← Product Code [MQB](/productcode/MQB) · K161942

# PerkinElmer, XRpad2 3025 HWC-M Flat Panel Detector (K161942)

_Perkinelmer, Inc. · MQB · Sep 1, 2016 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K161942

## Device Facts

- **Applicant:** Perkinelmer, Inc.
- **Product Code:** [MQB](/productcode/MQB.md)
- **Decision Date:** Sep 1, 2016
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1680
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The XRpad2 3025 HWC-M, when used with a radiographic imaging system, is indicated for use in generating radiographic images of human anatomy for diagnostic X-ray procedures, wherever conventional screen-film (SF), digital radiography (DR), or computed radiography (CR) systems may be used. It is not intended for mammographic use.

## Device Story

XRpad2 3025 HWC-M is a flat panel X-ray detector designed for integration into radiographic imaging systems. It captures X-ray radiation using a Cesium Iodide (CsI:Tl) scintillator, converting it into digital images. The device operates as a component within a clinical imaging workflow, typically used in hospitals or clinics by radiologic technologists. It supports both wired (GigE) and wireless (802.11n WiFi) connectivity to transmit image data to a workstation. The detector features automatic exposure detection (AED) or dedicated trigger interfaces to synchronize with X-ray generators. By providing high-resolution digital images, it enables clinicians to perform diagnostic examinations efficiently, replacing traditional film-based or older digital radiography systems. The device is a smaller, lighter version of the predicate, intended to maintain comparable image quality while offering improved portability and handling for various anatomical imaging tasks.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance was validated through internal non-clinical testing, comparing Modulation Transfer Function (MTF) and Detective Quantum Efficiency (DQE) against the predicate device to ensure equivalent image quality.

## Technological Characteristics

Flat panel detector with CsI:Tl scintillator; 100 µm pixel pitch; 2508 x 3004 pixel matrix; aluminum and carbon-fiber housing; 282 mm x 332 mm x 15.5 mm dimensions; 1.8 kg weight. Connectivity via Gb Ethernet or 802.11n WiFi (5.1-5.9 GHz). Power via external supply or battery. X-ray synchronization via AED or trigger lines.

## Regulatory Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- PerkinElmer XRpad 4336 MED Flat Panel Detector ([K140551](/device/K140551.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around the perimeter. In the center is a stylized graphic of what appears to be a human figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 1, 2016

PerkinElmer, Inc. % Ms. Dawn M. Spooner Associate Director, Regulatory Affairs 940 Winter Street WALTHAM MA 02451

Re: K161942

Trade/Device Name: XRpad2 3025 HWC-M Flat Panel Detector Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: July 14, 2016 Received: July 15, 2016

Dear Ms. Spooner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D'Hara

For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known)

### K161942

Device Name

XRpad2 3025 HWC-M Flat Panel Detector

### Indications for Use (Describe)

The XRpad2 3025 HWC-M, when used with a radiographic imaging system, is indicated for use in generating radiographic images of human anatomy for diagnostic X-ray procedures, wherever conventional screen-film (SF), digital radiography (DR), or computed radiography (CR) systems may be used. It is not intended for mammonfions .

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## olsand

əsinə aid bərində aid bir və qarşına qalındığı var. Azərbaycan filmlərinin siyahıya əlin qırmızı viri

The assigned 510(k) number is: X161942

910Z 'SI Isnany

| Submitted By:     | PerkinElmer Medical Imaging<br>2175 Mission College Blvd.<br>Santa Clara, CA 95054<br>U.S.A.                                                                                                                   |
|-------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:   | Dawn Spooner<br>Associate Director, Regulatory Affairs<br>Tel. 781-663-6071<br>Fax. 781-663-5969                                                                                                               |
| Device Name:      | PerkinElmer XRpad2 3025 HWC-M Flat Panel Detector                                                                                                                                                              |
| Classification:   | Product Code: MQB<br>Classification Name: Stationary X-ray System<br>Classification Regulation: 21 CFR 892.1680                                                                                                |
| Predicate Device: | PerkinElmer XRpad 4336 MED Flat Panel Detector<br>510(k) Clearance: K140551; August 1, 2014<br>Product Code: MQB<br>Classification Name: Stationary X-ray System<br>Classification Regulation: 21 CFR 892.1680 |

{4}------------------------------------------------

## evice Description

standard X-ray cassette Bucky ﻳﻮﻧﻴﻮ ﻟﻮﻃﻨﻴﻦ ﺍﻟﺴﻴﺎﺳﻲ ﻟﻠﻤﺮﺍﺗﺐ ﺍﻟﻤﺘﻮﺳﻂ ﺍﻟﻤﺘﻮﺍﺿﻊ ﺍﻟﻤﺘﺎﺣﺔ ﺍﻟﻤﺘﺎﺣﺔ ﺍﻟﻌﺎﻡ ﺍﻟﻤﺘﺎﺣﺪ ﺍﻟﻤﺘﺎﺣﺪ ﺍﻟﻌﺎﺩﻟﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺎﺣﺪ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺎﺑﻌﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺎﺑﻌﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻘﺎﺩ ﺍﻟﻤﺘﺤﺪﺓ in sug doily .unu c.cl x mm 0.282 × um 0.282 × um 0.282 × mm winioni

- Xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx
- topad IPU-2 (Interface and Power Un
- KRpad LPT2 Detector Cable
- XRpad Protective Insert 307
- OV Cable IEC 60320 C13
- ........ Cable IEC 60320 CLS
- frigger Cable 5 m/16.5
- ger Cable 20 m/65
- 
- GigE Interface Cable 15.25 m/ Interface Cable 7.6 m/2
- E Interface Cable 30.5 m/100

