K161650 · Neurotronics, Inc. · OLV · Jan 19, 2017 · Neurology
Device Facts
Record ID
K161650
Device Name
Polysmith Sleep System, Model NTI7593
Applicant
Neurotronics, Inc.
Product Code
OLV · Neurology
Decision Date
Jan 19, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1400
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Polysmith Sleep System, Model NTI7593 is intended to measure, amplify, and record physiological signals acquired from a patient for archival in a sleep study. The physiological signals are recorded and conditioned for analysis and display. The data may be analyzed in real-time or offline on dedicated polysomnography software running on a personal computer by a qualified sleep clinician to aid in the diagnosis of Sleep Disorders. The Polysmith Sleep System, Model NTI7593 is intended for use by healthcare professionals within a hospital, laboratory, clinic, or nursing home; or outside of a medical facility under direct supervision of a medical professional. The Polysmith Sleep System, Model NTI7593 is intended for use on both adults and children only under the direction of a physician or qualified sleep technician. The Polysmith Sleep System, Model NTI7593 includes automatic detection of some arrhythmias (including some potentially life threatening arrhythmias), however detection of an arrhythmia may be performed with 30 seconds or more delay, and is based on a single ECG lead only. The Polysmith Sleep System, Model NTI7593, or any accessory, is not intended for the life monitoring of high risk patients, does not include or trigger alarms, and is not intended to be used alone as, or a critical component of, an alarm or alarm system; an apnea monitor or apnea monitoring system; or a life monitor or life monitoring system.
Device Story
Polysmith Sleep System (Model NTI7593) records/amplifies physiological signals (EEG, ECG, EMG, SpO2, respiration, ocular motility) for sleep studies. System comprises amplifier, head box, PC, and sensors. Software processes data for real-time or offline review by clinicians to score polysomnograms and diagnose sleep disorders. Includes integrated arrhythmia detection algorithm (single-lead ECG) with 30-second delay. Used in hospitals, clinics, or supervised home settings. Output displayed on PC for clinician analysis; aids diagnosis but does not provide direct treatment or life-critical monitoring. Benefits include automated event detection to assist clinician workflow.
Clinical Evidence
Bench testing only. Arrhythmia analysis performance verified per AAMI ANSI EC57:2012 using MIT-BIH, AHA, and NST databases. Reported QRS sensitivity 99.56% (MIT/AHA) and 98.10% (NST). Ventricular Ectopic Beat sensitivity 85.46%-91.85%. Results demonstrate performance comparable to reference device BSM-9100A.
Technological Characteristics
PC-based software system; compatible with various amplifiers (e.g., Neurotronics Sphinx/Nomad, Nihon Kohden PSG/EEG series). Connectivity via USB/network. Software runs on Windows (32/64-bit). Arrhythmia algorithm integrated from Nihon Kohden BSM-9100A. Standards: AAMI ANSI IEC 62304 (software), 14971 (risk), 62366 (usability), AAMI ANSI EC57 (cardiac algorithms).
Indications for Use
Indicated for adults and children requiring polysomnography or sleep studies to aid in the diagnosis of sleep disorders. Used by healthcare professionals in hospitals, clinics, labs, nursing homes, or under professional supervision outside medical facilities. Contraindicated for life monitoring of high-risk patients or as a critical component of alarm, apnea, or life-support systems.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
Nihon Kohden Bedside Monitor BSM 9100A Series (K082785)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an emblem featuring three stylized human profiles facing to the right, with flowing lines suggesting movement or connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 19, 2017
Neurotronics, Inc. David Pezet Quality Manager 3600 NW 43rd Street, STE F1 Gainesville, Florida 32606
Re: K161650
Trade/Device Name: Polysmith Sleep System, Model NTI7593 Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OLV, DQA, MLO, OLZ Dated: December 20, 2016 Received: December 21, 2016
Dear Mr. Pezet:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michael J. Hoffmann -S
for
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K161650
Device Name Polysmith Sleep System, Model NTI7593
#### Indications for Use (Describe)
The Polysmith Sleep System, Model NT17593 is intended to measure, amplify, and record physiological signals acquired from a patient for archival in a sleep study. The physiological signals are recorded and conditioned for analysis and display. The data may be analyzed in real-time or offline on dedicated polysomnography software running on a personal computer by a qualified sleep clinician to aid in the diagnosis of Sleep Disorders.
