← Product Code [GEI](/productcode/GEI) · K161804

# LigaSure Retractable L-Hook Laparoscopic Sealer/Divider (K161804)

_Covidien · GEI · Sep 16, 2016 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K161804

## Device Facts

- **Applicant:** Covidien
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Sep 16, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The LigaSure Retractable L-Hook Laparoscopic Sealer/Divider is a 5mm bipolar/monopolar electrosurgical instrument intended for use in minimally invasive surgical procedures where ligation of vessels, tissue bundles, and lymphatics is desired. The LigaSure Sealer/Divider can be used on vessels (arteries and veins) up to and including 7mm in diameter. The monopolar L-Hook can be used to dissect through tissue planes and to create enterotomies or gastrotomies. It is indicated for use in general surgery and surgical specialties as urologic, vascular, thoracic, colorectal, bariatric, and gynecologic. Procedures may include, but are not limited to, gastric bypass, hysterectorny, Nissen fundoplication, colectomy, cholecystectomy, adhesiolysis, oophorectomy, etc. The LigaSure System has not been shown to be effective for tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures.

## Device Story

5mm sterile, single-use, hand-held bipolar/monopolar electrosurgical instrument; used with compatible Covidien generators. Device features fine curved jaw for vessel sealing and extendable/retractable monopolar L-Hook for tissue dissection/enterotomies. Operated by surgeons in minimally invasive procedures (OR setting). Generator delivers RF energy to tissue for controlled fusion; minimizes sticking, charring, and thermal spread. Surgeon controls energy activation via device; output provides hemostasis and tissue division. Benefits include permanent tissue fusion and versatile functionality (sealing and dissection) in one instrument.

## Clinical Evidence

No clinical data. Evidence consists of bench testing (biocompatibility per ISO 10993-1, electrical safety/EMC per IEC 60601 series) and pre-clinical animal studies (porcine models). Studies evaluated ex-vivo renal burst pressure, lymphatic duct sealing, acute hemostasis, lateral thermal spread, and enterotomy formation compared to predicate devices. Usability engineering performed per IEC 62366.

## Technological Characteristics

5mm diameter shaft (37cm/44cm lengths); bipolar/monopolar electrosurgical energy. Materials: stainless steel, ceramic, ETFE, PTFE, fiberglass, polyamide, polyester, polypthalamide, HDPE. Features extendable/retractable L-Hook. Compatible with Covidien electrosurgical generators. Complies with IEC 60601-1, 60601-2-2, 60601-2-18, and 60601-1-2. Sterilization: EO.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- LigaSure™ Advance Monopolar Tip Sealer/Divider, Pistol Gri ([K150835](/device/K150835.md), [K063195](/device/K063195.md))
- LigaSure™ 5mm Maryland Jaw Sealer/Divider ([K141153](/device/K141153.md), [K133338](/device/K133338.md))
- Valleylab Laparoscopic Handswitch/Electrodes ([K904560](/device/K904560.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 16, 2016

Covidien Ms. Dawn D. Tindall Senior Regulatory Affairs Product Specialist 5920 Longbow Drive Boulder, Colorado 80301

Re: K161804 Trade/Device Name: LigaSure™ Retractable L-Hook Laparoscopic Sealer Divider (LF5637. LF5644) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: September 1, 2016 Received: September 2, 2016

Dear Ms. Tindall:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR Part 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Christopher J. Ronk -S

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K161804

Device Name

LigaSure™ Retractable L-Hook Laparoscopic Sealer Divider (LF5637, LF5644)

#### Indications for Use (Describe)

The LigaSure Retractable L-Hook Laparoscopic Sealer/Divider is a 5mm bipolar/monopolar electrosurgical instrument intended for use in minimally invasive surgical procedures where ligation of vessels, tissue bundles, and lymphatics is desired. The LigaSure Sealer/Divider can be used on vessels (arteries and veins) up to and including 7mm in diameter. The monopolar L-Hook can be used to dissect through tissue planes and to create enterotomies or gastrotomies.

It is indicated for use in general surgery and surgical specialties as urologic, vascular, thoracic, colorectal, bariatric, and gynecologic. Procedures may include, but are not limited to, gastric bypass, hysterectorny, Nissen fundoplication, colectomy, cholecystectomy, adhesiolysis, oophorectomy, etc.

The LigaSure System has not been shown to be effective for tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D) | <div> </div> |
|----------------------------------------------|--------------|
| Over-The-Counter Use (21 CFR 801 Subpart C)  | <div> </div> |

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the word "COVIDIEN" in a sans-serif font. The word is in a dark blue color. To the left of the word is a blue and white square logo. 

