← Product Code [ELL](/productcode/ELL) · K162043

# NobelProcera HT ML FCZ Bridge (K162043)

_Nobel Biocare AB · ELL · Nov 22, 2016 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K162043

## Device Facts

- **Applicant:** Nobel Biocare AB
- **Product Code:** [ELL](/productcode/ELL.md)
- **Decision Date:** Nov 22, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3920
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

NobelProcera® HT ML FCZ Bridge is indicated for use as core structure of an artificial prosthesis for partially edentulous patients in the need of prosthetic oral reconstruction in order to restore chewing function. NobelProcera® HT ML FCZ Bridge is indicated for use as a bridge that will be cemented on natural teeth or artificial abutments.

## Device Story

Individualized cement-retained dental bridge; fabricated from translucent multi-layered yttria-stabilized tetragonal zirconia. Designed in dental laboratory/clinic using CAD software (NobelDesign or third-party); electronic order sent to centralized milling center for fabrication. Used by clinicians to replace natural teeth; suitable for cut-back (veneering) or stain and glaze techniques. Restores chewing function and esthetics for partially edentulous patients. Available in 6 shades; 2-14 units.

## Clinical Evidence

Bench testing only. No clinical data provided. Biocompatibility and performance data leveraged from predicate devices (K153534, K143439) due to identical material and manufacturing methods.

## Technological Characteristics

Material: Y-TZP Zirconium Oxide (Katana Zirconia). Manufacturing: CAD/CAM milling. Form factor: Full anatomic contour bridge (2-14 units) with optional cut-back. Minimum thickness: 0.6 mm. Connectivity: Electronic order transmission to centralized facility.

## Regulatory Identification

A porcelain tooth is a prefabricated device made of porcelain powder for clinical use (§ 872.6660) intended for use in construction of fixed or removable prostheses, such as crowns and partial dentures.

## Predicate Devices

- NobelProcera HT ML Full Contour Zirconia Crown ([K153534](/device/K153534.md))

## Reference Devices

- Procera Bridge Zirconia ([K071182](/device/K071182.md))
- Katana Zirconia ([K143439](/device/K143439.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, representing medicine and healing. The symbol is enclosed within a circular border. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 22, 2016

Nobel Biocare AB c/o Charlemagne Chua Senior Regulatory Affairs Manager Nobel Biocare USA LLC 22715 Savi Ranch Parkway Yorba Linda, California 92887

Re: K162043

Trade/Device Name: NobelProcera® HT ML FCZ Bridge Regulation Number: 21 CFR 872.3920 Regulation Name: Porcelain Tooth Regulatory Class: Class II Product Code: ELL Dated: October 25, 2016 Received: October 26, 2016

Dear Charlemagne Chua:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Michael J. Ryan -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| DEPARTMENT OF HEALTH AND HUMAN SERVICES |
|-----------------------------------------|
| Food and Drug Administration            |
| <b>Indications for Use</b>              |

| Form Approved: OMB No. 0910-0120  |
|-----------------------------------|
| Expiration Date: January 31, 2017 |
| See PRA Statement below.          |

| 510(k) Number ( <i>if known</i> ) | K162043                                                                                                                                                                                                                 |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name                       | Nobel Procera® HT ML FCZ Bridge                                                                                                                                                                                         |
| Indications for Use (Describe)    | NobelProcera® HT ML FCZ Bridge is indicated for use as core structure of an artificial prosthesis for partially edentulous patients in the need of prosthetic oral reconstruction in order to restore chewing function. |
|                                   | NobelProcera® HT ML FCZ Bridge is indicated for use as a bridge that will be cemented on natural teeth or artificial abutments.                                                                                         |

| Type of Use ( <i>Select one or both, as applicable</i> )                                           |                                                                                    |
|----------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <label><input checked="true" type="checkbox"/>Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/>Over-The-Counter Use (21 CFR 801 Subpart C)</label> |

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

| Department of Health and Human Services |
|-----------------------------------------|
| Food and Drug Administration            |
| Office of Chief Information Officer     |
| Paperwork Reduction Act (PRA) Staff     |
| PRAStaff@fda.hhs.gov                    |

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

| FORM FDA 3881 (1/14) | Page 1 of 1 | PSC Publishing Services (301) 443-6740 | EF |
|----------------------|-------------|----------------------------------------|----|
|----------------------|-------------|----------------------------------------|----|

网站

9Bbi18 ΣΟΠ JMTH ธารวดา¶l9doM :(វ))0) ۲ ข้อ
2019 10:10 ครั้ง

{3}------------------------------------------------

# A.4. 510(k)

## Summary

I. SUBMITTER

Nobel Biocare AB Vastra Hamngatan 1 Goteborg, SE-411 17 Sweden

Submitted by: Nobel Biocare USA LLC 22715 Savi Ranch Parkway Yorba Linda, CA 92887

Contact Person: Charlemagne Chua, Senior Regulatory Affairs Manager Phone: (714) 282-4800 x 7830 Fax: (714) 998-9348