The Polysmith Sleep System, Model NT17593 is intended for use by healthcare professionals within a hospital, laboratory, clinic, or nursing home; or outside of a medical facility under direct supervision of a medical professional.
The Polysmith Sleep System, Model NT17593 is intended for use on both adults and children only under the direction of a physician or qualified sleep technician.
The Polysmith Sleep System, Model NTT7593 includes automatic detection of some arrhythmias (including some potentially life threatening arrhythmias), however detection of an arrhythmia may be performed with 30 seconds or more delay, and is based on a single ECG lead only.
The Polysmith Sleep System, Model NTI7593, or any accessory, is not intended for the life monitoring of high risk patients, does not include or trigger alarms, and is not intended to be used alone as, or a critical component of,
- an alarm or alarm system:
- · an apnea monitor or apnea monitoring system; or
- · a life monitor or life monitoring system.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="padding-right: 5px;"> <div style="display: inline-block; vertical-align: middle;"> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <path d="M13.6 2.4L5.6 10.4L2.4 7.2L0 9.6L5.6 15.2L16 4L13.6 2.4Z" fill="black"></path> </svg> </div> Prescription Use (Part 21 CFR 801 Subpart D)</span> |
| <span style="padding-right: 5px;"> <div style="display: inline-block; vertical-align: middle;"> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> </svg> </div> Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Neurotronics, a Nihon Kohden Company. The logo features a blue circle with a white line running through it, resembling a heartbeat. To the right of the circle is the word "Neurotronics" in blue, and below that is the text "A Nihon Kohden Company".
# K161650 510(k) Summary
# Summary Date:
January 19, 2017
# 510(k) Submitter:
| Neurotronics, Inc.<br>3600 NW 43rd Street, Suite F1<br>Gainesville, FL, 32606 | Phone: | 352.372.9955 |
|-------------------------------------------------------------------------------|--------|--------------|
| PRIMARY CONTACT: | Fax: | 815.550.2871 |
| Name: | David Pezet | Title: | Quality Manager |
|--------|--------------------------|--------|----------------------|
| Email: | quality@neurotronics.com | Phone: | 352.372.9955 Ext 355 |
## Common Name:
Polysomnography Monitoring System
#### Trade Name:
Polysmith Sleep System, Model NTI7593
# 510(k) Review Panel:
Neurology
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Neurotronics, a Nihon Kohden Company. The logo features a blue circle with a white line running through it, resembling a heartbeat. To the right of the circle is the company name, "Neurotronics," in a blue sans-serif font. Below the company name is the text "A Nihon Kohden Company" in a smaller font size.
# Classification:
#### PRIMARY PRODUCT CODE
| Product<br>Code | Device | Regulation<br>Description | Regulation<br>Number |
|-----------------|-------------------------------------------------------|---------------------------|----------------------|
| OLV | Standard Polysomnograph With<br>Electroencephalograph | Electroencephalograph | 882.1400 |
## Subsequent Product Codes:
| Product<br>Code | Device | Regulation<br>Description | Regulation<br>Number |
|-----------------|----------------------------------------------------------------------------------------|-----------------------------------|----------------------|
| OLZ | Automatic Event Detection Software For<br>Polysomnograph With<br>Electroencephalograph | Electroencephalograph | 882.1400 |
| DQA | Oximeter | Oximeter | 870.2700 |
| MLO | Electrocardiograph, Ambulatory, With<br>Analysis Algorithm | Medical magnetic tape<br>recorder | 870.2800 |
#### PREDICATE DEVICE
| Submitter/Holder | Device Name | Model | Product Codes | 510(K) Number |
|--------------------|------------------------|---------|---------------|---------------|
| Neurotronics, Inc. | Polysmith Sleep System | NTI6600 | OLV, DQA, OLZ | K142774 |
#### REFERENCE DEVICE
| Submitter/Holder | Device Name | Model | Product Code | 510(K)<br>Number |
|--------------------|---------------------------------|---------------------|--------------|------------------|
| Nihon Kohden Corp. | Nihon Kohden<br>Bedside Monitor | BSM 9100A<br>Series | MHX | K082785 |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Neurotronics, which is a company. The logo features a blue circle with a white line running through it, resembling a heartbeat. The text "Neurotronics" is written in blue next to the circle, and below that, it says "A Nihon Kohden Company".
