 270°F (132°C)                   | 4 Minutes                | 30 Minutes       |
| Pre-Vacuum<br>(wrapped)                                     | 275°F (135°C)                   | 3 Minutes                | 16 Minutes       |
| Pre-Vacuum<br>(unwrapped for<br>immediate use)              | 273°F (134°C)                   | 3 Minutes                | None             |

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Sybron Dental Specialties. The logo consists of the letters "sds" in a bold, sans-serif font, with a curved line above the letters. Below the letters, the words "Sybron Dental Specialties" are written in a smaller, sans-serif font. The logo is blue and white.

# 510(k) SUMMARY for Steri-Cassette and Steri-Cage

# K162063

In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR §807.92, the following summary of information is provided:

#### A. Applicant Information:

Sybron Dental Specialties 1717 W. Collins Ave. Orange CA, 92687 USA Fax: 909-962-5694

### Correspondent Contact Information:

Jennifer Dzidrums, MS, RAC Regulatory Affairs Associate II Tel: 909-962-5650 Fax: 909-962-5694

Date Prepared: April 17, 2017

#### B. Subject Device:

| Trade Name          | Steri-Cassette and Steri-Cage Sterilization Packaging System      |
|---------------------|-------------------------------------------------------------------|
| Classification Name | Sterilization wrap containers, trays, cassettes & other accessory |
| Regulation Number   | 21 CFR § 880.6850                                                 |
| Common Name         | Instrument Cassette, Sterilization Cassette                       |
| Device Class        | II                                                                |
| Product Code        | KCT                                                               |
| Panel               | General Hospital                                                  |

#### Predicate Device: C.

| Trade Name          | PolyVac Surgical Instrument Delivery System's Instrument<br>Cassettes |
|---------------------|-----------------------------------------------------------------------|
| 510(k) Holder       | Symmetry Medical, Inc. (previously PolyVac, Inc.)                     |
| 510(k) #            | K012105 (FDA-cleared on August 02, 2002)                              |
| Classification Name | Sterilization wrap containers, trays, cassettes & other accessory     |
| Regulation Number   | 21 CFR § 880.6850                                                     |
| Common Name         | Instrument Cassette/ Tray, Sterilization Cassette/ Tray               |
| Device Class        | II                                                                    |
| Product Code        | KCT                                                                   |
| Panel               | General Hospital                                                      |

{4}------------------------------------------------

#### D. Description of Device:

Steri-Cassettes and Steri-Cages are used to hold instruments in a secure way for cleaning, sterilization and presentation at point of use. They keep instrumentation organized in an efficient way. They are available in a variety of sizes to accommodate instruments of various sizes and are offered in an assortment of colors to aid in office organization.

### Design

Steri-Cassettes and Steri-Cages have perforations consisting of open-slots on the lid and the base which allows exposure of the device to the sterilant during sterilization. All the five variations of the stackable Steri-Cassettes and Steri-Cages have the same design of open-slot perforations on their lid and base (1) Steri-Cassette AA - Shallow - High Heat, (2) Steri-Cassette AA - Shallow - Standard Heat, (3) Steri-Cassette AB - Medium -High Heat, (4) Steri-Cassette BB - Deep - High Heat and (5) Steri-Cage - Standard Heat. Steri-Cassettes include instrument holder inserts, which are removable and individually hold instruments. Refer to Table 5.1 for the various colors, measurements, construction and heat resistance for these hinged sterilization cassettes and cages. Standard Heat Steri-Cassettes and Steri-Cages and High Heat Steri-Cassettes can withstand 275°F and 320°F per performance and Sterilization testing, and resin specifications).

Since the proposed devices are perforated, an FDA-cleared sterilization wrap must be used to maintain sterility of its contents. The evenly-spaced open slots are sufficient to support sterilant penetration and drying as per ANSI/AAMI ST 77.

### Composition and Manufacturing Process

Cassettes and cages are manufactured by Injection Molding process from either high-heat resin or polypropylene with color - materials widely used in the medical field. The colors are already pre-mixed and vary by color in the supplier's proprietary resin mixtures: Steri-Cassettes High-Heat: RoHs-compliant, high-heat resin - Ultem Resin 1000 and Steri-Cassettes Standard Heat and Steri-Cages: polypropylene with color. The manufacturer name and UDI direct part marking will be added to the mold during the Injection Molding process.

