K161970 · Vitrolife Sweden AB · MQE · Jun 14, 2017 · Obstetrics/Gynecology
Device Facts
Record ID
K161970
Device Name
Follicle Aspiration Set
Applicant
Vitrolife Sweden AB
Product Code
MQE · Obstetrics/Gynecology
Decision Date
Jun 14, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.6100
Device Class
Class 2
Indications for Use
Intended for flushing and/or aspiration of oocytes from ovarian follicles.
Device Story
Follicle Aspiration Set consists of stainless steel needles, aspiration/flushing tubing, and silicone corks. Used for ultrasound-guided transvaginal oocyte retrieval. Physician inserts needle into ovarian follicle; vacuum pump (connected via tubing/cork) aspirates oocytes; flushing capability allows follicle irrigation to remove blockages. Echogenic needle tips aid ultrasound visualization. Single-use, radiation-sterilized. Output is collected oocytes for laboratory processing. Benefits include efficient, minimally invasive gamete retrieval for assisted reproduction.
Clinical Evidence
Bench testing only. Studies included: Sterilization validation (ISO 11137-2), Endotoxin testing (USP <85>, ≤1.2 EU/device), 1-Cell Mouse Embryo Assay (>80% expanded blastocysts at 96h), vacuum pressure testing (-500 mmHg), flow rate testing, mechanical needle testing (ISO 9626), tensile strength, and biocompatibility (ISO 10993-5, -10). Accelerated shelf-life studies (3 years) confirmed continued compliance with MEA, seal strength (ASTM F88), dye penetration (ASTM F1929), and visual inspection (ASTM F1886) criteria.
Technological Characteristics
Materials: stainless steel, TPE-O, silicone, medical grade adhesive, polypropylene, methyl methacrylate acrylonitrile butadiene styrene. Sensing: echogenic needle tip for ultrasound. Energy: vacuum-driven aspiration. Connectivity: none. Sterilization: radiation. Software: none.
Indications for Use
Indicated for flushing and/or aspiration of oocytes from ovarian follicles in patients undergoing assisted reproduction procedures.
Regulatory Classification
Identification
Assisted reproduction needles are devices used in in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures to obtain gametes from the body or introduce gametes, zygote(s), preembryo(s) and/or embryo(s) into the body. This generic type of device may include a single or double lumen needle and component parts, including needle guides, such as those used with ultrasound.
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a series of human profiles, rendered in a dark color, possibly black or a dark shade of blue.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 14, 2017
Vitrolife Sweden AB % Greg Holland Sr. Partner Regulatory Specialists, Inc. 3722 Ave. Sausalito Irvine, CA 92606
Re: K161970 Trade/Device Name: Follicle Aspiration Set Regulation Number: 21 CFR§ 884.6100 Regulation Name: Assisted Reproduction Needles Regulatory Class: II Product Code: MOE Dated: April 28, 2017 Received: May 1, 2017
Dear Greg Holland:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Charles Viviano -S
For Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K161970
Device Name Follicle Aspiration Set
Indications for Use (Describe) Intended for flushing and/or aspiration of oocytes from ovarian follicles
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
# 510(k) Summary – K161970
| Submitter | Vitrolife Sweden AB Gustaf Werners gata 2 SE - 421 32 Västra Frölunda SWEDEN Nina Arvidsson TEL: (46) 31 721 80 00 FAX: (46) 31 721 80 99 |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Greg Holland Regulatory Consultant to Vitrolife Sweden AB Regulatory Specialists, Inc. 3722 Ave. Sausalito Irvine, CA 92606 TEL: 949.422.3853 FAX: 949.552.2821 EMAIL: greg@regulatoryspecialists.com |
| Summary Date | June 14, 2017 |
| Trade Name | Follicle Aspiration Set (See the Device Description section below for device versions included in this submission) |
| Common Name | Oocyte Retrieval Needle |
| Classification Name | Assisted Reproduction Needles |
| Regulation Number | 884.6100 |
| Class | II |
| Panel | Obstetrics/Gynecology |
| Product Code | MQE (Needle, Assisted Reproduction) |
| Predicate Device | K991273, Swemed Follicle Aspiration Set, Double Lumen Swemed Follicle Aspiration Set, Single Lumen Swemed Follicle Aspiration Luer Needle |
The predicate device has not been subjected to a recall.
{4}------------------------------------------------
### Device Description:
Follicle Aspiration Sets included in this submission are intended for ultrasound-guided transvaginal aspiration and flushing of oocytes from ovarian follicles. Four types of Follicle Aspiration Sets are included in this submission, and are described in more detail below.