#### Device Description
The following description of The Polysmith Sleep System, Model NTI7593 remains unchanged from the previous model Polysmith Sleep System, Model NTI6600.
The Polysmith Sleep System, Model NT17593 is intended to amplify and record physiologic potentials used for Polysomnography (PSG) or Sleep Studies. The device consists of a compatible amplifier, head box, PC, patient sensors, and may include optional external devices, USB DC Box, and audio/video input devices.
Compatible amplifiers may use commercially available sensors and electrodes, an internal SpO2 module, and internal pressure transducers to collect, digitize, and send physiological signals to the host PC.
The Polysmith software may record from video, speaker and microphone equipment. The Polysmith software may also record auxiliary signals from compatible amplifiers or USB DC Box which allow for data inputs from compatible sources.
Polysmith records and displays the data for online or offline review. Qualified practitioners use the information to score polysomnograms and diagnose Sleep Disorders.
Image /page/5/Figure/7 description: The image is a block diagram of a polysomnography sleep system. The diagram shows the patient area, which includes the patient, speaker/mic (optional), and patient sensors. The patient sensors are connected to an amplifier, which is also connected to the speaker/mic and external auxiliary devices (optional). The amplifier is connected to a PC, which is also connected to the audio/video (optional), sleep clinician, and USB DC box (optional).
Polysmith Sleep System Basic Block Diagram
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for Neurotronics, a Nihon Kohden Company. The logo features a blue circle with a white line that resembles a heartbeat. To the right of the circle is the word "Neurotronics" in blue, and below that is the text "A Nihon Kohden Company".
#### Intended Use
The Polysmith Sleep System. Model NTI7593 is intended to measure, amplify, and record physiological signals acquired from a patient for archival in a sleep study. The physiological signals are recorded and conditioned for analysis and display. The data may be analyzed in realtime or offline on dedicated polysomnography software running on a personal computer by a qualified sleep clinician to aid in the diagnosis of Sleep Disorders.
The Polysmith Sleep System, Model NTI7593 is intended for use by healthcare professionals within a hospital, laboratory, clinic, or nursing home; or outside of a medical facility under direct supervision of a medical professional.
The Polysmith Sleep System. Model NTI7593 is intended for use on both adults and children only under the direction of a physician or qualified sleep technician.
The Polysmith Sleep System, Model NTI7593 includes automatic detection of some arrhythmias (including some potentially life threatening arrhythmias), however detection of an arrhythmia may be performed with 30 seconds or more delay, and is based on a single ECG lead only.
The Polysmith Sleep System, Model NT17593, or any accessory, is not intended for the life monitoring of high risk patients, does not include or trigger alarms, and is not intended to be used alone as, or a critical component of,
- an alarm or alarm system;
- an apnea monitor or apnea monitoring system; or
- life monitor or life monitoring system. ●
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Neurotronics, a Nihon Kohden Company. The logo features a blue circle with a white line running through it, resembling a heartbeat. To the right of the circle is the company name, "Neurotronics", in blue text, and below that is the text "A Nihon Kohden Company".
# Predicate Comparison
#### INTENDED USE COMPARISON
The Polysmith Sleep System, Model NTI7593 improves upon the original model's current capabilities of data analysis for arrhythmia related events. These improvements do not change the indications for use for the devices which it is an accessory. This device continues to be intended to aid in diagnosis of sleep disorders, and other related disorders, as arrhythmia events are identified as related disorders by the American Academy of Sleep Medicine. The change clarifies the functionality of the arrhythmia detection feature and that the feature is not intended for life monitoring of high-risk patients.
