The O2Vent T is a removable medical device that is fitted in the patient's mouth and is intended to reduce or alleviate snoring and mild to moderate obstructive sleep apnea (OSA). The device is indicated for use during sleep to aid in the treatment of these conditions.
Device Story
O2Vent T is a custom-fitted, removable intraoral mandibular advancement device used at home or in sleep labs. It consists of an upper tray (medical-grade titanium) and a lower tray (medical-grade polymer) connected by a screw-based mechanism. The device repositions the mandible forward to increase pharyngeal space, preventing soft tissue collapse and airway obstruction. A dedicated breathing port in the upper tray allows for oral breathing. Clinicians or patients use an adjustment key to titrate mandibular protrusion, eliminating the need for new impressions or laboratory procedures required by the OVENT predicate. The device is prescribed by dentists to treat snoring and mild-to-moderate OSA, benefiting patients by improving airway patency during sleep.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, risk assessment, and design comparison to legally marketed predicate devices.
Technological Characteristics
Custom-fitted intraoral appliance. Materials: medical-grade titanium, medical-grade polymers, 316 stainless steel. Mechanism: screw-based mandibular advancement. Features: dedicated breathing port, two-piece tray design. Non-sterile. Biocompatibility established via similarity to predicate materials.
Indications for Use
Indicated for adult patients 18 years and older with snoring and mild to moderate obstructive sleep apnea (OSA) for use during sleep.
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized human figure with three faces in profile, representing the department's focus on people. The figure is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 23, 2016
Oventus Manufacturing Pty Ltd % M.W. Anderson, Ph.D. Senior Principal Advisor Regulatory and Clinical Research Institute, Inc. 5353 Wayzata Boulevard, #505 Minneapolis, Minnesota 55416
Re: K161832
Trade/Device Name: O2 Vent T Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices For Snoring And Intraoral Devices For Snoring And Obstructive Sleep Apnea Regulatory Class: Class II Product Code: LRK Dated: August 23, 2016 Received: August 25, 2016
Dear Dr. M.W. Anderson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{2}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K161832
Device Name O2VENT T
Indications for Use (Describe)
The O2Vent T is a removable medical device that is fitted in the patient's mouth and is intended to reduce or alleviate snoring and mild to moderate obstructive sleep apnea (OSA). The device is indicated for use during sleep to aid in the treatment of these conditions.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Oventus Medical. The logo features a stylized blue circle with a wave-like design inside, suggesting a connection to breathing or airflow. Above the circle are several blue dots of varying sizes, arranged in a semi-circular pattern. Below the circle, the word "OVENTUS" is written in a stylized font.
## 510(k) SUMMARY FOR O₂VENT T
prepared in accordance to 21CFR 807.92
| Date Prepared | 23rd September 2016 |
|---------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510 (k) Owner Details | |
| Name: | Oventus Manufacturing Pty Ltd |
| Address | 1 Swann Road, Indooroopilly, QLD 4068, Australia. |
| Phone: | +61 7 3180 3196 |
| Name of Contact Person | Hemangi Malde |
| Device Trade name | O2Vent T |
| Common Name | Oral Appliance - Anti snoring device |
| Classification | Class II |
| Classification Name | Intraoral devices for snoring and intraoral devices for snoring<br>and obstructive sleep apnea. |
| Classification Code | LRK |
| Classification Regulation | 21 CFR 872.5570 |
| Device Model Number(s) | O2VTDL, O2VTOV |
| Predicate Device | OVENT – Oventus Manufacturing Pty Ltd<br>TAP-III – Airway Management |
| Device Description | The O2Vent T is a removable medical device that is fitted in the<br>patient's mouth and is intended to reduce or alleviate snoring<br>and mild to moderate obstructive sleep apnea (OSA).<br>The O2Vent T is made up of three parts:<br>1. The Upper Tray fitted over the upper teeth, with the breathing<br>port at the front leading to the airways on each side to the rear of<br>the appliance, is made from medical grade titanium<br>2. The Lower Tray fitted over the lower teeth, positions the jaw<br>for mandibular advancement.<br>The Upper and Lower Trays have a "landing area" customized to<br>the patient's teeth and made from medical grade polymers. |
| | 3. The Upper and Lower Trays are connected together with a<br>connector. A hook mechanism attached to the Upper Tray fits<br>into a socket attached to the Lower Tray and positions the lower<br>jaw forward, preventing the soft tissue of the throat from<br>collapsing and obstructing the airway. An Adjustment Key for<br>the <i>O</i> 2 <i>Vent T</i> allows the patient and their Clinician to adjust the<br>protrusion of the lower jaw to the most effective and<br>comfortable position. |
| Intended Use of the Device | The <i>O</i> 2 <i>Vent T</i> is a removable medical device that is fitted in the<br>patient's mouth and is intended to reduce or alleviate snoring<br>and mild to moderate obstructive sleep apnea (OSA). The device<br>is indicated for use during sleep to aid in the treatment of these<br>conditions. |
