← Product Code [GEX](/productcode/GEX) · K161670

# Lucid Q-PTP / HWA 55 (K161670)

_Bison Medical Co, Ltd. · GEX · Nov 2, 2016 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K161670

## Device Facts

- **Applicant:** Bison Medical Co, Ltd.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Nov 2, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The LUCID O-PTP Nd : YAG Laser System is indicated for the incision, excision, ablation, vaporization of soft tissues for general dermatologic and general surgical procedures for coagulation and hemostasis. 532nm Wavelength (nominal delivered energy of 585 nm with optional dye hand pieces): Tattoo removal light ink (red, tan, purple, orange, sky blue, green), Removal of Epidermal Pigmented Lesions, Minor Vascular Lesions, Talangiectasias Treatment of Lentigines, Cafe-Au-Lait, Sebrorrheic Keratoses, Post Inflammatory Hyper-Pigmentation, Treatment of Becker's Nevi, Freckles and Nevi Spilus 1064nm Wavelength: Tattoo removal: dark ink ( black, blue, brown) Removal of Nevus of Ota Removal or lightening of unwanted hair with or without adjuvant preparation Treatment of Common Nevi, Melasma, Skin resurfacing procedures for the treatment of acne scars, wrinkle

## Device Story

Q-switched Nd:YAG laser system; produces 1064nm and 532nm wavelengths; optional dye handpieces convert 532nm to 585nm and 650nm. Device consists of laser tube, resonator, articulated arm, multi-spot handpiece, cooling system, and LCD control panel. Operated by physician in clinical setting; laser emission activated via foot switch. Physician controls energy pulse and spot size to optimize treatment. Laser energy delivered to soft tissue for surgical/dermatologic procedures; enables coagulation, hemostasis, and tissue ablation. Benefits include precise removal of pigmented lesions, tattoos, and skin resurfacing.

## Clinical Evidence

No clinical or non-clinical performance test data provided. Substantial equivalence supported by adherence to safety standards including IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, and IEC 60825-1.

## Technological Characteristics

Q-switched Nd:YAG laser; 1064nm and 532nm wavelengths; optional dye handpieces for 585nm/650nm. Features articulated arm delivery, Xe-gas lamp, and cooling system. Non-contact mode. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, and IEC 60825-1 standards.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Spectra ([K113588](/device/K113588.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 2, 2016

Bison Medical Co., Ltd % Young Chi President Bio-med Usa Inc 27 New England Dr Ramsey, New Jersey 07446

Re: K161670

Trade/Device Name: Lucid Q-ptp / Hwa 55 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: September 27, 2016 Received: October 4, 2016

Dear Young Chi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

# Jennifer R. Stevenson -A

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

#### K161670

Device Name LUCID Q-PTP/ HWA 55

Indications for Use (Describe)

The LUCID O-PTP Nd : YAG Laser System is indicated for the incision, excision, ablation, vaporization of soft tissues for general dermatologic and general surgical procedures for coagulation and hemostasis.

532nm Wavelength (nominal delivered energy of 585 nm with optional dye hand pieces): Tattoo removal light ink (red, tan, purple, orange, sky blue, green), Removal of Epidermal Pigmented Lesions, Minor Vascular Lesions, Talangiectasias Treatment of Lentigines, Cafe-Au-Lait, Sebrorrheic Keratoses, Post Inflammatory Hyper-Pigmentation, Treatment of Becker's Nevi, Freckles and Nevi Spilus

1064nm Wavelength:

Tattoo removal: dark ink ( black, blue, brown) Removal of Nevus of Ota Removal or lightening of unwanted hair with or without adjuvant preparation Treatment of Common Nevi, Melasma, Skin resurfacing procedures for the treatment of acne scars, wrinkle

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510 (K) Summary

As required by CFR 807.92(c)

## 1. Manufacturer.

Prepared Sept 26, 2016

Reg Nr: 3011555967

BISON Medical Co., Ltd. Ace High-end tower 6, #1801/1802 234 Beotkkot-ro. Gasan dong GeumCheon -gu, Seoul 153-798, Rep of Korea t: 822865 7121, f: 82 2 865 7131