The cassettes and cage have a latching lid to contain the products. The proposed devices are designed to fit any standard autoclave and ultrasonic machine (Table 5.1), which allows it to be effective for sterilization and be able to withstand the environment of repeated steam sterilization cvcles. Since the proposed cassettes and cages have open slots, a legally-marketed and FDA-cleared sterilization wrap must be used for sterilization purposes to maintain the sterility of the contents.

The proposed Steri-Cassettes and Steri-Cages are packaged in non-sterile, plastic shrinkwrap, which is then appropriately labeled and includes a validated Cleaning and Sterilization/ Dry Time Instructions for use.

Table 5.2 describes the validated, FDA-cleared sterilization pouch required for the sterilization procedure.

{5}------------------------------------------------

# Table 5.1: Design Description: Subject Device – Size, Color and Heat-Resistance Type

| Proposed Device<br>Name and Heat-<br>Resistance   | Part Number                                                               | PeelVue+<br>Part Number                        | Color                                                                                           | Outer Dimensions<br>(when closed)<br>Width x Depth x Height |
|---------------------------------------------------|---------------------------------------------------------------------------|------------------------------------------------|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------|
| Steri-Cassette AA –<br>Shallow – High<br>Heat     | 31591                                                                     | 31615                                          | Beige                                                                                           |                                                             |
| Steri-Cassette AA –<br>Shallow – Standard<br>Heat | 32520, 32521,<br>32522, 32523,<br>32524, 32525                            | 31615                                          | French Vanilla,<br>Sand, Light<br>Mauve, Seafoam,<br>Lilac, Baby Blue                           | 7.5" x 5.5" x 1.5"                                          |
| Steri-Cassette AB –<br>Medium – High<br>Heat      | 31592                                                                     | Baby Blue<br>31643 or<br>Forest Green<br>31647 | Beige                                                                                           | 7.5" x 5.5" x 2.25"                                         |
| Steri-Cassette BB –<br>Deep – High Heat           | 31593                                                                     | Baby Blue<br>31643 or<br>Forest Green<br>31647 | Beige                                                                                           | 7.5" x 5.5" x 3.0"                                          |
| Steri-Cage –<br>Standard Heat                     | 32506, 32507,<br>32508, 32509,<br>32510, 32511,<br>32512, 32513,<br>32514 | 31645<br>31615                                 | French Vanilla,<br>Sand, Light<br>Mauve, Seafoam,<br>Lilac, Baby Blue,<br>White, Gray,<br>Beige | 8.00" x 1.75" x 1.75"                                       |

Note: Coordinating PeelVue+ size and shape identified by Part Number.

{6}------------------------------------------------

| Description and<br>Specific Recommendation                                                        | FDA Clearance Information                                                                         | Manufacturer                                                                                               |
|---------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|
| Validated, FDA-cleared<br>sterilization pouch –<br>PeelVue+ sterilization pouches<br>recommended* | • 510(k) #: K894437<br>• Product Code: FRG<br>• Regulation #: 21 CFR §<br>880.6850<br>• Class: II | Kerr Corporation<br>1717 W. Collins Ave.<br>Orange, CA 92867 USA<br>Establishment<br>Registration: 2024312 |

Table 5.2: Recommended System Components - Used in Conjunctions with Subject 01710000

*Specific brand used in Sterilization Validation testing for the subject devices.

#### E. Statement of Indications for Use:

The Steri-Cassettes and Steri-Cages are intended to contain dental instruments for cleaning, sterilization, organization, storage and handling. The Steri-Cassettes and Steri-Cages are suitable for pre-vacuum and gravity displacement steam sterilization methods. The cassettes and cages are not intended to maintain sterility; they are intended to be used in conjunction with validated, FDA-cleared sterilization wrap in order to maintain sterility of the enclosed devices.

The validated steam sterilization cycle parameters are as follows:

| Steam<br>Sterilization Cycle                                | Minimum Exposure<br>Temperature | Minimum<br>Exposure Time | Minimum Dry Time |
|-------------------------------------------------------------|---------------------------------|--------------------------|------------------|
| Gravity<br>Displacement<br>(wrapped)                        | 250°F (121°C)                   | 30 Minutes               | 15 Minutes       |
| Gravity<br>Displacement<br>(wrapped)                        | 270°F (132°C)                   | 15 Minutes               | 15 Minutes       |
| Gravity<br>Displacement<br>(wrapped)                        | 275°F (135°C)                   | 10 Minutes               | 30 Minutes       |
| Gravity<br>Displacement<br>(unwrapped for<br>immediate use) | 132°C (270°F)                   | 3 Minutes                | None             |
| Pre-Vacuum<br>(wrapped)                                     | 270°F (132°C)                   | 4 Minutes                | 30 Minutes       |
| Pre-Vacuum<br>(wrapped)                                     | 275°F (135°C)                   | 3 Minutes                | 16 Minutes       |
| Pre-Vacuum<br>(unwrapped for<br>                            | 273°F (134°C)                   | 3 Minutes                | None             |

