Salter Labs Bubble Humidifier

K161719 · Salter Labs · BTT · Sep 21, 2017 · Anesthesiology

Device Facts

Record IDK161719
Device NameSalter Labs Bubble Humidifier
ApplicantSalter Labs
Product CodeBTT · Anesthesiology
Decision DateSep 21, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5450
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The bubble humidifier is intended to add moisture to breathing gases for administration to patients >1 month in homecare, hospital, extended care and hospice. The bubble humidifier is a non-sterile device indicated for single-patient usage. The device is indicated for patients who require humidification of high flow supplemental breathing gases.

Device Story

Passive, non-heated, non-powered bubble humidifier; used in-line between breathing gas source (oxygen concentrator, cylinder, wall outlet) and patient circuit (face mask or cannula). Device consists of humidifier bottle, lid, and PVC diffuser tube. Gas enters bottle, passes through water via diffuser for humidification, and exits to patient. Includes 6 PSI (410 mbar) audible pressure relief safety valve for downstream occlusion. Used in homecare, hospital, extended care, and hospice; operated by clinicians or patients. Provides humidified gas to improve patient comfort and airway hydration. Material change from DEHP-plasticized PVC to DOTP-plasticized PVC for diffuser components.

Clinical Evidence

No clinical data. Bench testing only. Testing included bond strength verification for the new material, biocompatibility (ISO 10993-5, 10993-10, 10993-18), and a useful life study verifying performance up to 120 cleaning cycles.

Technological Characteristics

Passive, non-heated, non-powered humidifier. Materials: Polyethylene bottle, polystyrene lid, PVC (DOTP) diffuser tubing. Dimensions: Bottle 5.12" length, 2.75" base diameter. Connectivity: In-line between gas source and patient circuit. Safety: 6 PSI (410 mbar) brass pop-off valve. Non-sterile, single-patient use.

Indications for Use

Indicated for patients >1 month requiring humidification of high flow supplemental breathing gases in homecare, hospital, extended care, and hospice settings.

