The indications for use of the product is to remove the cold from the bucky by placing this sheet between patient's breast and the image receptor plate allowing the patient to feel more comfortable and less cold without interfering with image quality.
Device Story
Comfort Cover is a single-use, disposable, non-sterile fabric sheet with adhesive backing; applied by radiology technologists to the mammographic image receptor plate (bucky) before patient positioning. Device acts as a thermal barrier between patient breast and receptor plate; reduces cold sensation during mammography. Designed for use in hospitals and radiology centers; compatible with both Full Field Digital Mammography (FFDM) and Digital Breast Tomosynthesis (DBT). Product is removed and discarded after examination; adhesive allows clean removal without residue. Benefits include improved patient comfort without interfering with image quality, radiation dose, or patient positioning.
Clinical Evidence
Bench testing only. No clinical or animal studies performed. Performance validated via mechanical testing (adhesion, thermal conductivity, abrasion resistance) and radiological phantom studies per MQSA and IEC 62220-1-2:2007/61223-3-2. Results confirmed artifact-free imaging, equivalent radiation dose, and equivalent image quality compared to no-cover baseline. Usability studies involving radiologists, technologists, and patients confirmed design requirements for comfort and positioning were met.
Technological Characteristics
Single-use, non-sterile, disposable woven fabric sheet with adhesive backing and printed ink. Dimensions: 22.9cm x 27.9cm or 29.2cm x 33.0cm. Radiolucent. Mechanical testing per ASTM D5470-12 (thermal conductivity), ASTM D5264-98 (abrasion resistance), and ASTM D6252-98 (peel adhesion). Biocompatibility per ISO 10993-5 and ISO 10993-10.
Indications for Use
Indicated for women over age 35 undergoing mammography to provide comfort by reducing cold sensation from the image receptor plate.
Regulatory Classification
Identification
A mammographic x-ray system is a device intended to be used to produce radiographs of the breast. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 15, 2016
Precision Dynamics Corporation % Mr. Dave Yungvirt CEO Third Party Review Group, LLC The Old Station House 24 Lackawanna Place MILLBURN NJ 07041
Re: K161920
Trade/Device Name: Comfort Cover Regulation Number: 21 CFR 892.1710 Regulation Name: Mammographic x-ray system Regulatory Class: II Product Code: IZH Dated: July 10, 2016 Received: July 13, 2016
Dear Mr. Yungvirt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D.'Hara
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health For
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
K161920
Device Name Comfort Cover
Indications for Use (Describe)
The indications for use of the product is to remove the cold from the bucky by placing this sheet between patient's breast and the image receptor plate allowing the patient to feel more comfortable and less cold without interfering with image quality.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for PDC, which is a Brady business. The logo features a series of vertical bars in shades of blue and green, followed by the letters "pdc" in a bold, blue font. Below the letters, the words "A Brady Business" are written in a smaller font, followed by the Brady logo.
27770 N. Entertainment Dr. Ste. 200 Valencia, CA 91355 800.847.0670 818.897.1111 Email info@pdcorp.com
pdcorp.com
### SECTION 5: 510(k) SUMMARY
The following information is provided in accordance with 21 CFR 807.92, for the Traditional Premarket 510(k) Summary:
#### (1) SUBMITTER INFORMATION
Company Name and Address:
Contact Name: Contact Telephone: Contact E-mail: Date Prepared:
#### (2) DEVICE INFORMATION
Trade Name: Name of Device: Common Name: Classification Name: Regulatory Classification: Regulation Number: Product Code:
Precision Dynamics Corporation (PDC) 27770 N. Entertainment Drive, Suite 200, Valencia, California, 91355 Swarna Mukund Ph. D. (661) 481-8857 swarna_mukund@bradycorp.com June 10, 2016
Comfort Cover™ Comfort Cover Mammography Cover System, X-ray, mammographic Class II 892.1710 IZH
#### (3) PREDICATE DEVICE
K073262, Bella, Bella Blankets: Classification Name: Regulatory Classification: Regulation Number: Product Code:
Beekley Corporation System, X-ray, mammographic Class II 892.1710 IZH
# (4) DEVICE DESCRIPTION
Comfort Cover is a single-use disposable cover placed between the patient's breast and the image receptor plate during mammography. The purpose of this product is to allow the patient to feel more comfortable and less cold during the mammographic examination without interfering with image quality. Comfort Cover will allow for artifact-free images in both Field Digital Mammography (FFDM) and Digital Breast Tomosynthesis (DBT), patient radiation dose is equivalent to that when using no Comfort Cover and the Image quality is equivalent to images exposed without Comfort Cover.