{7}------------------------------------------------

| immediate use) |  |  |  |
|----------------|--|--|--|
|                |  |  |  |

{8}------------------------------------------------

| Element                                             | Subject Device<br>Steri-Cassette and Steri-Cage                                                                                                                                          | Predicate Device (K012105)<br>PolyVac Surgical Instrument Delivery System's<br>Instrument Cassettes                                                                          | Subject Device<br>vs.<br>Predicate Device |
|-----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|
| Trade/ Proprietary<br>Name                          | Steri-Cassette and Steri-Cage Sterilization Packaging<br>System                                                                                                                          | PolyVac Surgical Instrument Delivery System's<br>Instrument Cassettes                                                                                                        | N/A                                       |
| Fundamental<br>Scientific Technology                | Sterilization Cassette                                                                                                                                                                   | Sterilization Cassette                                                                                                                                                       | Same                                      |
| Product Code.<br>Regulation No. and<br>Device Class | KCT (21 CFR § 880.6850), Class II                                                                                                                                                        | KCT (21 CFR § 880.6850), Class II                                                                                                                                            | Same                                      |
| Legal Manufacturer                                  | Kerr Corporation<br>1717 W. Collins Ave.<br>Orange, CA 92867 USA<br>Registration Number: 2024312                                                                                         | Symmetry Medical, Inc. (previously PolyVac, Inc.)<br>220 West Market Street<br>Warsaw, Indiana 46582 USA<br>Registration Number: Not stated in predicate's 510(k)<br>summary | N/A                                       |
| Contract Manufacturer                               | SDS de Mexico S. de R.L. de C.V. (subsidiary of Kerr)<br>Circuito Sur Num. 31<br>Parque Ind. Nelson<br>Mexicali, B.C. C.P. Baja California, MEXICO 21395<br>Registration Number: 9680845 | Not stated in predicate's 510(k) summary                                                                                                                                     | N/A                                       |
| Intended Use                                        | Perforated trays with lids to hold surgical instruments in<br>place during transport, steam sterilization, and storage                                                                   | Perforated trays with lids to hold surgical instruments in<br>place during transport, steam sterilization, and storage                                                       | Same                                      |

