K161588 · GE Medical Systems Ultrasound & Primary Care Diagnostics, LLC · IYN · Aug 31, 2016 · Radiology
Device Facts
Record ID
K161588
Device Name
Vscan Extend
Applicant
GE Medical Systems Ultrasound & Primary Care Diagnostics, LLC
Product Code
IYN · Radiology
Decision Date
Aug 31, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Vscan Extend is a general purpose diagnostic ultrasound system for use by qualified and trained healthcare professionals enabling visualization and measurement of anatomical structures and fluid. It's pocket-sized portability and simplified user interface enables integration into examination and training sessions indoors and in other environments described in the user manual. The information can be used for basic focused assessments and adjunctively with other medical data for clinical diagnosis purposes during routine, periodic monitoring, and triage.
Device Story
Vscan Extend is a pocket-sized, battery-powered diagnostic ultrasound system; consists of handheld unit with 5-inch touchscreen and permanently attached probe (sector phased array or dual-headed phased/linear array). Device performs digital acquisition, processing, and display of ultrasound images. Used by trained healthcare professionals in clinical settings for basic focused assessments, triage, and procedure guidance (e.g., needle/catheter placement). Supports wireless image transfer to DICOM servers/Windows shares; data export via USB. Includes access to GE Marketplace for software application downloads. Provides clinicians with real-time visualization of anatomical structures and fluid to assist in clinical diagnosis and monitoring. Benefits include portability and integration into routine examinations.
Clinical Evidence
No clinical studies were required to support substantial equivalence. Evidence is based on bench testing, including acoustic output, biocompatibility, cleaning/disinfection, thermal, electrical, electromagnetic, and mechanical safety testing in accordance with AAMI/ANSI ES60601-1, IEC60601-1-2, IEC60601-2-37, NEMA UD 2/3, and ISO 10993-1.
Technological Characteristics
Pocket-sized, battery-powered ultrasound system. Transducers: phased array (deep) and linear array (shallow). Connectivity: WiFi, DICOM, USB. Software: supports downloadable applications via GE Marketplace. Safety standards: AAMI/ANSI ES60601-1, IEC60601-1-2, IEC60601-2-37, NEMA UD 2/3, ISO 10993-1, ISO 14971. Biocompatible materials. Internal storage encryption for data privacy.
Indications for Use
Indicated for qualified healthcare professionals for diagnostic ultrasound imaging, measurement, and analysis of anatomical structures and fluid. Applications include: Cardiac, Abdominal, Renal, OB/GYN, Urology, Fetal, needle/catheter guidance, peripheral vascular (arteries/veins), thoracic/lung, adult cephalic, pediatric, small organs (thyroid), and musculoskeletal (long bone, hip, shoulder, elbow, knee).
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
GE Medical Systems Ultrasound & Primary Care Diagnostics LLC % Ms. Tracey Ortiz Regulatory Affairs Director 9900 W. Innovation Drive WAUWATOSA WI 53226
Re: K161588 Trade/Device Name: Vscan Extend Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: August 10, 2016 Received: August 11, 2016
Dear Ms. Ortiz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
August 31, 2016
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular border. The background and the letters are a light blue color. The border has a swirling design, giving the logo a classic and recognizable appearance.
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
510(k) Number (if known)
K161588 Device Name Vscan Extend
Indications for Use (Describe)
V scan Extend is a general purpose diagnostic ultrasing system for use by qualified and trained healthcare professionals enabling visualization and measurement of anatomical structures and fluid. It's pocket-sized portability and simplified user interface enables integration into examination and training sessions indoors and in other environments described in the user manual. The information can be used for basic focused assessments and adjunctively with other medical data for clinical dagnosis purposes during routine, periodic monitoring, and triage.
With the plased array transducer on the specific clinical applications and exam types include: Cardia: Abdominal: Renal: OB/GYN; Urology: Fetal, Evaluation of Presence of Fluid; Imaging Guidance for Needle Catheter Placement (e.g. paracentesis, pericardiocentesis, anniocentesis): Peripheral Vascular Imaging (e.g. arteries and veins); Thoracic Lung (e.g. pleural motion/sliding, line artifacts); Adult Cephalic; and Pediatrics.
With the addition of the linear array transducer on the single dual headed probe solution, the specific clinical applications and exam types are expanded to include: Peripheral vascular inaging (e.g. lower extremity, carotid); Procedure Guidance for Arterial or Venous Vessels (e.g. central lines, upper extremity); Snall Organs (e.g. thyroid); and Musculoskeletal (Long Bone; Hip. shoulder, elbow and Knee Joints); Evaluation of Presence of Fluid; Thoracic/Lung (e.g. pleural motion/sliding, line artifacts); and Pediativs.
Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized script inside. The circle has a decorative border with swirling patterns. The logo is simple and recognizable, and it is associated with a well-known company.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are swirling lines around the letters, giving the impression of movement or energy. The logo is simple and recognizable, and it is associated with a well-known and established company.
#### Diagnostic Ultrasound Indications for Use Form Vscan Extend Ultrasound System
| Clinical Application | Mode of Operation | | | | | | | | | | |
|----------------------------------|-------------------|---|----|----|-------|---------|-------|-------------------|---------------------|----------------|-------|
| Anatomy/Region of Interest | B | M | PW | CW | Color | Color M | Power | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Fetal/OB | N | | | | N | | | N | N | | 7 |
| Abdominal[1] | N | | | | N | | | N | N | | 7 |
| Pediatric | N | | | | N | | | N | N | | 7 |
| Small Organ[2] | N | | | | N | | | N | | | 7 |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | N | | | | N | | | N | N | | |
| Cardiac Adult | N | | | | N | | | N | N | | 7 |
| Cardiac Pediatric | N | | | | N | | | N | N | | 7 |
| Peripheral Vascular[3] | N | | | | N | | | N | N | | 7 |
| Musculo-skeletal Conventional[4] | N | | | | N | | | N | | | 7 |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Thoracic/Pleural[5] | N | | | | N | | | N | N | | 7 |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | |
| Vascular Access[6] | N | | | | N | | | N | | | 7 |
| Nonvascular | N | | | | N | | | N | N | | 7 |
Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:
N = new indication; P= previously cleared by FDA in K140693
Notes:
[1] Abdominal includes Gynecology, Renal and Urology;
[2] Small Organ includes thyroid;
[3] Peripheral Vascular includes arteries and veins, lower extremity, carotid;
[4] Musculo-skeletal Conventional includes long bone, hip, shoulder, elbow and knee joint visualization;
[5] Thoracic/Pleural is pleural motion/sliding, line artifacts as well as fluid detection;
[6] Vascular Access includes intravenous, intra-arterial, central and peripheral lines, upper extremity;
[7] Image guidance for freehand needle/catheter placement;
[*] Combined mode is B/Color.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a blue circle. The circle has a double-line border, with the outer line being slightly thicker than the inner line. The logo is simple and recognizable, representing the multinational conglomerate corporation.
### Diagnostic Ultrasound Indications for Use Form Vscan Extend with deep scanning-phased array transducer (G3S)
| Clinical Application | Mode of Operation | | | | | | | | | | |
|----------------------------------|-------------------|---|----|----|---------------|------------|-------|-------------------|---------------------|----------------|-------|
| | B | M | PW | CW | Doppler Modes | Color<br>M | Power | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Fetal/OB | P | | | | P | | | P | P | | 7 |
| Abdominal[1] | P | | | | P | | | P | P | | 7 |
| Pediatric | P | | | | P | | | P | P | | 7 |
| Small Organ[2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | N | | | | N | | | N | N | | |
| Cardiac Adult | P | | | | P | | | P | P | | 7 |
| Cardiac Pediatric | P | | | | P | | | P | P | | 7 |
| Peripheral Vascular[3] | P | | | | P | | | P | P | | 7 |
| Musculo-skeletal Conventional[4] | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Thoracic/Pleural[5] | P | | | | P | | | P | P | | 7 |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | |
| Vascular Access[6] | | | | | | | | | | | |
| Nonvascular | P | | | | P | | | P | P | | 7 |
Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:
N = new indication; P= previously cleared by FDA in K140693
Notes:
[1] Abdominal includes Gynecology, Renal and Urology;
[2] Small Organ includes thyroid;
[3] Peripheral Vascular includes arteries and veins, lower extremity, carotid;
[4] Musculo-skeletal Conventional includes long bone, hip, shoulder, elbow and knee joint visualization;
[5] Thoracic/Pleural is pleural motion/sliding, line artifacts as well as fluid detection;
[6] Vascular Access includes intravenous, intra-arterial, central and peripheral lines, upper extremity;
[7] Image guidance for freehand needle/catheter placement;
[*] Combined mode is B/Color.
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OlR) Prescription Use (Per 21 CFR 801.109)
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular frame. The color of the logo is a light blue.
