Xpert TV, Xpert Urine Specimen Collection Kit, GeneXpert Dx Systems (GX-I, GX-II, GX-IV, GX-XVI), GeneXpert Infinity-48, Genxpert Infinity-48s and GeneXpert Infinity-80 Systems
Applicant
Cepheid
Product Code
OUY · Immunology
Decision Date
Aug 29, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3860
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K161619 · Aug 29, 2016
Xpert TV, Xpert Urine Specimen Collection Kit, GeneXpert Dx Systems (GX-I, GX-II, GX-IV, GX-XVI), GeneXpert Infinity-48, Genxpert Infinity-48s and GeneXpert Infinity-80 Systems
Cepheid
Clinical specimens (male urine, endocervical swabs, patient-collected vaginal swabs) from patients in routine clinical settings (OB/GYN, STD, and family planning clinics)
The clinical study evaluated the performance (sensitivity, specificity, PPV, NPV) of the Xpert TV Assay in a multi-site prospective study of symptomatic and asymptomatic individuals in routine clinical practice.
Multi-site prospective investigational study; Prospective multi-site clinical study
Symptomatic and asymptomatic, sexually active males and females seen at OB/GYN, STD, and family planning clinics; Sample Size: 10,017 tests performed; Number of Sites: 17 clinical sites
Patient infected status (PIS) algorithm (culture and validated bidirectional sequencing for male urine; FDA cleared NAAT and culture for female specimens)
Sensitivity, specificity, PPV, and NPV
Indications for Use
The Cepheid Xpert TV Assay, performed on the GeneXpert® Instrument Systems, is a qualitative in vitro diagnostic test for the detection of Trichomonas vaginalis genomic DNA. The test utilizes automated real-time polymerase chain reaction (PCR) to detect Trichomonas vaginalis genomic DNA. The Xpert TV Assay uses female and male urine specimens, endocervical swab specimens, and patient-collected vaginal swab specimens (collected in a clinical setting). The Xpert TV Assay is intended to aid in the diagnosis of trichomoniasis in symptomatic or asymptomatic individuals. Ancillary Collection Kits: Xpert Vaginal/Endocervical Specimen Collection Kit The Cepheid® Xpert® Vaginal/Endocervical Specimen Collection Kit is designed to collect, preserve, and transport Chlamydia trachomatis, Neisseria gonorrhoeae, and Trichomonas vaginalis DNA in endocervical swab specimens (collected by a clinician) and patient-collected vaginal swab specimens (collected in a clinical setting) from symptomatic and asymptomatic women prior to analysis with the Xpert CT/NG Assay and the Xpert TV Assay. Xpert Urine Specimen Collection Kit The Cepheid® Xpert® Urine Specimen Collection Kit is designed to preserve and transport Chlamydia trachomatis, Neisseria gonorrhoeae, and Trichomonas vaginalis DNA in first-catch female and male urine specimens from symptomatic and asymptomatic individuals prior to analysis with the Xpert CT/NG Assay and the Xpert TV Assay.
Device Story
The Xpert TV Assay is an automated, qualitative in vitro diagnostic test for Trichomonas vaginalis genomic DNA. It uses single-use, multi-chambered fluidic cartridges containing PCR reagents. The user places a specimen (urine or swab) into the cartridge and inserts it into a GeneXpert Instrument System (GeneXpert Dx, Infinity-48, Infinity-48s, or Infinity-80). The system automates sample preparation, real-time PCR amplification, and detection within 70 minutes. The instrument uses an ultrasonic horn for cell lysis (though not used for this specific assay) and a proprietary I-CORE thermocycler. Results are automatically generated and displayed on the system interface. The device is used in clinical settings by laboratory personnel. It aids in the diagnosis of trichomoniasis, providing clinicians with rapid, automated molecular results to guide patient management.
Clinical Evidence
Prospective multicenter study of 4,611 eligible male subjects (symptomatic and asymptomatic). Performance evaluated against Patient Infected Status (PIS) using InPouch TV culture and bi-directional sequencing. Overall sensitivity 89.6% (95% CI: 83.0%-93.8%); overall specificity 99.3% (95% CI: 99.0%-99.5%). Assay success rate 99.7%.