| Target Population | Adult patients 18 years and older. |
| Environment of Use | Home Use and Sleep laboratories. |
| Summary of Comparison<br>of Technological<br>Characteristics of the<br>Device to Predicate Device | As mentioned in the following pages. |
{5}------------------------------------------------
{6}------------------------------------------------
| Feature | O2Vent T | OVENT | TAP-III |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510k Number | Not yet received. | K160234 | K062951 |
| Device Photo | Image: O2Vent T device | Image: OVENT device | Image: TAP-III device |
| Intended Use | The O2Vent T is a<br>removable medical<br>device that is fitted in<br>the patient's mouth<br>and is intended to<br>reduce or alleviate<br>snoring and mild to<br>moderate obstructive<br>sleep apnea (OSA).<br>The device is<br>indicated for use<br>during sleep to aid in<br>the treatment of these<br>conditions. | The OVENT is a<br>removable medical<br>device that is fitted in<br>the patient's mouth and<br>is intended to reduce or<br>alleviate snoring and<br>mild to moderate<br>obstructive sleep apnea,<br>OSA. The device is<br>indicated for use during<br>sleep to aid in the<br>treatment of these<br>conditions. | To reduce or alleviate<br>night time snoring and<br>mild to moderate<br>obstructive sleep apnea<br>(OSA). |
| Product Codes | LRK | LRK | LRK |
| Regulation | 21CFR 872.5570 | 21CFR 872.5570 | 21CFR 872.5570 |
| Use of device | Removable intraoral<br>device. Multiple use.<br>Prescription use only. | Removable intraoral<br>device. Multiple use.<br>Prescription use only. | Removable intraoral<br>device. Multiple use.<br>Prescription use only. |
| Target Population | People over 18 years<br>of age who snore<br>and/or have sleep<br>apnea. | People over 18 years<br>of age who snore<br>and/or have sleep<br>apnea. | People over 18 years<br>of age who snore<br>and/or have sleep<br>apnea. |
| Environment of<br>Use | To be used in the<br>patient's home or in<br>sleep laboratories. | To be used in the<br>patient's home or in<br>sleep laboratories. | To be used in the<br>patient's home or in<br>sleep laboratories. |
| Cleaning<br>Instructions | Should be cleaned<br>daily in soap and<br>water. | Should be cleaned<br>daily in soap and<br>water. | Should be cleaned<br>daily in soap and<br>water. |
| | Can be periodically<br>cleaned in an<br>ultrasonic cleaner with<br>an effervescent<br>denture cleaning<br>tablet. | Can be periodically<br>cleaned in an<br>ultrasonic cleaner with<br>an effervescent<br>denture cleaning<br>tablet. | Can be periodically<br>cleaned in an<br>ultrasonic cleaner with<br>an effervescent<br>denture cleaning<br>tablet. |
| Feature | O₂Vent T | OVENT | TAP-III |
| Device<br>Functionality | Repositions the lower<br>jaw forward.<br>Acts by increasing the<br>pharyngeal space to<br>improve the patient's<br>ability to exchange air.<br>Does not have a<br>lingual flange for<br>holding the mandible<br>forward.<br>Permits patient to<br>breathe through their<br>mouth2.<br>Retains the top and<br>bottom teeth using<br>rigid trays. | Repositions the lower<br>jaw forward.<br>Acts by increasing the<br>pharyngeal space to<br>improve the patient's<br>ability to exchange air.<br>Has a lingual flange<br>for holding the<br>mandible forward1.<br>Permits patient to<br>breathe through their<br>mouth2.<br>Retains the top and<br>bottom teeth using<br>rigid trays. | Repositions the lower<br>jaw forward.<br>Acts by increasing the<br>pharyngeal space to<br>improve the patient's<br>ability to exchange air.<br>Does not have a<br>lingual flange for<br>holding the mandible<br>forward.<br>Permits patient to<br>breathe through their<br>mouth2.<br>Retains the top and<br>bottom teeth using<br>rigid trays. |
| Device Design | Customized for each<br>patient in a dental<br>laboratory. | Customized made for<br>each patient in a dental<br>laboratory. | Customized made for<br>each patient in a dental<br>laboratory. |
| Adjustment | Can be adjusted by the<br>clinician and patient4. | Cannot be adjusted by<br>the clinician or<br>patient4. | Can be adjusted by the<br>clinician and patient4. |
| Supplied<br>Sterile/Non-Sterile | Non-Sterile | Non-sterile | Non-sterile |
| Materials Use - 1 | Medical grade metal<br>used5. | Medical grade metal<br>used5. | Medical grade metals<br>used5. |
| Materials Use - 2 | Dental plastic<br>laminates and acrylics<br>used for upper and<br>lower trays which is in<br>contact with the<br>patient's teeth. | Dental plastic<br>laminates and acrylics<br>used for upper and<br>lower trays which is in<br>contact with the<br>patient's teeth. | Dental plastic<br>laminates and acrylics<br>used for upper and<br>lower trays which is in<br>contact with the<br>patient's teeth. |
| Biocompatibility* | Not performed as the<br>materials are identical<br>as in the OVENT. | Passes Part 5 and Part<br>10 of ISO10993. | Passes Part 5 and Part<br>10 of ISO10993. |
| Mechanical safety | Can withstand<br>mechanical forces<br>without significant<br>degradation. | Can withstand<br>mechanical forces<br>without significant<br>degradation. | Can withstand<br>mechanical forces<br>without significant<br>degradation. |
{7}------------------------------------------------
* This is assumed as the devices are FDA 510(k) cleared and are currently on the market. Only the OVENT device was tested as part of this submission. Biocompatibility was not performed on the O₂Vent T device as the materials used in the device were similar to that of OVENT.