# 2. Submitter and Contact person

Bio-Med USA Inc. Young Chi, President. 27 New England Drive, Ramsey, NJ 07446. U.S.A. t: 1-973 278 5222 f: 1 201 934 6030 e mail: biomedusa(@msn.com

# 3. Name of Device

| Trade name           | LUCID Q-PTP                        |
|----------------------|------------------------------------|
| Classification name  | Powered, Laser surgical instrument |
| Common name          | Nd:YAG Q-switched Surgical Laser   |
| Regulation           | 878.4810 Class II                  |
| Classification Panel | General and Plastic Surgery.       |
| Product Code         | GEX                                |
| Submission type      | Traditional                        |

# 4. Legally marketed Predicate Device

| K113588 | Spectra | Nd:YAG | Lutronic Corp |
|---------|---------|--------|---------------|
|---------|---------|--------|---------------|

LUCID Q-PTP Q-Switched Nd:YAG laser system produce same two wave length (1064nm, 532nm), and same characteristics such as Design, Construction, Energy rate, Pulse Duration, applied optional Dye hand piece, Cooling system and intended use as already cleared predicate device K113588 by Lutronics.

# 5. Device Description

The LUCID O-PTP O-Switched Nd: Y AG laser system produces a two pulsed beam, 1064 nm Infrared and 532nm long pulse laser, and optional 2 dye Handpieces are available that convert the 532nm wave length to 585nm and 650nm, using different Handpiece able to control various treatment fluence.

this device is non-contacted mode and consists of main function,

laser tube ; placed in the mixed crystals of copper pipe to the heater and produces a laser beam,

Resonator : amplifies the beam, through the Xe-gas contained lamp : Xe-gas contains high pressure lamp to increase specific laser beam lamp

{4}------------------------------------------------

This converted light energy creates the ND:YAG crystal and exhaust from the crystal is amplified into a specific wave length. Laser energy produced is delivered to the Tissue by means of an articulated arm and a specially designed multi spot Hand Piece.

The Physician can optimize the effect for different applications by controlling the energy of the laser pulse and the spot size of the treatment beam, and is able to activate laser emission using Foot Switch. Laser parameters and other system features are controlled from a display panel located on the front of the power supply unit.

## This system also consist of

Optic main Bench assembly, Articulated Arm Hand pieces, LCD control panel, Cooling system, Foot Pedal Switch

#### 6. Performance test

Clinical and Non-Clinical performance test data was not provided in this submission. But, manufactured in accordance with both mandatory and voluntary standard

IEC60601-1 part 1 : General requirement for basic safety and essential performance. IEC60601-1-2: 2007 E M C test IEC60601-2-22 Part 2, Particular requirements for safety of diagnostic and Therapeutic laser IEC60825-1 :2nd ED, Equipment classification and requirement.

Lucid Q-PTP, demonstrates no significant different compare to the predicate device

#### 7. Indication for use

The LUCID Q-PTP Nd : YAG Laser System is indicated for the incision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis.

#### 532mm Wavelength

(nominal delivered energy of 585 nm and 650 nm with optional dye handpieces) Tattoo removal light ink (red, tan, purple, orange, sky blue, green), Removal of Epidermal Pigmented Lesions, Minor Vascular Lesions, Talangiectasias Treatment of Lentigines, Cafe-Au-Lait, Sebrorrheic Keratoses,

Post Inflammatory Hyper-Pigmentation,

Treatment of Becker's Nevi, Freckles and Nevi Spilus

1064nm Wavelength:

Tattoo removal: dark ink (black, blue, green )

Removal of Nevus of Ota

Removal or lightening of unwanted hair with or without adjuvant preparation Treatment of Common Nevi, Melasma,

Skin resurfacing procedures for the treatment of acne scars, wrinkle

{5}------------------------------------------------

# 8. Biocompatibility.

This device are non-contacted mode. Hand piece tips is made by same material as predicate device.

### 9.Conclusion.

LUCID Q-PTP, Nd: YAG laser system, in this submission, is substantially equivalent to several already cleared predicate device in respect to the Intended use, Main function, Technology, Principal operation and performance. And every Safety test report show it as safe and effective as predicate device and it does not raise any additional issues for safety and effectiveness.

---

**Source:** [https://fda.innolitics.com/device/K161670](https://fda.innolitics.com/device/K161670)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
