The EZstim*III is a battery powered Peripheral Nerve Stimulator/Nerve Locator with two (2) indications for use: (1) on the high (0.05 - 80 mA) output current range, the device is used to monitor the effects of skeletal muscle relaxants during general anesthesia. (2) on the low (0.05-5.0 mA) output current range, it is used as a Nerve Locator to help locate the tip of a hypodermic needle near the nerve to which local anesthesia is to be delivered in Regional Nerve Block procedures.
Device Story
Battery-powered, hand-held peripheral nerve stimulator/locator; delivers constant current electrical pulses to patient via lead cables. Output range (high 0.05-80mA or low 0.05-5.0mA) automatically determined by connected lead cable type. High range supports monitoring neuromuscular junction (twitch, tetanus, double-burst, train-of-four modes); low range assists needle placement for regional anesthesia. Operated by clinicians in clinical settings. Device provides visual feedback via LCD. Output current amplitude and pulse patterns controlled by user to facilitate nerve stimulation or localization. Benefits include improved accuracy in nerve block needle placement and objective monitoring of muscle relaxation levels during anesthesia.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including electrical safety, EMC, software verification/validation, hardware functional testing, packaging distribution simulation (ASTM D4169-14), electrosurgery immunity, and battery life testing.
Technological Characteristics
Hand-held, battery-powered (9V) constant current stimulator. High range: 0.05-80mA; Low range: 0.05-5.0mA. High contrast LCD. Connectivity via specific patient lead cables (NSL-5 or RBW-5U). Safety/EMC standards: IEC 60601-1 (3.1 ed), IEC 60601-1-2 (4th ed). Software class: IEC 62304 Class B (Moderate Level of Concern). Non-sterile.
Indications for Use
Indicated for monitoring skeletal muscle relaxant effects during general anesthesia (high current range) and locating hypodermic needle tips near nerves for regional nerve block procedures (low current range).
Regulatory Classification
Identification
An electrical peripheral nerve stimulator (neuromuscular blockade monitor) is a device used to apply an electrical current to a patient to test the level of pharmacological effect of anesthetic drugs and gases.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 5, 2016
Halyard Health - Irvine Maria Pronina Technical Leader, Regulatory Affairs 43 Discovery, Suite 100 Irvine, California 92618
Re: K162048
Trade/Device Name: EZstim* III Peripheral Nerve Stimulator/Nerve Locator Regulation Number: 21 CFR 868.2775 Regulation Name: Electrical Peripheral Nerve Stimulator Regulatory Class: Class II Product Code: KOI Dated: November 1, 2016 Received: November 2, 2016
Dear Maria Pronina:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Tejashri Purohit-Sheth, M.D. Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162048
Device Name
EZstim*III Peripheral Nerve Stimulator/Nerve Locator
Indications for Use (Describe)
The EZstim*III is a battery powered Peripheral Nerve Stimulator/Nerve Locator with two (2) indications for use:
(1) on the high (0.05 - 80 mA) output current range, the device is used to monitor the effects of skeletal muscle relaxants during general anesthesia.
(2) on the low (0.05-5.0 mA) output current range, it is used as a Nerve Locator to help locate the tip of a hypodermic needle near the nerve to which local anesthesia is to be delivered in Regional Nerve Block procedures.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Halyard logo. The logo consists of a blue circle with two stylized sails inside, one white and one green. To the right of the circle is the word "HALYARD" in a bold, sans-serif font, also in blue.