510(k) Summary Date summary prepared: June 28, 2016

# 510(k) Submitter/Holder

Covidien 5920 Longbow Drive Boulder, CO 80301

# Contact

Dawn D. Tindall Sr. Regulatory Affairs Product Specialist Telephone: 303-581-7034 Fax: 303-530-6313 Email: dawn.d.tindall@medtronic.com

# Name of Device

| Trade Name:          | LigaSure™ Retractable L-Hook Laparoscopic Sealer/Divider                                          |
|----------------------|---------------------------------------------------------------------------------------------------|
| Catalog Numbers:     | LF5637 and LF5644                                                                                 |
| Common Name:         | Bipolar and Monopolar Electrosurgical Instrument                                                  |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories (21 CFR § 878.4400, Class II, GEI) |
| Predicate Devices:   | Primary                                                                                           |
| Trade Name:          | LigaSure™ Advance Monopolar Tip Sealer/Divider, Pistol Gri                                        |
| Common Name:         | Bipolar and Monopolar Electrosurgical Instrument                                                  |
| Catalog Number:      | LF5544                                                                                            |
| 510(k) Number:       | K150835 (cleared 10/15/15); K063195 (clearance 11/29/06)                                          |
| Manufacturer:        | Covidien                                                                                          |
| Recalls:             | This device has not been subject to a design-related recall                                       |
|                      | Secondary                                                                                         |
| Trade Name:          | LigaSure™ 5mm Maryland Jaw Sealer/Divider                                                         |
| Common Name:         | Bipolar and Monopolar Electrosurgical Instrument                                                  |
| Catalog Number:      | LF1723, LF1737, LF1744                                                                            |
| 510(k) Number:       | K141153 (cleared 08/14/14); K133338 (cleared 12/20/13)                                            |
| Manufacturer:        | Covidien                                                                                          |
| Recalls:             | This device has not been subject to a design-related recall                                       |
|                      | Secondary                                                                                         |
| Trade Name:          | Valleylab Laparoscopic Handswitch/Electrodes                                                      |
| Common Name:         | Monopolar Electrodes                                                                              |
| Catalog Number:      | E2773-36                                                                                          |
| 510(k) Number:       | K904560 (cleared 12/31/1990)                                                                      |
| Manufacturer:        | Covidien                                                                                          |
| Recalls:             | This device has not been subject to a design-related recall                                       |

{4}------------------------------------------------

No reference devices were used in this submission.

#### Device Description

The LigaSure™ Retractable L-Hook Laparoscopic Sealer/Divider is a 5mm sterile, single-use, hand-held bipolar electrosureical instrument designed for use with compatible Covidien generators that include vessel sealing capabilities to ligate (seal) and divide (cut) vessels, tissue bundles, and lymphatics during general procedures (as indicated) and monopolar capabilities to electrically dissect through tissue planes and to create openings in bowel (enterotomies) and stomach (gastrotomies). Covidien electrosurgical generators that include vessel sealing capabilities deliver precise energy through the device to tissue for a controlled tissue response to achieve complete and permanent tissue fusion while producing minimal sticking, charring, and thermal spread to adjacent tissue. The new device will be provided EO sterilized with a 5mm diameter shaft, in shaft lengths of 37 cm and 44 cm with a fine curved jaw featuring an extendable and retractable monopolar L-Hook. The proposed devices do not contain software.

# Indications for use:

The LigaSure™ Retractable L-Hook Laparoscopic Sealer/Divider is a 5mm bipolar/monopolar electrosurgical instrument intended for use in minimally invasive surgical procedures where ligation and division of vessels, tissue bundles, and lymphatics is desired. The LigaSure Sealer/Divider can be used on vessels (arteries and veins) up to and including 7mm in diameter. The monopolar L-Hook can be used to dissect through tissue planes and to create enterotomies or gastrotomies.

It is indicated for use in general surgery and such surgical specialties as urologic, vascular, thoracic, colorectal, bariatric, and gynecologic. Procedures may include, but are not limited to gastric bypass, hysterectomy, Nissen fundoplication, colectomy, cholecystectomy, adhesiolysis, oophorectomy, etc.

The LigaSure System has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures.

# Technological Characteristics

The subject device and the predicate devices are based on the following same technological elements:

- LigaSure Advance Bipolar
	- Electrical RF Bipolar Energy o
	- Pistol grip and momentary pistol trigger O
	- Integrated and independent cutting blade o
	- Sterilized, Single use o
	- Intended use for Advance Bipolar O
	- Generator Compatibility O
- Monopolar
	- Monopolar Energy O
	- Monopolar electrode O
	- Sterilized, single use O
	- Intended use for monopolar O
	- Generator Compatibility C

The following technological differences exist between the subject and predicate devices:

- . Mechanical Deployment/Retraction Feature

{5}------------------------------------------------

- Monopolar Electrode Shape ●
- Monopolar Power Settings
- Monopolar Operating Voltage ●
- Inline Energy Activation
- Shaft/Jaw Rotation

# Performance Data:

The following performance data were provided in support of the substantial equivalence determination.

