One Drop Blood Glucose Monitoring System

K161834 · Agamatrix, Inc. · NBW · Nov 30, 2016 · Clinical Chemistry

Device Facts

Record IDK161834
Device NameOne Drop Blood Glucose Monitoring System
ApplicantAgamatrix, Inc.
Product CodeNBW · Clinical Chemistry
Decision DateNov 30, 2016
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Intended Use

The One Drop Blood Glucose Monitoring System is intended for the quantitative measurement of blood glucose (sugar) levels in fresh capillary whole blood samples drawn from the fingertip. It is intended to be used by a single patient and should not be shared. The One Drop Blood Glucose Monitoring System is intended for self-testing outside the body (In vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The One Drop Blood Glucose Monitoring System is not for the diagnosis of, or screening for diabetes, and is not intended for use with neonates. The One Drop Blood Glucose Test Strips are for use with the One Drop Blood Glucose Meter to quantitatively measure blood glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.

Device Story

System measures glucose concentration in capillary blood; user inserts test strip into meter; blood applied to strip tip; chemical reaction generates electrical current; meter measures current magnitude to determine glucose concentration. Used at home by patients with diabetes for self-monitoring. Meter displays results in mg/dL. Data transmitted via Bluetooth to smart devices (iOS/Android) for tracking. Benefits include convenient, portable monitoring of diabetes control effectiveness.

Clinical Evidence

No clinical data provided; substantial equivalence based on identical fundamental scientific technology and performance specifications to the predicate device.

Technological Characteristics

Glucose oxidase electrochemical biosensor; non-coding; powered by two CR2032 3V lithium batteries. Dimensions: 2.76" x 0.47"; weight 20g. Housing: ABS polymer with aluminum plating. Connectivity: Bluetooth for data transfer to smart devices. Measurement range: 20-600 mg/dL. Operating conditions: 50-104°F, 10-90% humidity, up to 10,000 ft altitude.