The Follicle Aspiration Sets are radiation sterilized and are for single use only. The materials used in these devices include: stainless steel, TPE-O tubing, silicone, medical grade adhesive, polypropylene, and methyl methacrylate acrylonitrile butadiene styrene.
Follicle Aspiration Set, Single Lumen: This device consists of a single-lumen stainless steel needle, aspiration tubing, and a silicone rubber cork for attachment to a collection tube. One version (17132) also includes vacuum tubing attached to the silicone rubber cork for attachment to the vacuum pump used for aspiration. This device also includes an adapter for flushing of ovarian follicles and removal of blockages. The needles for the majority of the versions of this device type include echogenic markings at the tip of the needle to aid in visualization during aspiration procedures. The table below shows the specific features of the different versions of this device:
| REF | Needle | | | Tubing | | |
|--------|---------|---------|-------------|---------|---------|-------------|
| | OD [mm] | ID [mm] | Length [mm] | OD [mm] | ID [mm] | Length [mm] |
| 17103 | 1.4 | 1.0 | 300 | 2.22 | 1.33 | 400 |
| 17107 | 1.4 | 1.0 | 300 | 2.22 | 1.33 | 600 |
| 17112 | 1.4 | 1.0 | 300 | 2.22 | 1.33 | 900 |
| 17125 | 1.4 | 1.0 | 330 | 2.22 | 1.33 | 900 |
| 17132 | 1.4 | 1.0 | 350 | 2.22 | 1.33 | 600/600 |
| 17156 | 1.4 | 1.0 | 350 | 2.22 | 1.33 | 600 |
| 17130 | 1.4 | 1.0 | 350 | 2.22 | 1.33 | 900 |
| 17119 | 1.6 | 1.1 | 250 | 2.32 | 1.53 | 600 |
| 17126 | 1.6 | 1.1 | 250 | 2.32 | 1.53 | 900 |
| 17113 | 1.6 | 1.1 | 300 | 2.32 | 1.53 | 400 |
| 17116 | 1.6 | 1.1 | 300 | 2.32 | 1.53 | 600 |
| 17120 | 1.6 | 1.1 | 300 | 2.32 | 1.53 | 900 |
| 17145 | 1.6 | 1.1 | 350 | 2.32 | 1.53 | 400 |
| 17157 | 1.6 | 1.1 | 350 | 2.32 | 1.53 | 600 |
| 17104 | 1.6 | 1.1 | 350 | 2.32 | 1.53 | 900 |
| 17183* | 1.4 | 1.0 | 350 | 2.22 | 1.33 | 750 |
| 17185 | 1.2 | 1.0 | 350 | 2.32 | 1.43 | 600 |
| 17186 | 1.2 | 1.0 | 350 | 2.32 | 1.43 | 900 |
*This version of the device does not include echogenic markings.
Follicle Aspiration Set, Double Lumen: This device consists of a double lumen stainless steel needle, tubing for flushing of ovarian follicles and removal of blockages, tubing for aspiration, and a silicone rubber cork for attachment to a collection tube. One version (17129) also includes vacuum tubing attached to the silicone rubber cork for attachment to the vacuum pump used for aspiration. The needles for these versions of the device include echogenic markings at the tip of the needle to aid in visualization
{5}------------------------------------------------
| REF | Needle | | | | | Tubing | | | | | | | | | |
|-------|---------|---------|--------|---------|---------|--------|-----------------|---------|--------|------------|---------|--------|-------------|---------|--------|
| | Outer | | | Inner | | | Aspiration tube | | | Flush tube | | | Vacuum tube | | |
