Aerin Medical Stylus

K161994 · Aerin Medical, Inc. · GEI · Sep 22, 2016 · General, Plastic Surgery

Device Facts

Record IDK161994
Device NameAerin Medical Stylus
ApplicantAerin Medical, Inc.
Product CodeGEI · General, Plastic Surgery
Decision DateSep 22, 2016
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Aerin Medical Stylus is indicated for use in otorhinolaryngology (ENT) surgery for the coagulation of soft tissue in the nasal airway, to treat nasal airway obstruction by shrinking submucosal tissue.

Device Story

Handheld bipolar radiofrequency (RF) probe for ENT surgery; comprises handle, shaft, and treatment tip with bipolar electrode array and temperature sensor. Connects to external RF generator (ORA-50 S) via flexible cable. Clinician inserts tip into nostril; delivers low-power RF energy (3-5W) to target submucosal tissue. Temperature sensor monitors tissue heating (50-70°C) to ensure controlled coagulation. RF energy creates coagulative lesion; tissue fibroses and retracts during healing, reducing nasal airway obstruction. Used in clinical settings by ENT surgeons.

Clinical Evidence

No clinical data. Bench testing only, including biocompatibility (ISO 10993-1), sterilization validation (ISO 11137), electrical safety (IEC 60601-1, 60601-2-2), electromagnetic compatibility (IEC 60601-1-2), and packaging integrity (ASTM standards). Performance testing included cadaver study for usability, force load testing, and thermocouple accuracy/response time in water bath and tissue models.

Technological Characteristics

Bipolar RF probe; 460 kHz (± 5 kHz) operating frequency; 3-5W power output. Features integrated temperature sensor for 50-70°C control. Sterilized via radiation (ISO 11137). Electrical safety per IEC 60601-1/60601-2-2. Packaging tested per ASTM D4169, F1980, F2096, F88.

