K162023 · Getemed Medizin- Und Informationstechnik AG · DQK · Mar 24, 2017 · Cardiovascular
Device Facts
Record ID
K162023
Device Name
CardioDay V2.5
Applicant
Getemed Medizin- Und Informationstechnik AG
Product Code
DQK · Cardiovascular
Decision Date
Mar 24, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1425
Device Class
Class 2
Attributes
Software as a Medical Device, Pediatric
Indications for Use
The CardioDay V2.5 Holter Analysis Software is designed for the acquisition, analysis, edit, review, report, and storage of ambulatory and multi-parameter ECG data. Results of the automated analysis are intended to assist the physician in the interpretation of the recorded data. This information is not intended to serve as a substitute for the physician overread of the recorded ECG data. The CardioDay V2.5 system is intended to be used by trained operators under the direct supervision of a licensed healthcare practitioner in a hospital or clinic environment. Patient population includes both adult and pediatric human patients. CardioDay V2.5 provides the user arrhythmia study and Holter analysis capabilities. Data acquired may be used for the following indications: Evaluation of symptoms that may be caused by cardiac arrhythmia and/or conduction disturbances, Evaluation of symptoms that may be due to myocardial ischemia, Detection of ECG events that alter prognosis in certain forms of heart disease, Detection and analysis of pacemaker function and failure, Determination of cardiac response to lifestyle, Evaluation of therapeutic interventions, Investigations in epidemiology and clinical trials. Prescription Device Statement: United States federal law restricts CardioDay to sale by or on the order of a physician.
Device Story
CardioDay V2.5 is a software package for Holter ECG analysis; processes continuous ECG data acquired from compatible recorders (e.g., CardioMem, SEER series). Software displays ECG morphologies, heart rate trends, RR variability, and statistical graphs; allows physician review, editing, and report generation. Used in hospitals or clinics by trained operators under physician supervision. Does not perform automated diagnosis; assists physician interpretation. Supports data import/export via MUSE and Carescape Gateway. Benefits include improved diagnostic workflow for cardiac rhythm and conduction disorders.
Clinical Evidence
No clinical testing was performed. Substantial equivalence is supported by non-clinical bench testing, including ECG performance testing per AAMI/ANSI/IEC 60601-2-47, software lifecycle validation (IEC 62304), risk management (ISO 14971), usability engineering (IEC 62366-1), and cybersecurity/interoperability testing.
Technological Characteristics
Software-based Holter analysis system. Operates on standard PC hardware. Connectivity includes Bluetooth for recorder hook-up, MUSE import/export, and Carescape Gateway integration. Complies with AAMI/ANSI/IEC 60601-2-47 for ECG performance. Implements single sign-on and error logging. No hardware components included in this submission.
Indications for Use
Indicated for adult and pediatric patients requiring ambulatory or multi-parameter ECG analysis for evaluation of cardiac arrhythmias, conduction disturbances, myocardial ischemia, pacemaker function/failure, cardiac response to lifestyle, therapeutic interventions, and clinical research.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of an eagle with three heads, representing the department's commitment to health, human services, and science. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle image.
March 24, 2017
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
GETEMED Medizin- und Informationstechnik AG Bert Schadow Regulatory Affairs Manager Oderstrasse 77 Teltow, 14513 DE
Re: K162023
Trade/Device Name: CardioDay V2.5 Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DQK Dated: July 19, 2016 Received: July 22, 2016
Dear Bert Schadow:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Mude Jellman
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
### K162023
Device Name CardioDay V2.5
Indications for Use (Describe)
The CardioDay V2.5 Holter Analysis Software is designed for the acquisition, analysis, edit, review, report, and storage of ambulatory and multi-parameter ECG data.
Results of the automated analysis are intended to assist the physician in the interpretation of the recorded data. This information is not intended to serve as a substitute for the physician overread of the recorded ECG data.
The CardioDay V2.5 system is intended to be used by trained operators under the direct supervision of a licensed healthcare practitioner in a hospital or clinic environment.
