The EVA (Enhanced Visual Assessment) System is intended to provide magnified viewing of the vagina, cervix and external genitalia in order to aid in diagnosing abnormalities and selecting areas for biopsy. The device is intended for use in hospitals, doctor's offices, and remote and rural clinics.
Device Story
Handheld colposcope; utilizes smartphone (Motorola G2/G3) as imaging/display unit. Device comprises colposcope body, lens, smartphone brackets, LED electronics. Operator captures images via CervDx app; images transferred to online portal for remote viewing. Used in hospitals, clinics, rural settings. Provides magnified view (up to 16x digital) of vagina/cervix; aids clinician in identifying abnormalities and biopsy site selection. Features include acetic acid timer, LED illumination, image annotation, and cloud-based storage. Benefits include portability, ease of use, and remote access to clinical images for decision-making.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via electrical safety (AAMI/ANSI 60601-1), EMC (IEC 60601-1-2), and optical standards (ISO 8600-3, ISO 8600-5). Additional bench testing included power consumption, battery life, image distortion, cervical model image quality, and illuminance. Software validated per IEC 62304; usability validated per IEC 62366-1.
Technological Characteristics
Handheld colposcope; materials: PU (Biresin RG53 FR), 6061/5052 Aluminum, Stainless Steel. Optical magnification up to 4.0x (16x digital). Light source: 3W 6500k star LED. Working distance: 22.5-42.5 cm. Connectivity: Smartphone-based (Motorola G2/G3) with cloud portal. Power: Battery-operated (10-hour life). Software: CervDx app for image capture/annotation. Standards: AAMI/ANSI 60601-1, IEC 60601-1-2, ISO 8600-3, ISO 8600-5, IEC 62304, IEC 62366-1.
Indications for Use
Indicated for magnified viewing of vagina, cervix, and external genitalia to aid in diagnosing abnormalities and selecting biopsy sites. Intended for use in hospitals, doctor's offices, and remote/rural clinics.
Regulatory Classification
Identification
A colposcope is a device designed to permit direct viewing of the tissues of the vagina and cervix by a telescopic system located outside the vagina. It is used to diagnose abnormalities and select areas for biopsy. This generic type of device may include a light source, cables, and component parts.
Special Controls
*Classification.* Class II (special controls). The device, when it is a standard colposcope (or colpomicroscope) that uses only a white light source, does not use filters other than a green filter, does not include image analysis software, and is not smartphone-based, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 19, 2016
MobileODT Ltd. Jen Acobas RA/QA Manager Ben Avigdor 8, 3rd Floor Tel Aviv, 6721832 Israel
Re: K161871
> Trade/Device Name: EVA (Enhanced Visual Assessment) System Regulation Number: 21 CFR 884.1630 Regulation Name: Colposcope Regulatory Class: Class II Product Code: HEX Dated: November 29, 2016 Received: November 29, 2016
Dear Jen Acobas,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{2}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K161871
Device Name
EVA (Enhanced Visual Assessment) System
Indications for Use (Describe)
The EVA (Enhanced Visual Assessment) System is intended to provide magnified viewing of the vagina, cervix and external genitalia in order to aid in diagnosing abnormalities and selecting areas for biopsy. The device is intended for use in hospitals, doctor's offices, and remote and rural clinics.
Type of Use (Select one or both, as applicable)
| <span style="font-size:12px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------|
| <span style="font-size:12px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
# K161871
#### Submitter Information 1.
Manufacturer Name and Address MobileODT Ltd. Ben Avigdor 8, 3rd Floor Tel Aviv 6721832 Israel Phone: (617) 454-4687
### Official Correspondent
Jennifer Acobas, VP Regulatory MobileODT Ltd. Ben Avigdor 8, 3rd Floor Tel Aviv 6721832 Israel Phone: (617) 454-4687
#### 2. Date Prepared: December 14, 2016
#### 3. Device Information
| Proprietary Name: | EVA (Enhanced Visual Assessment) System |
|--------------------------|-----------------------------------------|
| Common Name: | Colposcope |
| FDA Classification Name: | 21 CFR 884.1630; Colposcope |
| FDA Classification: | Class II, Product Code: HEX |
#### 4. Predicate Device
The EVA System is substantially equivalent to the following device:
| Manufacturer | Device | 510(k) | Date Cleared |
|-----------------------|------------------------------------|---------|---------------|
| MedGyn Products, Inc. | MedGyn Digital<br>Video Colposcope | K122973 | June 12, 2013 |
#### ട. Device Description
The EVA (Enhanced Visual Assessment) System is intended to provide magnified viewing of the vagina, cervix and external genitalia in order to aid in diagnosing abnormalities and
{5}------------------------------------------------
selecting areas for biopsy. The EVA System is a handheld device in which a smartphone is fitted and consists of a colposcope body, lens, smartphone upper and lower brackets, LED electronics, CervDx app for image capture, and online portal for image transfer and remote viewing. The device also includes the strap, carrier case, and instruction manual. During colposcopy, the operator may mount the EVA System on a tripod for hands free imaging, turn on the LED light for improved viewing, see a magnified view of the vagina and cervix on a smartphone screen resting at the back of the device, annotate and save images on the CervDx app, and upload images to the portal. The EVA System is compatible with Motorola G2 and Motorola G3 smartphones.