Regulatory Classification

Identification

A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 21, 2017 Salter Labs Mara Caler Regulatory Affairs 2365 Camino Vida Roble Carlsbad, California 92011 Re: K161719 Trade/Device Name: Salter Labs Bubble Humidifier (6-15 LPM) with 6 PSI (410mbars) safety valve Regulation Number: 21 CFR 868.5450 Regulation Name: Respiratory Gas Humidifier Regulatory Class: Class II Product Code: BTT Dated: September 14, 2017 Received: September 18, 2017 Dear Mara Caler: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, # Tara A. Ryan -S for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K161719 Device Name Salter Labs Bubble Humidifier (6-15 LPM) with 6 PSI (410mbars) safety valve #### Indications for Use (Describe) The bubble humidifier is intended to add moisture to breathing gases for administration to patients >1 month in homecare, hospital, extended care and hospice. The bubble humidifier is a non-sterile device indicated for single-patient usage. The device is indicated for patients who require humidification of high flow supplemental breathing gases. | Type of Use (Select one or both, as applicable) | |-----------------------------------------------------------------------------------------| | <span style="padding-right: 5px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | | <span style="padding-right: 5px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 510(k) Summary In accordance with the Safe Medical Devices Act (SMDA) of 1990 and Title 21 of the Code of Federal Regulations Part 807 (21 CFR §807), and in particular §807.92, the following summary of substantial equivalence information is provided: ### 510 (k) Summary #### A. Submitter: Salter Labs 2365 Camino Vida Robles Carlsbad, CA 92011 Telephone: 760-795-7100 760-683-6797 Fax: Contact Person: Mara Caler Regulatory Affairs Date Prepared: 14 Sept 2017 #### Device Names: B. | Trade Name | Common Name | Classification Name | Product Code | |---------------------------------------------------------------------------------------|-------------------------------|-------------------------------|--------------| | Salter Labs Bubble<br>Humidifier (6-15 LPM) with<br>6 PSI (410mbars) safety<br>valve. | Respiratory gas<br>humidifier | Respiratory gas<br>humidifier | BTT | | Regulation Number: | 868.5450 | |-----------------------|----------------------| | Classification: | II | | Classification Panel: | Anaesthesiology (73) | #### C. Predicate Devices: This submission demonstrates substantial equivalence to the predicate device: K113542, Salter Labs Bubble Humidifier #### D. Device Descriptions The Salter Labs Bubble Humidifier (6-15 LPM) with 6 PSI (410 mbars) safety valve is an empty, disposable, non-sterile, not made with natural rubber latex, device intended to {4}------------------------------------------------ humidify breathing gas prior to delivery to a patient. The Salter Labs Bubble Humidifier (6-15 LPM) with 6 PSI (410mbars) safety valve is provided with a 6 pounds per square inch (PSI) safety valve and can operate within flow rates of 6 to 15 liters per minute (LPM). The device is used with various breathing gas sources (i.e., oxygen concentrators, gas cylinders and wall outlets) and provides connection for delivery of humidified breathing gas via face masks or cannulas, and use of optional oxygen tubing and water traps (face masks, 21 CFR 868.5580: nasal cannulas, 21 CFR 868.5340; oxygen tubing, 21 CFR 868.5860 and water traps, 21 CFR 868.5995 are 510(k) exempt). This device is a passive device and is not a cascade humidifier, is not heated and is not prefilled. The device is made of a humidifier bottle which is used to hold water during use, a lid which seals the humidifier bottle, an audible pressure relief mechanism to notify the user of a downstream occlusion and a diffuser located at the end of a PVC diffuser tube inside the humidifier bottle. The diffusor is designed to uniformly disperse the gas throughout the water. Both the bottle and lid are constructed to be easy to grip and reduce the chance of cross threading. The bottle is permanently marked with "minimum/maximum" water levels. The lid is marked with minimum source pressure, flow ranges and pressure value of the safety valve. #### E. Indications for Use: The bubble humidifier is intended to add moisture to breathing gases for administration to patients >1 month in homecare, hospital, extended care and hospice. The bubble humidifier is a non-sterile device indicated for single-patient usage. The device is indicated for patients who require humidification of high flow supplemental breathing gases. ## E.1: Usage The bubble humidifier is indicated for use with oxygen concentrators or gas sources in homecare, hospital, extended care facilities and hospice environments. The bubble humidifier is a non-sterile device indicated for single-patient usage. The device is indicated for patients who require humidification of high flow supplemental breathing gases. ## E.2: Comparison of intended use to the predicate device and separate it from the Indications for Use Statement. The predicate device is indicated for any patient population requiring humidification of high flow supplemental breathing gases. There are no excluded patient populations. For clarity, a more detailed description of the patient population is defined in the indications for use statement. This device is intended for the same intended population, including patients >1 month in homecare, hospital, extended care and hospice environments. {5}------------------------------------------------ #### F. Substantial Equivalence: | | | Comparison of Technological Characteristics with the Predicate Device | |--|--|-----------------------------------------------------------------------| | Specification | Predicate: Salter Labs<br>Bubble Humidifier (High<br>Flow) | Test device: Salter Labs<br>Bubble Humidifier (6-15<br>LPM) with 6 PSI<br>(410mbars) safety valve | Changes | |-------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------| | Device Use & General Characteristics | | | | | Indications for<br>use | The bubble humidifier is a<br>device that is intended to<br>add moisture to breathing<br>gases for administration to a<br>patient. The bubble<br>humidifier is a non-sterile<br>device indicated for single-<br>patient usage. The device is<br>indicated for patients who<br>require humidification of<br>high flow supplemental<br>breathing gases. | The bubble humidifier is<br>intended to add moisture to<br>breathing gases for<br>administration to<br>patients >1 month in<br>homecare, hospital, extended<br>care and hospice.