| Polysmith Sleep System, NTI7593 (NEW) | Polysmith Sleep System, NTI6600 (K142774) |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| The Polysmith Sleep System, Model NTI7593 is intended to<br>measure, amplify, and record physiological signals acquired<br>from a patient for archival in a sleep study. The<br>physiological signals are recorded and conditioned for<br>analysis and display. The data may be analyzed in real-time<br>or offline on dedicated polysomnography software running<br>on a personal computer by a qualified sleep clinician to aid<br>in the diagnosis of Sleep Disorders. | The device is intended to measure, amplify, and<br>record physiological signals acquired from a patient<br>for archival in a sleep study. The physiological<br>signals are recorded and conditioned for analysis and<br>display. The data may be analyzed in real-time or<br>offline on dedicated polysomnography software<br>running on a personal computer by a qualified sleep<br>clinician to aid in the diagnosis of sleep disorders. |
| The Polysmith Sleep System, Model NTI7593 is intended<br>for use by healthcare professionals within a hospital,<br>laboratory, clinic, or nursing home; or outside of a medical<br>facility under direct supervision of a medical professional. | The device is intended for use by healthcare<br>professionals within a hospital, laboratory, clinic, or<br>nursing home; or outside of a medical facility under<br>direct supervision of a medical professional. |
| The Polysmith Sleep System, Model NTI7593 is intended<br>for use on both adults and children only under the direction<br>of a physician or qualified sleep technician. | The device is intended for use on both adults and<br>children only under the direction of a physician or<br>qualified sleep technician. |
| The Polysmith Sleep System, Model NTI7593 includes<br>automatic detection of some arrhythmias (including some<br>potentially life threatening arrhythmias), however detection<br>of an arrhythmia may be performed with 30 seconds or | The device, or any accessory, does not include<br>alarms, and is not intended to be used as a critical<br>component of an alarm system. |
| more delay, and is based on a single ECG lead only. | The device, or any accessory, is not to be used alone |
| The Polysmith Sleep System, Model NTI7593, or any<br>accessory, is not intended for the life monitoring of high risk | as an apnea monitor or as a component in an apnea |
| patients, does not include or trigger alarms, and is not<br>intended to be used alone as, or a critical component of, | monitoring system. |
| an alarm or alarm system;an apnea monitor or apnea monitoring system; orlife monitor or life monitoring system. | The device, or any accessory, is not to be used alone<br>as a life support device or as a critical component of<br>a life support system. |
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATES
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for Neurotronics, a Nihon Kohden Company. The logo features a blue circle with a white waveform inside, followed by the company name "Neurotronics" in blue, sans-serif font. Below the company name, the text "A Nihon Kohden Company" is written in a smaller, sans-serif font, underlined by a curved line.
The Polysmith Sleep System, Model NT17593 includes and improves upon the same feature set as found in the predicate software. Therefore, Neurotronics believes the Polysmith Sleep System, Model NT17593 is substantially equivalent to the Polysmith Sleep System, Model NTI6600 (K142774), and the arrhythmia analysis performance of the reference device, BSM 9100A Series Nihon Kohden Bedside Monitor (K082785).
#### Software Technical Comparison
| Software Technical Comparison | Polysmith Sleep<br>System, NTI7593<br>Polysmith<br>(NEW) | Polysmith Sleep<br>System, NTI6600<br>Polysmith<br>(K142774) |
|-----------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|--------------------------------------------------------------|
| Display electrical activity of the brain and other<br>physiological signals on a monitor | Yes | Yes |
| Device uses Windows on a 32-bit or 64-bit<br>computer | 32-bit & 64-bit | 32-bit & 64-bit |
| Display waveforms in acquisition and review | Yes | Yes |
| Maximum number of waveforms on screen | 32 | 32 |
| Simultaneously view waveforms from multiple<br>EEG/PSG instruments and display values from<br>external instruments | On-line & Off-line | On-line & Off-line |
