21.3 inch (54 cm) Color 2M pixel LCD Monitor, CCL214 (CL21214) is intended to be used in displaying and viewing medical images for diagnosis by trained medical practitioners. It is not meant to be used in digital mammography.
Device Story
21.3-inch color LCD monitor; displays medical images for diagnostic review by trained medical practitioners. Inputs via DVI-D or DisplayPort digital interfaces; processes signals to render images on screen. Features LED backlight; front luminance sensor for stabilization; DICOM GSDF compliance via internal lookup table. Used in clinical environments; operated by physicians/radiologists. Output provides visual representation of medical imaging data; supports clinical decision-making by enabling diagnostic interpretation. Benefits include high-resolution visualization and luminance stability for consistent image quality.
Clinical Evidence
Bench testing only. Performance verified against AAPM-TG18 standards for luminance uniformity and chromaticity. Safety and EMC standards validated by third-party certifiers. No clinical data presented.
Technological Characteristics
21.3-inch color LCD monitor; 1600x1200 resolution; LED backlight; DVI-D and DisplayPort connectivity. Luminance uniformity <30% (AAPM-TG18 4.4); Chromaticity Delta (u', v') <= 0.01 (AAPM-TG18 4.8.4). Pixel defects meet ISO13406-2 Class II. Front luminance sensor for stabilization. Power supply 100-240V AC. Complies with ANSI/AAMI ES60601-1, CAN/CSA C22.2 No. 60601-1, FCC Class B.
Indications for Use
Indicated for trained medical practitioners to display and view medical images for diagnostic purposes. Not for use in digital mammography.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
21.3 inch (54 cm) Color LCD Monitor CCL210 (CL21210) (K151134)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 8, 2016
JVC KENWOOD Corporation % Mr. Tsukasa Tashiro Engineering Specialist 3-12, Moriya-cho, Kanagawa-ku Yokohama. Kanagawa 221-0022 JAPAN
Re: K161895
Trade/Device Name: 2MP Color LCD Monitor CCL214 (CL21214) Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: PGY Dated: July 8, 2016 Received: July 11, 2016
Dear Mr. Tashiro:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K161895
Device Name
2MP Color LCD Monitor CCL214 (CL21214)
Indications for Use (Describe)
2MP Color LCD Monitor CCL214 (CL21214)
is intended to be used in displaying and viewing medical images for diagnosis by trained medical practitioners. It is not meant to be used in digital mammography.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
| <span> <span style="font-family: 'Arial Unicode MS', sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
ion of
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for JVC KENWOOD. The text "JVC KENWOOD" is in a bold, dark blue font. Below the company name, the text "creates excitement & peace of mind" is in a lighter blue font.
# 510(k) SUMMARY
| Submitted Information: | JVC KENWOOD Corporation<br>3-12, Moriya-cho, Kanagawa-ku,<br>Yokohama-shi, Kanagawa, 221-0022 Japan |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Tsukasa Tashiro, Engineering Specialist<br>Email: tashiro.tsukasa@jvckenwood.com<br>Tel: +81-258-24-6611<br>Fax: +81-258-24-6617 |
| Date Prepared: | July 08, 2016 |
| Device Name: | 2MP Color LCD Monitor CCL214<br>(CL21214) |
| Common Name: | CCL214 (CL21214) |
| Classification Name: | Class II<br>(Part 892 Radiology Devices<br>Sec. 892.2050 Picture Archiving and Communication System) |
| Predicate Device: | 21.3 inch (54 cm) Color LCD Monitor CCL210 (CL21210)<br>(K151134) |
| Device Description: | CCL214 (CL21214) is a 21.3-inch (54 cm) Color LCD monitor<br>whose display resolution is 1600 x 1200 (landscape), 1200 x 1600<br>(portrait) supporting DVI-D (digital interface) and DisplayPort<br>(digital interface). |
| Intended Use: | 21.3 inch (54 cm) Color 2M pixel LCD Monitor, CCL214<br>(CL21214) is intended to be used in displaying and viewing<br>medical images for diagnosis by trained medical practitioners.<br>It is not meant to be used in digital mammography. |
| Substantial Equivalence: | CCL214 (CL21214) shares the same technical characteristics,<br>application, and intended use as our predicate device CCL210<br>(CL21210 / K151134). |
{4}------------------------------------------------
| Technical Specification |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. Luminance uniformity<br>[SPEC] Less than 30% based on AAPM-TG18 4.4. |
| 2. Pixel Defects / Fault<br>[SPEC] Class II or more. ISO13406-2 |
| 3. Artifacts<br>• phase/clock issues flicker<br>• miscellaneous including ringing, ghosting, image sticking<br>[SPEC] By visible check, no flicker, ringing, ghosting and image sticking. |