# Table 5.3: Device Comparison Table Demonstrating Substantial Equivalence

{9}------------------------------------------------

| Element                  | Subject Device<br>Steri-Cassette and Steri-Cage                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Predicate Device (K012105)<br>PolyVac Surgical Instrument Delivery System's<br>Instrument Cassettes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Subject Device<br>vs.<br>Predicate Device |                     |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|---------------------|
| Description              | Steri-Cassettes and Steri-Cages are used to hold instruments<br>in a secure way for cleaning, sterilization and presentation at<br>point of use. They keep instrumentation organized in an<br>efficient way. They are available in a variety of sizes to<br>accommodate instruments of various sizes and are offered in<br>an assortment of colors to aid in office organization.                                                                                                                                                                              | PolyVac Delivery Systems consist of different sizes of the<br>same basic configuration. All systems consist of a<br>minimum of a plastic or metal base and lid. All lids can be<br>fastened to the base by means of assembled hardware or by<br>a locking tab, designed as part of the lid. Accessories may<br>be used with systems to organize or separate contents to be<br>placed in them for use.<br>The Delivery Systems are designed using plastic and metal<br>materials that can be reused with steam sterilization<br>methods. Each tray and lid has an evenly distributed hole<br>pattern in relation to its size. | Similar                                   |                     |
| Element                  | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Predicate Device (K012105)<br>PolyVac Surgical Instrument Delivery System's<br>Instrument Cassettes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Subject Device<br>vs.<br>Predicate Device |                     |
|                          | Steri-Cassette and Steri-Cage                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                           |                     |
|                          | The Steri-Cassettes and Steri-Cages are intended to contain dental instruments for cleaning, sterilization, organization, storage and handling. The Steri-Cassettes and Steri-Cages are suitable for pre-vacuum and gravity displacement steam sterilization methods. The cassettes and cages are not intended to maintain sterility; they are intended to be used in conjunction with validated, FDA-cleared sterilization wrap in order to maintain sterility of the enclosed devices.<br>The validated steam sterilization cycle parameters are as follows: | PolyVac's delivery systems consist of perforated trays with lids, which are intended to enclose and protect medical device instrumentation, and to facilitate the sterilization process by allowing steam penetration and air removal. When used in conjunction with an approved sterilization wrap, sterility of the enclosed medical device is maintained until used.<br>PolyVac's delivery systems are to be sterilized in one of the following cycles:<br>Prevacuum Steam: 132°C - 4 minutes minimum<br>Gravity Steam: 132°C - 30 minutes minimum<br>Gravity Steam: 121°C - 55 minutes minimum                           | Similar                                   |                     |
| Indications for Use      | Steam<br>Sterilization<br>Cycle                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Minimum<br>Exposure<br>Temperature                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Minimum<br>Exposure<br>Time               | Minimum<br>Dry Time |
|                          | Gravity<br>Displacement<br>(wrapped)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 250°F<br>(121°C)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 30 Minutes                                | 15 Minutes          |
|                          | Gravity<br>Displacement<br>(wrapped)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 270°F<br>(132°C)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 15 Minutes                                | 15 Minutes          |
|                          | Gravity<br>Displacement<br>(wrapped)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 275°F<br>(135°C)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 10 Minutes                                | 30 Minutes          |
|                          | Gravity<br>Displacement<br>(unwrapped<br>for immediate<br>use)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 132°C<br>(270°F)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 3 Minutes                                 | None                |
|                          | Pre-Vacuum<br>(wrapped)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 270°F<br>(132°C)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 4 Minutes                                 | 30 Minutes          |
|                          | Pre-Vacuum<br>(wrapped)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 275°F<br>(135°C)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 3 Minutes                                 | 16 Minutes          |
| Element                  | Subject Device<br>Steri-Cassette and Steri-Cage                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Predicate Device (K012105)<br>PolyVac Surgical Instrument Delivery System's<br>Instrument Cassettes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Subject Device<br>vs.<br>Predicate Device |                     |
| - DESIGN -               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                           |                     |
| Material Composition     | Cassettes – High Heat: High-Resistant Plastic - Ultem Resin<br>1000 (proprietary mixture with pigments) (thermoplastic)<br>Cassettes - Standard Heat: polypropylene with color<br>(proprietary mixture with pigments) (thermoplastic)<br>Cages: polypropylene with color (proprietary mixture with<br>pigments) (thermoplastic)                                                                                                                                                                                                                                | Trays & Lids: Aluminum, 300 Series stainless steel,<br>biomedical grade silicone; or Radal R Plastic<br>Instrument Cassettes: Radal R Plastic (thermoplastic)                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Similar                                   |                     |
| Design/ Configuration    | Cassettes and cages are manufactured from one of two<br>thermoplastic mixtures used in injection molding.<br>Perforated (slotted) base and lid hinge together and are<br>secured with a latch mechanism (tab on lid).<br>Cassettes have built-in inserts used to stabilize the<br>instruments. These inserts are removable and composed of<br>the same thermoplastic as the body of the cassette.<br>Cassettes and cages are stackable.                                                                                                                        | Instrument Cassettes:<br>Manufactured from a thermoplastic mixture (Radal R) used<br>in injection molding.<br>Perforated (slots and holes) base and lid hinge together and<br>are secured with a latch mechanism (tab on lid).<br>Cassettes have built-in slots (Radal R) used to stabilize the<br>instruments or may use a non-slip, silicone mat to secure<br>the instruments while providing ventilation to facilitate<br>drying.<br>Cassettes are stackable.                                                                                                                                                             | Similar                                   |                     |
| Manufacturing<br>Process | Injection Molding                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Injection Molding                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                      |                     |

