K162025 · Philips Medical Systems Nederlands B.V. · LLZ · Oct 18, 2016 · Radiology
Device Facts
Record ID
K162025
Device Name
IntelliSpace Portal Platform
Applicant
Philips Medical Systems Nederlands B.V.
Product Code
LLZ · Radiology
Decision Date
Oct 18, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
Philips IntelliSpace Portal Platform is a software medical device that allows multiple users to remotely access clinical applications from compatible computers on a network. The system allows networking, selection, processing and filming of multimodality DICOM images. This software is for use with off-the-shelf PC computer technology that meets defined minimum specifications. IntelliSpace Portal Platform is intended to be used by trained professionals, including but not limited to physicians and medical technicians. This medical device is not to be used for mammography. The device is not intended for diagnosis of lossy compressed images.
Device Story
IntelliSpace Portal Platform is a software-based PACS that enables remote access to clinical applications via networked, off-the-shelf PCs. It ingests multimodality DICOM images; processes, displays, and manages these images; and supports clinical workflows including patient directory, reporting, filming, and worklist management. Operated by physicians and medical technicians in clinical environments, the system provides tools for 2D/3D visualization (MPR, Volume Rendering, MIP, MiniP), fusion viewing, segmentation, and image manipulation (pan, zoom, rotate, measurement). It integrates with hospital information systems (HIS, RIS, EMR) and PACS. By centralizing image access and providing advanced visualization tools, the device assists clinicians in diagnostic review and decision-making. It supports speech recognition for reporting and allows for background processing and data pre-fetching to improve workflow efficiency.
Clinical Evidence
No clinical data was required or submitted. Substantial equivalence was established through non-clinical verification and validation testing, including system-level performance tests and risk analysis, demonstrating compliance with recognized consensus standards.
Technological Characteristics
Software-based PACS platform for off-the-shelf PC hardware. Supports DICOM-3 standard for multimodality image communication. Features include thin client-server architecture, VMware support, multi-monitor support, and integration with HIS/RIS/EMR. Implements image processing algorithms for 2D/3D visualization, segmentation, and fusion. Complies with ISO 14971 (risk management), IEC 62304 (software lifecycle), and NEMA-PS 3.1-3.20 (DICOM).
Indications for Use
Indicated for use by trained professionals, including physicians and medical technicians, for networking, selection, processing, and filming of multimodality DICOM images. Not for use in mammography. Not intended for diagnosis of lossy compressed images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or fabric.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
PHILIPS MEDICAL SYSTEMS NEDERLANDS B.V. C/O MARK JOB RESPONSIBLE THIRD PARTY OFFICIAL 1394 25TH STREET, NW BUFFALO MN 55313
Re: K162025
Trade/Device Name: IntelliSpace Portal Platform Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: October 7, 2016 Received: October 11, 2016
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
October 18, 2016
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D.'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162025
Device Name IntelliSpace Portal Platform
| Indications for Use (Describe) | |
|--------------------------------|--|
|--------------------------------|--|
Philips IntelliSpace Portal Platform is a software medical device that allows multiple users clinical applications from compatible computers on a network.
The system allows networking, selection, processing and filming of multimodality DICOM images.
This software is for use with off-the-shelf PC computer technology that meets defined minimum specifications .
Philips IntelliSpace Portal Platform is intended to be used by trained professionals, including but not limited to physicians and medical technicians.
This medical device is not to be used for mammography.
The device is not intended for diagnosis of lossy compressed images.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. The background is white.
# 510(k) Summary
#### IntelliSpace Portal Platform
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
| Date prepared: | September 18, 2016 |
|----------------|--------------------|
|----------------|--------------------|
#### I. Submitter's name and address
| Establishment name: | Philips Medical Systems Nederland B.V. |
|--------------------------------|--------------------------------------------------------------------------------------------------------------|
| Establishment address: | Veenpluis 4-6<br>5684 PC Best<br>The Netherlands |
| Establishment<br>registration: | 3003768277 |
| Contact person: | Ilana Ben Moshe<br>Regulatory Affairs Expert<br>Phone: +972 525233496<br>Email: Ilana. Ben-Moshe@philips.com |
#### II. Device information
| Trade name: | IntelliSpace Portal Platform |
|------------------------|------------------------------------------------------------------------------------|
| Common name: | Image and Information Management Platform |
| Classification name: | Picture Archiving and Communications System (21 CFR 892.2050,<br>Product Code LLZ) |
| Device classification: | Class II |
# III. Predicate device information
| Trade name: | AW Server |
|--------------------------|---------------------------------------------|
| Common name: | Picture Archiving and Communication System |
| Manufacturer: | GE Medical System, LLC |
| Classification name: | Picture Archiving and Communications System |
| Device classification: | Class II |
| 510(k) clearance number: | K081985 |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and fill most of the frame. The background is white, providing a strong contrast to the blue text.