Technological Characteristics
Real-time PCR assay; uses single-use, multi-chambered fluidic cartridges. Platform: GeneXpert Instrument Systems (GeneXpert Dx, Infinity-48, Infinity-48s, Infinity-80). Features: Sample Processing Control (SPC), Sample Adequacy Control (SAC), and Probe Check Control (PCC). Automated sample prep and thermocycling via I-CORE module. Connectivity: Standalone instrument system.
Indications for Use
Indicated for the detection of Trichomonas vaginalis genomic DNA in male and female urine, endocervical swabs, and patient-collected vaginal swabs to aid in the diagnosis of trichomoniasis in symptomatic or asymptomatic individuals.
Regulatory Classification
Identification
A Trichomonas vaginalis nucleic acid assay is a device that consists of primers, probes, enzymes, and controls for the amplification and detection of trichomonas nucleic acids in endocervical swabs, vaginal swabs, and female urine specimens, from women symptomatic for vaginitis, cervicitis, or urethritis and/or to aid in the diagnosis of trichomoniasis in asymptomatic women. The detection of trichomonas nucleic acids, in conjunction with other laboratory tests, aids in the clinical laboratory diagnosis of trichomoniasis caused by Trichomonas vaginalis.
Special Controls
*Classification.* Class II (special controls). The special controls are set forth in FDA's guideline document entitled: “Class II Special Controls Guideline: Nucleic Acid Amplification Assays for the Detection of*Trichomonas vaginalis;* Guideline for Industry and Food and Drug Administration Staff.” See § 866.1(e) for information on obtaining this document.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION MEMORANDUM
A. 510(k) Number:
K161619
B. Purpose for Submission:
The purpose of this submission is to add male urine specimens for use with the Xpert® Trichomonas vaginalis (TV) assay, which has already been cleared for female urine, endocervical swabs, and patient collected vaginal swabs under K151565. The original clinical study in K151565 included both male and female clinical specimens, but the male and female claims are addressed in two separate submissions. The male urine clinical data is submitted in the present file (K161619). All analytical studies on the urine specimen type were submitted and reviewed under K151565 unless otherwise specified.
C. Measurand:
Trichomonas vaginalis (TV) DNA
D. Type of Test:
Nucleic acid amplification test using real-time polymerase chain reaction (PCR)
E. Applicant:
Cepheid
F. Proprietary and Established Names:
Proprietary Name: Xpert® TV
Common names: Xpert® TV assay, Xpert® Trichomonas Assay
G. Regulatory Information:
1. Regulation section:
21 CFR 866.3860, Trichomonas vaginalis nucleic acid assay
2. Classification:
Class II
{1}
2
3. Product code:
OUY- Trichomonas vaginalis nucleic acid amplification test
OOI- Real time nucleic acid amplification system
4. Panel:
Microbiology (83)
H. Intended Use:
1. Intended use(s):
The Cepheid Xpert TV Assay, performed on the GeneXpert® Instrument Systems, is a qualitative *in vitro* diagnostic test for the detection of *Trichomonas vaginalis* genomic DNA. The test utilizes automated real-time polymerase chain reaction (PCR) to detect *Trichomonas vaginalis* genomic DNA. The Xpert TV Assay uses female and male urine specimens, endocervical swab specimens, and patient-collected vaginal swab specimens (collected in a clinical setting). The Xpert TV Assay is intended to aid in the diagnosis of trichomoniasis in symptomatic or asymptomatic individuals.
Ancillary Collection Kits:
Xpert Vaginal/Endocervical Specimen Collection Kit
The Cepheid® Xpert® Vaginal/Endocervical Specimen Collection Kit is designed to collect, preserve, and transport *Chlamydia trachomatis*, *Neisseria gonorrhoeae*, and *Trichomonas vaginalis* DNA in endocervical swab specimens (collected by a clinician) and patient-collected vaginal swab specimens (collected in a clinical setting) from symptomatic and asymptomatic women prior to analysis with the Xpert CT/NG Assay and the Xpert TV Assay.
Xpert Urine Specimen Collection Kit
The Cepheid® Xpert® Urine Specimen Collection Kit is designed to preserve and transport *Chlamydia trachomatis*, *Neisseria gonorrhoeae*, and *Trichomonas vaginalis* DNA in first-catch female and male urine specimens from symptomatic and asymptomatic individuals prior to analysis with the Xpert CT/NG Assay and the Xpert TV Assay.