<br>The bubble humidifier is a<br>non-sterile device indicated<br>for single-patient usage. The<br>device is indicated for<br>patients who require<br>humidification of high flow<br>supplemental breathing<br>gases. | The patient<br>population<br>and the use<br>environment<br>has been<br>defined | | Target Patient<br>Population | The device is indicated for<br>patients who require<br>humidification of high flow<br>supplemental breathing<br>gases. | The device is indicated for<br>patients >1 month who<br>require humidification of<br>high flow supplemental<br>breathing gases. | The patient<br>population<br>has been<br>defined | | Intended use | To humidify breathing gases | To humidify breathing gases | Unchanged | | Environment of<br>use | For use in homecare,<br>hospital, acute care facilities,<br>extended care facilities and<br>hospice environments. | For use in homecare,<br>hospital, acute care facilities,<br>extended care facilities and<br>hospice environments. | Unchanged | | Use | Single patient use | Single patient use | Unchanged | | Supplied as | Non-sterile | Non-sterile | Unchanged | | Principal of<br>Operation/<br>Mechanisms of<br>Action | The bubble humidifier is<br>indicated for use with<br>oxygen concentrators or gas<br>sources to humidify (add<br>water to) the oxygen | The bubble humidifier is<br>indicated for use with<br>oxygen concentrators or gas<br>sources to humidify (add<br>water to) the oxygen | Unchanged | | Technology | Air is channelled through a | Air is channelled through a | Unchanged | | Specification | Predicate: Salter Labs<br>Bubble Humidifier (High<br>Flow) | Test device: Salter Labs<br>Bubble Humidifier (6-15<br>LPM) with 6 PSI<br>(410mbars) safety valve | Changes | | Device Use & General Characteristics | | | | | | water-containing bottle. The<br>air then becomes humidified<br>before exiting to the patient | water-containing bottle. The<br>air then becomes humidified<br>before exiting to the patient | | | Flow rate | 6-15 liters per minute | 6-15 liters per minute | Unchanged | | Pop-off safety<br>valve | At 6 PSI | At 6 PSI | Unchanged | | Power Source | None - this is a passive<br>device | None - this is a passive<br>device | Unchanged | | Prescription Use /<br>Over-The-<br>Counter Use | Prescription Devices | Prescription Devices | Unchanged | | Sterile / Non-<br>Sterile | Non-Sterile | Non-Sterile | Unchanged | | Use | Single patient use | Single patient use | Unchanged | | Required<br>preparations prior<br>to use | None | None | Unchanged | | Manufacturing<br>Process of Device<br>from Material | Molded (bottle, lid)<br>Extruded (diffuser tube) | Molded (bottle, lid)<br>Extruded (diffuser tube) | Unchanged | | Patient-Contact<br>Components | Device does not touch the<br>patient. | Device does not touch the<br>patient. | Unchanged | | Labeling | Instructions for Use are<br>unchanged except for the<br>removal of the DEHP<br>symbol. | Instructions for Use are<br>unchanged except for the<br>removal of the DEHP<br>symbol. | SE | | Environmental<br>Limitations | Storage temperature:<br>-40°C (-40°F) to +60°C<br>(140 °F) | Storage temperature:<br>-40°C (-40°F) to +60°C<br>(140 °F) | Unchanged | | Materials | | | | | K113542 | PVC (DEHP) | PVC (DOTP) | SE | | Biocompatibility | ISO 10993 compliant | ISO 10993 compliant | Unchanged | {6}------------------------------------------------ {7}------------------------------------------------ ### Similarities and Differences: #### Similarities: The similarities between the Salter Labs Bubble Humidifier and the modified Salter Labs Bubble Humidifier (6-15 LPM) with 6 PSI (410mbars) safety valve are: Design: The exact same design is used; there are no changes to the Salter Labs Bubble Humidifier (6-15 LPM) with 6 PSI (410mbars) safety valve except the materials used in the line/diffusor. - Both are passive, non-heated, non-powered humidifier bottles. There is no external ● energy source. - Both are in-line between the air source and patient circuit (nasal cannula) ● - Both are intended for single-patient use. ● - Both are supplied non-sterile - Both use identical pop-off safety valve (6 PSI) made from identical material (brass) ● - Both have identical dimensions. ● | | Parameter (inch) | Comparison to predicate | |-------------------|------------------|-------------------------| | Bottle | | | | • Length | 5.12 | No change | | • Diameter (base) | 2.75 | No change | | • Diameter (top) | 2.84 | No change | | Lid | | | | • Height | 2.26 | No change | | • Diameter | 3.08 | No change | | Tubing | | | | • Length | 5.50 | No change | | • Diameter | 0.29 | No change | | Diffusor | | | | • Length | 0.51 | No change | | • Diameter (Base) | 0.40 | No change | | • Diameter (Top) | 0.33 | No change | - Both have identical performance characteristics ● - o Flow rate: 6-15 LPM - o 6 PSA (410 mbars) for pop-off safety valve ### Materials - . Both use identical materials: polyethylene bottle with a polystyrene lid (except as identified in the "differences" table below). {8}------------------------------------------------ ## Energy source - Both use are non-heated, non-powered in-line humidifier bottles. There is no . external energy source. ### Differences The differences between the Salter Labs Bubble Humidifier and the modified Salter Labs Bubble Humidifier (6-15 LPM) with 6 PSI (410mbars) safety valve is a material modification from PVC (DEHP) to PVC (DOTP). The materials used in the remainder of the device are unchanged. | Features | Predicate<br>K113542 | Modified<br>device | Performance<br>Testing | |----------------------------------------------------------------------------------------------------------|----------------------|--------------------|-------------------------------------| | Material Formulation<br>of the diffusor tubing<br>and diffusor (all other<br>materials are<br>unchanged. | PVC, DEHP | PVC, DOTP | Biocompatibility<br>and Performance | {9}------------------------------------------------ ### Standards Standards used for this submission: | Standard | Test | Results | |--------------|--------------------------------------------------------------------------------------------------------|------------| | ISO 10993-5 | Biological Evaluation of Medical Devices<br>Part 5: Tests for in-vitro cytotoxicity | Acceptable | | ISO 10993-10 | Biological evaluation of medical devices -<br>Part 10: Tests for irritation and skin<br>sensitization. | Acceptable | | ISO 10993-18 | Biological evaluation of medical devices -<br>Part 18: Chemical characterization of<br>materials | Acceptable | #### Performance Data G. The modified Salter Labs Bubble Humidifier (6-15 LPM) with 6 PSI (410mbars) safety valve was tested to verify that the bond strength of the new material used for the different tube and diffusor met the pre-defined performance specifications. The test results demonstrate that the modified Salter Labs Bubble Humidifier (6-15 LPM) with 6 PSI (410mbars) safety valve is substantially equivalent to the predicate devices. No clinical testing was required for this submission. A cleaning study was performed which verified the cleaning outlined in the instructions for use. A useful life study was performed which verified performance up to 120 cleaning cycles. ### H. Conclusions The Salter Labs Bubble Humidifier (6-15 LPM) with 6 PSI (410mbars) safety valve data and test results (bond strength testing) demonstrate that the devices are substantially equivalent to the predicate devices. Based on the performance testing performed on the modified device safety valve and bond strength, Salter Labs Bubble Humidifier (6-15 LPM) with 6 PSI (410mbars) we have demonstrated that the modified device continue to perform substantially equivalent to the predicate devices.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...