| Variable waveform display duration | On-line & Off-line | On-line & Off-line |
| Change waveform parameters (pattern, montage,<br>amplifier conditions, AV induction) | On-line & Off-line | On-line & Off-line |
| The device can record and measure vital signs<br>(including ECG, EMG, Respiration, Ocular<br>Motility, SpO2, and CO2) | On-line & Off-line | On-line & Off-line |
| Nihon Kohden Amplifiers Compatibility | Yes | Yes |
| Lifelines Trackit Series Amplifiers Compatibility | Yes | Yes |
| USB DC Box Compatibility | Yes | Yes |
| Sleep Stage Analysis | On-line & Off-line | On-line & Off-line |
| Arrhythmia Analysis | On-line & Off-line | Off-line<br>(integrated K012686) |
| Respiratory Event Analysis | On-line & Off-line | On-line & Off-line |
| SpO2 Desaturation Analysis | On-line & Off-line | On-line & Off-line |
| Limb Movement Event Analysis | On-line & Off-line | On-line & Off-line |
| Snore Event Analysis | On-line & Off-line | On-line & Off-line |
| FFT Analysis | Off-line | Off-line |
| Manual Event Scoring | On-line & Off-line | On-line & Off-line |
| Manual Custom Event Scoring | On-line & Off-line | On-line & Off-line |
| Trend Plots | On-line & Off-line | On-line & Off-line |
| Display patient video along with waveforms | On-line & Off-line | On-line & Off-line |
| Record Patient Audio and Video | Yes | Yes |
| Software Technical Comparison | Polysmith Sleep<br>System, NTI7593<br>Polysmith<br>(NEW) | Polysmith Sleep<br>System, NTI6600<br>Polysmith<br>(K142774) |
| Integrated Intercom Interface | Yes<br>(with PMU710) | Yes<br>(with PMU710) |
| For use by medical personnel in a medical facility,<br>physician's office, laboratory, clinic, or nursing<br>home | Yes | Yes |
| Any patient population including adults and<br>children as determined by a trained professional | Yes | Yes |
| Patient Database and Scheduling | Yes | Yes |
| Reporting and Custom Report Capabilities | Yes | Yes |
| HL7 Interface Capabilities | Yes | Yes |
| Remote Access Capabilities | On-line & Off-line | On-line & Off-line |
| Display Resolution | >=1600 x 1200 | >=1600 x 1200 |
| File Format | XDF (XML), EDF | XDF (XML), EDF |
| File Management | Yes | Yes |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Neurotronics, a Nihon Kohden Company. The logo consists of a blue circle with a white line running through it, resembling a heartbeat. To the right of the circle is the word "Neurotronics" in blue, and below that is the text "A Nihon Kohden Company".
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Neurotronics. The logo consists of a blue circle with a white line running through it, resembling a heartbeat. To the right of the circle is the company name, "Neurotronics," in blue font.
A Nihon Kohden Company
| Software Technical Comparison | Polysmith Sleep<br>System, NT17593<br>Polysmith<br>(NEW) | Polysmith Sleep<br>System, NTI6600<br>Polysmith<br>(K142774) |
|-------------------------------------|-----------------------------------------------------------------------|-----------------------------------------------------------------------|
| Compatible Amplifiers for Recording | Neurotronics Sphinx PMU710 (K142774) | Neurotronics Sphinx PMU710 (K142774) |
| | Neurotronics Sphinx PMU700 (K062943) | Neurotronics Sphinx PMU700 (K062943) |
| | Neurotronics Nomad PMU800 (K092699) | Neurotronics Nomad PMU800 (K092699) |
| | Nihon Kohden PSG-1100 (K120888) | Nihon Kohden PSG-1100 (K120888) |
| | Nihon Kohden EEG-1200a Amplifiers (K113117, K080546) | Nihon Kohden EEG-1200a Amplifiers (K113117, K080546) |
| | Nihon Kohden Wireless Input Unit, Model WEE-1000a Series (K033475) | Nihon Kohden Wireless Input Unit, Model WEE-1000a Series (K033475) |
| | Nihon Kohden Electrode Junction Box, Model JE-921a (K050833) | Nihon Kohden Electrode Junction Box, Model JE-921a (K050833) |
| | Nihon Kohden PSG Input Box, Model JE-912ak (K022121) | Nihon Kohden PSG Input Box, Model JE-912ak (K022121) |
| | Nihon Kohden Neurofax, Models EEG-1100a, EEG-9100a (K011204, K992742) | Nihon Kohden Neurofax, Models EEG-1100a, EEG-9100a (K011204, K992742) |
| | Lifelines Trackit Series amplifiers (K010460) | Lifelines Trackit Series amplifiers (K010460) |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for Neurotronics, a Nihon Kohden Company. The logo features a blue circle with a white line running through it, resembling a heartbeat. To the right of the circle is the company name, "Neurotronics," in blue text. Below the company name is the text "A Nihon Kohden Company" in a smaller font size.
