← Product Code [KPS](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPS) · K161740

# D-SPECT Scanner, D-SPECT L Scanner (K161740)

_Spectrum Dynamics Medical, Ltd. · KPS · Nov 4, 2016 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPS/K161740

## Device Facts

- **Applicant:** Spectrum Dynamics Medical, Ltd.
- **Product Code:** [KPS](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPS.md)
- **Decision Date:** Nov 4, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1200
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The D-SPECT Cardiac Scanner System is an emission computed tomography system intended for detection of radioisotope tracer uptake in the patient's body and produce crosssectional images through computer reconstruction of the data. The system uses a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality. The scanning modes include planar mode (Static, Multi-gated and Dynamic) and tomographic mode (Static, Multi-gated and Dynamic). The acquisition types include single and multi-isotope/multi peak frame/list mode single-photon imaging. The imaging-enhancement features include gating by physiological signals, real-time body movement control, and low count rate (low dose) acquisition without loss of image quality. The D-SPECT Cardiac Scanner System may consist analysis and display equipment contain data and image processing software to produce images in a variety of trans-axial and reformatted planes. To perform analysis and uptake quantitation and to apply the appropriate filters. The system utilizes combined images for attenuation corrected imaging as well as functional and anatomical mapping imaging (localization, registration and fusion). The D-SPECT Cardiac Scanner System is intended for use by the appropriately trained healthcare professionals to aid in detecting, localizing and diagnosing of (but not limited) cardiac or individual organs diseases. The system output can be used for planning, guiding, and monitoring therapy.

## Device Story

SPECT system detects radioisotope tracer uptake in patient body; produces cross-sectional images via computer reconstruction. Input: single/multi-isotope, multi-peak frame/list mode single-photon imaging data. Features: planar/tomographic scanning (static, multi-gated, dynamic), physiological gating, real-time body movement control, low-dose acquisition. System comprises detector head, gantry, patient supports, UPS, display/processing equipment. Available in 9-detector (D-SPECT) or 6-detector (D-SPECT L) configurations. Used in clinical settings by trained professionals. Software performs analysis, uptake quantitation, filtering, attenuation correction, and functional/anatomical mapping (fusion/registration). Output aids diagnosis, therapy planning, and monitoring. Benefits: high-performance, compact imaging for cardiac and small organ evaluation.

## Clinical Evidence

Bench testing only. Performance validation conducted per NEMA NU-1:2012 standards. Device designed in accordance with EN 60601 series (IEC 60601-1, 1-2). All testing results met predetermined acceptance values.

## Technological Characteristics

CZT detectors (40-170 keV range); 9-detector or 6-detector configurations; gantry-based system. Includes gating and real-time movement control. Software for image reconstruction, analysis, quantitation, and fusion. Complies with IEC 60601-1 and 1-2 standards. NEMA NU-1:2012 performance validation.

## Regulatory Identification

An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.

## Predicate Devices

- D-SPECT Cardiac Scanner ([K110507](/device/K110507.md))

## Reference Devices

- D-SPECT Processing and Reviewing Workstation ([K160120](/device/K160120.md))
- Discovery NM/CT 670 CZT ([K153402](/device/K153402.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of a caduceus, which is a symbol often associated with healthcare and medicine.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 4, 2016

Spectrum Dynamics Medical Ltd % Mr. Igor Naroditsky Director QA/RA 22 Bareket Street Caesarea, 3088900 ISRAEL

Re: K161740

Trade/Device Name: D-SPECT® Cardiac Scanner System D-SPECT® L Cardiac Scanner System Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: II Product Code: KPS Dated: August 1, 2016 Received: August 11, 2016

Dear Mr. Naroditsky:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Robert Oels

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K161740

Device Name D-SPECT® Cardiac Scanner D-SPECT® L Cardiac Scanner

#### Indications for Use (Describe)

The D-SPECT® Cardiac Scanner is an emission computed tomography system intended for detection of radioisotope tracer uptake in the patient's body and produce cross-sectional images through computer reconstruction of the data. The system uses a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality. The scanning modes include planar mode (Static, Multi-gated and Dynamic) and tomographic mode (Static, Multi-gated and Dynamic). The acquisition types include single and multi-isotope/multi peak frame/list mode single-photon imaging. The imaging-enhancement features include gating by physiological signals, real-time body movement control, and low count rate (low dose) acquisition without loss of image quality.