2. Indication(s) for use:
Same as intended use
{2}
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
The Cepheid Xpert TV Assay uses PCR technology on the GeneXpert® Instrument Systems, which extract, amplify, and detect the target DNA
I. Device Description:
The Xpert TV Assay is an automated real-time polymerase chain reaction (PCR) in vitro diagnostic test for the qualitative detection of genomic DNA from Trichomonas vaginalis. The Xpert TV Assay is intended as an aid in the diagnosis of trichomoniasis.
The Xpert TV Assay is performed on the Cepheid GeneXpert® Instrument Systems (GeneXpert Dx, GeneXpert Infinity-48, GeneXpert Infinity-48s, and GeneXpert Infinity-80 systems). The GeneXpert Instrument System consists of an instrument, personal computer, and preloaded software for running tests and viewing the results. The GeneXpert Instrument System requires single-use, disposable cartridges (the Xpert TV cartridges) that hold the PCR reagents and host the PCR process. The cartridges are self-contained, so specimens never come into contact with the working parts of the instrument.
A Sample Processing Control (SPC), Sample Adequacy Control (SAC), and a Probe Check Control (PCC) are controls utilized by the GeneXpert Instrument System. The SPC is present to control for adequate processing of the target trichomonads and to monitor the presence of inhibitors in the real-time PCR reaction to reduce the possibility of false negative results. The SAC reagents detect the presence of a single copy human gene and monitor whether the specimen contains human cells. The PCC verifies reagent rehydration, real-time PCR tube filling in the cartridge, probe integrity, and dye stability.
The ancillary specimen collection kits for use with the Xpert TV Assay are the Cepheid Xpert Vaginal/Endocervical Specimen Collection Kit and the Cepheid Xpert Urine Specimen Collection Kit. The swab and/or urine specimens are collected from asymptomatic or symptomatic patients and placed into a specimen transport tube containing preservative. After transferring the specimen to the sample chamber of the Xpert TV cartridge, the user initiates a test, and places the cartridge into the GeneXpert Instrument System. Results are automatically generated by the instrument at the end of the process in a report that can be viewed and printed.
J. Substantial Equivalence Information:
1. Predicate device name(s):
{3}
Cepheid Xpert TV assay
2. Predicate 510(k) number(s):
K151565
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Subject Device: Cepheid Xpert TV Assay (K161619) | Predicate Device: Cepheid Xpert TV Assay (K151565) |
| Intended Use | The Cepheid Xpert TV Assay, performed on the GeneXpert® Instrument Systems, is a qualitative in vitro diagnostic test for the detection of Trichomonas vaginalis genomic DNA. The test utilizes automated real-time polymerase chain reaction (PCR) to detect Trichomonas vaginalis genomic DNA. The Xpert TV Assay uses female and male urine specimens, endocervical swab specimens, and patient-collected vaginal swab specimens (collected in a clinical setting). The Xpert TV Assay is intended to aid in the diagnosis of trichomoniasis in symptomatic or asymptomatic individuals. Ancillary Collection Kits: Xpert Vaginal/Endocervical Specimen Collection Kit The Cepheid® Xpert® Vaginal/Endocervical Specimen Collection Kit is designed to collect, preserve, and transport Chlamydia trachomatis, | The Cepheid Xpert TV Assay, performed on the GeneXpert® Instrument Systems, is a qualitative in vitro diagnostic test for the detection of Trichomonas vaginalis genomic DNA. The test utilizes automated real-time polymerase chain reaction (PCR) to detect Trichomonas vaginalis genomic DNA. The Xpert TV Assay uses female urine specimens, endocervical swab specimens, or patient-collected vaginal swab specimens (collected in a clinical setting). The Xpert TV Assay is intended to aid in the diagnosis of trichomoniasis in symptomatic or asymptomatic individuals. Ancillary Collection Kits: Xpert Vaginal/Endocervical Specimen Collection Kit The Cepheid® Xpert® Vaginal/Endocervical Specimen Collection Kit is designed to collect, preserve, and transport Chlamydia trachomatis, Neisseria gonorrhoeae, and |
| | the Xpert TV Assay. The Xpert TV Assay is intended to aid in the diagnosis of trichomoniasis in symptomatic or asymptomatic individuals. Cervical Swab Specimens: The Xpert TV Assay is intended to aid in the diagnosis of trichomoniasis in symptomatic or asymptomatic individuals. Cervical Swab Specimens: The Xpert TV Assay is intended to aid in the diagnosis of trichomoniasis in symptomatic or asymptomatic individuals. | the Xpert TV Assay is intended to aid in the diagnosis of trichomoniasis in symptomatic or asymptomatic individuals. Cervical Swab Specimens: The Xpert TV Assay is intended to aid in the diagnosis of trichomoniasis in symptomatic or asymptomatic individuals. Cervical Swab Specimens: The Xpert TV Assay is intended to aid in the diagnosis of trichomoniasis in symptomatic or asymptomatic individuals. |
{4}
| Similarities | | |
| --- | --- | --- |
| Item | Subject Device: Cepheid Xpert TV Assay (K161619) | Predicate Device: Cepheid Xpert TV Assay (K151565) |
| | Neisseria gonorrhoeae, and Trichomonas vaginalis DNA in endocervical swab specimens (collected by a clinician) and patient-collected vaginal swab specimens (collected in a clinical setting) from symptomatic and asymptomatic women prior to analysis with the Xpert CT/NG Assay and the Xpert TV Assay.
