The SLIT LAMP is an AC-powered slit lamp biomicroscope that is intended for use in examining the anterior eye segment, from the corneal epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma, which affect the structural properties of the anterior segment of the eye.
Device Story
The YF-100 Slit Lamp is an AC-powered biomicroscope used by clinicians to examine the anterior eye segment. The device features a high-intensity LED light source, binocular microscope, and mechanical positioning components (base plate, chin rest, fixation light). The clinician manually adjusts the slit width, aperture, and angle to shine a thin sheet of light into the patient's eye. The patient rests their chin and forehead on the support for stability. The clinician views the eye through the binocular eyepieces to obtain a stereoscopic, magnified view of eye structures. This visual information aids in the diagnosis of trauma or disease affecting the anterior segment. The device is intended for clinical use.
Clinical Evidence
Bench testing only. The device was evaluated for electrical safety (IEC 60601-1:2005), EMC compliance (IEC 60601-1-2), environmental tolerance (ISO 15004-1), and photobiological safety (ISO 15004-2, ISO 10939). All tests passed.
Technological Characteristics
AC-powered slit lamp biomicroscope; Galileo binocular converging microscope; LED light source (3V); manual zoom magnification (6.4x to 40x); mechanical base with chin rest; filters include heat absorption, red-free, and cobalt blue; powered by external adapter (MDS-060BAS12A) compliant with ANSI/AAMI ES 60601-1.
Indications for Use
Indicated for examination of the anterior eye segment (corneal epithelium to posterior capsule) to aid in the diagnosis of diseases or trauma affecting the structural properties of the anterior segment in patients requiring ophthalmic evaluation.
Regulatory Classification
Identification
An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.
Special Controls
*Classification.* Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
Predicate Devices
APPASAMY SLIT LAMP, model:A1A-11 AND A1A-12 SERIES (K082031)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 16, 2017
Chongqing Yeasn Science-Technology Co., Ltd. % Ms. Iris Fung Official Correspondent SGS-CTSC Standards Technical Services Co., Ltd. 198 Kezhu Road, Scientech Park Guangzhou Economic & Technology Development District Guangzhou, CN Guangdong
Re: K161764
Trade/Device Name: Slit Lamp, Model: YF-100 Regulation Number: 21 CFR 886.1850 Regulation Name: AC-powered slitlamp biomicroscope Regulatory Class: Class II Product Code: HJO Dated: December 3, 2016 Received: January 3, 2017
Dear Ms. Fung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Page 2 - Ms. Iris Fung
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Denise L. Hampton -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K161764
Device Name SLIT LAMP
Indications for Use (Describe)
The SLIT LAMP is an AC-powered slit lamp biomicroscope that is intended for use in examining the anterior eye segment, from the corneal epithelium to the posterior capsule. It is used to aid in the diagnosis of trauma, which affect the structural properties of the anterior segment of the eye.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
## Date of the summary prepared: February 10, 2017
## 1. Submitted Information
- � 510(k) Owner's Name: CHONGQING YEASN SCIENCE-TECHNOLOGY CO., LTD
- � Address: 5 Danlong Road, Nan An District, Chongqing Chongqing, CHINA 400060
- Phone:(86 23)62797666-8050 �
- Fax:(8623)62766200 �
- Contact Person: Bing Hu/Managing Director �
- Email: tech@yeasn.com �
#### 2. Application Correspondent:
- � Company Name: SGS-CSTC Standards Technical Services Co., Ltd.