## Indications For Use

əsn ondare o (ЯП) үүдбэлэгийн төрий (ЧЗ) шүү-пээлэг төлөлтөлтөг тэлэгээгүй хэлийн хэлийн хэлийн хэлийн and and

{5}------------------------------------------------

## Comparison Chart

| Characteristics                          |                                                                                                                                                                                                                                                                                                                                                                   |                                                    |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|
| Intended Use /<br>Indications for<br>Use | The XRpad 4336 MED when used<br>with a radiographic imaging system, is<br>indicated for use in generating<br>radiographic images of human<br>anatomy for diagnostic X-ray<br>procedures, wherever conventional<br>screen-film (SF), digital radiography<br>(DR), or computed radiography (CR)<br>systems may be used. It is not intended<br>for mammographic use. | Same                                               |
| Model                                    | PerkinElmer<br>XRpad 4336 MED<br>(K140551)                                                                                                                                                                                                                                                                                                                        | Proposed Model<br>PerkinElmer<br>XRpad2 3025 HWC-M |
| Panel                                    |                                                                                                                                                                                                                                                                                                                                                                   |                                                    |
| Scintillator                             | Direct deposition CsI:Tl                                                                                                                                                                                                                                                                                                                                          | Same                                               |
| Pixel pitch                              | $100 \mu m$                                                                                                                                                                                                                                                                                                                                                       | Same                                               |
| Pixel matrix                             | 3556 × 4320 pixels                                                                                                                                                                                                                                                                                                                                                | 2508 × 3004                                        |
| Limiting<br>resolution                   | 5 lp/mm                                                                                                                                                                                                                                                                                                                                                           | Same                                               |
| Binning<br>capable                       | 2 x 2 binning for 200 $\mu m$                                                                                                                                                                                                                                                                                                                                     | Same                                               |
| Data transmit<br>area                    | 355 mm x 432 mm                                                                                                                                                                                                                                                                                                                                                   | 250 mm x 300 mm                                    |
| External<br>dimensions<br>(w x l x h)    | 384 mm x 460 mm × 15 mm                                                                                                                                                                                                                                                                                                                                           | 282 mm x 332 mm × 15.5 mm                          |
| Weight                                   | 3.8 kg                                                                                                                                                                                                                                                                                                                                                            | 1.8 kg                                             |
| Housing<br>material                      | Aluminum with carbon-fiber                                                                                                                                                                                                                                                                                                                                        | Same                                               |
| Finish                                   | Matte carbon-fiber front with white<br>silk-screen active area markers                                                                                                                                                                                                                                                                                            | Same                                               |
| Communication<br>interface               | Gb Ethernet or 802.11n WiFi                                                                                                                                                                                                                                                                                                                                       | Same                                               |
| WiFi band                                | 5.1-5.3 GHz                                                                                                                                                                                                                                                                                                                                                       | 5.1-5.9 GHz                                        |
| X-ray<br>synchronization<br>interface    | Dedicated trigger in/out signal lines or<br>Automatic Exposure Detection                                                                                                                                                                                                                                                                                          | Same                                               |
| Power                                    | External power supply or battery                                                                                                                                                                                                                                                                                                                                  | Same                                               |
| Software<br>library                      | Windows OS                                                                                                                                                                                                                                                                                                                                                        | Windows and Linux OS                               |

{6}------------------------------------------------

## Summary of Studies

compared the producite Personal internet Association are similar. unsterials, are similar in design and construction, and have the same intended uss Contraction and physics of the PerkinElment Kepth 2015 320 F pus yaoloudsat ultimiz sziliyu soivel sissipsing brokes sinning similaring suit lates to the decrease in dimensions and does not impact image quality

vs w . ອີກາ່າວະ 1821 ເມກາວເປັນ ລາວ. internal no-climical testing, complies with standards and regulations such as U duros Allufessans san rollosted to MCC MH OMH ST0C Thed Ax and Huiyan and A x (CEM ESA SSAI CAX) add this the states and leasters now a osh are and and and and of aldesting the (CEM ESA CC91 CAX) SCAZZTX Trodums As a 6 MED device.

# many of Design Control and Risk Managemen

not impact image quality MED (X 140551). The modifications are related to the decrease in dimensions and sque 2025 HWC-M flat panel X-ray detector is modification of the Reppe

A min and and and a more a fit de descreative de server many and be by Free cessfully mitigated and accepted is modification were reviewed and implemented as part of product desi suditiones suobrezed bue system befinant in tall boblyono tusursesss The specific risk control supprective measures to mitigate th

## Summary of Non-Clinical Data

ət qarının ədən və və və və və və və və və və və və və və və və və və və və və və və və və və və və və və və və və və və və v U se your suoiitelus ar pue spreparis parders partis souring un onels vince lettings on attensured and status submittee and with the the sentively of the entineling of the Park in the Promise of the Promotion consisted for sonly be ine min physis in the super of the RPports of Sevice, PerkinElmer has performed internal non-clinical testing an BE and MTF are comparable to the predicate device. The non-clinics

{7}------------------------------------------------

## Substantist Equiivalency

ən səsinə qalınmış var. məlimis ərə salarının bir yazırlığında olunur. Bu mənis məsl od de detected and and the read on the segent only of the seguires in asn tot bepared signing software of the X-ray uniti

### Conclusion.

 its intended use. supeps si bur single sousides sus states the sprepatis and is and is a a pure

in fəsiləsindən filmlərinə aid olunan filmlərinin filmlərində vasitəsinə aid (123041) CIN əsinə aid bir və bulunun məqli (xi)01 ç sidir ui bərldqus qortanından ən bir mənələrini əsasən

---

**Source:** [https://fda.innolitics.com/device/K161942](https://fda.innolitics.com/device/K161942)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
