Steerable Guide Catheter

K161985 · Abbott Vascular · DRA · Aug 7, 2016 · Cardiovascular

Device Facts

Record IDK161985
Device NameSteerable Guide Catheter
ApplicantAbbott Vascular
Product CodeDRA · Cardiovascular
Decision DateAug 7, 2016
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1280
Device ClassClass 2

Indications for Use

The Steerable Guide Catheter is used for introducing various catheters into the left side of the heart through the interatrial septum.

Device Story

Steerable Guide Catheter facilitates introduction of cardiovascular catheters into left heart via interatrial septum. System comprises guide catheter and dilator. Guide features distal/proximal shafts, radiopaque tip ring, steering handle, hemostasis valve, luer lock flush port, and atraumatic distal tip. Central lumen allows air aspiration, fluid infusion, and passage of ancillary devices up to 0.204" diameter. Dilator includes radiopaque shaft, echogenic distal tip, hemostasis valve, and lumen for devices up to 0.035" diameter. Used in clinical settings under fluoroscopic visualization. Physician-operated. Provides conduit for interventional procedures; benefits patient by enabling access to left heart structures.

Clinical Evidence

Bench testing only. No clinical data presented. Device demonstrated to meet performance specifications equivalent to predicate devices.

Technological Characteristics

Single-use, EtO sterilized. Components: guide catheter (shafts, steering handle, hemostasis valve, radiopaque tip) and dilator (radiopaque shaft, echogenic tip). Materials include radiopaque markers for fluoroscopic visualization. Mechanical steering mechanism via handle knob. No software or electronic components.

Indications for Use

Indicated for introducing various cardiovascular catheters into the left side of the heart via the interatrial septum.

Regulatory Classification

Identification

A steerable catheter is a catheter used for diagnostic and monitoring purposes whose movements are directed by a steering control unit.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 07, 2016 Abbott Vascular Makena Mc Gowan Regulatory Affairs Associate 3200 Lakeside Drive Santa Clara, California 95054 Re: K161985 Trade/Device Name: Steerable Guide Catheter Regulation Number: 21 CFR 870.1280 Regulation Name: Steerable Catheter Regulatory Class: Class II Product Code: DRA Dated: July 18, 2016 Received: July 19, 2016 Dear Ms. Mc Gowan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, M.A. Hilleman for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K161985 Device Name Steerable Guide Catheter Indications for Use (Describe) The Steerable Guide Catheter is used for introducing various catheters into the left side of the heart through the interatrial septum. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary The 510(k) Summary is submitted in accordance with 21 CFR §807.1280 and the requirements of the Safe Medical Device Act (SMDA) of 1990. | 1. | Submitter's Name | Abbott Vascular | |-----|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 2. | Submitter's Address | 3200 Lakeside Drive Santa Clara, CA 95054 | | 3. | Telephone | (408) 845-0937 | | 4. | Fax | (408) 845-3734 | | 5. | Contact Person | Makena Mc Gowan | | 6. | Date Prepared | August 4, 2016 | | 7. | Device Trade Name | Steerable Guide Catheter | | 8. | Device Common Name | Steerable Catheter | | 9. | Device Classification Name | Catheter, Steerable | | 10. | Predicate Device Name | K083793 Steerable Guide Catheter<br>K091596 Steerable Guide Catheter<br>K093866 Steerable Guide Catheter<br>K100789 Steerable Guide Catheter<br>K112239 Steerable Guide Catheter | ## 11. Device Description The Steerable Guide Catheter consists of a Guide and a Dilator provided EtO sterile and for single use only. The Steerable Guide Catheter consists of a distal and proximal catheter shaft, a radiopaque tip ring, a handle with a steering knob, a hemostasis valve with a luer lock flush port, an atraumatic distal tip, and a Dilator with a single central lumen. The central lumen of the Guide allows for aspiration of air and infusion of fluids such as saline, and serves as a conduit during introduction and or exchange of the Dilator and ancillary devices (e.g. catheters) that have a maximum diameter of .204". The atraumatic distal tip of the Steerable Guide Catheter is radiopaque to allow visualization under fluoroscopy. The Dilator consists of a radiopaque shaft, an echogenic feature at the distal tip, a hemostasis valve with a flush port and an internal lumen designed to accept ancillary devices that have a maximum diameter of 0.035" (e.g. needles or guidewires). The Steerable Guide Catheter, Dilator and accessories are packaged in a tray enclosed in a sealed Tyvek pouch and boxed in a cardboard shelf-carton. {4}------------------------------------------------ # 12. Indication for Use The Steerable Guide Catheter is used for introducing various cardiovascular catheters into the left side of the heart through the interatrial septum. # 13. Technological Characteristics Comparison of the new device and predicate device demonstrate that the technological characteristics such as design, material, and indications for use are substantially equivalent to the current marketed predicate device. # 14. Performance Data The Steerable Guide Catheter is equivalent in fundamental scientific design to the predicate Steerable Guide Catheter, and therefore, meets the same performance specifications as the predicate device. # 15. Conclusions The Steerable Guide Catheter has the same indications for use and technological characteristics and is substantially equivalent to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...