Indications for Use

Indicated for single-patient use for the quantitative measurement of glucose in fresh capillary whole blood from the fingertip. Intended for use by people with diabetes mellitus to monitor glucose levels.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the HHS logo, which consists of a stylized caduceus with three human profiles facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the logo. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 AGAMATRIX, INC. DAVID OLSEN VP QUALITY & REGULATORY AFFAIRS 7C RAYMOND AVE. SALEM NH 03079 November 30, 2016 Re: K161834 Trade/Device Name: One Drop Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: NBW, CGA Dated: November 01, 2016 Received: November 02, 2016 Dear Mr. David Olsen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Katherine Serrano -S For : Courtney H. Lias, Ph.d. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K161834 Device Name One Drop Blood Glucose Monitoring System #### Indications for Use (Describe) The One Drop Blood Glucose Monitoring System is intended for the quantitative measurement of blood glucose (sugar) levels in fresh capillary whole blood samples drawn from the fingertip. It is intended to be used by a single patient and should not be shared. The One Drop Blood Glucose Monitoring System is intended for self-testing outside the body (In vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The One Drop Blood Glucose Monitoring System is not for the diagnosis of, or diabetes, and is not intended for use with neonates. The One Drop Blood Glucose Test Strips: The One Drop Blood Glucose Test Strips are for use with the One Drop Blood Glucose Meter to quantitatively measure blood glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| Prescription Use (Part 21 CFR 801 Subpart D) |X | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510 (k) Section 15 # 510(k) Summary Rev. 092716 This summary of 510(k) substantial equivalence is being submitted in accordance with the requirements of 21 CFR 807.92. - 1. Prepared: June 29, 2016 - 2. Submitter: AgaMatrix, Inc. - 3. Address: 7C Raymond Ave. Salem, NH 03079 Phone: (603) 328-6000 - 4. Contact: David M. Olsen VP Quality Assurance / Regulatory Affairs dolsen@agamatrix.com Phone: 603-328-6049 Fax: 603-836-4025 - 5. Device Name: Trade/Proprietary Name: One Drop Blood Glucose Monitoring System Common Name: Glucose Test System #### 6. Device Classification: | Product Code | Classification | Regulation | Pan | |--------------------------------------------------------|----------------|-----------------|------------------------| | CGA – glucose Oxidase | Class II | 21 CFR 862.1345 | 75, Clinical Chemistry | | NBW – system, test, blood<br>glucose, over the counter | Class II | 21 CFR 862.1345 | 75, Clinical Chemistry | #### 7. Predicate Device: AgaMatrix Jazz Wireless 2 Blood Glucose Monitoring System, 510(k) K152365 {4}------------------------------------------------ #### 8. Device Description: The One Drop Blood Glucose Monitoring System (BGMS) is intended for the quantitative measurement of blood glucose levels in fresh capillary whole blood samples drawn from the fingertip. It is intended to be used by a single patient and should not be shared. The One Drop BGMS is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The System is not for the diagnosis of, or screening for diabetes, and is not intended for use with neonates. The system measures the concentration of glucose in a drop of blood obtained from the fingertip. The test strip is inserted into the meter and the blood is then introduced to the tip of the test strip. The blood reacts with chemicals on the test strip and this reaction generates an electrical current which is measured by the meter. The magnitude of the current is related to the blood glucose concentration. The fundamental scientific technology and the electronics that sends the electrical signals to the assay algorithm are all the same as the predicate device AgaMatrix Jazz Wireless 2 BGMS (K152365). The unit of measure displayed on the meter screen is fixed in mg/dL and cannot be modified by the user. The One Drop Blood Glucose Monitoring System (BGMS) consists of: - 1 One Drop Blood Glucose Meter ● - 1 One Drop Owner's Guide ● - 1 Carry Case - 1 Lancing Device - 10 Lancets - 100 One Drop Glucose Test Strips ● - AgaMatrix Glucose Control Solution (available separately) ● #### 9. Intended Use & Indications for Use: #### 9.1. Intended Use: Blood Glucose Monitoring #### 9.2. Indications for Use: The One Drop Blood Glucose Monitoring System is intended for the quantitative measurement of blood glucose (sugar) levels in fresh capillary whole blood samples drawn from the fingertip. It is intended to be used by a single patient and should not be shared. The One Drop Blood Glucose Monitoring System is intended for self-testing outside the body (In vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The One Drop Blood Glucose Monitoring System is not for the diagnosis of, or screening for diabetes, and is not intended for use with neonates. {5}------------------------------------------------ The One Drop Blood Glucose Test Strips are for use with the One Drop Blood Glucose Meter to quantitatively measure blood glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. | Similarities | | | |----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Item | Predicate Device<br>AgaMatrix Jazz Wireless 2<br>Blood Glucose Monitoring<br>System<br>(K152365) | Candidate Device<br>One Drop Blood Glucose<br>Monitoring System | | Intended<br>Use/Indications<br>for Use | The AgaMatrix Jazz Wireless 2 Blood<br>Glucose Monitoring System is<br>intended for the quantitative<br>measurement of blood glucose levels<br>in fresh capillary whole blood<br>samples drawn from the fingertip. It is<br>intended to be used by a single patient<br>and should not be used for testing<br>multiple patients. The AgaMatrix Jazz<br>Wireless 2 Blood Glucose Monitoring<br>System is intended for self-testing<br>outside the body (In vitro diagnostic<br>use) by people with diabetes at home<br>as an aid to monitor the effectiveness<br>of diabetes control. The AgaMatrix<br>Jazz Wireless 2 Blood Glucose<br>Monitoring System is not for the<br>diagnosis of or screening for diabetes,<br>and is not intended for use with<br>neonates.<br>The AgaMatrix Jazz Wireless 2 Blood<br>Glucose Test Strips are for use with the<br>AgaMatrix Jazz Wireless 2 Glucose<br>Meter to quantitatively measure blood<br>glucose (sugar) in fresh capillary whole<br>blood samples drawn from the fingertips. | The One Drop Blood Glucose Monitoring<br>System is intended for the quantitative<br>measurement of blood glucose levels in<br>fresh capillary whole blood samples drawn<br>from the fingertip. It is intended to be used<br>by a single patient and should not be used<br>for testing multiple patients. The One<br>Drop Blood Glucose Monitoring System is<br>intended for self-testing outside the body<br>(In vitro diagnostic use) by people with<br>diabetes at home as an aid to monitor the<br>effectiveness of diabetes control. The One<br>Drop Blood Glucose Monitoring System is<br>not for the diagnosis of or screening for<br>diabetes, and is not intended for use with<br>neonates.<br>The One Drop Blood Glucose Test Strips<br>are for use with the One Drop Blood<br>Glucose Meter to quantitatively measure<br>blood glucose (sugar) in fresh capillary<br>whole blood samples drawn from the<br>fingertips.<br>(Same as Predicate) | | Similarities | | | | Item | Predicate Device<br>AgaMatrix Jazz Wireless 2<br>Blood Glucose Monitoring<br>System<br>(K152365) | Candidate Device<br>One Drop Blood Glucose<br>Monitoring System | | Sample Type | Capillary whole blood | Same | | Sample Sites | Finger tip | Same | | Sample Volume | 0.5 µL | Same | | Average Test Time | 5 seconds | Same | | Test range | 20 - 600 mg/dL | Same | | Operating<br>Temperature | 50° - 104°F | Same | | Measurement<br>Range | 20 to 600 mg/dL | Same | | Calibration | Plasma Equivalent | Same | | Control Levels | Two Levels | Same | | Operating<br>Humidity | 10% to 90% | Same | | Hematocrit Range | 20 - 60% | Same | | Altitude Study | Up to 10,000 feet | Same | | Similarities | | | | Item | Predicate Device<br>AgaMatrix Jazz Wireless 2<br>Blood Glucose Monitoring<br>System<br>(K152365) | Candidate Device<br>One Drop Blood Glucose<br>Monitoring System | | Power Source | (2) replaceable CR2032 3 volt,<br>lithium batteries | Same | | Connectivity | Meter can send BGM results via<br>Bluetooth to smart devices using the<br>appropriate iOS or Android software<br>application. | Same | | Detection method | Dynamic Electrochemistry | Same | | Enzyme | Glucose Oxidase | Same | | Calibration<br>Coding | Non-Coding | Same | ## 10. Comparison to Predicate {6}------------------------------------------------ {7}------------------------------------------------ {8}------------------------------------------------ | Differences | | | |-----------------------|------------------------------------------------------------------------------|-------------------------------------------------------| | Item | Predicate Device | Candidate Device | | | AgaMatrix Jazz Wireless 2<br>Blood Glucose Monitoring<br>System<br>(K152365) | One Drop Blood Glucose Monitoring<br>System | | Meter size | L-2.56 inches, H-0.39 inches | L-2.76inches, H-0.47inches | | Weight | 18g | 20g | | Meter Outer<br>Finish | ABS polymer blend with color<br>pigment added to polymer | ABS polymer blend covered with an<br>aluminum plating | ### 11. Technological Characteristics: The device modifications described in this premarket notification do not affect or alter the fundamental scientific technology of the predicate device, the AgaMatrix Jazz Wireless 2 Blood Glucose Monitor. #### 12. Intended Use: The modifications described in this premarket notification do not alter or change the intended use of the modified device from its predicate the AgaMatrix Jazz Wireless 2 Blood Glucose Monitor.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...