| Reference device no 1 | |
|--------------------------|------------------------------------------------------------------------------------|
| Trade name: | Extended Brilliance Workplace a component of Brilliance iCT<br>(Brilliance Volume) |
| Common name: | Workstation |
| Manufacturer: | Philips Medical Systems (Cleveland), Inc. |
| Classification name: | System, X-Ray, Tomography, Computed (21 CFR 992.1750, Product<br>Code JAK) |
| Device classification: | Class II |
| 510(k) clearance number: | K060937 |
| Reference device no 2 | |
| Trade name: | MR Permeability Application |
| Common name: | Picture Archiving and Communication System |
| Manufacturer: | Philips Medical Systems Nederland B.V. |
| Classification name: | Picture Archiving and Communication System (21 CFR 892.2050,<br>Product Code LLZ) |
| Device classification: | Class II |
| 510(k) clearance number: | K130278 |
| Reference device no 3 | |
| Trade name: | ViewForum 2003 |
| Common name: | Picture Archiving and Communication System |
| Manufacturer: | Philips Medical Systems Nederland B.V. |
| Classification name: | Picture Archiving and Communication System (21 CFR 892.2050,<br>Product Code LLZ) |
| Device classification: | Class II |
| 510(k) clearance number: | K032096 |
| Reference device no 4 | |
| Trade name: | IntelliSpace PACS 4.x |
| Common name: | Picture Archiving and Communication System |
| Manufacturer: | Philips Healthcare Informatics, Inc. |
| Classification name: | Picture Archiving and Communication System (21 CFR 892.2050,<br>Product Code LLZ) |
| Device Classification | Class II |
| Clearance Number | K111804 |
| Trade name: | EBW NM 2.0 |
| Common name: | Picture Archiving and Communication System |
| Manufacturer: | Philips Medical Systems (Cleveland), Inc. |
| Classification name: | Picture Archiving and Communication System (21 CFR 892.2050,<br>Product Code LLZ) |
| Device classification: | Class II |
| 510(k) clearance number: | K111336 |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and aligned horizontally. A thin black line is present beneath the word, spanning its entire length.
# Reference device no 5
# IV. Device description
Philips IntelliSpace Portal Platform is a software medical device that allows multiple users to remotely access clinical applications from compatible computers on a network. The system allows networking, selection, processing and filming of multimodality DICOM images. This software is for use with offthe-shelf PC computer technology that meets defined minimum specifications.
The IntelliSpace Portal Platform communicates with imaging systems of different modalities using the DICOM-3 standard.
# V. Intended use
Philips IntelliSpace Portal Platform is a software medical device that allows multiple users to remotely access clinical applications from compatible computers on a network. The system allows networking, selection, processing and filming of multimodality DICOM images. This software is for use with offthe-shelf PC computer technology that meets defined minimum specifications.
IntelliSpace Portal Platform is intended to be used by trained professionals, including but not limited to physicians and medical technicians.
This medical device is not to be used for mammography.
The device is not intended for diagnosis of lossy compressed images.
# VI. Technological characteristics
Philips believes that IntelliSpace Portal Platform product is substantially equivalent to the predicate GE A W Server legally marketed product with additional/enhanced functionality deriving from the reference devices:
- (K032096) ViewForum2003
- (K060937) Brilliance iCT (Brilliance Volume) EBW Workstation component
- (K130278) MR Permeability including details on IntelliSpace Portal component the next generation of the EBW product,
- I (K111804) IntelliSpace PACS 4.4
- l (K111336) EBW NM 2.0
Below is the comparison table to summarize the similarities and differences of the subject, predicate, and reference devices.