- Address: 198 KEZHU Road, SCIENTECH Park Guangzhou Economic & �
Technology Development District, Guangzhou, Guangdong, CHINA
- Contact Person: Ms. Iris Fung �
- Tel: +86-20-32136908 �
- Email: Iris.Fung@sgs.com �
## 3. Subject Device Information
- Trade Name: SLIT LAMP, Model: YF-100 �
- Slit Lamp Common Name: �
- AC Powered Slit Lamp Bio-microscope � Classification name:
- Review Panel: Ophthalmic �
- HJO Product Code: �
{5}------------------------------------------------
- Regulation Class: II �
- 21 886.1850 � Regulation Number:
#### 4. Predicate Device Information
| Sponsor | APPASAMY ASSOCIATES |
|----------------------|-------------------------------------------------------|
| Device Name | APPASAMY SLIT LAMP, model:A1A-11 AND A1A-12<br>SERIES |
| 510(k) Number | K082031 |
| Product Code | HJO |
| Regulation<br>Number | 21 886.1850 |
| Regulation Class | 2 |
#### 5. Device Description
The slit lamp is an instrument consisting of a high-intensity light source that can be focus to shine a thin sheet of light into the eye. The binocular slit-lamp examination provides a stereoscopic magnified view of the eye structure in detail, enabling diagnosis of diseases or trauma, which affect the structural properties of the anterior segment of the eye. The device consists of a base plate, chin rest, fixation light, illumination light, slit project, focusing ring and other accessories that in aid of examination. The patients can rest their chin and forehead on a support to keep the head steady.
It is powered by the specified adapter (model: MDS-060BAS12A) which is certificated by ANSI/AAMI ES 60601-1 and considered as part of this equipment. Appliance coupled with US Plug.
#### 6. Intended Use / Indications for Use
The SLIT LAMP is an AC-powered slit lamp biomicroscope that is intended for use in examining the anterior eye segment, from the corneal epithelium to the posterior capsule.
{6}------------------------------------------------
It is used to aid in the diagnosis of diseases or trauma, which affect the structural properties of the anterior segment of the eye.
## 7. Test Summary for SLIT LAMP
YF-100 has been evaluated the safety and performance by lab bench testing as following:
- � The YF-100 was evaluated to IEC 60601-1:2005 (3rd edition) for electrical safety, and passed.
- � The YF-100 was evaluated to IEC 60601-1-2 for EMC compliance, and passed at Class A radiated emissions limits.
- � The YF-100 was evaluated to ISO 15004-1 for environmental tolerance of ophthalmic devices and passed.
- � The YF-100 was evaluated to ISO 15004-2 for assessment of photo-biological safety of light intensity limits, and passed.
- The YF-100 was evaluated to ISO 10939, which requires assessment to ISO 15004-1, � 15004-2, and IEC 60601-1, and passed.
#### 8. Comparison to predicate device and conclusion
The technological characteristics, features, specifications, materials, mode of operation, and intended use of SLIT LAMP is substantially equivalent to the predicate devices quoted above.
The differences between the subject device and predicate device do not raise new issues of safety or effectiveness.
| Elements of<br>Comparison | Subject Device | Predicate Device | Remark |
|-----------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Basic Unit Characteristics | | | |
| Device Name and<br>Model | SLIT LAMP, Model: YF-100 | APPASAMY SLIT LAMP, model:A1A-11 AND A1A-12 SERIES | -- |
| Elements of<br>Comparison | Subject Device | Predicate Device | Remark |
| 510 (K) Number | K161764 | K082031 | -- |