43 Discovery, Suite 100 Irvine, CA 92618 USA
Tele: (800) 448-3569 (949) 923-9400 (949) 923-2401 Fax:
Visit us on the web at:
www.halyardhealth.com www.myON-Q.com
## 510(k) Summary
This summary of 510(k) information is being submitted in accordance with the requirements of 21 CFR 807.92
The assigned 510(k) number is:
| Submitted by: | Halyard Health<br>43 Discovery, Suite 100<br>Irvine, CA 92618 | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment Registration<br>Number: | 2026095 | |
| Contact Person: | Maria Pronina<br>Technical Leader, Regulatory Affairs<br>Phone: 949.923.2325<br>Fax: 678.389.9791<br>e-mail: maria.pronina@hyh.com | |
| Date Summary Prepared: | December 5, 2016 | |
| Reason for Premarket<br>Notification: | New Device | |
| Trade Name: | EZstim* III Peripheral Nerve Stimulator/Nerve Locator | |
| Common/Usual Name: | Stimulator, Nerve, Peripheral, Electric | |
| Regulation Name: | Electrical peripheral nerve stimulator | |
| Classification Regulation: | 21 CFR 868.2775 | |
| Product Code: | KOI | |
| Regulatory Class: | Class II | |
| Panel: | 73 - Anesthesiology | |
| Predicate Device: | Life-Tech, Inc. received clearance for the Electrical Peripheral<br>Nerve Stimulator, EZstim ES100 (K954505). Ownership of<br>K954505 was transferred to Halyard Health (formerly Kimberly-<br>Clark Healthcare) in 2013. | |
| Characteristic | Predicate Device | Subject Device |
| | EZstim (ES100)<br>(K954505) | EZstim* III (ES500)<br>(K162048) |
| Characteristic | Predicate Device<br>EZstim (ES100)<br>(K954505) | Subject Device<br>EZstim* III (ES500)<br>(K162048) |
| Regulation<br>Product Code | KOI | SAME |
| Manufacturer | Currently Halyard Health | SAME |
| Indications for Use | The EZstim has two indications for use:<br>(1) On the high (0-80mA) current range<br>the device is used to monitor the effects<br>of skeletal muscle relaxants during<br>general anesthesia. (2) On the low (0-<br>8mA) it is used as a nerve locator in<br>Regional Nerve Block procedures. | The EZstim* III has two (2) indications for<br>use: (1) On the high (0.05 – 80mA) output<br>current range, the device is used to monitor<br>the effects of skeletal muscle relaxants during<br>general anesthesia. (2) On the low (0.05 -<br>5.0mA) output current range, it is used as a<br>nerve Locator to help locate the tip of a<br>hypodermic needle near the nerve to which<br>local anesthesia is to be delivered in Regional<br>Nerve Block procedures. |
| Physical<br>Configuration | Hand-Held | SAME |
| Size | 3.7" (9.4cm) W x 6.2" (15.8cm) L x<br>1.4" (3.6cm) H | 3.5" (8.5 cm) W x 6.5" (16.3 cm) L x 2"<br>(5 cm) H (including knob) |
| Weight | 11.5 oz (326 g) | 9.6 oz (275gm) |
| Output Current | HIGH Range: 0 to 80 mA ± 2 mA.<br>LOW Range: 0.0 to 8.0 mA ± 0.2mA. | High Range: adjustable high current mode<br>regulated output 0.05 to 80 mA ± 5% into<br>3.74 k ohms or less<br>Low Range: adjustable current regulated 0.05<br>to 5.0 mA ± 1% into 11.5 k ohms or less |
| Stimulation<br>Categories | Double Burst, Train-of-Four, Twitch, or<br>Tetanus | SAME |
| Display | Single-line alphanumeric LCD with 16<br>characters | High contrast LCD approximately 1.0" high<br>x 2.45" wide |
| Mounting<br>Bracket | Mounting Bracket to an IV pole is<br>available as an optional accessory. | SAME |
| Expected<br>Service Life | 5 Years | SAME |
| Performance<br>Testing Data | • EMC Compatibility per "FDA<br>Reviewers Guidance for Pre-<br>market Notification Submission for<br>the anesthesiology and respiratory<br>device branch." (Based on IEC<br>801.)<br>• Software Verification and<br>Validation Testing per FDA<br>Guidance for software with Minor<br>Level of Concern<br>• Hardware mechanical and<br>functional Testing<br>• Packaging - None | • IEC 60601-1 (3.1 edition) standard for<br>safety and the IEC 60601-1-2 (4th<br>edition) standard for EMC.<br>• Software Verification and Validation<br>Testing per FDA Guidance for software<br>with Moderate Level of Concern (IEC<br>62304 Class B).<br>• Hardware mechanical and functional<br>Testing<br>• Packaging verification simulation<br>testing |
| Biocompatibility | Not Applicable (device is non-<br>patient contacting) | SAME |
| Power Source | One 9 Volt Alkaline Battery | SAME |
| Sterilization | Non-Sterile | SAME |
| Test Name and Description | Pass/Fail | |
| IEC 60601 Safety Testing Electrostatic Discharge Radiated<br>Immunity, Conducted RD Immunity, Magnetic Field Immunity<br>Description:<br>Electrical Safety and Electromagnetic Compatibility (EMC) were<br>tested and found to comply with IEC 60601-1 (3.1 Edition)<br>standard and the IEC 60601-1.2 (4th Edition) standard for EMC:<br>o Radiated Emissions<br>o Electrostatic Discharge Immunity<br>o Radiated RF Electromagnetic Fields Immunity<br>o Immunity to proximity field from RF wireless<br>communication equipment<br>o Conducted Disturbances, Induced by RF Fields Immunity<br>o Power Frequency Magnetic Fields Immunity<br>o Leakage Current<br>o Dielectric Voltage Withstand<br>o Excessive Temperature<br>o Humidity Preconditioning<br>o Abnormal Operation and Fault Conditions | Pass | |