See Conclusion statement at the end of this section for substantial equivalency assessment of the differences indicated in the table.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. A thin black line is visible beneath the word, possibly indicating the bottom edge of a logo or design element.
| # | Feature/ | IntelliSpace Portal | AW Server | Various (see details |
|----|------------------------------------------------------------------------------------|---------------------------------------------------|-----------------------------------------------|----------------------------------------------------------------------|
| | Capability | Platform | Predicate Device | below) |
| | | Subject Device | (K081985) | Reference Devices |
| | System Configuration | | | |
| 1 | Standard off the<br>Shelf Hardware<br>and Operating<br>System (OS)<br>requirements | Yes | Yes | Not supported |
| 2 | VMware<br>Support | Yes | Yes | Not supported |
| 3 | Thin client-<br>server<br>technology | Yes | Yes | Yes<br>(K032096) ViewForum |
| 4 | Dual monitor<br>support | 1024x768 and up to<br>3MP | 1024x768 and up to<br>3MP | Not supported |
| 5 | Client multi-<br>resolution<br>display support | 1024x768 and up to<br>3MP | 1024x768 and up to<br>3MP | Not supported |
| | | System Applications and Functionality | | |
| 6 | Multiple<br>concurrent user<br>support | Yes | Yes | Yes<br>(K032096) ViewForum |
| 7 | Patient<br>directory /<br>Report/ Film /<br>Worklist<br>application | Yes | Not supported | Yes<br>(K032096) ViewForum |
| 8 | Presentation<br>State Support | Yes | Yes | Yes<br>(K032096) ViewForum |
| 9 | Pre-processing<br>and background<br>processing | Yes | Yes | Yes<br>(K032096) ViewForum |
| 10 | CD/DVD<br>burning | Yes | Yes | Yes<br>(K032096) ViewForum |
| 11 | USB device<br>support | Yes | Not supported | Yes<br>(K111804) IntelliSpace<br>PACS 4.4 |
| # | Feature/<br>Capability | IntelliSpace<br>Portal Platform<br>Subject Device | AW Server<br>Predicate<br>Device<br>(K081985) | Various (see details<br>below)<br>Reference Devices |
| 12 | Speech Recognition<br>software<br>(PowerScribe) for<br>reporting | Yes | Unknown | Yes<br>(K032096) ViewForum |
| 13 | Bookmarks: creating,<br>saving, and loading | Yes | Yes | Not supported |
| 14 | Support of Lossy<br>Compression for<br>improved<br>performance | Yes | Yes | Yes<br>(K032096) ViewForum |
| 15 | Licensing | Yes | Unknown | Yes<br>(K130278) MR<br>Permeability including<br>IntelliSpace Portal |
| 16 | Support of Data Pre-<br>Fetch from remote<br>devices | Yes | Yes | Yes<br>(K032096) ViewForum |
| 17 | Management tool | Yes | Yes | Yes<br>(K032096) ViewForum |
| | Review Modes and Image Manipulation | | | |
| 18 | Support multi-<br>modality data | Yes | Yes | Yes<br>(K032096) ViewForum |
| 19 | 2D slice view, Thin<br>slab/MPR view,<br>3D Volume<br>Rendering/MIP view<br>/MiniP | Yes | Unknown | Yes<br>(K032096) ViewForum |
| 20 | Fusion viewing | Yes | Unknown | Yes<br>(K032096) ViewForum<br>(K111336) EBW NM 2.0 |
| 21 | MR Segmentation | Yes | Unknown | Yes<br>(K032096) ViewForum |
| 22 | Cine viewing | Yes | Yes | Yes<br>(K032096) ViewForum |
| 23 | Image<br>Manipulation:<br>scroll/pan/<br>zoom/rotate/<br>comparison/ merge | Yes | Yes | Yes<br>(K032096) ViewForum |
| # | Feature/ | IntelliSpace Portal | AW Server | Various (see details |
| | Capability | Platform | Predicate Device | below) |
| | | Subject Device | (K081985) | Reference Devices |
| 24 | Send images | Yes | Yes | Yes |
| | and views to | | | (K032096) ViewForum |
| | film application | | | |
| 25 | Send images | Yes | Yes | Yes |
| | and views to a | | | (K032096) ViewForum |
| | reporting<br>application | | | |
| 26 | Create image | Yes | Yes | Yes |
| | annotations | | | (K032096) ViewForum |
| 27 | Measurement | Yes | Yes | Yes |
| | and graphics: | | | (K032096) ViewForum |
| | ROI, angle, text | | | |
| | Communication and Interoperability | | | |
| 28 | Integration with | Yes | Yes | Yes |
| | institution's | | | (K032096) ViewForum |
| | HIS, RIS, EMR | | | |
| 29 | PACS | Yes | Yes | Not supported |
| | integration | | | |
| 30 | Data Streaming | Yes | Unknown | Not supported |
| | tool integration | | | |
| 31 | Multi-vendor | Yes | Yes | Not supported |
| | Application | | | |
| 32 | plug-in<br>Enterprise | Yes | Yes | Not supported |
| | deployment | | | |
| | (scalability | | | |
| | feature) | | | |