Xpert Urine Specimen Collection Kit
The Cepheid® Xpert® Urine Specimen Collection Kit is designed to preserve and transport Chlamydia trachomatis, Neisseria gonorrhoeae, and Trichomonas vaginalis DNA in first-catch female and male urine specimens from symptomatic and asymptomatic individuals prior to analysis with the Xpert CT/NG Assay and the Xpert TV Assay. | Trichomonas vaginalis DNA in endocervical swab specimens (collected by a clinician) and patient-collected vaginal swab specimens (collected in a clinical setting) from symptomatic and asymptomatic women prior to analysis with the Xpert CT/NG Assay and the Xpert TV Assay.
Xpert Urine Specimen Collection Kit
The Cepheid® Xpert® Urine Specimen Collection Kit is designed to preserve and transport Chlamydia trachomatis, Neisseria gonorrhoeae, and Trichomonas vaginalis DNA in first-catch urine specimens from symptomatic and asymptomatic individuals prior to analysis with the Xpert CT/NG Assay and the Xpert TV Assay. The Xpert Urine Specimen Collection Kit is intended for use with male (Xpert CT/NG Assay) and female (Xpert CT/NG Assay and Xpert TV Assay) urine. |
| Assay Targets | Same | T. vaginalis genomic DNA |
| Nucleic Acid Extraction | Yes | Yes |
| Assay Results | Same | Qualitative |
| Collection Kit | Same | Urine collection kit
Swab collection kit |
{5}
| Similarities | | |
| --- | --- | --- |
| Item | Subject Device: Cepheid Xpert TV Assay (K161619) | Predicate Device: Cepheid Xpert TV Assay (K151565) |
| Technology/Detection | Same | Real-time polymerase chain reaction (PCR) |
| Instrument System | Same | Cepheid GeneXpert Instrument System |
| Laboratory Users | Same | Operators in Moderate and High Complexity Labs |
| Early Assay Termination function | Same | Yes (for positive Samples) |
| Differences | | |
| Item | Subject Device: Cepheid Xpert TV Assay (K161619) | Predicate Device: Cepheid Xpert TV Assay (K151565) |
| Specimen Types | Endocervical Swabs
Patient-collected Vaginal Swabs
Female Urine
Male Urine | Endocervical Swabs
Patient-collected Vaginal Swabs
Female Urine |
| Urine Collection Kit | Intended use:
The Cepheid® Xpert® Urine Specimen Collection Kit is designed to preserve and transport Chlamydia trachomatis, Neisseria gonorrhoeae, and Trichomonas vaginalis DNA in first-catch female and male urine specimens from symptomatic and asymptomatic individuals prior to analysis with the Xpert CT/NG Assay and the Xpert TV Assay. | Intended use:
The Cepheid® Xpert® Urine Specimen Collection Kit is designed to preserve and transport Chlamydia trachomatis, Neisseria gonorrhoeae, and Trichomonas vaginalis DNA in first-catch urine specimens from symptomatic and asymptomatic individuals prior to analysis with the Xpert CT/NG Assay and the Xpert TV Assay. The Xpert Urine Specimen Collection Kit is intended for use with male (Xpert CT/NG Assay) and female (Xpert CT/NG Assay and Xpert TV Assay) urine. |