Date Prepared: October 25, 2016

II. DEVICE

Name of Device: NobelProcera HT ML FCZ Bridge Common or Usual Name: Porcelain tooth Classification Name: Porcelain tooth (21 CFR 872.3920) Regulatory Class: II Product Code: ELL

III. PREDICATE DEVICE

Primary predicate: Nobel Biocare - NobelProcera HT ML Full Contour Zirconia Crown (K153534)

Reference predicates: Nobel Biocare - Procera Bridge Zirconia (K071182) Kuraray Noritake Dental Inc. - Katana Zirconia (K143439)

## IV. DEVICEDESCRIPTION

NobelProcera®HT ML FCZ Bridge is an individualized cement retained dental restoration manufactured from translucent multi-layered zirconia material.

NobelProcera® HT ML FCZ Bridge is intended to be a replacement for a natural teeth. After finalizing the NobelProcera®HT ML FCZ Bridge in the dental laboratory, it is cemented or bonded on prepared teeth or artificial abutments by a clinician, to provide a natural tooth like appearance and to restore chewing functionality in the patient's mouth. To achieve esthetics and required value and chroma of the surrounding natural teeth the NobelProcera® HT ML FCZ Bridge is suitable for cut-back (veneering) or stain and glaze techniques.

The design of the NobelProcera® HT ML FCZ Bridge is determined in a dental laboratory, hospital or dental practice by scanning, designing and ordering the bridge using NobelDesign or supported third party CAD systems. The bridge, once ordered, is sent electronically to one of NobelProcera's centralized milling centers for fabrication.

{4}------------------------------------------------

The NobelProcera® HT ML FCZ Bridge is manufactured from a solid piece of yitria-stabilized tetragonal zirconia. It is available in 6 shades and for bridges between 2 and 14 units.

# V. INDICATIONS FOR USE

NobelProcera® HT ML FCZ Bridge is indicated for use as core structure of an artificial prosthesis for partially edentulous patients in the need of prosthetic oral reconstruction in order to restore chewing function.

NobelProcera® HT ML FCZ Bridge is indicated for use as a bridge that is cemented on natural teeth or artificial abutments.

{5}------------------------------------------------

# Comparison of Technological Characteristics VI.

| Characteristic                             |                         | Candidate                                                                                                                                                                                                                                                                                                                                                                               | Primary Predicate                                                                                                                                                                                                                                                                                                                                                                                                                      | Reference Predicate                                                                                                                                                                                                                                                                                                                                                          | Reference Predicate                                                                                                                                            |
|--------------------------------------------|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                            |                         | NobelProcera HT ML FCZ<br>Bridge                                                                                                                                                                                                                                                                                                                                                        | NobelProcera HT ML Full<br>Contour Zirconia Crown<br>(K153534)                                                                                                                                                                                                                                                                                                                                                                         | Procera Bridge Zirconia<br>Extended Units<br>(K071182)                                                                                                                                                                                                                                                                                                                       | Katana Zirconia<br>(K143439)                                                                                                                                   |
| Features                                   | Material                | Y-TZP Zirconium Oxide<br>(Katana Zirconia K143439)                                                                                                                                                                                                                                                                                                                                      | Y-TZP Zirconium Oxide<br>(Katana Zirconia K143439)                                                                                                                                                                                                                                                                                                                                                                                     | Y-TZP Zirconium Oxide                                                                                                                                                                                                                                                                                                                                                        | Y-TZP Zirconium Oxide                                                                                                                                          |
|                                            | Shape                   | Full anatomic contour with<br>partial cut-back option                                                                                                                                                                                                                                                                                                                                   | Full anatomic contour with<br>partial cut-back option                                                                                                                                                                                                                                                                                                                                                                                  | Core structure (framework)<br>for a bridge                                                                                                                                                                                                                                                                                                                                   | N/A                                                                                                                                                            |
|                                            | Design<br>Method        | CAD                                                                                                                                                                                                                                                                                                                                                                                     | CAD                                                                                                                                                                                                                                                                                                                                                                                                                                    | CAD                                                                                                                                                                                                                                                                                                                                                                          | N/A                                                                                                                                                            |
|                                            | Manufacturing<br>Method | Nobel Biocare in-house<br>CAM                                                                                                                                                                                                                                                                                                                                                           | Nobel Biocare in-house CAM                                                                                                                                                                                                                                                                                                                                                                                                             | Nobel Biocare in-house<br>CAM                                                                                                                                                                                                                                                                                                                                                | N/A                                                                                                                                                            |
|                                            | Number of<br>units      | 2-14 units                                                                                                                                                                                                                                                                                                                                                                              | Individual crown                                                                                                                                                                                                                                                                                                                                                                                                                       | 2-14 units                                                                                                                                                                                                                                                                                                                                                                   | N/A                                                                                                                                                            |
|                                            | Minimum<br>Thickness    | 0.6 mm                                                                                                                                                                                                                                                                                                                                                                                  | 0.4 mm (Anterior)<br>0.7 mm (Pre-molar and Molar)                                                                                                                                                                                                                                                                                                                                                                                      | 0.6 mm                                                                                                                                                                                                                                                                                                                                                                       | N/A                                                                                                                                                            |
| Intended<br>Use/Principles of<br>Operation |                         | The NobelProcera HT ML<br>FCZ Bridge is an<br>individualized cement<br>retained dental restoration.<br>It is intended to be a<br>replacement for natural<br>teeth.                                                                                                                                                                                                                      | The NobelProcera HT ML Full<br>Contour Zirconia Crown is an<br>individualized cement retained<br>dental restoration.<br>It is intended to be a<br>replacement for natural teeth.                                                                                                                                                                                                                                                       | The Procera Bridge Zirconia<br>is an individualized cement<br>retained dental restoration.<br>It is intended to be a<br>replacement for natural teeth                                                                                                                                                                                                                        | Katana Zirconia is used for<br>the fabrication of the all-<br>ceramic restorations<br>(frameworks, FCZ crowns,<br>FCZ bridges, inlays, onlays<br>and veneers.) |
| Indications for Use                        |                         | NobelProcera HT ML FCZ<br>Bridge is indicated for use<br>as core structure of an<br>artificial prosthesis for<br>partially edentulous<br>patients in the need of<br>prosthetic oral<br>reconstruction in order to<br>restore chewing function.<br>NobelProcera HT ML FCZ<br>Bridge is indicated for use<br>as a bridge that is<br>cemented on natural teeth<br>or artificial abutments. | NobelProcera HT ML Full<br>Contour Zirconia Crown is<br>indicated for use as core<br>structure of an artificial<br>prosthesis for partially<br>edentulous patients in the need<br>of prosthetic oral reconstruction<br>in order to restore chewing<br>function.<br>NobelProcera HT ML Full<br>Contour Zirconia Crown is<br>indicated for use as single<br>crown that will be cemented to<br>a natural or artificial tooth<br>abutment. | Nobel Biocare's Procera<br>Bridge Zirconia is indicated<br>for use as the core structure<br>of an artificial prosthesis for<br>partially edentulous patients<br>in the need of prosthetic oral<br>reconstruction in order to<br>restore chewing function.<br>The Procera Bridge Zirconia<br>may be cemented or bonded<br>to either natural or artificial<br>tooth abutments. | KATANA Zirconia is used<br>for the fabrication of the all-<br>ceramic restorations<br>(frameworks, FCZ crowns,<br>FCZ bridges, inlays, onlays<br>and veneers.) |