| 4. Chromaticity Measurement of 5%, 50%, 95% Level |
| 5. Chromaticity<br>[SPEC] Delta (u', v') $\leq$ 0.01 measured at 80% Lmax based on AAPM-TG18 4.8.4 |
| 6. Power On Luminance Drift<br>[SPEC] $\triangle$ Lmax≤±10% within 60 seconds.<br>$\triangle$ Lmax : the deviation between the target maximum luminance and the measured luminance. |
{5}------------------------------------------------
| | CCL210 (CL21210) | CCL214 (CL21214) |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K151134 | Not Known |
| Display Area | Horizontal: 432.0mm, Vertical: 324.0mm | Horizontal: 432.0mm, Vertical: 324.0mm |
| Input Signal | DVI-I 29-pin connector, DisplayPort connector | DVI-D 24-pin connector, DisplayPort connector |
| Maximum Resolution | 1200 x 1600 at portrait display<br>1600 x 1200 at landscape display | 1200 x 1600 at portrait display<br>1600 x 1200 at landscape display |
| Pixel Pitch | 0.270 mm x 0.270 mm | 0.270 mm x 0.270 mm |
| Scanning Frequency | Horizontal: 30 - 75kHz<br>Vertical: 55 - 60Hz | DVI<br>Horizontal: 74.1KHz, Vertical: 60Hz (Landscape)<br>Horizontal:98.1KHz, Vertical: 60Hz (Portrait)<br>DisplayPort<br>Horizontal: 75.0KHz, Vertical: 60Hz (Landscape)<br>Horizontal:99.0KHz, Vertical: 60Hz (Portrait) |
| Maximum Image Clock | 162MHz | 162MHz |
| Maximum Luminance | 250 cd/m² DICOM calibrated<br>440 cd/m² typ. As LCD component | 250 cd/m² DICOM calibrated<br>500 cd/m² typ. As LCD component |
| Luminance Calibration<br>(Optional) | Software<br>Photo Sensor (optional): X-Rite Chroma 5 | Software<br>Photo Sensor (optional): X-Rite i1 Display |
| Contrast Ratio | Typ 1500:1 | Typ 1200:1 |
| Serial Communication | USB | USB |
| Safety Standards | Medical Safety: ANSI/AAMI ES60601-1, CAN/CSA C22.2<br>No. 60601-1, FCC (Class B), MDD/CE, VCCI-B (Class B),<br>ICES-003 (Class B) | Medical Safety: ANSI/AAMI ES60601-1, CAN/CSA C22.2<br>No. 60601-1, FCC (Class B), MDD/CE, VCCI-B (Class B),<br>ICES-003 (Class B) |
| Weight & Dimension | Net: 10.7kg<br>473(w) x 476.4 – 537.9(H) x 220(D) mm<br>Packed: 15.5kg<br>470(w) x 675(H) x 347(D) mm | Net: 11.1kg<br>367.0(w) x 521.9 – 583.4(H) x 220(D) mm<br>Packed: Approx.14.0kg<br>470(w) x 670(H) x 340.0(D) mm |
| Power Supply | 100-240V AC 50/60Hz | 100-240V AC 50/60Hz |
# antial Equivalence Compa
The LCD panel, the power supply, and the main board for CCL214 (CL21214) are different from those for CCL210 (Predicati
device). Differences
tilt stand for CCL214 (CL21214) is the same as CCL210 (K151134).
the sizes of front bezel and the rear cover are smaller than ones of CCL210 only sli Similarities:
{6}------------------------------------------------
CCL214 (CL21214) can be considered to have equivalent performances to those of the predicate device CCL210 (K151134) due to the following reasons:
- a. CCL214 (CL21214) and the predicate device CCL210 (K151134) have the same active area whose size is 432.0mm x 324.0mm. Also, the maximum display resolution for both models is 1600 x 1200.
- b. As for the maximum luminance, CCL214 (CL21214) is superior to CCL210 (K151134). (CCL214 (CL21214): 500cd/m2 and CCL210 (K151134):440cd/m2) The maximum luminance by DICOM calibration at the factory default is 250cd/m2 for both CCL214 (CL21214) and the predicate device CCL210 (K151134).
- The LED backlight is equipped with CCL214 (CL21214) and the predicate device ﻥ CCL210 (K151134).
- As the factory default status, the both devices display images in accordance with DICOM ರ. GSDF which is stored in the lookup table inside the device.
- e. CCL214 (CL21214) supports DVI and DisplayPort as well as the predicate device CCL210 (K151134) .
As for the maintenance, the same QC software is used for both devices. CCL210 (K151134) has the backlight sensor to stabilize the luminance, but CCL214 (CL21214) has the front luminance sensor, which means the higher performance, to do so.
The overall design of the CCL214 (CL21214) has been validated in accordance with internationally recognized Safety and EMC standards by third-party certifiers. Besides, JVC KENWOOD Corporation performed the system and performance tests to verify that the CCL214 (CL21214) performs in accordance with its specifications. None of the tests revealed behaviors which is inconsistent with the expected performance.
## Conclusion
The 2M pixel Color LCD Monitor, CCL214 (CL21214) is substantially equivalent to the predicate device with respect to technical characteristics, its application, its intended use, and the specifications of the primary components.
And other differences have been independently validated.
But any differences between these two devices do not affect safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.