| Intended Use | The SLIT LAMP is intended<br>for use in eye examination of<br>the anterior eye segment,<br>from the cornea epithelium to<br>the posterior capsule. It is<br>used to aid in the diagnosis of<br>diseases or trauma which<br>affects the structural<br>properties of the anterior eye<br>segment. | Slit Lamp is an AC-powered<br>slit lamp biomicroscope<br>intended for use in eye<br>examination of the anterior<br>eye segment, from the cornea<br>epithelium to the posterior<br>capsule. It is used to aid in the<br>diagnosis of diseases or<br>trauma which affects the<br>structural properties of the<br>anterior eye segments. | SE |
| Microscope type | Galileo binocular converging | Galilean type | SE |
| Exposure<br>parameters | 1. Magnification Change:<br>Continuous by manual zoom<br>2. Eyepieces:12.5×<br>3. PD Range:55mm to 80mm<br>4. Slit Width:0mm to 14mm<br>continuous (become a circle at<br>14mm)<br>5. Slit Apertures: φ0.3mm、φ<br>5.5mm、φ9mm、φ14mm<br>6. Slit Angles: 0° to 180°<br>continuous adjustable from<br>vertical to horizontal direction<br>7. Diopter adjustment:-5.00D<br>to +5.00D<br>8. Working Distance: 100mm<br>9. Longitudinal(In/Out) :<br>Movement 100mm<br>10. Lateral(Left/Right) : | 1. Magnification Change:<br>Continuous by manual zoom<br>2. Eyepiece: 12.5x<br>3. PD Range: 55mm to 75mm<br>4. Slit Width: 0mm to 14mm<br>5. Slit Apertures: 0mm to 14mm,<br>φ0.2, φ1, φ3, φ4, φ6, φ10, φ<br>14mm<br>6. Slit Angles: 0° to 180°<br>7. Diopter adjustment: -6D to<br>+6D<br>8.Working Distance: 100mm<br>9. Longitudinal(In/Out) :<br>Movement 99mm<br>10. Lateral(Left/Right) :<br>Movement 118mm<br>11. Vertical(Up/Down) :<br>Movement 30mm<br>12. Chin Rest vertical:55mm | Note 1 |
| Elements of<br>Comparison | Subject Device | Predicate Device | Remark |
| Movement 100mm<br>11.Vertical(Up/Down) :<br>Movement 30mm<br>12. Horizontal:<br>Movement 10mm<br>13.Chin Rest Elevation :<br>70mm | | | |
| Filters | Heat aborption, redfree, cobalt blue | Heat absorbing, Green,<br>Cobalt Blue | SE<br>Note 1 |
| Total Magnification | $6.4 \times,10 \times,16 \times,25 \times,40 \times$ | 5.5x to 35.0x | SE<br>Note 1 |
| Fixation Light | Red LED (620nm~720nm) | LED | SE<br>Note 1 |
| Data collection<br>and/or display<br>system | N/A | N/A | SE |
| Flammability of<br>materials | This instrument is not suitable<br>for use in a flammable<br>atmosphere. Do not use this<br>instrument if any flammable<br>gases are present. | This instrument is not suitable for<br>use in a flammable atmosphere.<br>Do not use this instrument if any<br>flammable gases are present. | SE |
| Max. temperature<br>of parts of the<br>device held by the<br>operator or<br>accessible to the<br>patient | No parts of device with<br>patient/operator contact is<br>energized and remains at<br>ambient temperature. | No parts of device with<br>patient/operator contact is<br>energized and remains at<br>ambient temperature. | SE |
| Brightness Control | 3V LED lamp | 12V 30W halogen bulb | SE<br>Note 1 |
| Duration of<br>illumination | Maximum examination times<br>according to ISO 15004-2 and<br>ISO 10939 | Maximum examination times<br>according to ISO 15004-2 and<br>ISO 10939 | SE |
| Compliance<br>with | IEC 60601-1-2 | IEC 60601-1 | -- |
| Elements of<br>Comparison | Subject Device | Predicate Device | Remark |
| Safety Standards | IEC 60601-1:2005 | ISO 10939:2007 | |
| | ISO 15004-1 | ISO 15004-2:2007 | |
| | ISO 15004-2 | | |
| | ISO 10939 | | |
| Input Voltage AC | 230 / 110V - 50 / 60 Hz | 230 / 110V - 50 / 60 Hz | SE |
| Power Rating | 60W | 45W | SE<br>Note1 |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
#### Note 1:
Although the Brightness Control, Power Rating, Filters, Total Magnification Light of subject device are a little different from the predicate devices, they are all compliant with requirements of IEC 60601-1, ISO 15004-2, ISO 10939 and Guidance for Slit Lamp. So the differences of the function specifications will not raise any safety or effectiveness issue.
#### Finial Conclusion:
The subject device "SLIT LAMP" is Substantially Equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.