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

| Element                                                | Subject Device<br>Steri-Cassette and Steri-Cage                                     | Predicate Device (K012105)<br>PolyVac Surgical Instrument Delivery System's<br>Instrument Cassettes                                                                                                                                                                                                                                                         | Subject Device<br>vs.<br>Predicate Device |  |  |  |  |  |  |  |  |  |  |  |  |  |  |         |
|--------------------------------------------------------|-------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|---------|
| Dimensions                                             | Various - refer to Table 5.1 for outer dimensions                                   | Various depending on model of Instrument Cassette:<br>Device Name Part Number Outer Dimensions (when closed)<br>Width x Depth x Height   Shallow Cassette w/ Silicone Mat 7-6050 8.0" x 4.5" x 0.61"   Deep Cassette w/ Silicone Mat 7-1250 8.0" x 4.5" x 1.2"   6-Position Slots 7-6000 8.0" x 4.5" x 0.61"   12-Position Slots 7-1200 8.0" x 4.5" x 0.61" |                                           |  |  |  |  |  |  |  |  |  |  |  |  |  |  | Similar |
| Air Permeance                                          | Yes                                                                                 | Yes                                                                                                                                                                                                                                                                                                                                                         | Same                                      |  |  |  |  |  |  |  |  |  |  |  |  |  |  |         |
| Percent Perforation                                    | Each base and lid contains evenly distributed slot-pattern in relation to its size. | Each base and lid contains evenly distributed slot and hole pattern in relation to its size.                                                                                                                                                                                                                                                                | Same                                      |  |  |  |  |  |  |  |  |  |  |  |  |  |  |         |
| - PERFORMANCE CHARACTERISTICS and VALIDATION TESTING - |                                                                                     |                                                                                                                                                                                                                                                                                                                                                             |                                           |  |  |  |  |  |  |  |  |  |  |  |  |  |  |         |
| Cleaning Instructions for Reusable Devices             | Yes - Per Cleaning Validations conducted.                                           | Yes - Per Cleaning Validations conducted.                                                                                                                                                                                                                                                                                                                   | Same                                      |  |  |  |  |  |  |  |  |  |  |  |  |  |  |         |
| Sterilization Method                                   | • Pre-Vacuum<br>• Gravity Displacement                                              | • Pre-Vacuum<br>• Gravity Displacement                                                                                                                                                                                                                                                                                                                      | Same                                      |  |  |  |  |  |  |  |  |  |  |  |  |  |  |         |

{13}------------------------------------------------

| Element                                                                                  | Subject Device<br>Steri-Cassette and Steri-Cage                                                                      |                                    |                             |                     | Predicate Device (K012105)<br>PolyVac Surgical Instrument Delivery System's<br>Instrument Cassettes                  |                                    |                             |                               | Subject Device<br>vs.<br>Predicate Device |
|------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|------------------------------------|-----------------------------|---------------------|----------------------------------------------------------------------------------------------------------------------|------------------------------------|-----------------------------|-------------------------------|-------------------------------------------|
| Sterilization<br>Parameters per US<br>Requirements<br>(wrapped)*                         | Steam<br>Sterilization<br>Cycle                                                                                      | Minimum<br>Exposure<br>Temperature | Minimum<br>Exposure<br>Time | Minimum<br>Dry Time | Steam<br>Sterilization<br>Cycle                                                                                      | Minimum<br>Exposure<br>Temperature | Minimum<br>Exposure<br>Time | Minimu<br>m Dry<br>Time       | Similar                                   |
|                                                                                          | Gravity<br>Displacement                                                                                              | 250°F (121°C)                      | 30<br>Minutes               | 15<br>Minutes       | Gravity<br>Displacement                                                                                              | 250°F<br>(121°C)                   | 55<br>Minutes               | 20-50<br>Minutes<br>as needed |                                           |
|                                                                                          | Gravity<br>Displacement                                                                                              | 270°F (132°C)                      | 15<br>Minutes               | 15<br>Minutes       | Gravity<br>Displacement                                                                                              | 270°F<br>(132°C)                   | 30<br>Minutes               | 20-50<br>Minutes<br>as needed |                                           |
|                                                                                          | Gravity<br>Displacement                                                                                              | 275°F (135°C)                      | 10<br>Minutes               | 30<br>Minutes       | Pre-Vacuum                                                                                                           | 270°F<br>(132°C)                   | 4 Minutes                   | 20-40<br>Minutes<br>as needed |                                           |
|                                                                                          | Pre-Vacuum                                                                                                           | 270°F (132°C)                      | 4 Minutes                   | 30<br>Minutes       |                                                                                                                      |                                    |                             |                               |                                           |
| Pre-Vacuum                                                                               | 275°F (135°C)                                                                                                        | 3 Minutes                          | 16<br>Minutes               |                     |                                                                                                                      |                                    |                             |                               |                                           |
| *Refer to Indications for Use for immediate use sterilization<br>parameters (unwrapped). |                                                                                                                      |                                    |                             |                     |                                                                                                                      |                                    |                             |                               |                                           |
| Reusable                                                                                 | Yes, validated to be reused at least five (5) times.                                                                 |                                    |                             |                     | Yes                                                                                                                  |                                    |                             |                               | Similar                                   |
| Material Compatibility<br>with Sterilization<br>Method                                   | Yes - Materials are compatible with sterilization method                                                             |                                    |                             |                     | Yes - Materials are compatible with sterilization method                                                             |                                    |                             |                               | Same                                      |
| Sterilant Penetration<br>Studies                                                         | Yes - Sterilant (steam) penetration through perforations in<br>base and lid - Steam Sterilization Validation testing |                                    |                             |                     | Yes - Sterilant (steam) penetration through perforations in<br>base and lid - Steam Sterilization Validation testing |                                    |                             |                               | Same                                      |