The EVA System is capable of working for ten hours before recharging. The EVA System weighs about 0.5 kg and is light enough that the device can be operated for 10 minutes at a time without any fatiguing.
The device operates at a working distance of 22.5 - 42.5 cm with an optical magnification up to 4.0x and ultimately up to 16x, with digital zoom.
#### 6. Indications for Use
The EVA (Enhanced Visual Assessment) System is intended to provide magnified viewing of the vagina, cervix and external genitalia in order to aid in diagnosing abnormalities and selecting areas for biopsy. The device is intended for use in hospitals, doctor's offices, and remote and rural clinics.
#### Predicate Comparison 7.
The following table compares the EVA System to the predicate device with respect to indications for use, technological characteristics, and materials:
| Device & Predicate Device(s): | Subject Device: | Predicate Device: |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | EVA System | MedGyn Digital Video Colposcope |
| | K161871 | K122973 |
| Indications for Use | The EVA (Enhanced Visual<br>Assessment) System is intended to<br>provide magnified viewing of the<br>vagina, cervix and external genitalia in<br>order to aid in diagnosing abnormalities | Medgyn's digital video colposcope<br>is intended for magnified viewing<br>of the tissues of the vagina, cervix<br>and external genitalia in order to<br>assist doctors in diagnosing |
| | and selecting areas for biopsy. The<br>device is intended for use in hospitals,<br>doctor's offices, and remote and rural<br>clinics. | abnormalities such as lesions or<br>cancer, and selecting areas for<br>biopsy. The images from the digital<br>video colposcope are to be viewed<br>on a color monitor. The digital<br>video colposcope is intended for |
| | | use in hospitals, clinics, and<br>doctor's offices. |
| Device Characteristics | | |
| Working Distance | 225-425mm | 200-300mm |
| Focusing Mechanism | Manual | Autofocus, manual |
| Magnification | 225:16x<br>425: 15x | 1-40x |
| Digital Magnification | 1x-4x | 1-32x (model 106A: 1-40x) |
| Optical magnification | 225: 4.0x<br>425: 3.8x | 1-18x (model 106A: 1-36x) |
| Depth of Field | 17mm (at 225mm) - 34mm (at 425mm) | 5-120mm |
| Field of View | 45mm (at 225mm) – 100mm (at<br>425mm) | 1x: ≥60mm<br>32x: ≥6mm<br>At working distance 200mm<br>1x: 195mm (52°)<br>32x: 6.98mm (2°) |
| On-axis Spatial Resolution | 11.78 line-pairs/mm | 11.31 line-pairs/mm |
| On-axis Angular Resolution | 0.022° | 0.02534° |
| Distortion | ≤2.5% | 2.49% |
| Light Source | 3W 6500k star LED | Double circular LED group |
| Color | Digital green filter, polarizing/glare<br>reducing filter | 3 Grade filter function |
| Image Output | Image capture, cloud patient recording<br>and image exporting | S-Video |
| Image Freeze | Full image capturing support | Image Freeze Control |
| Illuminance | ≥1000lux at working distance 425mm | When working distance is 300mm,<br>≥ 1600 lx<br>When working distance is 200mm |
| | | $\ge$ 2000 1x |
| Controls | Interactive GUI mobile application<br>Timers – Acetic Acid Timer | "Multi-function" button control<br>Timers- Magnification & Acetic<br>Acid Timer |
| Materials | PU – Biresin RG53 FR<br>6061 Aluminum<br>5052 Aluminum<br>Stainless Steel | Not Available |
{6}------------------------------------------------
{7}------------------------------------------------
The EVA System has the same intended use but different technological characteristics compared to the predicate device. The differences in technological characteristics do not raise different questions of safety and effectiveness.
#### 8. Non-Clinical Performance Testing
Non-clinical tests were conducted to verify that the EVA System met all design specifications and is substantially equivalent to the predicate. The test results demonstrated that the proposed device complies with the following standards:
- AAMI/ANSI 60601-1 (2012), Medical Electrical Equipment Part 1: General ● Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005, Mod).
- IEC 60601-1-2 (Edition 3.0, 2007), Medical Electrical Equipment Part 1-2: General . Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests.
- ISO 8600-3:1997 Optics and Optical instruments - Medical endoscopes and endoscopic accessories - Part 3: Determination of field of view and direction of view of endoscopes with optics.
- . ISO 8600-5:2005 Optics and photonics - Medical endoscopes and endotherapy devices Part 5: Determination of optical resolution of rigid endoscopes with optics.
- The EVA System was further tested for the following:
- Power Consumption O
- Battery Lifetime O
- Image Distortion o
- Cervical Model Image Quality O
- Image Pairs Evaluation O
- Working Distance о
{8}------------------------------------------------
- o Illuminance
- Validation of the device software was performed according to the clause 14 of IEC . 60601-1 (third edition) Software requirements, IEC 62304:2004 to IEC 62304-2006/AC 2008, Medical device, and Software life-cycle processes standards.
- Usability testing was conducted according to the IEC 62366-1:2015 standard and FDA ● guidelines for Applying Human Factors and Usability Engineering to Medical Devices.
#### Clinical Performance Data 9.
Not Applicable
#### 10. Conclusion
Based on the comparison and analysis above, the EVA System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.