#### Arrhythmia Analysis Event Comparison
Events detected by the third-party holter analysis software used with Polysmith Sleep System model NTI6600 compared to the integrated algorithm used in the new Polysmith software in model NTI7593. The newly integrated algorithm is the same algorithm implemented in the reference device, BSM-9100A (K082785) for applicable event types. This software allows the subject device to perform real-time arrhythmia analysis (with a 30 second or more delay for the applicable event types listed below).
| Analysis Events | Polysmith<br>Sleep<br>Svstem,<br>NTI6600<br>(K142774) | Polysmith<br>Sleep<br>System,<br>NTI7593<br>(NEW) | Nihon Kohden<br>Bedside<br>Monitor<br>BSM-9100A<br>(K082785) |
|--------------------------------------------------------------|-------------------------------------------------------|---------------------------------------------------|--------------------------------------------------------------|
| Pause | Yes | Yes | Yes |
| Sinus Tachycardia | Yes | Yes | Yes |
| Extreme Tachycardia | No | No | Yes |
| Bradycardia | Yes | Yes | Yes |
| Extreme Bradycardia | No | No | Yes |
| Wide-Complex Tachycardia<br>(Ventricular tachycardia) | Yes | Yes | Yes |
| Narrow-Complex Tachycardia<br>(Supraventricular tachycardia) | Yes | Yes | Yes |
| Atrial Fibrillation (Irregular R-R) | No | Yes | Yes |
| Asystole | Yes | Yes | Yes |
| PVC | Yes | Yes | Yes |
| Early PVC | Yes | Yes | Yes |
| PVC Run | No | No | Yes |
| Frequent PVC | No | No | Yes |
| Multiform PVC | No | No | Yes |
| Idioventricular<br>Rhythm (Ventricular<br>Bradycardia) | No | Yes | Yes |
| Accelerated Idioventricular<br>Rhythm (Ventricular Rhythm) | No | Yes | Yes |
| Bigeminy | Yes | Yes | Yes |
| Trigeminy | No | Yes | Yes |
| Couplet | Yes | Yes | Yes |
| Ventricular Fibrillation | No | Yes | Yes |
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for Neurotronics, which is a company of Nihon Kohden. The logo features a blue circle with a white line running through it, resembling a heartbeat. The text "Neurotronics" is written in blue next to the circle, and below that is the text "A Nihon Kohden Company".
| Analysis Events | Polysmith<br>Sleep<br>System,<br>NTI6600<br>(K142774) | Polysmith<br>Sleep<br>System,<br>NTI7593<br>(NEW) | Nihon Kohden<br>Bedside<br>Monitor<br>BSM-9100A<br>(K082785) |
|--------------------------------|-------------------------------------------------------|---------------------------------------------------|--------------------------------------------------------------|
| Prolonged R-R | No | No | Yes |
| No Pacer Pulse | No | No | Yes |
| Pacer Non-Capture | No | No | Yes |
| ST Level | Yes | No | Yes |
| SupraVentricular Ectopic (SVE) | Yes | No | No |
# Non-Clinical Performance Data
#### APPLIED STANDARDS
The following applied standards are related to the integration of the arrhythmia analysis algorithm within the Polysmith software. All other applied standards from the previously registered Polysmith Sleep System, Model NTI6600 (K142774) remain unchanged.
| Standards<br>Number | Standards<br>Organization | Standards Title |
|---------------------|---------------------------|-------------------------------------------------------------------------------------------------------|
| 62304 | AAMI ANSI IEC | Medical device software - Software life cycle processes |
| 14971 | AAMI ANSI ISO | Medical devices - Applications of risk management to<br>medical devices |
| 62366 | AAMI ANSI IEC | Medical devices - Application of usability engineering to<br>medical devices |
| EC57 | AAMI ANSI | Testing And Reporting Performance Results of Cardiac<br>Rhythm And St-Segment Measurement Algorithms. |
#### PERFORMANCE RESULTS
Arrhythmia Analysis verification was performed in a manner consistent with the applicable portions of the AAMI ANSI EC57:2012, Testing and Reporting Performance Results of Cardiac Rhythm and St-Segment Measurement Algorithms standard.
The arrhythmia analysis results are to be reviewed by a qualified clinician and are not to serve as results for direct diagnosis or treatment of arrhythmia conditions. The arrhythmia analysis function is only to provide indicators for further investigation.
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the logo for Neurotronics, a Nihon Kohden Company. The logo features a blue circle with a white waveform inside, followed by the word "Neurotronics" in blue. Below the word is a blue line and the text "A Nihon Kohden Company".