{6}
K. Standard/Guidance Document Referenced (if applicable):
1. ASTM D4169-09, Standard Practice for Performance Testing of Shipping Containers and Systems.
2. CLSI EP5-A2, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline –Second Edition; 2004.
3. CLSI EP7-A2, Interference Testing in Clinical Chemistry; Approved Guideline – Second Edition; 2005.
4. CLSI EP15-A2, User Verification of Performance for Precision and Trueness; Approved Guideline – Second Edition; 2006.
5. CLSI EP17-A2, Protocols for Determination of Limits of Detection and Limits of Quantitation; Approved Guideline – Second Edition; 2006.
6. CLSI MM3-A2, Molecular Diagnostic Methods for Infectious Disease; Approved Guideline –Second Edition; 2006.
7. BS EN ISO 23640, Evaluation of Stability Testing of in vitro Diagnostic Reagent; 2013.
8. EN ISO 11135-1, Sterilization of health care products –Ethylene oxide – Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices; 2007.
9. FDA. Class II Special Controls Guideline: Nucleic Acid Amplification Assays for the Detection of Trichomonas vaginalis. Guideline for Industry and FDA Staff; 2015.
10. FDA. General Principles of Software Validation. Guidance for Industry and FDA Staff; 2002.
11. FDA. Format for Traditional and Abbreviated 510(k). Guidance for Industry and FDA Staff; 2005.
12. FDA. Class II Special Controls Guidance Document: Instrumentation for Clinical Multiplex Test Systems. Guidance for Industry and FDA Staff; 2005.
13. FDA. Content of Premarket Submissions for Management of Cybersecurity in Medical Devices. Guidance for Industry and FDA Staff; 2014.
14. FDA. Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices; Guidance for Industry and FDA Staff; 2005.
15. FDA. Cybersecurity for Networked Medical Devices Containing Off-the-Shelf (OTS) Software; Guidance for Industry and FDA Staff; 2005.
16. FDA. Guidance for Off-the-Shelf Software Use in Medical Devices. Guidance for Industry, FDA Reviewers and Compliance; 1999.
17. FDA. Guidance on Informed Consent for In Vitro Diagnostic Device Studies Using Leftover Human Specimens that are Not Individually Identifiable. Guidance for Industry and FDA Staff; 2006.
18. ISO 10993-1: Biological Evaluation of Medical Devices, Part 1: Evaluation and Testing; 2009.
19. ISO 10993-7: Biological Evaluation of Medical Devices, Part 7: Ethylene oxide sterilization residuals; 2008.
20. ISO 10993-10: Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization; 2010.
21. ISO 11737-2: Sterilization of Medical Devices – Microbiological Methods – Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization
7
{7}
process; 2009.
22. NF EN ISO 10993-5: Biological evaluation of medical devices Part 5: Tests for in vitro Cytotoxicity; 2010.
23. UNI EN 556-1: Sterilization Of Medical Devices - Requirements For Medical Devices To Be Designated "sterile" - Requirements for Terminally Sterilized Medical Devices; 2002.
## L. Test Principle:
This assay detects and amplifies Trichomonas vaginalis genomic DNA using real time polymerase chain reaction technology. The probe used in the assay is conjugated to a fluorescent reporter dye and quencher. Upon amplification, the reporter dye is released from the quencher, resulting in a fluorescent signal that is quantified using the GeneXpert instrument system.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
a. Precision/Reproducibility:
Reproducibility and within laboratory precision studies were previously reviewed and described in K151565.
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability: Not applicable
Stability testing of urine (both unprocessed and preserved in transport reagent) was previously reviewed and described in K151565.
The external and internal controls were previously reviewed and described in K151565.
d. Detection limit:
Limit of detection studies were previously reviewed and described in K151565.
e. Analytical specificity:
Inclusivity, microbial interference (cross-reactivity) and carryover/cross contamination studies were previously reviewed and described in K151565.
{8}
# Potentially Interfering Substances
Studies analyzing potentially interfering substances were previously reviewed and described in K15165. Due to the addition of the male urine specimen type in this submission, an additional interfering substance, seminal fluid, was also tested using the same protocol as in K151565. The presence of seminal fluid did not interfere with detection of Trichomonas vaginalis in the Xpert Trichomonas Assay. Table 1 lists the potentially interfering substances tested in urine specimen matrix from K151565 with the addition of seminal fluid.
