Poseidon OCT Spinal Fixation System

K161637 · Medyssey USA, Inc. · NKG · Jan 5, 2017 · Orthopedic

Device Facts

Record IDK161637
Device NamePoseidon OCT Spinal Fixation System
ApplicantMedyssey USA, Inc.
Product CodeNKG · Orthopedic
Decision DateJan 5, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3075
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Poseidon OCT Spinal Fixation System is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1 to C7) and the thoracic spine (T1 to T3): traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g., pseudoarthrosis); tumors involving the cervical/thoracic spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The Poseidon OCT Spinal Fixation System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion. In order to achieve additional levels of fixation, the Poseidon OCT Spinal Fixation System may be connected to the Medyssey Zenius, Iliad or Kora Spinal Systems using parallel connectors and taper rods.

Device Story

Posterior spinal fixation system for occipito-cervical-thoracic use; provides rigid, single or multilevel stabilization. System components include screws, rods, and connectors; used to immobilize spinal segments as adjunct to fusion or for temporary stabilization in advanced tumor cases. Operates via mechanical fixation of spinal segments; allows connection to Medyssey Zenius, Iliad, or Kora systems via parallel connectors and taper rods for extended fixation. Used by surgeons in clinical/OR settings. Output is mechanical stabilization of the spine; assists in treating fractures, deformities, and degenerative conditions; benefits patients by restoring spinal integrity and reducing pain/instability.

Clinical Evidence

Bench testing only. No clinical data provided. Mechanical testing performed per ASTM F2706 (static/dynamic axial compression, static torsion, static tension) and ASTM F1798 (static/dynamic interconnection mechanism and subassemblies).

Technological Characteristics

Posterior spinal fixation system; rigid, multi-component assembly. Materials and design compared to predicates. Mechanical testing conducted per ASTM F2706 and ASTM F1798. System allows modular connectivity to other Medyssey spinal systems via parallel connectors and taper rods.

Indications for Use

Indicated for patients with acute/chronic instabilities of craniocervical junction, cervical spine (C1-C7), and thoracic spine (T1-T3) requiring immobilization as an adjunct to fusion. Conditions include traumatic fractures/dislocations, instability/deformity, failed fusions, tumors, and degenerative disease (radiculopathy, myelopathy, discogenic pain, facet instability). Also indicated for spinal column integrity restoration in patients with advanced cervical tumors with limited life expectancy where fusion is not achievable.

Regulatory Classification

Identification

Posterior cervical screw systems are comprised of multiple, interconnecting components, made from a variety of materials that allow an implant system to be built from the occiput to the upper thoracic spine to fit the patient's anatomical and physiological requirements, as determined by preoperative cross-sectional imaging. Such a spinal assembly consists of a combination of bone anchors via screws (i.e., occipital screws, cervical lateral mass screws, cervical pedicle screws, C2 pars screws, C2 translaminar screws, C2 transarticular screws), longitudinal members (e.g., plates, rods, including dual diameter rods, plate/rod combinations), transverse or cross connectors, interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors), and closure mechanisms (e.g., set screws, nuts). Posterior cervical screw systems are rigidly fixed devices that do not contain dynamic features, including but not limited to: non-uniform longitudinal elements or features that allow more motion or flexibility compared to rigid systems.Posterior cervical screw systems are intended to provide immobilization and stabilization of spinal segments in patients as an adjunct to fusion for acute and chronic instabilities of the cervical spine and/or craniocervical junction and/or cervicothoracic junction such as: (1) Traumatic spinal fractures and/or traumatic dislocations; (2) deformities; (3) instabilities; (4) failed previous fusions (e.g., pseudarthrosis); (5) tumors; (6) inflammatory disorders; (7) spinal degeneration, including neck and/or arm pain of discogenic origin as confirmed by imaging studies (radiographs, CT, MRI); (8) degeneration of the facets with instability; and (9) reconstruction following decompression to treat radiculopathy and/or myelopathy. These systems are also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