| | OD [mm] | ID [mm] | L [mm] | OD [mm] | ID [mm] | L [mm] | OD [mm] | ID [mm] | L [mm] | OD [mm] | ID [mm] | L [mm] | OD [mm] | ID [mm] | L [mm] |
| 17151 | 1.5 | 1.3 | 350 | 1.1 | 0.90 | 400 | 2.32 | 1.43 | 600 | 2.22 | 1.33 | 600 | - | - | - |
| 17102 | 1.5 | 1.3 | 350 | 1.1 | 0.90 | 400 | 2.32 | 1.43 | 1000 | 2.22 | 1.33 | 1000 | - | - | - |
| 17150 | 1.65 | 1.45 | 250 | 1.2 | 1.0 | 300 | 2.32 | 1.43 | 400 | 2.22 | 1.33 | 400 | - | - | - |
| 17105 | 1.65 | 1.45 | 300 | 1.2 | 1.0 | 350 | 2.32 | 1.43 | 900 | 2.22 | 1.33 | 900 | - | - | - |
| 17108 | 1.65 | 1.45 | 300 | 1.2 | 1.0 | 350 | 2.32 | 1.43 | 400 | 2.22 | 1.33 | 400 | - | - | - |
| 17131 | 1.65 | 1.45 | 300 | 1.2 | 1.0 | 350 | 2.32 | 1.43 | 400 | 2.22 | 1.33 | 900 | - | - | - |
| 17158 | 1.65 | 1.45 | 300 | 1.2 | 1.0 | 350 | 2.32 | 1.43 | 600 | 2.22 | 1.33 | 400 | - | - | - |
| 17111 | 1.65 | 1.45 | 300 | 1.2 | 1.0 | 350 | 2.32 | 1.43 | 600 | 2.22 | 1.33 | 600 | - | - | - |
| 17129 | 1.65 | 1.45 | 300 | 1.2 | 1.0 | 350 | 2.32 | 1.43 | 900 | 2.22 | 1.33 | 900 | 2.32 | 1.43 | 900 |
| 17161 | 1.65 | 1.45 | 350 | 1.2 | 1.0 | 400 | 2.32 | 1.43 | 400 | 2.22 | 1.33 | 400 | - | - | - |
| 17109 | 1.65 | 1.45 | 350 | 1.2 | 1.0 | 400 | 2.32 | 1.43 | 600 | 2.22 | 1.33 | 600 | - | - | - |
| 17148 | 1.65 | 1.45 | 350 | 1.2 | 1.0 | 400 | 2.32 | 1.43 | 900 | 2.22 | 1.33 | 600 | - | - | - |
| 17114 | 1.65 | 1.45 | 350 | 1.2 | 1.0 | 400 | 2.32 | 1.43 | 900 | 2.22 | 1.33 | 900 | - | - | - |
| 17168 | 1.65 | 1.45 | 350 | 1.2 | 1.0 | 400 | 2.32 | 1.43 | 1000 | 2.22 | 1.33 | 1000 | - | - | - |
during aspiration procedures. The table below shows the specific features of the different versions of this device:
Follicle Aspiration Set, Single Lumen Luer with Tubing: This device consists of a single lumen stainless steel needle, and aspiration tubing with a luer connection at the end. The needle is provided with echogenic markings at the tip of the needle to aid in visualization during aspiration procedures. The table below shows the specific features of this device:
| REF | Needle | | | Tubing | | |
|-------|---------|---------|-------------|---------|---------|-------------|
| | OD [mm] | ID [mm] | Length [mm] | OD [mm] | ID [mm] | Length [mm] |
| 17123 | 1.5 | 1.2 | 300 | 2.32 | 1.43 | 600 |
Follicle Aspiration Set, Single Lumen, Luer: This device consists of a single lumen stainless steel needle with a luer connection at the base of the needle. The needle is provided with echogenic markings at the tip of the needle to aid in visualization during aspiration procedures. The table below shows the specific features of the different versions of this device:
| | Dimension | | |
|-------|-----------|---------|-------------|
| REF | OD [mm] | ID [mm] | Length [mm] |
| 17154 | 1.2 | 0.8 | 300 |
| 17152 | 1.4 | 1.2 | 250 |
| 17140 | 1.4 | 1.2 | 300 |
| 17147 | 1.6 | 1.3 | 300 |
| 17137 | 1.6 | 1.3 | 350 |
{6}------------------------------------------------
## Indication for Use:
Intended for flushing and/or aspiration of oocytes from ovarian follicles.