{8}------------------------------------------------
Note (1) The OVENT has a choice of lingual flange (to be decided by the clinician) although for most cases adequate retention (in advanced position) is achieved by the molded polymer insert.
Note (2) The O2Vent T and OVENT both have a dedicated breathing port (delivering air to the rear of the mouth) to allow the user to breathe through their mouth if they want, the TAP-III allows users to open their mouth to a degree.
Note (4) The OVENT requires a new insert to be made for mandibular adjustment, the O2Vent T and TAP device can be adjusted by the clinician or patient.
Note (5) The O2Vent T and OVENT uses titanium and the TAP-III uses stainless steel.
| Summary of Non Clinical Tests | The non-clinical testing included risk assessment of the physical<br>properties of the O2Vent T and its ability to achieve its intended<br>use. The only new material that was incorporated in the O2Vent<br>T is the connector assembly made from special grade 316<br>stainless steel. The material and design specifications of the<br>connector assembly is already used in as the predicate TAP-III<br>device. No other risks are introduced in the device. The O2Vent<br>T met the specifications as set.<br>A biocompatibility assessment of the device was performed. The<br>purpose of the biocompatibility assessment was to ensure that<br>biocompatibility had been established for the device. No new<br>materials are being used in the device; all materials are already<br>used in the predicate device (OVENT). No new risks are<br>introduced in the new device apart from the connector assembly,<br>which are not present in the predicate OVENT device, however<br>is present in the TAP-III predicate device and the new material<br>has no patient tissue contact in the new device.<br>The device is deemed to be biocompatible, based on the<br>similarity of the materials of constructions to the predicate<br>devices (OVENT & TAP-III). |
|-----------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summary of Clinical Tests | Human clinical studies were not deemed necessary to evaluate<br>the substantial equivalence of the O2Vent T as the O2Vent T<br>does not:<br>• Use designs dissimilar from the predicate device and other<br>previously-cleared devices under a 510(k);<br>• Use new technologies different from legally-marketed intra-<br>mandibular repositioning devices for snoring and/or obstructive<br>sleep apnea;<br>• Deviate from the indications for use identified in the predicate<br>devices: the OVENT and TAP-III.<br><br>In lieu of human clinical testing, the risks and mitigating<br>controls associated with the use of mandibular repositioning<br>devices, as identified by the FDA, have been addressed in the |
| | |
| | "Risk Assessment". The function of mandibular repositioning<br>devices required that the prescribing dentist be aware of the<br>potential for soreness caused by mandibular advancement.<br>Management of these risks is achieved by advising the patients<br>and dentists in the "Instructions for Use". The contraindications,<br>warnings, precautions, storage, fitting and adjustment directions<br>are written to avoid potential problems from arising or persisting<br>caused by Mandibular Advancement Devices. |
| Differences between the<br>proposed and predicate<br>device(s). | As discussed in the comparison chart above, the main functional<br>difference between the proposed O2Vent T and predicate device<br>OVENT is the method of adjustment. O2Vent T allows protrusion<br>of the mandible from the original treatment position via a screw<br>mechanism without requiring additional impressions or laboratory<br>procedures as required for OVENT. The adjustments can be made<br>by the clinician or patient for the O2Vent T device, however it is<br>not possible in case of the OVENT device.<br>Additionally, in order to facilitate the ability to titrate the device,<br>the O2Vent T has separate upper and lower trays connected via the<br>screw mechanism, while the OVENT has a monobloc design.<br>No new risks are introduced in the new device apart from the<br>connector assembly and separate trays, which are not present in the<br>predicate OVENT device, however are present in the TAP-III<br>predicate device.<br>The treatment outcomes for O2Vent T and OVENT are expected to<br>be equivalent as the final levels of mandibular protrusion and the<br>provision for oral breathing via the customized airway is same for<br>each device.<br>Hence the differences mentioned are unlikely to adversely or<br>otherwise affect the expected clinical outcomes for patients using<br>the O2Vent T device as compared to those using the OVENT, thus<br>not affecting the substantial equivalency of the proposed device. |
| Conclusion on Non Clinical<br>and Clinical Tests | The O2Vent T is considered to be substantially equivalent to the<br>predicate devices based on the following:<br>• It has the same intended use and is indicated for the same user<br>population.<br>• It has equivalent technological characteristics to the predicates.<br>• The device is as safe, as effective and performs as well as or<br>better than the legally marked devices identified above. |
{9}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.