Special Controls

*Classification.* Class II (special controls). The special controls for posterior cervical screw systems are:(1) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (2) Nonclinical performance testing must demonstrate the mechanical function and durability of the implant. (3) Device components must be demonstrated to be biocompatible. (4) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (5) Labeling must include the following: (i) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (ii) Intended use and indications for use including levels of fixation; (iii) Device specific warnings, precautions, and contraindications that include the following statements: (A) “Precaution: Preoperative planning prior to implantation of posterior cervical screw systems should include review of cross-sectional imaging studies ( *e.g.,* CT and/or MRI) to evaluate the patient's cervical anatomy including the transverse foramen, neurologic structures, and the course of the vertebral arteries. If any findings would compromise the placement of these screws, other surgical methods should be considered. In addition, use of intraoperative imaging should be considered to guide and/or verify device placement, as necessary.”(B) “Precaution: Use of posterior cervical pedicle screw fixation at the C3 through C6 spinal levels requires careful consideration and planning beyond that required for lateral mass screws placed at these spinal levels, given the proximity of the vertebral arteries and neurologic structures in relation to the cervical pedicles at these levels.” (iv) Identification of magnetic resonance (MR) compatibility status; (v) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user, and; (vi) Detailed instructions of each surgical step, including device removal.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a symbol and text arranged in a circular format. The symbol features a stylized representation of three human profiles facing right, stacked on top of each other, with flowing lines suggesting movement or connection. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper half of the circle, following its curvature. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 5, 2017 Medyssey USA, Inc. % Rich Jansen, Pharm.D. Consultant Silver Pine Consulting, LLC 11821 Bramble Cove Drive Fort Myers, Florida 33905 Re: K161637 Trade/Device Name: Poseidon OCT Spinal Fixation System Regulatory Class: Unclassified Product Code: NKG, KWP Dated: December 2, 2016 Received: December 5, 2016 Dear Dr. Jansen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely. # Lori A. Wiggins -S for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K161637 Device Name Poseidon OCT Spinal Fixation System #### Indications for Use (Describe) The Poseidon OCT Spinal Fixation System is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1 to C7) and the thoracic spine (T1 to T3): traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g., pseudoarthrosis); tumors involving the cervical/thoracic spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The Poseidon OCT Spinal Fixation System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion. In order to achieve additional levels of fixation, the Poseidon OCT Spinal Fixation System may be connected to the Medyssey Zenius, Iliad or Kora Spinal Systems using parallel connectors and taper rods. Type of Use (Select one or both, as applicable) | <div> <span> <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | |--------------------------------------------------------------------------------------------------------------------| | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary | Date Prepared: | January 4, 2016 | |---------------------|------------------------------------------| | Submitter: | Shawn Kim | | | Medyssey USA, Inc. | | | 1550 E. Higgins Road | | | Suite 123 | | | Skokie, IL 60077 | | | 847-427-0200 | | | 888-518-9070 (fax) | | Regulatory Contact: | Rich Jansen, Pharm. D. | | | Silver Pine Consulting, LLC | | Product | | | Trade Names: | Poseidon OCT Spinal Fixation System | | Product Class: | Unclassified | | Classification: | Unclassified | | Common Name: | Posterior, Cervical Pedicle Screw System | | Product Codes: | NKG, KWP | | Panel Code: | 87 | #### Device Description: The Poseidon OCT Spinal Fixation System is a posterior fixation system designed for occipito-cervicalthoracic use for single and/or multilevel rigid fixation to treat occipital and/or cervical conditions of the spine. #### Indications for Use: The Poseidon OCT Spinal Fixation System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1 to C7) and the thoracic spine (T1 to T3): traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g., pseudoarthrosis); tumors involving the cervical/thoracic spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The Poseidon OCT Spinal Fixation System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion. In order to achieve additional levels of fixation, the Poseidon OCT Spinal Fixation System may be connected to the Medyssey Zenius, Iliad or Kora Spinal Systems using parallel connectors and taper rods. #### Predicate Device(s): Medyssey believes these devices are substantially equivalent to a predicate device previously cleared by FDA. The primary predicate device is the Synapse OCT System from DePuy/Synthes (K142838). One additional predicate device is the Vertex Reconstruction System from Medtronic (K143471). #### Performance Standards: The pre-clinical testing was conducted per ASTM F2706 and ASTM F1798. Testing included: - Static and dynamic axial compression, static torsion and static tension per ASTM F2706. - Static and dynamic interconnection mechanism and subassemblies per ASTM F1798. {4}------------------------------------------------ #### Technological Characteristics: Medyssey has compared the Poseidon OCT Spinal Fixation System to the predicate devices in regards to indications for use, materials, function, sizes and mechanical test results. These comparisons demonstrate substantial equivalence to the predicate devices. #### Conclusion: Medyssey concludes that the Poseidon OCT Spinal Fixation System is substantially equivalent to the predicate devices and raise no new questions of safety or effectiveness.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%