| Software Verification | Pass | |
| Description: | | |
| The device's software has been validated in accordance with the<br>requirements set forth in the FDA Guidance for the Content of<br>Premarket Submissions for Software Contained in Medical<br>Devices (May 11, 2005). The software validation tests<br>demonstrated that the software version meets the design<br>requirements.<br>Hardware Verification | | |
| Description:<br>Hardware Verification test was conducted to ensure that the<br>ES500 Hardware and Software system functions as intended and<br>included the following:<br>o Output current test<br>o Beeper test<br>o Battery removal test<br>o Weight test<br>o Internal temperature test<br>The Hardware Verification test demonstrates that the design<br>outputs meet the design input<br>Packaging Verification | Pass | |
| Description:<br>Packaging distribution simulation testing was conducted per<br>ASTM D4169-14 (Standard Practice for Performance Testing of<br>Shipping Containers and Systems). The ES500 went through the<br>following distribution simulation:<br>o Temperature and Humidity Conditioning<br>o Schedule A - Manual Handling (First Drop Sequence)<br>o Schedule C - Vehicle Stacking<br>o Schedule F - Loose Load Vibration<br>o Schedule E - Vehicle Vibration<br>o Schedule J - Concentrated Impact<br>o Schedule A – Manual Handling (Second Drop Sequence)<br>Electrosurgery Test | Pass | |
| Description:<br>Electrosurgery Immunity testing was successfully completed to<br>ensure the ES500 device is electrosurgery compatible, where the<br>device operates as intended in High Output Current mode. The<br>electrosurgery immunity test demonstrated that the design output<br>meets the design input requirements.<br>Battery Life Test | Pass | |
| Description: | Pass | |
| Battery Life Test was conducted to ensure the duration of single<br>9V battery to function in accordance with the established<br>acceptance criteria for the ES500 device. The battery life tests<br>demonstrated that the design output meets the design input<br>requirement. | | |
| Battery Door Seal Integrity (Adhesive) Test | | |
| Description:<br>The Battery Door Seal Integrity (Adhesive) Test was conducted<br>to ensure the adhesive gasket in the battery door compartment<br>maintains integrity after battery replacements. The battery<br>adhesive test demonstrated that the design output meets the<br>design input requirement. | Pass | |
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Halyard logo. The logo consists of a blue circle with two triangles inside, one white and one green. To the right of the circle is the word "HALYARD" in bold, blue letters.
#### 5.1 Description of the Device
The EZstim* III (Model ES500) is both a constant current Peripheral Nerve Stimulator (HIGH output current range) and a Peripheral Nerve Locator (LOW output current range). Output current range is determined by the model of patient lead cable connected to the unit.
- . When the Halyard NSL-5 patient lead cable is connected, the unit automatically sets to the HIGH output current range (0.05 to 80 mA), with no stimulus mode selected (null mode). When the user selects a stimulus mode (i.e., 1 or 2 Hz Twitch, 50 Hz Tetanus, 100 Hz Tetanus, Double-Burst, or Train-of-Four), the unit functions as a nerve stimulator for use in monitoring the effects of skeletal muscle relaxants on the neuromuscular junction.
- . When the Halyard RBW-5U patient lead cable is connected, the unit automatically sets to the LOW output current range (0.05 to 5.0 mA). In this range, the unit functions as a nerve locator for use in regional nerve block procedures.
#### Indications for Use 5.2
EZstim* III Peripheral Nerve Stimulator/Nerve Locator is a battery powered Peripheral Nerve Stimulator / Nerve Locator with two (2) indications for use:
- (1) on the high (0.05 80 mA) output current range, the device is used to monitor the effects of skeletal muscle relaxants during general anesthesia.
- (2) on the low (0.05 5.0 mA) output current range, it is used as a Nerve Locator to help locate the tip of a hypodermic needle near the nerve to which local anesthesia is to be delivered in Regional Nerve Block procedures.
#### 5.3 Summary of Substantial Equivalence:
The intended use, materials, performance, and technological principles of operation of the EZstim* III Peripheral Nerve Stimulator/Nerve Locator (Model ES500) are substantially equivalent to the predicate device, EZstim (Model ES100) cleared under K954505. Just as with the predicate, energy is delivered to the patient via lead cable accessories that connect to the ES500 output connectors. As with the predicate device, stimulus control functions on the ES500 allow control of the output current amplitude and electrical pulse pattern to deliver the…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.