Patient population includes both adult and pediatric human patients.
CardioDay V2.5 provides the user arrhythmia study and Holter analysis capabilities.
Data acquired may be used for the following indications:
- · Evaluation of symptoms that may be caused by cardiac arrhythmia and/or conduction disturbances,
- · Evaluation of symptoms that may be due to myocardial ischemia,
- · Detection of ECG events that alter prognosis in certain forms of heart disease.
- · Detection and analysis of pacemaker function and failure,
- · Determination of cardiac response to lifestyle,
- · Evaluation of therapeutic interventions,
- · Investigations in epidemiology and clinical trials.
Prescription Device Statement: United States federal law restricts CardioDay to sale by or on the order of a physician.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| DHF - Technical Documentation | @getemed |
|--------------------------------------------------|-----------------|
| CardioDay V2.5 | 1450-TD-0033-00 |
| 510(K) SUMMARY - Assigned 510(k) number: K162023 | Rev 04 |
#### 1 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is: K162023
| Submitted By: | GETEMED Medizin- und Informationstechnik AG<br>Oderstrasse 77, 14513 Teltow, Germany<br>Tel: + 49 - 3328 3942 0 |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Dr. Bert Schadow<br>Regulatory Affairs Manager<br>GETEMED Medizin- und Informationstechnik AG<br>Oderstrasse 77, 14513 Teltow, GERMANY<br>Tel: + 49 – 3328 3942 70<br>Fax: + 49 - 3328 3942 99 |
| Manufacturing Facility: | GETEMED Medizin- und Informationstechnik AG<br>Oderstrasse 77, 14513 Teltow, GERMANY |
| Date of Preparation: | March 22, 2017 |
| Name of the Device: | CardioDay® V2.5 |
| Common or Usual Name: | Holter evaluation software |
| Classification Name: | Computer, Diagnostic, Programmable |
| Product Classification: | 21 CFR 870.1425, Class II |
| Product Code: | DQK |
# Predicate Device Information:
| Device | Manufacturer | 510(k) Number |
|--------------------------------------------------------------------|------------------------------------------------|---------------|
| Primary Predicate Device:<br>CardioDay® version 2.4 | GETEMED Medizin- und<br>Informationstechnik AG | K130516 |
| Secondary Predicate Device:<br>MARS Holter Analysis<br>Workstation | GE Medical Systems<br>Information Technologies | K132437 |
#### 2 Reason for Submission
Premarket notification for CardioDay® V2.5, a New Device, seeking authority to market the device under Section 510(k) as a device that is substantially equivalent to the evaluation software CardioDay version 2.4 (K130516, GETEMED Medizin- und Informationstechnik AG) and MARS (K132437, GE Medical Systems Information Technologies).
{4}------------------------------------------------
| DHF - Technical Documentation | @getemed |
|--------------------------------------------------|-----------------|
| CardioDay V2.5 | 1450-TD-0033-00 |
| 510(K) SUMMARY - Assigned 510(k) number: K162023 | Rev 04 |
#### 3 Device Description
CardioDay® is a software package that allows a trained physician or health care professional knowledgeable in Holter interpretation, after having performed a long-term continuous electrocardiographic (ECG) recording, to download and analyze the data, review it and produce printed reports. CardioDay® does not perform any diagnosis of data by itself but only displays ECG morphologies and associated, calculated graphs such as heart rate trends, RR variability, and other statistical values in graphical form. The physician will be able to review, edit, and print the data collected.
The following Interfacing Devices (not part of this 510(k)) are supported by CardioDay V2.5:
- CardioMem CM 3000 (identical to CardioDay V2.4) ●
- . CardioMem CM 3000Z (identical to CardioDay V2.4)
- CardioMem CM 3000-12BT (identical to CardioDay V2.4) ●
- CardioMem CM 4000(B) (identical to CardioDay V2.4) ●
- SEER 12 (identical to CardioDay V2.4) ●
- SEER Light (new in CardioDay V2.5)
- SEER Light Extend (new in CardioDay V2.5) ●
- SEER 1000 (24h, 48h and 7-day models) (identical to CardioDay V2.4) ●
- . Care Scape Gateway (new in CardioDay V2.5)
#### ব Intended Use
The CardioDav V2.5 Holter Analysis Software is designed for the acquisition, analysis, edit, review, report, and storage of ambulatory and multi-parameter ECG data.