Table 1: Potentially Interfering Substances Tested in Urine
| Substance/Class | Product | Active Ingredient | Concentration Tested |
| --- | --- | --- | --- |
| Blood | Blood | Blood | 0.3%, 1.0 v/v |
| Seminal Fluid | Seminal Fluid | Seminal Fluid | 5.0% v/v |
| Mucus | Mucin | Mucin | 0.8% w/v |
| Analgesics & Antibiotics | Aspirin | Acetylsalicylic acid 500mg | 40 mg/mL |
| | Tylenol | Acetaminophen | 3.2 mg/mL |
| | Azithromycin | Azithromycin | 1.8 mg/mL |
| | Doxycycline | Doxycycline | 3.6 mg/mL |
| OTC Deodorant & Powders | Norforms Feminine Deodorant Suppositories | PEG-20; PEG-32; PEG-20 Stearate | 0.25% w/v |
| | Vagisil Feminine Deodorant Powder | Nanoxynol – 9 | 0.25% w/v |
| Albumin | Albumin | BSA | 10 mg/ml |
| Glucose | Glucose | Glucose | 10 mg/ml |
| Bilirubin | Bilirubin | Bilirubin | 1 mg/ml |
| Acidic urine pH 4.0 | Urine | Urine + N Acetyl-LCysteine | pH 4.0 |
| Alkaline Urine (pH 9.0) | Urine | Urine + Ammonium Citrate | pH 9.0 |
| Leukocytes | Leukocytes | Leukocytes | 10^{5} cells/mL |
| Intraveginal Hormones | Estradiol and Progesterone | Estradiol; Progesterone | 7 mg/mL
Progesterone + 0.07 mg/mL
Beta Estradiol |
f. Assay cut-off:
The studies determining the assay cut-off were previously reviewed and described in K151565.
2. Comparison studies:
a. Method comparison with predicate device:
Not applicable.
{9}
b. Matrix comparison:
Not Applicable
3. Clinical studies:
a. Clinical Sensitivity and Specificity:
The clinical performance of the Xpert TV assay on female urine, endocervical swabs, and patient collected vaginal swabs has been previously reviewed under K151565.
To assess the clinical performance characteristics of the Xpert TV assay on male urine, a multicenter prospective study was performed to collect urine specimens from asymptomatic males and males presenting with symptoms potentially associated with TV infection. The performance of the Xpert TV Assay was evaluated relative to a Patient Infected Status (PIS) algorithm. The designation of a subject as being infected was based on the results of the reference tests, which consisted of InPouch TV culture and validated bi-directional sequencing. A positive result from either of these two reference tests signified a TV positive subject. The subject was considered to be not infected when both reference test results were negative.
The clinical study was performed between August 2014 and January 2015 and enrolled both male and female subjects. The clinical performance of the Xpert TV assay on the female specimen types (urine, endocervical swabs and patient-collected vaginal swabs) was previously reviewed and cleared in K151565. For the collection of male urine specimens, a total of 4465 male subjects were enrolled, of which 4458 were eligible for inclusion. The seven ineligible subjects included one subject with improper or incomplete informed consent, two subjects previously enrolled in the study, two subjects who had been treated with antibiotics within 21 days prior to study enrollment, one subject who was not sexually active, and one subject who was <14 years of age.
The clinical study was extended for male subjects only between October 2015 and March 2016 at three of the original study sites. The inclusion/exclusion criteria, study procedures, enrollment, assay procedures, quality control, bias elimination, data collection and sample size calculations were the same as described in K151565. A total of 333 additional male subjects were enrolled and eligible for inclusion.
A total of 4791 eligible male subjects (4458 from the original study and 333 subjects from the extended study) were enrolled. Of these, 165 were excluded due to 96 sample shipping delays, 36 samples tested more than 72 hours after collection, 21 Xpert modules out of calibration, and 12 samples not tested. This resulted in a total of 4626 specimens from eligible male subjects that were tested with the Xpert TV assay. Xpert TV results for 97.7% (4521/4626) of samples were successful on the first attempt; 105 results were indeterminate. One hundred of the 105 indeterminate samples were retested, and 90 of those cases yielded valid results. The overall rate of assay success was 99.7% (4611/4626).