Indications for Use

Indicated for patients with chronic nasal airway obstruction requiring ENT surgery for coagulation of soft tissue in the nasal airway to shrink submucosal tissue.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three lines forming its body and wings. The eagle is positioned to the right of a circular emblem that contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 22, 2016 Scott Wolf, M.D. President and Chief Medical Officer Aerin Medical, Inc. 690 W. Fremont Avenue, Unit 3 Sunnyvale, CA 94087 Re: K161994 Trade/Device Name: Aerin Medical Stylus Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: August 19, 2016 Received: August 23, 2016 Dear Dr. Wolf: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Page 2 - Scott Wolf, M.D. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Eric A. Mann -S for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K161994 Device Name Aerin Medical Stylus Indications for Use (Describe) The Aerin Medical Stylus is indicated for use in otorhinolaryngology (ENT) surgery for the coagulation of soft tissue in the nasal airway, to treat nasal airway obstruction by shrinking submucosal tissue. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW * The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Aerin Medical. The word "Aerin" is in a large, blue, sans-serif font, with the word "MEDICAL" in a smaller, gray, sans-serif font below it. To the right of the word "Aerin" are several green circles that are arranged in a gradient pattern. This 510(k) Summary is being submitted in accordance with the requirements of 21CFR 807.92. ## General Information | Submitter Information | | |--------------------------------------|-----------------------------------------------------------------------------------------------------| | Company: | Aerin Medical, Inc. | | Submitter's Address: | 690 W. Fremont Avenue, Unit 3<br>Sunnyvale, CA 94087 | | Contact Person: | Scott Wolf, M.D.<br>President and Chief Medical Officer<br>Phone: 650-434-3247<br>Fax: 408-716-2438 | | Establishment<br>Registration Number | 3011625895 | | Date Prepared: | September 21, 2016 | | Name of the Device | | | Proprietary Name: | Aerin Medical Stylus | | Common Name: | Radiofrequency wand | | Classification Name: | Electrosurgical cutting and coagulation device and<br>accessories | | Classification Panel: | General and Plastic Surgery | | Device Class: | Class II | | Product Code: | GEI | | CFR Section: | 21 CFR 878.4400 | | Predicate Device: | Aerin Medical Wand (K150637) | ## Device Description The Aerin Medical Stylus is a handheld bipolar radiofrequency (RF) probe designed for use in otorhinolaryngology (ENT) surgery. The Stylus comprises a handle, shaft and treatment tip. The treatment tip consists of an array of bipolar electrodes and a temperature sensor that allows for monitoring of tissue temperature during radiofrequency energy delivery. The stylus is connected, via a flexible cable, to a commercially-available RF generator (the ORA-50 S, cleared under K993854) meeting the following requirements: RF operating frequency of 460 kHz (± 5 kHz); bipolar low power (3 watts to 5 watts) RF energy delivery; and sensitive temperature control with low overshoot (50°C to 70°C). {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Aerin Medical. The word "Aerin" is in a large, blue, sans-serif font. Below "Aerin" is the word "MEDICAL" in a smaller, sans-serif font. To the right of "Aerin" are several green circles of varying sizes, arranged in a pattern. The Aerin Medical Stylus is used in the treatment of patients experiencing chronic nasal airway obstruction. During a treatment procedure, the clinician inserts the tip of the Aerin Medical Stylus into a patient's nostril to deliver low power RF energy to the target tissue of the nasal airway. Radiofrequency treatment of tissue creates a coaqulative lesion which fibroses and retracts as it heals, thereby shrinking the tissue to lessen the degree of obstruction. #### Indications for Use The Aerin Medical Stylus is indicated for use in otorhinolaryngology (ENT) surgery for the coagulation of soft tissue in the nasal airway, to treat nasal airway obstruction by shrinking submucosal tissue. #### Summary of the technological characteristics of the device compared to the predicate device The Aerin Medical Stylus was found to be equivalent to the predicate device in design and intended use to deliver bipolar RF energy to coagulate tissue in otorhinolaryngology procedures. | Characteristic | Aerin Medical Stylus<br>(Models FG011 & FG174)<br>Subject Device | Aerin Medical Wand<br>(Model FG011)<br>Predicate Device<br>K150637 | |----------------------|------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for Use | Same | Indicated for use in<br>otorhinolaryngology (ENT)<br>surgery for the coagulation<br>of soft tissue in the nasal<br>airway, to treat nasal<br>airway obstruction by<br>shrinking submucosal<br>tissue. | | Design configuration | Same | Integrated cable, handle<br>and electrode | | Energy type | Same | Bipolar radiofrequency | | Tissue temperature | Same | 50 — 70 °C<br>(temperature controlled) | ## Summary of non-clinical tests Device safety is supported through biocompatibility testing, sterilization and packaging validation activities, electrical safety testing and electromagnetic Biocompatibility of the Aerin Medical Stylus was compatibility testing. demonstrated through testing performed in accordance with AAMI / ANSI / ISO standards 10993-1 (2009/(R) 2013). Radiation sterilization was validated per AAMI / ANSI / ISO 11137-1 (2006/(R) 2010) and 11137-2 (2013). Electrical safety was demonstrated through testing to IEC 60601-1 (2005/(R) 2012) and {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Aerin Medical. The word "Aerin" is in a large, blue, sans-serif font. Below "Aerin" is the word "MEDICAL" in a smaller, sans-serif font. To the right of "Aerin" are several green circles of varying sizes. 60601-2-2 (2009) and electromagnetic compatibility testing was performed according to IEC 60601-1-2 (2007). The packaging systems were subjected to transit testing, visual inspection, accelerated aging, gross leak detection and seal strength testing in accordance with the applicable ASTM standards (ASTM D4169-14, ASTM F1980-07, ASTM F2096-11, and ASTM F88/F88M-15), and met all acceptance criteria. Device performance testing included usability factors as well as factors related to system responsiveness and effectiveness for use in a clinical setting. Usability factors, such as ergonomics, visualization, orientation and force application, were evaluated as part of a cadaver study, Additionally, force load testing was conducted to verify adequate shaft strength. The efficacy of the subject device is supported by thermocouple accuracy and response time testing via water bath immersion and tissue heating time testing. Tissue heating time testing was performed using tissues representative of the submucosal tissues of the nasal airway. The subject device met all performance testing requirements, thereby supporting its equivalency to the predicate device in terms of delivery of RF energy. #### Summary of clinical tests Not applicable. No clinical tests were necessary. #### Conclusion Testing demonstrates that the Aerin Medical Stylus is substantially equivalent to the predicate in terms of both indications for use and delivered RF treatment and is as safe and effective for its intended use.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%