Bindex BI-2

K161971 · Bone Index Finland , Ltd. · MUA · Jan 9, 2017 · Radiology

Device Facts

Record IDK161971
Device NameBindex BI-2
ApplicantBone Index Finland , Ltd.
Product CodeMUA · Radiology
Decision DateJan 9, 2017
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 892.1180
Device ClassClass 2

Indications for Use

Bindex measures apparent cortical bone thickness at the proximal tibia and can be used in conjunction with other clinical risk factors or patient characteristics as an aid to the physician in the diagnosis and other medical conditions leading to reduced bone strength and in the determination of fracture risk.

Device Story

Bindex BI-2 is a handheld ultrasound bone sonometer used in clinical settings by healthcare providers. The system consists of a handheld ultrasound transducer connected to a computer via USB. The operator applies gel to the patient's skin and manually places the transducer perpendicularly on the proximal tibia (1/3 length). The device measures apparent cortical bone thickness and calculates a Density Index (DI), which estimates hip bone density as measured by DXA. The output is displayed via computer software to assist physicians in diagnosing reduced bone strength and assessing fracture risk. The device requires manual transducer orientation and five repeated measurements per location. The transducer is disinfected between uses with isopropyl alcohol.

Clinical Evidence

No new clinical data provided. The device relies on the clinical and pre-clinical data previously established for the predicate Bindex BI-100. Performance testing included bench-level biocompatibility (cytotoxicity, sensitization, irritation per ISO 10993-1), electrical safety, EMC, and mechanical/acoustic testing (drop test, ball pressure test, moulding stress relief).

Technological Characteristics

Handheld ultrasound transducer; 3.0 MHz nominal center frequency. Powered via PC USB port. Operates in temperature range +10°C to +40°C. Complies with IEC 60601-1, IEC 60601-2-37, and IEC 60601-1-2. Biocompatibility per ISO 10993-1. Software-controlled measurement of apparent cortical bone thickness.

Indications for Use

Indicated for patients requiring assessment of cortical bone thickness at the proximal tibia to aid physicians in diagnosing conditions leading to reduced bone strength and determining fracture risk.