Comfort Cover is designed to fit small and large image receptor plates of all leading mammographic equipments. The product is constructed of a fabric material with a purple printed design and has adhesive backing with a paper liner. Both the material and adhesive backing of the product do not affect imaging.
{4}------------------------------------------------
The product is directly applied wrinkle free to the image receptor plate of the mammographic machine by the adhesive backing exposed by peeling off the liner. After the patient's mammogram, the product is peeled off the image receptor plate and disposed. As the product removes cleanly without leaving behind any residue, the image receptor plate does not need to be cleaned between Comfort Covers nor between patients.
# (5) INDICATIONS FOR USE
The indications for use of the product is to remove the cold from the bucky by placing this sheet between patient's breast and the image receptor plate allowing the patient to feel more comfortable and less cold without interfering with image quality.
### (6) COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The following table compares the Comfort Cover to the predicate device with respect to indications for use, principles of operation, technological characteristics, materials, and performance testing. The comparison of the devices provides more detailed information regarding the basis for the determination of substantial equivalence. The subject device does not raise any new issues of safety or effectiveness based on the similarities to the predicate device.
| Device Characteristics | Comfort Cover | Bella | Significant Differences |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The indications for use of the<br>product is to remove the cold<br>from the bucky by placing this<br>sheet between patient's<br>breast and the imaging<br>receptor plate allowing the<br>patient to feel more<br>comfortable and less cold<br>without interfering with image<br>quality. | The intended use of the<br>product is to remove the cold<br>from the bucky by placing this<br>sheet between patient's<br>breast and the imaging<br>receptor plate allowing the<br>patient to feel more<br>comfortable and less cold<br>without interfering with image<br>quality. | None |
| Target<br>population | Women over the age of 35<br>having mammograms. | Women over the age of 35<br>having mammograms | None |
| Where used | Hospitals, radiology centers | Hospitals, radiology centers | None |
| Product Size | 22.9cm x 27.9cm designed<br>for use with 18cm x 24cm<br>image receptor plate<br>29.2cm x 33.0cm designed<br>for use with 24cm x 30cm<br>image receptor plate | 22.9cm x 27.9cm designed<br>for use with 18cm x 24cm<br>image receptor plate<br>29.2cm x 33.0cm designed<br>for use with 24cm x 30cm<br>image receptor plate | None |
| Material | Fabric (woven), adhesive<br>backing, purple ink. | Fabric (non-woven), adhesive<br>backing, pink ink. | Different. Safety and<br>performance testing<br>results have shown<br>that the slight<br>differences in<br>material composition<br>do not affect safety<br>and effectiveness |
# Table 5-1: Comparison of Technological Characteristics with Predicate Device
{5}------------------------------------------------
| Device<br>Characteristics | Comfort Cover | Bella | Significant<br>Differences |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|----------------------------|
| Packaging and<br>Labeling | Not individually wrapped.<br>Fifty covers packaged in one<br>box of Comfort cover. | Not individually wrapped.<br>Fifty covers packaged in one<br>box of Comfort cover. | None |
| | Prescription use only. | Prescription use only. | None |
| | Labeling same as Bella. | NA | NA |
| Biocompatibility | Yes | Yes | None |
| Performance<br>Testing | No visible defects. | No visible defects. | None |
| | Repositionable Adhesive | Repositionable Adhesive | None |
| | Provides comfort and<br>removes cold during a<br>mammographic examination<br>without affecting quality of the<br>images. | Provides comfort and<br>removes cold during a<br>mammographic examination<br>without affecting quality of the<br>images. | None |
| | Allows for artifact-free images<br>and image quality equivalent<br>to images exposed without<br>Comfort Cover. | Not tested side-by-side with<br>Comfort Cover. Bella testing<br>results can be found in their<br>510k (K073262). | None |
| | Removes cleanly from the<br>mammographic equipment<br>without leaving behind any<br>residue. | Removes cleanly from the<br>mammographic equipment<br>without leaving behind any<br>residue. | None |