| 33 | Multimodality | Yes | Not supported | Not supported |
| | Web Viewer | | | |
| | (with any | | | |
| | computer) | | | |
| 34 | Multimodality | Yes | Not supported | Not supported |
| | Web Viewer -<br>Mobile non- | | | |
| | diagnostic | | | |
| | viewing | | | |
| | (Collaboration | | | |
| | Viewer) | | | |
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. The background is white.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and fill most of the frame. The background is white, providing a strong contrast to the blue text.
Listed above technological differences are considered low risk, providing further support to clinicians in visualization or administrative functions. The majority of these functionalities are derived from reference devices and were verified and validated, and do not raise new questions on safety and/or effectiveness.
These features have not changed the intended use and operational principles of the device.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, which makes the blue letters stand out. The word is horizontally oriented and centered.
Therefore, the IntelliSpace Portal Platform is substantially equivalent to the currently marketed predicate device GE AW Server (K081985) in terms of technological characteristics.
#### Conclusion:
The IntelliSpace Portal Platform is substantially equivalent to the currently marketed predicate device, AW Server (K081985) in terms of indication for use, design features, fundamental scientific technology, and safety and/or effectiveness.
#### VII. Brief discussion of the nonclinical tests submitted, referenced or relied on
No performance standards for PACS systems or components have been issued under the authority of Section 514. Non-clinical performance testing has been performed on ISPP and demonstrates compliance with the following International and FDA-recognized consensus standards and FDA guidance document:
- ı ISO 14971 Medical devices – Application of risk management to medical devices
- IEC 62304 Medical device software - Software life cycle processes
- . NEMA-PS 3.1- PS 3.20 Digital Imaging and Communications in Medicine (DICOM)
- . Guidance for Industry and FDA Staff - Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
The ISPP system was tested in accordance with Philips verification processes. Verification and Validation tests have been performed to address intended use, the technological characteristics claims, requirement specifications and the risk management results.
The test results in this 510(k) premarket notification demonstrate that ISPP:
- complies with the aforementioned international and FDA-recognized consensus standards and l FDA guidance document, and
- . Meets the acceptance criteria and is adequate for its intended use.
Therefore, ISPP is substantially equivalent to the currently marketed predicate device GE AW Server (K081985) in terms of safety and effectiveness.
# VIII, Brief discussion of clinical tests submitted referenced or relied on
The subject of this premarket submission, ISPP does not require clinical studies to support equivalence.
# IX. The conclusions drawn from the nonclinical and clinical tests
Verification and Validation activities required to establish performance and functionality of IntelliSpace Portal Platform were performed. Testing involved system level tests, performance tests, and safety testing from Risk Analysis. Testing performed demonstrated the IntelliSpace Portal Platform meets all defined functionality requirements and performance claims.
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, providing a strong contrast that makes the word stand out. The overall impression is clean and modern.
| X. Overall conclusion: | The IntelliSpace Portal Platform is substantially equivalent to the predicate device AW Server (K081985) in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness.<br>Additionally, substantial equivalence was demonstrated with non-clinical performance testing, which complied with the requirements specified in the international and FDA-recognized consensus standards. The non-clinical performance tests provided in this 510(k) premarket notification demonstrate that the subject IntelliSpace Portal Platform is as safe and effective as its predicate device without raising any new safety and/or effectiveness concerns. |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.