The D-SPECT® Cardiac Scanner may consist and display equipment contain data and image processing software to produce images in a variety of trans-axial and reformatted planes. To perform analysis and uptake quantitation and to apply the appropriate filters. The system utilizes combined images for attenuation corrected imaging as well as functional and anatomical mapping imaging (localization, registration and fusion).

The D-SPECT® Cardiac Scanner is intended for use by the appropriately trained healthcare professionals to aid in detecting, localizing and diagnosing of (but not limited) cardiac or individual organs diseases. The system output can be used for planning, guiding, and monitoring therapy.

| Type of Use (Select one or both, as applicable)                                       |                                                                                      |
|---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| <div> <span>☑</span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span>☐</span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Spectrum Dynamics Medical. The logo consists of a blue triangular prism on the left, followed by the company name in blue text. The words "SPECTRUM" and "DYNAMICS MEDICAL" are on separate lines, with the latter being smaller and underlined. Below that, it says "A BIOSENSORS INTERNATIONAL GROUP COMPANY" in smaller, lighter blue text, also underlined.

D-SPECT ™ Cardiac Scanner

# 510(k) Summary

# D-SPECT® Cardiac Scanner System

# 510(k) Number K161740

| <b>Identification of the Submitter</b>                  |                                                                                                               |               |
|---------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|---------------|
| Submitter:                                              | Igor Naroditsky, Director QA/RA<br>Spectrum Dynamics Medical Ltd.<br>22 Bareket St. Caesarea, Israel 3088900. |               |
| Telephone Number:                                       | + (972) 73-737-4500,                                                                                          |               |
| Fax Number:                                             | + (972) 73-737-4501                                                                                           |               |
| Name / Address of Manufacturer:                         | Spectrum Dynamics Medical Ltd.<br>22 Bareket St. Caesarea, Israel 3088900.                                    |               |
| Date of Submission:                                     | 30 May 2016                                                                                                   |               |
| <b>Identification of the product</b>                    |                                                                                                               |               |
| Device Proprietary Names:                               | D-SPECT® Cardiac Scanner System<br>D-SPECT® L Cardiac Scanner System                                          |               |
| Common/Classification Names:                            | Emission computed tomography system                                                                           |               |
| Class:                                                  | II                                                                                                            |               |
| Product Code:                                           | KPS                                                                                                           |               |
| Classification Panel:                                   | Radiology                                                                                                     |               |
| Regulation No:                                          | 21 CFR 892.1200                                                                                               |               |
| <b>Marketed Devices to which Equivalence is claimed</b> |                                                                                                               |               |
| Device                                                  | Manufacturer                                                                                                  | 510(k) Number |

Spectrum Dynamics Medical Ltd.

K110507

{5}------------------------------------------------

### Reference Devices:

| Device                                           | Manufacturer                   | 510(k) Number |
|--------------------------------------------------|--------------------------------|---------------|
| D-SPECT® Processing and Reviewing<br>Workstation | Spectrum Dynamics Medical Ltd. | K160120       |
| Discovery NM/CT 670 CZT                          | GE Healthcare                  | K153402       |

## Device Description:

Spectrum Dynamics D-SPECT® Cardiac Scanner System is a single photon emission computing tomography system intended for detection of radioisotope tracer uptake in the body and produce cross-sectional images through computer reconstruction of the data.

The device uses a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality. System's scanning modes include planar mode (Static, Multi-Gated and Dynamic) and tomographic mode (Static, Multi-Gated and Dynamic). The acquisition types include single and multi-isotope/multi peak frame/list mode single-photon imaging. The imaging-enhancement features include gating by physiological signals, real-time body movement control. The device may proceed a low count rate (low dose) acquisition without loss of image quality. The device may utilize variate modalities to create attenuation corrected images along with functional and anatomical mapping imaging (localization, registration and fusion).

The device is a high performance and compact Single Photon Emission Computed Tomography system intended for imaging of the breast and additional small organs in order to aid in the evaluation of lesions.

The system detectors support radionuclides within the energy range of 40 -170 Kev.

