K161702 · Vascular Solutions, Inc. · DQX · Dec 6, 2016 · Cardiovascular
Device Facts
Record ID
K161702
Device Name
Endura guidewire
Applicant
Vascular Solutions, Inc.
Product Code
DQX · Cardiovascular
Decision Date
Dec 6, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1330
Device Class
Class 2
Indications for Use
The Endura guidewire is intended for use in percutaneous procedures to introduce and position catheters and other interventional devices within the coronary and/or peripheral vasculature.
Device Story
Endura guidewire is a straight-tip, 0.014" diameter wire (190 cm or 300 cm lengths) used to navigate coronary and peripheral vasculature. It functions as a track for catheters and interventional devices. The device features a bi-metal stainless steel and nitinol core, a distal radiopaque coil, a distal hydrophilic coating, and a proximal PTFE coating. It is used in clinical settings (e.g., cath lab) by physicians during percutaneous interventions. The device provides mechanical support and guidance for interventional tools, facilitating access to target vascular sites. It is a sterile, single-use device.
Clinical Evidence
Bench testing only. Verification testing included tensile strength, torque strength, torque response, support profile, tip load, fracture resistance, flex resistance, coating adherence/integrity (HPC and PTFE), corrosion resistance, track force, friction force, fluoroscopy visualization, and length extension. Biocompatibility testing included cytotoxicity, sensitization, irritation/intracutaneous reactivity, systemic toxicity, pyrogenicity, and hemocompatibility. All tests met acceptance criteria.
Technological Characteristics
Bi-metal stainless steel and nitinol core wire; 0.014" O.D.; 190 cm/300 cm lengths; distal radiopaque coil; hydrophilic coating (distal); PTFE coating (proximal). Sterilized via Ethylene Oxide (EO) to SAL 10^-6. Mechanical device; no software or electronics.
Indications for Use
Indicated for patients undergoing percutaneous procedures requiring catheter or interventional device positioning within coronary or peripheral vasculature.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure, with three stylized profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 6, 2016
Vascular Solutions, Inc. Beka Vite Sr. Regulatory Product Specialist 6464 Sycamore Court North Minneapolis, Minnesota 55369
Re: K161702
Trade/Device Name: Endura Guidewire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: November 3, 2016 Received: November 4, 2016
Dear Beka Vite:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely.
Brian D. Pullin -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name Endura guidewire
Indications for Use (Describe)
The Endura guidewire is intended for use in percutaneous procedures to introduce and position catheters and other interventional devices within the coronary and/or peripheral vasculature.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
[As required by 21 CFR 807.92]
Date Prepared: December 2, 2016
510(k) Number: K161702
## Submitter's Name / Contact Person
# Manufacturer Vascular Solutions, Inc. 6464 Sycamore Court North Minneapolis, MN 55369 USA Establishment Registration # 2134812
# Contact Person Beka Vite Sr. Regulatory Product Specialist
Tel: 763-656-4300 Fax: 763-656-4253
| General Information | |
|---------------------|--------------------------------------------------------------------------------------------------------------|
| Trade Name | Endura guidewire |
| Common / Usual Name | Guidewire |
| Classification Name | 21 CFR 870.1330, Catheter Guidewire (DQX) |
| Predicate Device | K151234 - R350 guidewire (Vascular Solutions, Inc.) |
| Reference Device | K063695 - Runthrough NS guidewire (Terumo Medical Corp.)<br>K052339 - MiracleBros 6 (Asahi Intecc Co., Ltd.) |
## Device Description
The Endura guidewire is a straight tip guidewire with a maximum outer diameter of 0.014" available in 190 cm and 300 cm lengths. The Endura guidewire consists of a bi-metal stainless steel and nitinol core wire covered on the distal end by a radiopaque coil at the distal end of the Endura guidewire has a hydrophilic coating (HPC) and the proximal end is coated with polytetrafluoroethylene (PTFE). The 190 cm length model is compatible with a guidewire extension.
## Intended Use
The Endura guidewire is intended for use in percutaneous procedures to introduce and position catheters and other interventional devices within the coronary and/or peripheral vasculature.
{4}------------------------------------------------
#### Technological Characteristics Comparison
| Characteristic | Subject Device | Predicate Device |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|
| | Endura Guidewire | R350 Guidewire<br>(K151234, VSI) |
| Indications for Use | ...for use in percutaneous procedures to introduce and position catheters and other<br>interventional devices within the coronary and/or peripheral vasculature. | |
| Anatomical Location | Coronary and peripheral vasculature | |
| Dimensions | Maximum O.D.: 0.014"<br>Length: 190 cm and 300 cm | Maximum O.D.: 0.013"<br>Length: 350 cm |
| Core Wire | Nitinol and Stainless Steel | Nitinol |
| Distal Tip | 3 cm radiopaque coil | 5 cm radiopaque coil |
| Lubricious Coatings | Hydrophilic (distal end) and PTFE (proximal end) | |
| Sterility | Ethylene Oxide (EO) – Sterility Assurance Level (SAL) 10-6 | |
The table below compares the technological characteristics of the Endura guidewire.
### Substantial Equivalence and Summary of Studies
The technological differences between the subject and predicate devices have been evaluated through performance and biocompatibility tests and results did not raise new questions of safety or effectiveness. The Endura guidewire is substantially equivalent to the specified predicate device based on comparisons of the device functionality, technological characteristics, and indications for use. The device design has been verified through the following tests:
- Tensile Strength ●
- Torque Strength
- Torque Response (Torqueability) ●
- Guidewire Support Profile
- Tip Load (Tip Flexibility)
- Fracture Resistance
- Flex Resistance
- Hydrophilic Coating Particulate (Coating Adherence/Integrity)
- PTFE Coating Adhesion (Coating Adherence/Integrity)
- Corrosion Resistance
- Track Force
- Friction Force (HPC and PTFE) .
- Fluoroscopy Visualization
- Length Extension
{5}------------------------------------------------
The only evaluation that was not performed on the modified guidewire was package integrity as it is identical to the predicate device. Device samples passed the following biocompatibility tests performed in accordance with ISO 10993-1:
- Cytotoxicity
- Sensitization
- . Irritation/Intracutaneous Reactivity
- Systemic Toxicity
- Pyrogenicity
- Hemocompatibility
The results of the verification tests met the specified acceptance criteria and did not raise new safety or performance issues. Therefore the Endura guidewire, a modified version of the R350, is considered substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.