| | Printed ink does not smear. | Printed ink does not smear. | None |
| | Remains where placed. | Remains where placed. | None |
| Sterility | Non-sterile | Non-sterile | None |
| Dispensing<br>method | Dispenser Box | Dispenser Box | None |
| Environment<br>Compatibility | Disposable, Non-recyclable | Disposable, Non-recyclable | None |
| Radiation<br>Safety | Does not image, no artifacts,<br>quality of images not<br>affected, radiolucent. | Not tested side by side with<br>Comfort Cover. Bella testing<br>results can be found in their<br>510k (K073262). | None |
# (7) PERFORMANCE DATA
As part of demonstrating safety and effectiveness of Comfort Cover and in showing substantial equivalence to the predicate device, Precision Dynamics Corporation completed a number of non-clinical performance tests. The Comfort Cover met all the requirements for overall design, biocompatibility, and radiologic safety confirming that the design output meets the design inputs and specifications for the device.
# Summary and Conclusions of Biocompatibility Testing
Biocompatibility testing was conducted in accordance with ISO 10993-5, Biological evaluation of medical devices - Part 5: Tests for in vitro Cytotoxicity, and ISO 10993-10, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.
The subject device passed all of the above tests.
{6}------------------------------------------------
#### Summary and Conclusions of Comparative Testing
The following mechanical tests were conducted on both the subject and the predicate devices.
- Visual and Dimensional examination ●
- Thermal Conductivity Test
- This testing was performed per ASTM D5470-12 which is a test method for O measurement of thermal conductivity of materials.
- Ink Smear Test
- This test was performed as per ASTM D5264-98 which is a standard test method o for abrasion resistance of printed materials by the Sutherland Rub Tester to test if the ink from the printed graphics smears.
- . Peel Adhesion Test
- This test was performed as per ASTM D6252-98 which is a standard test method O for peel adhesion of pressure sensitive label Stocks at a 90° Angle to test the peel adhesion of the device.
- Removes Cleanly Test
- o This test was an in-house developed method to test if the product removes cleanly from the mammographic equipment without leaving behind any residue.
The subject device passed all of the above tests.
#### Summary and Conclusions of Radiological Studies
Phantom image testing was performed as per MQSA regulations and with relevant International Electro technical Commission (IEC) standards (62220-1-2:2007 and 61223-3-2) to verify that,
- Comfort Cover will allow for artifact-free images in both FFDM and DBT.
- . Patient radiation dose is equivalent to that when using no Comfort Cover.
- . Image quality is equivalent to images exposed when using no Comfort Cover.
The subject device passed all of the above tests.
#### Summary and Conclusions of Usability Studies
Usability studies included radiologists, radiology technologists and patients testing all of the user need requirements. The Comfort Cover fulfilled its design requirements of making the mammography examination more comfortable and did not negatively impact patient positioning.
#### Summary of Animal Studies
No animal testing was necessary.
#### Summary of Clinical Studies
No clinical testing was necessary.
{7}------------------------------------------------
### 8) CONCLUSIONS
The subject device Comfort Cover and the predicate device Bella have the same technological characteristics and intended use.
Both devices have similar design of having rectangular shape with adhesive backing which allows easy application and removal from the image receptor plate and with floral graphics printed on the top of the sheet using FDA approved ink. Both devices are presented as single use, non-sterile and non-compressible.
Both devices remove the cold from the image receptor plate of the mammographic equipment and provide comfort during routine mammographic examinations without affecting image quality. Both devices do not negatively affect patient positioning.
Minor difference in the Comfort Cover material as compared to the predicate device. Bella does not affect safety and effectiveness of the device when used as labeled as shown by safety and efficacy studies. Therefore, Precision Dynamics Corporation believes that Comfort Cover is substantially equivalent to the existing legally marketed predicate device, Bella.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.