D-SPECT® Cardiac Scanner System comprising detector head, gantry, patient supports, uninterruptible power supply (UPS), image display and processing equipment, gating and real-time body movement control tools, interconnecting cables and related appurtenances.

The device is available in two models, D-SPECT with nine detectors configuration and the D-SPECT L with six detectors configuration.

D-SPECT® Cardiac Scanner System consist integrated signal analysis and display equipment or may use an FDA cleared D-SPECT® Processing and Reviewing Workstation (K160120) for image processing.

{6}------------------------------------------------

D-SPECT Cardiac Scanner System is intended for use by the appropriately trained healthcare professionals to aid in detecting, localizing and diagnosing of (but not limited) cardiac diseases. The system output can be used for planning, guiding, and monitoring therapy.

#### Intended Use Statement:

The D-SPECT Cardiac Scanner System is an emission computed tomography system intended for detection of radioisotope tracer uptake in the patient's body and produce crosssectional images through computer reconstruction of the data. The system uses a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality. The scanning modes include planar mode (Static, Multi-gated and Dynamic) and tomographic mode (Static, Multi-gated and Dynamic). The acquisition types include single and multi-isotope/multi peak frame/list mode single-photon imaging. The imaging-enhancement features include gating by physiological signals, real-time body movement control, and low count rate (low dose) acquisition without loss of image quality.

The D-SPECT Cardiac Scanner System may consist analysis and display equipment contain data and image processing software to produce images in a variety of trans-axial and reformatted planes. To perform analysis and uptake quantitation and to apply the appropriate filters. The system utilizes combined images for attenuation corrected imaging as well as functional and anatomical mapping imaging (localization, registration and fusion).

The D-SPECT Cardiac Scanner System is intended for use by the appropriately trained healthcare professionals to aid in detecting, localizing and diagnosing of (but not limited) cardiac or individual organs diseases. The system output can be used for planning, guiding, and monitoring therapy.

#### Predicate Devices:

| Device        | D-SPECT <sup>tm</sup> Cardiac Scanner |
|---------------|---------------------------------------|
| Manufacturer  | Spectrum Dynamics Medical Ltd.        |
| 510(k) Number | K110507                               |

{7}------------------------------------------------

## Reference Devices:

| Device                                           | Manufacturer                   | 510(k) Number |
|--------------------------------------------------|--------------------------------|---------------|
| D-SPECT® Processing and Reviewing<br>Workstation | Spectrum Dynamics Medical Ltd. | K160120       |
| Discovery NM/CT 670 CZT                          | GE Healthcare                  | K153402       |

## Substantial Equivalence to Predicate Devices

D-SPECT scanner is a modification of the D-SPECT Cardiac Scanner (K110507) that reorganizes the system's indications for use and introduces an additional device model. D-SPECT® Scanner and its predicate, the Spectrum Dynamics D-SPECT Cardiac Scanner (K110507) have the similar indication for use and both utilize same fundamental technology. Discovery NM/CT 670 CZT (K153402), in particular its NM part, is supporting modified intended use. D-SPECT® Processing and Reviewing Workstation (K160120) supports image post processing modalities.

Both devices the modified D-SPECT Cardiac Scanner System and its predicate D-SPECT Cardiac Scanner are intended for use by the appropriately trained healthcare professionals to aid in detecting, localizing and diagnosing of (but not limited) cardiac or individual organs diseases. The system output can be used for planning, guiding, and monitoring therapy. The intended use of the subject device has been reorqanized and extended to support and clarify a well-established imaging modalities.

Discovery NM/CT 670 CZT (K153402), especially its NM part, uses to support intended use reorganization.

### Technology characteristic

D-SPECT® Cardiac Scanner System based on the commercially available D-SPECT ™ Cardiac Scanner (K110507). The modified D-SPECT Cardiac Scanner System hardware based on the same fundamental technology and utilize same CZT detectors, collimators and image reconstruction tools. The device efficacy and safety as well as the performance specifications remain the same.

### Performance testing

D-SPECT® Cardiac Scanner System is designed in accordance with the EN 60601 series of standards, including all relevant collateral standards, i.e. IEC 60601-1, 1-2. etc. Device

{8}------------------------------------------------

performance validation testing conducted according to NEMA NU-1:2012. All testing results are met the predetermined acceptance values.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPS/K161740](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPS/K161740)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