Nobel Biocare Traditional 510(k) Notification NobelProcera HT ML FCZ Bridge 88-20196

{6}------------------------------------------------

Analysis of Differences Between Subject Device and Predicates

The NobelProcera HT ML FCZ Bridge differs from the primary predicate NobelProcera HT ML Full Contour Zirconia Crown (K153534) in the number of units available. As a crown the primary predicate is intended only to be used on a single tooth. The subject device is a bridge intended for 2 to 14 unit restorations.

The subject device uses the same high translucent multi laver Katana material (reference predicate K143439) as the primary predicate K153534. This material is cleared for the fabrication of full contour bridges such as the subject NobelProcera HT ML FCZ Bridge. The number of units available for the subject bridges is the same as the reference predicate Procera Bridge Zirconia (K071182).

### Intended Use:

The NobelProcera HT ML FCZ Bridge and primary predicate NobelProcera HT ML Full Contour Zirconia Crown have the same intended use. They are individualized cement retained dental restorations intended to be a replacement for natural teeth.

#### Summary:

The changes detailed above do not raise different questions of safety or effectiveness compared to the predicates.

## VII. PERFO RMANCE DATA

### Summary of Non-Clinical Testing:

Since the subject device does not represent a new worst case, data from the predicate device was leveraged in the following aspect of the 510(k).

- patibility - Biocom
	- The subject device is manufactured entirely from the zirconia cleared under o K143439. It is manufactured using the same method as the predicate (K153534). The subject device has the same intended use, patient contact type, and duration as the predicate. Therefore, no additional testing was required.

#### - Perform ance Data

- The NobelProcera HT ML FCZ Bridge is made entirely from the Katana Zirconia o material cleared under K143439. Since this predicate material is intended to be used in the manufacture of bridge such as the subject device, no additional performance testing is necessary.
#### VIII. CONCLUSIONS

The NobelProcera HT ML FCZ Bridge was evaluated for substantial equivalence. Based on technological characteristics, the NobelProcera HT ML FCZ Bridge has been shown to be substantially equivalent to the NobelProcera HT ML Full Contour Zirconia Crown (K153534).

Nobel Biocare Traditional 510(k) Notification NobelProcera HT ML FCZ Bridge October 2016

---

**Source:** [https://fda.innolitics.com/device/K162043](https://fda.innolitics.com/device/K162043)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