# Biocompatibility

The biocompatibility evaluation for the LigaSure™ Retractable L-Hook Laparoscopic Sealer Divider devices was conducted in accordance with ISO 10993-1, "Biological Evaluation of Medical Devices-Part 1: Evaluation and Testing Within a Risk Management Process". The testing included the following:

- Cytotoxicity
- Sensitization .
- . Irritation
- . Systemic Toxicity
- o Hemolysis

The direct patient-contacting materials are: stainless steel, ceramic, and polymers including ETFE, PTFE, fiberglass, polyamide, polyester, polypthalamide, and HDPE.

# Electrical Safety and Electromagnetic Compatibility (EMC)

Electrical safety testing and EMC testing were conducted on the LigaSure™ Retractable L-Hook Laparoscopic Sealer Divider devices. The system complies with relevant portions of the IEC 60601-1, IEC 60601-2-2, and IEC 60601-2-18 standards for electrical safety and IEC 60601-1-2 standard for EMC.

# Mechanical / Functional Testing

Mechanical, electrical, and functional testing was carried out to verify that the new devices perform as expected and conform to requirements defined in related design inputs and subsequent product specifications.

# LigaSure Advanced Bipolar

# Ex-vivo Renal Burst Pressure

Ex-vivo burst pressure testing of excised fresh porcine renal arteries was conducted using the LigaSure™ Retractable L-Hook Laparoscopic Sealer Divider devices in comparison to the Maryland LF17XX predicate device. Testing utilized the ForceTriad and the FT10 generators.

# Ex-Vivo / Lymphatic

A pre-clinical study using a porcine model was conducted to evaluate the lymphatic duct sealing performance of the LigaSure™ Retractable L-Hook Laparoscopic Sealer Divider devices in comparison to the predicate Maryland LF17XX devices. Testing utilized the ForceTriad and the FT10 generators.

{6}------------------------------------------------

#### In-Vivo Acute

A pre-clinical study using a porcine model was conducted to verify acute hemostasis and lateral thermal spread with the subject devices and the predicate Maryland LF17XX device. Testing utilized the ForceTriad and the FT10 generators.

#### Monopolar

#### Ex-vivo Monopolar Thermal Effect

Monopolar ex-vivo testing evaluated thermal effect resulting from monopolar energy application across the range of power setting and modes on LigaSure™ Retractable L-Hook Laparoscopic Sealer Divider (LF5637/LF5644) devices in comparison to the predicate laparoscopic L-Hook (E2773-36) attached to a Force TriVerse (FT3000) pencil. Testing utilized the ForceTriad and the FT10 generators. Monopolar thermal effects were evaluated on three different tissue types with three power settings for all three energy modes.

#### In-Vivo Acute

Enterotomy formation was verified with the subject device and the predicate controls of LF5544 and L-hook E2773-36 in a pre-clinical study using a porcine model. Testing utilized the ForceTriad and the FT10 generators

#### Chronic Animal Study

A Chronic Hemostasis Porcine study was conducted to verify the performance of the LigaSure™ Retractable L-Hook Laparoscopic Sealer Divider devices on the ForceTriad™ and FT10™ generators when used on vessels 7mm or smaller and tissue bundles during general surgical procedures.

#### Human Factors and Usability

The usability engineering process applied to these devices was in compliance with the requirements of IEC 62366 "Medical devices—Application of usability engineering to medical devices". The process included analysis of user needs and potential use errors followed by testing to demonstrate that representative users can use the instruments safely and effectively.

#### Clinical Studies

This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence.

#### Summarv

Based on the pre-clinical performance as documented in the performance testing, the LigaSure™ Retractable L-Hook Laparoscopic Sealer/Divider was found to have a safety and effectiveness profile that is similar to the predicate devices.

#### Conclusions

The comparison of device characteristics and the review of the performance data support the conclusion that the LigaSure™ Retractable L-Hook Laparoscopic Sealer/Divider devices are substantially equivalent to the predicate devices. The devices have the same intended use for their common fundamental technologies. The proposed device has some intentional design differences that enable it to be well suited for its indicated uses. Testing has demonstrated that these differences do not raise any new questions of safety or effectiveness.

---

**Source:** [https://fda.innolitics.com/device/K161804](https://fda.innolitics.com/device/K161804)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