## Comparison of Intended Use and Technological Characteristics of Subject and Predicate Devices:
| Device & Predicate Device(s): | K161970 | K991273 |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| General Device Characteristics | | |
| Indication for use | Intended for flushing<br>and/or aspiration of<br>oocytes from ovarian<br>follicles | The swemed follicle aspiration<br>set, double lumen, swemed<br>follicle aspiration set, single<br>lumen, and swemed follicle<br>aspiration luer needle and<br>intended to obtain gametes<br>from the body are specially<br>indicated for flushing and/or<br>aspiration of oocytes from<br>ovarian follicles. |
| Materials | | |
| Aspiration needle | Stainless steel | Stainless steel |
| Needle grip | Stainless steel | Stainless steel |
| Aspiration tubing | Polyolefin based<br>TPE (TPE-0) | Teflon |
| Cork | Silicone | Silicone |
| Cork adapter | Bent blunt cannula<br>mounted on the cork<br>Adapter/ blunt<br>cannula is not<br>needed for a syringe<br>with luer slip.<br>Syringe with luer<br>lock - new adapter is<br>needed to be able to<br>connect to the cork | Straight Blunt Stainless steel<br>cannula (can be used to connect<br>a vacuum pump)<br>An adapter (a blunt cannula) is<br>always needed to be able to<br>connect a syringe to the cork. |
| Cannula Luer | Methyl Methacrylate<br>Acrylonitrile<br>Butadiene Styrene | Polypropylene luer lock<br>connection |
| Cannula for flushing | Stainless steel | Stainless steel |
| Protective sheath | Polypropylene | Not provided |
{7}------------------------------------------------
K161970
| Bonding | Epoxy resin | Epoxy resin |
|------------------------------------------------------------|----------------------------------------------------------------|-------------------------------------------------------------|
| Packaging | Tyvek+<br>Polyethlene/<br>Polypropylene (clear<br>transparent) | Paper +<br>Polyethlene/Polypropylene<br>(green transparent) |
| Sterility | Radiation | Steam |
| Mouse Embryo Assay (MEA) | 1-Cell MEA: ≥ 80 %<br>expanded blastocysts<br>within 96 hours. | 2-Cell MEA |
| Endotoxins | < 1.2 EU/ device | < 1.2 EU/ device |
| Design Features | | |
| Follicle Aspiration set, single lumen | | |
| Needle length (straight) | 250,300, 330, 350<br>(mm) | 200, 250, 300, 350 (mm) |
| Needle Outer Diameter | 1.2, 1.4, 1.6 (mm) | 1.2, 1.4 , 1.5 , 1.6 (mm) |
| Needle Inner Diameter | 1.0, 1.1, 1.2 (mm) | 0.8-1.3 (mm) |
| Aspiration tube length | 400, 600, 750, 900<br>(mm) | 400, 600, 900 |
| Tubing Outer diameter | 2.22 & 2.32 (mm) | 1.67, 1.95 (mm) |
| Tubing Inner diameter | 1.33, 1.43, 1.53<br>(mm) | 1.07, 1.35 (mm) |
| Vacuum Tubing | 600 mm | NA |
| Follicle Aspiration set, double lumen | | |
| Needle length (Inner and outer needle)<br>-Outer<br>-Inner | 250, 300 & 350<br>(mm)<br>300, 350 & 400<br>(mm) | 250-350 (mm) |
| Outer Needle Diameter<br>-Outer<br>-Inner | - 1.5 & 1.65 (mm)<br>-1.3 & 1.45 (mm) | 1.2, 1.6 (mm)<br>0.8, 1.0 |
| Inner Needle Diameter<br>-Outer<br>-Inner | - 1.1 & 1.2 (mm)<br>- 0.9 & 1.0 (mm) | NA |
| Aspiration tube length | 400, 600, 900, 1000<br>(mm) | 400, 600, 900 |
| Aspiration Tubing Outer diameter | 2.32 (mm) | 1.67, 1.95 (mm) |
| Aspiration Tubing Inner diameter | 1.43 (mm) | 1.07 , 1.35 (mm) |
| End of Flushing tube | Blunt cannula OD/ID<br>1.5/1.3 mm | Hub with hole 1.2 mm |
| Flush Tube Length | 400, 600, 900, 1000 (mm) | 300, 400, 600, 900 |
| Flush Tube Outer Diameter | 2.22 (mm) | Not provided |
| Flush Tube Inner Diameter | 1.33 (mm) | Not provided |
| Vacuum Tube | 900 (mm) | Not provided |
| Vacuum Outer Diameter | 2.32 (mm) | Not provided |
| Vacuum Inner Diameter | 1.43 (mm) | Not provided |
| Follicle Aspiration Set, Single Lumen, Luer with Tubing | | |
| Needle Length | 300 (mm) | 200, 250, 300, 350 (mm) |
| Needle Outer Diameter | 1.5 (mm) | 1.2, 1.4, 1.5, 1.6 (mm) |
| Needle Inner Diameter | 1.2 (mm) | 0.8-1.3 (mm) |
| Tubing Length | 600 (mm) | 200, 250, 300, 350 (mm) |
| Tubing Outer Diameter | 2.32 (mm) | 1.2, 1.4, 1.5, 1.6 (mm) |
| Tubing Inner Diameter | 1.43 (mm) | 0.8-1.3 (mm) |
| Leur Connector on Aspiration Tubing | Yes | No |
| Follicle Aspiration Set, Single Lumen, Luer | | |
| Needle Length | 250,300,350 (mm) | 250, 300, 350 (mm) |
| Needle Outer Diameter | 1.2,1.4,1.6 (mm) | 1.2, 1.4, 1.6 (mm) |
| Needle Inner Diameter | 0.8,1.2,1.3 (mm) | 0.8, 1.0 1.2, 1.3 (mm) |
{8}------------------------------------------------
K161970 Page 6 of 9
The indications for use statements for the subject and predicate devices are similar in that they are both for use in flushing and retrieval of oocytes from ovarian follicles. Therefore, these devices have the same intended uses.