10
{10}
In total, 4611 specimens from male subjects were included in the clinical performance analysis (4626 eligible specimens tested minus 15 indeterminate results). Of these, 1088 were symptomatic and 3523 were asymptomatic. The average age of the male subjects was 36.2 years old, and the average age among symptomatic and asymptomatic subjects was 32.6 years and 37.4 years, respectively.
The clinical performance of the Xpert TV assay was assessed for the original and supplemental studies both separately and combined. Analysis was performed using Fisher's Exact p-value to ensure that no statistical difference was found between the results generated in the original and supplemental studies. The combined sensitivity for the urine specimen type was 87.5% and 90.3% for symptomatic and asymptomatic males, respectively. The overall sensitivity (including both symptomatic and asymptomatic subjects) was 89.6%. The combined specificity was 99.8% and 99.2% for symptomatic and asymptomatic males, respectively. The overall specificity (including both symptomatic and asymptomatic subjects) was 99.3%. The clinical performance data is reported in Table 2.
Table 2: Xpert TV Assay clinical study results vs. PIS
| Collection | Status | Total (n) | Sensitivity | 95% CI | Specificity | 95% CI | Prev (%) | PPV (%) | NPV (%) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Combined Study | Symp | 1088 | 87.5% (28/32) | 71.9%-95.0% | 99.8% (1054/1056) | 99.3%-99.9% | 2.9% | 93.3% | 99.6% |
| | Asymp | 3523 | 90.3% (84/93) | 82.6%-94.8% | 99.2% (3401/3430) | 98.8%-99.4% | 2.6% | 74.3% | 99.7% |
| | Overall | 4611 | 89.6% (112/125)^{a} | 83.0%-93.8% | 99.3% (4455/4486)^{b} | 99.0%-99.5% | 2.7% | 78.3% | 99.7% |
a: Testing results by additional sequencing: 10 of 13 false negatives were TV negative; 3 of 13 were TV positive
b: Testing results by additional sequencing: 27 of 31 false positives were TV positive; 4 of 31 were TV negative
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
The estimated positive predictive value (PPV) and negative predictive value (NPV) of the Xpert TV Assay across different hypothetical prevalence rates are shown for the male urine specimen type in Table 3. (The expected values for female urine, endocervical swabs, and vaginal swabs are reported in K151565). These calculations are based on the overall estimated sensitivity and specificity observed for male urine specimens during the Xpert TV multi-center clinical study. The overall sensitivity and specificity for male urine (UR) was 89.6% and 99.3%, respectively.
{11}
Table 3: Hypothetical PPV and NPV of the Xpert TV assay
| Specimen Type | Prevalence (%) | PPV (%) | NPV (%) |
| --- | --- | --- | --- |
| Male UR | 1 | 56.7% | 99.9% |
| | 2 | 72.6% | 99.8% |
| | 5 | 87.2% | 99.5% |
| | 10 | 93.5% | 98.9% |
| | 12 | 94.6% | 98.6% |
| | 15 | 95.8% | 98.2% |
| | 20 | 97.0% | 97.4% |
| | 25 | 97.7% | 96.6% |
N. Instrument Name:
GeneXpert® Instrument Systems (GeneXpert Dx, GeneXpert Infinity-48, GeneXpert Infinity-48s and GeneXpertInfinity-80 systems).
O. System Descriptions:
1. Modes of Operation:
Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?
Yes ______ or No ☐ X
Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?
Yes ______ or No ☐ X
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X or No ______
3. Specimen Identification:
The user can scan or type the patient ID.
4. Specimen Sampling and Handling:
Male urine is collected and tested using the Xpert urine collection kit. First-catch urine is collected in a urine collection cup free of any preservatives. A disposable pipet is used to
{12}
transfer approximately 7 mL of urine into the Xpert urine transport reagent tube. The tube is capped and inverted 3-4 times to ensure mixing.
5. Calibration:
The user is instructed to contact Cepheid regarding instrument calibration in each operator manual.
6. Quality Control:
The Xpert TV assay includes a sample adequacy control to verify that human DNA is present in the sample, a sample processing control to verify adequate processing of sample, and a probe check control to ensure adequate reagent rehydration.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
Not applicable
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
13
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.