QRS & VPC Detection Performance
| | | Polysmith Sleep System, NTI7593<br>(NEW) | | |
|--------------------------|-----|------------------------------------------|--------|--------|
| | | MIT | AHA | NST |
| QRS | Se | 99.56% | 99.56% | 98.10% |
| | +P | 99.67% | 99.80% | 84.78% |
| Ventricular Ectopic Beat | Se | 91.85% | 85.46% | 88.28% |
| | +P | 96.94% | 98.18% | 90.94% |
| | FPR | 0.220% | 0.163% | 0.881% |
MIT: The Massachusetts Institute of Technology–Beth Israel Hospital Arrhythmia Database AHA: The American Heart Association Database for Evaluation of Ventricular Arrhythmia Detectors NST: The Noise Stress Test Database +P: Predictivity Se: Sensitivity FPR: False Positive Rate
#### Arrhythmia Detection Statistical Summary
| | | Polysmith Sleep System,<br>NTI7593<br>(NEW) | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|---------------------------------------------|-----|
| Bigeminy | E Se | 90% | |
| | E +P | 91% | |
| Trigeminy | E Se | 87% | |
| | E +P | 64% | |
| R-on-T | Se | 68.17% | |
| | +P | 65.54% | |
| | FPR | 0.267% | |
| Asystole | E Se | 100% | |
| | | Polysmith Sleep System,<br>NTI7593<br>(NEW) | |
| | E +P | 100% | |
| Idioventricular Rhythm | E Se | 50% | |
| | E +P | 100% | |
| Accelerated Idioventricular Rhythm | E Se | 62% | |
| | E +P | 57% | |
| | E Se | 94% | |
| Ventricular Fibrillation | E +P | 88% | |
| | D Se | 83% | |
| | D +P | 94% | |
| Ventricular Tachycardia | E Se | 100% | |
| | E +P | 86% | |
| Supraventricular Tachycardia | E Se | 77% | |
| | E +P | 18% | |
| Pause | E Se | 100% | |
| | E +P | 92% | |
| Couplet | E Se | 83% | |
| | E +P | 96% | |
| | Avg E Se | 62.00% | |
| | Avg E +P | 70.00% | |
| Atrial Fibrillation | E Se | 91% | |
| | | Polysmith Sleep System,<br>NTI7593<br>(NEW) | |
| | E +P | 47% | |
| | | D Se | 75% |
| | | D +P | 57% |
| | | NST E Se | - |
| | | NST E +P | 0% |
| | | NST D Se | - |
| | | NST D +P | 0% |
| Short Run Premature Ventricular Contraction<br>(PVC)1 | | E Se | 78% |
| | | E +P | 95% |
| | | Avg E Se | 60% |
| | | Avg E +P | 80% |
| Long Run Premature Ventricular Contraction<br>(PVC)1 | | E Se | 39% |
| | | E +P | 94% |
| | | Avg E Se | 71% |
| | | Avg E +P | 91% |
| 1 PVC Runs are not displayed or reported in Polysmith Software. | | | |
| E: Episode<br>D: Duration<br>+P: Predictivity, the fraction of detections that are events<br>Se: Sensitivity, the fraction of events that are detected<br>Avg: Average<br>NST: Noise Stress Test Database | | | |
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the logo for Neurotronics, a Nihon Kohden Company. The logo consists of a blue circle with a white line running through it, resembling a heartbeat. To the right of the circle is the word "Neurotronics" in blue, and below that is the text "A Nihon Kohden Company".
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the logo for Neurotronics, a Nihon Kohden Company. The logo consists of a blue circle with a white line running through it, resembling a heartbeat. To the right of the circle is the word "Neurotronics" in blue, and below that is the text "A Nihon Kohden Company".
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image contains the logo for Neurotronics, a Nihon Kohden Company. The logo features a blue circle with a white line running through it, resembling a heartbeat. To the right of the circle is the word "Neurotronics" in blue, with the text "A Nihon Kohden Company" underneath.
#### Results Discussion
Testing shows that the arrhythmia detection results are comparable to the already registered reference device, BSM 9100A Series Nihon Kohden Bedside Monitor (K082785), which includes the same algorithm. Since the BSM-9100A is much more critical in nature, comparable results are considered acceptable for the scope and intended use of the Polysmith software.
#### Conclusion
Based on the results of the Intended Use Comparison, the Technical Comparison, and Testing Data, it is believed that the Polysmith Sleep System, Model NT17593 presents no new questions of safety and effectiveness and, is substantially equivalent to the features provided by the identified predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.