{14}------------------------------------------------

| Element                                                                        | Subject Device                                                                                                                                                        | Predicate Device (K012105)<br>PolyVac Surgical Instrument Delivery System's<br>Instrument Cassettes | Subject Device<br>vs.<br>Predicate Device |
|--------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|-------------------------------------------|
| Microbial Barrier<br>Studies (packaging<br>integrity to maintain<br>sterility) | To be used with a validated, FDA-cleared sterilization wrap                                                                                                           | To be used with a validated, FDA-cleared sterilization<br>wrap                                      | Same                                      |
| Material Compatibility<br>with Repeat<br>Sterilization                         | Yes - Materials are compatible with repeated sterilization<br>cycles per validation testing.<br>Materials of construction are compatible with steam<br>sterilization. | Yes - Materials are compatible with repeated sterilization<br>cycles.                               | Similar                                   |
| - SAFETY / BIOCOMPATIBILITY ASSESSMENT -                                       |                                                                                                                                                                       |                                                                                                     |                                           |
| Toxicological<br>Properties                                                    | Yes - Materials are known to be biocompatible and do not<br>come into direct contact with patient.                                                                    | Yes - Materials are known to be biocompatible and do not<br>come into direct contact with patient.  | Same                                      |
| Patient Contact                                                                | No                                                                                                                                                                    | No                                                                                                  | Same                                      |
| Blood/Fluid Contact                                                            | None                                                                                                                                                                  | None                                                                                                | Same                                      |
| - PACKAGING INTEGRITY -                                                        |                                                                                                                                                                       |                                                                                                     |                                           |
| Packaging                                                                      | Non-Sterile Plastic Wrap                                                                                                                                              | Non-Sterile - Specific packaging not stated in predicate's<br>510(k) summary.                       | Same                                      |
| Labeling                                                                       | Labeled Plastic Wrap Containing Device and Cleaning and<br>Sterilization IFU                                                                                          | Labeled Polybag Containing Device and Cleaning and<br>Sterilization IFU                             | Similar                                   |
| Distribution Process                                                           | Shipped Non-Sterile                                                                                                                                                   | Shipped Non-Sterile                                                                                 | Same                                      |

{15}------------------------------------------------

#### F. Non-clinical Performance Testing

Testing was conducting in accordance with the following standards:

- ISO 10993-1:2009 (Biological evaluation of medical devices Part 1: Evaluation . and testing within a risk management process)
- . ISO 7405:2008 (Dentistry - Evaluation of biocompatibility of medical devices used in dentistry)
- . ISO 17664:2004 (Sterilization of medical devices - Information to be provided by the manufacturer for the processing of sterilizable medical devices)
- . ANSI/AAMI/ISO 17665-1:2006 (Sterilization of health care products - Moist Heat - Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices)
- . ISO14937:2000 (Sterilization of health care products – General requirements for characterization of a sterilizing agent and the development, validation and routine control of a sterilization process for medical devices)
- . AAMI TIR12:2010 (Designing, testing, and labeling for reprocessing in health care facilities: A guide for medical device manufacturers)
- . AAMI TIR30:2011 (A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable medical devices)
- . AAMI ST77, Rev. 2013 (Containment devices for reusable medical device sterilization)
- . ANSI/AAMI ST79:2010 (Comprehensive guide to steam sterilization and sterility assurance in health care facilities)

#### H. Conclusion

Based on the intended use, indications for use, technological characteristics, performance data and comparison to the predicate device, the subject Steri-Cassette and Steri-Cage product line has been shown to be substantially equivalent to the legally marketed predicate device PolyVac Surgical Instrument Delivery System's Instrument Cassettes (K012105).

{16}------------------------------------------------

{17}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/device/K162063](https://fda.innolitics.com/device/K162063)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