The subject and predicate devices are similar in some technological features, including design (e.g., needle material, needle length, lumen type, tip shape, cork material, etc.). However, there are a number of differences between the subject and predicate device that are described below:
- . Device materials - Some of the device materials (e.g., tubing, luer, etc.) are not the same. Materials differences are common in Assisted Reproduction Technology (ART) devices and do not raise different questions of safety and effectiveness, and can be addressed through testing (e.g., MEA, biocompatibility, performance).
- Tubing Dimensions - Dimensions of all the design types are slightly different from the predicate device (e.g., length and diameter). Differences in dimensions noted in the table above do not raise different questions of safety and effectiveness (e.g., ability to withstand vacuum pressures during use, etc.) and differences can be assessed through performance testing.
{9}------------------------------------------------
- . Needle Dimensions - There is slight difference in needle length and diameter for all device types. Variations in needle length occur in ART oocyte aspiration needles and do not raise different questions of safety or effectiveness (e.g., sufficient length to aspirate oocytes, diameter compatible with dimensions of oocytes). Performance data can be used to support these types of differences
- . Device design - 1) The subject device has a bent (45 degrees) cannula mounted on cork for easier access for connection of a vacuum pump, whereas the predicate device has a straight cannula. 2) For the double lumen aspiration kit, the subject device has a blunt cannula at the end of the flushing tube, while the predicate device has a luer fitting. 3) The subject device has an additional device type, Single Lumen Luer with Tubing, while the corresponding predicate device is described as a Single Lumen (i.e., no leur at end of tubing) . The modification of adding leur fitting to the end of the tubing, modifying the cork and cannula design/geometry, and the addition of a blunt cannula do not raise different questions of safety and effectiveness and differences can be assessed through performance testing. . The Single Lumen Luer with tubing has not been previously cleared in the predicate; however, the technology is similar to the devices within this submission and previously cleared devices with the same intended use.
- . Packaging material- the subject device uses Tyvek and clear polyethylene/polypropylene, whereas the predicate device uses paper and green transparent material. The material differences in packaging are common and do not raise different questions of safety and effectiveness, and can be addressed through testing (e.g., package integrity).
- . The subject device MEA assay has an acceptance criterion of > 80% expanded blastocyst within 96 hours using the one-cell MEA, while the predicate device conducted a two-cell MEA. Both methods are acceptable to support ART devices and do not raise different questions of safety or effectiveness.
### Performance Testing:
The following studies have been performed to support substantial equivalence to the predicate device:
- Sterilization Validation per ISO 11137-2:2013.
- . Endotoxin testing (USP<85>, LAL) - ≤1.2 EU/ device
- . Mouse embryo assay (MEA) – 1-Cell MEA: >80% expanded blastocysts within 96 hours
One-cell mouse embryos were exposed to subject device extracts and cultured at 37°C in an atmosphere containing 5% CO2. The percent of embryos developed to the expanded blastocyst stage within 96 hours were assessed in comparison with the control group.
- . Vacuum Testing – Assessment of device function under vacuum pressure at -500 mmHg. Devices were shown to operate as intended without signs of damage.
- Flow Testing - Testing was conducted to assess that flow rates were within anticipated ranges under intended use conditions.
{10}------------------------------------------------
- . Mechanical Testing of Needles
- Breakage, Bend, Stiffness per ISO 9626: 2016. o
- Tensile Strength Testing
- . Biocompatibility Testing:
- Cytotoxicity testing per 10993-5:2009 o
- Guinea Pig Maximization Sensitization testing per ISO 10993-10:2010 O
- o Intracutaneous Reactivity testing per ISO 10993-10:2010
Testing showed that the device extracts assessed were non-cytotoxic, non-sensitizing, and non-irritating.
- . Shelf-life studies (accelerated) were conducted to ensure that the following acceptance criteria are met for the for devices over their 3 years shelf-life:
- 1-cell MEA ≥80% expanded blastocysts within 96 hours O
- Standard Test Method for Seal Strength of Flexible Barrier Materials per O ASTM F88 / F88M - 09
- Dye Penetration per ASTM F1929 12 о
- Visual Inspection per ASTM F1886 / F1886M 09 о
- Tensile Strength Testing o
#### Conclusions from Non-Clinical Performance Data
The results of the performance testing described above demonstrate that the Follicle Aspiration Set is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.