Results of the automated analysis are intended to assist the physician in the interpretation of the recorded data. This information is not intended to serve as a substitute for the physician overread of the recorded ECG data.
The CardioDay V2.5 system is intended to be used by trained operators under the direct supervision of a licensed healthcare practitioner in a hospital or clinic environment. Patient population includes both adult and pediatric human patients.
CardioDay V2.5 provides the user arrhythmia study and Holter analysis capabilities. Data acquired may be used for the following indications:
• Evaluation of symptoms that may be caused by cardiac arrhythmia and/or conduction disturbances,
- · Evaluation of symptoms that may be due to myocardial ischemia.
- · Detection of ECG events that alter prognosis in certain forms of heart disease.
- · Detection and analysis of pacemaker function and failure,
- · Determination of cardiac response to lifestyle,
- · Evaluation of therapeutic interventions,
- · Investigations in epidemiology and clinical trials.
Prescription Device Statement:
United States federal law restricts CardioDay® to sale by or on the order of a physician.
{5}------------------------------------------------
| DHF – Technical Documentation | | getemed |
|--------------------------------------------------|-----------------|---------|
| CardioDay V2.5 | 1450-TD-0033-00 | |
| 510(K) SUMMARY - Assigned 510(k) number: K162023 | Rev 04 | |
#### 5 Comparison to Predicate Devices
The primary predicate device is CardioDay version 2.4 (K130516). There have been no changes implemented in the modifications to CardioDay® V2.5 that impact the fundamental technology. The indications for use from CardioDay® version 2.4 (K130516) was adapted to the indications for use from MARS (K132437) in order to clarify the indications without affecting the substance or meaning of the indications.
As all indications for use cover the same diagnostic scope, i.e. patients suffering from cardiac related illnesses or complaining of potentially having such illnesses, as well as patients wearing a pacemaker, the Holter evalution software CardioDay® V2.5 can be regarded as substantially equivalent to CardioDay® version 2.4 (K130516) from GETEMED Medizin- und Informationstechnik AG and MARS Holter Analysis Workstation (K132437) from GE Medical Systems Information Technologies.
The following incremental changes were also made:
- Graphical user interface -> OEM Version for GE Healthcare;
- MUSE Import/Export Option (order management implemented);
- ECG Monitor Import Option (via Carescape Gateway from GE Healthcare) and support of SEER Light Holter recorders;
- Software Licensing Concept with Single Sign-On;
- Bluetooth Hook-Up Process for SEER 1000 Holter recorders;
- Error Logging implemented
A detailed discussion of similarities and differences is provided as a table of comparison in this 510(k) submission.
### 6 Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows
Non-clinical testing that has been conducted includes:
- a. ECG performance testing according to AAMI / ANSI / IEC 60601-2-47:
- b. Software development life cycle according to IEC 62304;
- c. Risk management process according to AAMI / ANSI / ISO 14971 and IEC /TR80002-1;
- d. Usability engineering process according to AAMI / ANSI / IEC 62366-1 and IEC 60601-1-6;
- e. Security tests to fulfill FDA Guidance for cybersecurity;
- f. Interface tests to fulfill FDA Guidance for interoperability.
None of the testing demonstrated that CardioDay® V2.5 brought up any issues of safety or effectiveness.
#### 7 Discussion of Clinical Tests Performed
No clinical testing was performed in order to support safety or effectiveness.
#### 8 Conclusion
CardioDay® V2.5 is identical to its predicate device in intended use / indications for use and operating principle. Verification and validation of CardioDay V2.5 demonstrated that the incremental changes do not raise any new questions of safety and effectiveness to the subject device and the subject device is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.