Regulatory Classification

Identification

A bone sonometer is a device that transmits ultrasound energy into the human body to measure acoustic properties of bone that indicate overall bone health and fracture risk. The primary components of the device are a voltage generator, a transmitting transducer, a receiving transducer, and hardware and software for reception and processing of the received ultrasonic signal.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Guidance for Industry and FDA Staff; Class II Special Controls Guidance Document: Bone Sonometers.” See § 892.1(e) for the availability of this guidance document.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines connecting them. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 9, 2017 Bone Index Finland, Ltd. % Janne Karjalainen Chief Technology Officer (CTO) Microkatu 1 Kuopio 70211 FINLAND Re: K161971 Trade/Device Name: Bindex BI-2 Regulation Number: 21 CFR 892.1180 Regulation Name: Bone sonometer Regulatory Class: II Product Code: MUA Dated: July 12, 2016 Received: December 15, 2016 Dear Janne Karjalainen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Michael D.'Hara For Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K161971 Device Name Bindex BI-2 Indications for Use (Describe) Bindex measures apparent cortical bone thickness at the proximal tibia and can be used in conjunction with other clinical risk factors or patient characteristics as an aid to the physician in the diagnosis and other medical conditions leading to reduced bone strength and in the determination of fracture risk. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------| | <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary #### SUBMITTER l. Bone Index Finland, Ltd. Microkatu 1 Kuopio, Finland 70211 Phone: +358 45 896 2650 Contact Person: Janne Karjalainen Date Prepared: 2 Dec, 2016 # II. DEVICE Name of Device: Bindex BI-2 Common or Usual Name: Bone sonometer Classification Name: Bone sonometer (21 CFR 892.1180) Regulatory Class: II Product Code: MUA ## III. PREDICATE DEVICE Bindex BI-100 bone sonometer, K152020. This predicate has not been subject to a design-related recall. {4}------------------------------------------------ # IV. DEVICE DESCRIPTION The Bindex Bl-2 system consists of handheld ultrasound transducer and software. Bindex Bl-2 is connected to the USB port of a computer and controlled with computer software. Bindex Bl-2 is used for measurement of cortical bone thickness and it provides Density Index (DI), a parameter which estimates bensity at the hip as measured with DXA. For measurements, gel is applied on skin and ultrasound transducer is manually placed on the measurement location. Standardized measurement location is at proximal tibia (1/3 length of tibia). Transducer is manually oriented perpendicularly to the cortical bone to achieve accepted measurement. Measurement is repeated five times at each measurement location. Finally, transducer is disinfected by wiping gel off with isopropyl alcohol moistened cloth. The associated accessories include: - Measurement stick - Ultrasound gel (optional) #### V. INDICATIONS FOR USE Bindex measures apparent cortical bone thickness at the proximal tibia and can be used in conjunction with other clinical risk factors or patient characteristics as an aid to the physician in the diagnosis and other medical conditions leading to reduced bone strength and in the determination of fracture risk. ### Indications for use discussion The Indications for Use statement for the Bindex Bl-2 is identical to the predicate device; thus raises no additional questions on safety and effectiveness. ### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE VI. DEVICE The predicate and subject devices are based similar ultrasound technology and are based on the same measurement principle. The main differences between the devices are structural in terms mechanics and electronics. Importantly, the transducer (responsible for conversion of electrical energy to kinetic mechanical energy), is the same for both devices. The predicate and subject devices are compared in table 1. Table 1. Device comparison table. | # | Item | Bindex BI-100 (Predicate) | Bindex BI-2 (subject) | |---|---------------------|------------------------------------------------------------------------|-----------------------| | 1 | Indications for use | Bindex measures apparent cortical bone thickness at the proximal tibia | Same | {5}------------------------------------------------ | | | and can be used in conjunction with<br>other clinical risk factors or patient<br>characteristics as an aid to the<br>physician in the diagnosis of<br>osteoporosis and other medical<br>conditions leading to reduced bone<br>strength and in the determination<br>of fracture risk. | | |---|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------| | 2 | Measurement mode | Apparent Cortical Thickness<br>(Ct.Th.). | Same | | 3 | Probe compatibility | One transducer, centre frequency =<br>3.0 MHz (nominal) | Same | | 4 | Electrical safety | IEC 60601-2-37:2001 including<br>Amendments 1 and 2; and IEC<br>60601-1:2005 | Same | | 5 | Electromagnetic<br>compatibility | Complies with IEC 60601-1-2:2007 | Same | | 6 | Power supply | PC USB port powered. | Same | | 7 | Operating<br>Environment | Temperature: +10°C to +40°C<br>Humidity: 5% to 85% RH, non-<br>condensing<br>Atmospheric Pressure: 600 to 1060<br>Pc | Same | | 8 | Storage Environment | Temperature: +10°C to +40°C<br>Humidity: 5% to 85% RH, non-<br>condensing | Same | | 9 | Calibration | Calibration before each use. | Same | # VII. PERFORMANCE DATA The following performance data were provided in support of the substantial equivalence determination. ## Biocompatibility testing The biocompatibility evaluation for the Bindex Bl-2 device was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,"May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices –Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing along the ISO 10993-1 would recommend the following tests: - Cytotoxicity - Sensitization - Irritation {6}------------------------------------------------ The Bindex Bl-2 is in skin contact for duration of less than 24 hours. Since the use of Bindex involves very short skin contact (typically less than 10 minutes) on a healthy skin and therefore poses a very low risk. Safety of the manufacturing and used materials has been further discussed in Biocompatibility report (Section 13). ## Electrical safety and electromagnetic compatibility (EMC) Electrical safety and EMC testing were conducted on the measurement system, consisting of the Bindex Bl-2 and a laptop computer (on battery use or connected to power supply with isolation transformer). The system complies with the IEC 60601-1, and IEC 60601-2-37 standards for safety and the IEC 60601-1-2 standard for EMC. ## Software Verification and Validation Testing Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern. ## Mechanical and acoustic testing The acoustic output and mechanical properties have been tested along the principles presented in harmonized standards IEC 62359 and IEC 60601-1: - . Acoustic Testing has been conducted on the predicate device and test documentation in this submission shows equivalence of the devices on behalf of acoustic output. - Drop test (hand piece dropped from 1m on hard surface). - . Ball pressure test (pressed with steel ball 5mm in diameter with force of 20N one hour at a temperature of 75±2°C) - Moulding stress relief (device placed in circulating air oven at 70 °C for 7 hours and let cool down) ## Clinical Studies The same clinical and pre-clinical data supports the safe and effective use of Bindex Bl-100, to which substantial equivalence is shown. The safety and effectiveness of Bindex Bl-2 device has been discussed in clinical evaluation document Section 16. #### CONCLUSIONS VIII. {7}------------------------------------------------ The related safety standards are the same for the predicate and subject device. The same clinical evidence applies to both devices. In the present submission, the performance testing comparing the two models shows that the subject device is substantially equivalent to the predicate device and does not raise additional questions on safety or effectiveness. Date: 12/02/2016 mm / dd / year signature Janne Karjalainen, CTO, Bone Index Finland Ltd.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...