# Diagnostic Ultrasound Indications for Use Form Vscan Extend with shallow scanning-linear array transducer (G8L)
| Clinical Application | Mode of Operation | | | | | | | | | | | | |
|----------------------------------|-------------------|---|----|----|---------------|------------|-------|-------|---------|-------------------------|-------|--|--|
| | | | | | Doppler Modes | | | | | Combined Harmonic Coded | | | |
| Anatomy/Region of Interest | B | M | PW | CW | Color | Color<br>M | Power | Modes | Imaging | Pulse | Other | | |
| Fetal/OB | | | | | | | | | | | | | |
| Abdominal[1] | | | | | | | | | | | | | |
| Pediatric | P | | | | P | | | P | | | 7 | | |
| Small Organ[2] | P | | | | P | | | P | | | 7 | | |
| Neonatal Cephalic | | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | | |
| Cardiac Adult | | | | | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | | | | | |
| Peripheral Vascular[3] | P | | | | P | | | P | | | 7 | | |
| Musculo-skeletal Conventional[4] | P | | | | P | | | P | | | 7 | | |
| Musculo-skeletal Superficial | | | | | | | | | | | | | |
| Thoracic/Pleural[5] | P | | | | P | | | P | | | 7 | | |
| Exam Type, Means of Access | | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | | | |
| Vascular Access[6] | P | | | | P | | | P | | | 7 | | |
| Nonvascular[7] | | | | | | | | | | | | | |
Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:
N = new indication; P= previously cleared by FDA in K140693
Notes:
[1] Abdominal includes Gynecology, Renal and Urology;
[2] Small Organ includes thyroid;
[3] Peripheral Vascular includes arteries and veins, lower extremity, carotid;
[4] Musculo-skeletal Conventional includes long bone, hip, shoulder, elbow and knee joint visualization;
[5] Thoracic/Pleural is pleural motion/sliding, line artifacts as well as fluid detection;
[6] Vascular Access includes intravenous, intra-arterial, central and peripheral lines, upper extremity;
[7] Image guidance for freehand needle/catheter placement;
[*] Combined mode is B/Color.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OlR) Prescription Use (Per 21 CFR 801.109)
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo. The logo is a blue circle with the letters "GE" in a stylized script inside. The letters are also blue, and they are surrounded by a swirling pattern. The logo is simple and recognizable, and it is often used to represent the company.
## 510(k) Summary
| | In accordance with 21 CFR 807.92 the following summary of information is provided: |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date:<br>Submitter: | June 7, 2016<br>GE Medical Systems Ultrasound and Primary Care Diagnostics,<br>LLC<br>9900 Innovation Drive<br>Wauwatosa, WI 53226 |
| Primary Contact Person: | Tracey Ortiz<br>Regulatory Affairs Director<br>GE Healthcare<br>T:(262) 676-6120<br>F:(414) 918-8275 |
| Secondary Contact Person: | Sadashiva Subraya<br>Regulatory Affairs Leader<br>GE Healthcare<br>T: +91 80 4088 2969 |
| Device:<br>Trade Name: | Vscan Extend |
| Common/Usual Name: | Ultrasound system |
| Classification Names:<br>Product Code: | Class II<br>Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550 90-IYN<br>Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO<br>Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX |
| Primary Predicate<br>Device(s): | Vscan with Dual Probe - K140693 |
| Reference Predicate<br>Device(s): | Venue 50 - K152758<br>Venue 40 - K112122 |
| Device Description: | Vscan Extend is a pocket-sized, battery powered general purpose,<br>track 3, diagnostic ultrasound system. The system consists of a<br>handheld unit with a 5 inch touch screen display and a<br>permanently attached probe. The probe is available in one of two<br>configurations: a sector probe with phased array transducer or<br>both phased and linear array transducers in the single probe (dual<br>probe). It has digital acquisition, processing and display<br>capabilities. The device specific battery can be charged either in<br>the system or independently. The system also may include an<br>AC/DC adapter with cable, case, Micro SD card, USB cable, or<br>other accessories. |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in white inside. The letters are stylized and connected, and there are decorative swirls around the outside of the circle. The logo is simple and recognizable, and it is associated with a well-known company.
The system is capable of transferring images wirelessly to a DICOM server or Windows share. Data can also be exported to the user's computer by using the USB export option and a standard USB/micro USB cable. Capabilities also include access to GE Marketplace. which shall allow the user to download software applications to the device.
Vscan Extend is a general purpose diagnostic ultrasound imaging Intended Use: system for use by qualified and trained healthcare professionals enabling visualization and measurement of anatomical structures and fluid. It's pocket-sized portability and simplified user interface enables integration into examination and training sessions indoors and in other environments described in the user manual. The information can be used for basic/focused assessments and adjunctively with other medical data for clinical diagnosis purposes during routine, periodic monitoring, and triage.
> With the phased array transducer on the sector probe, the specific clinical applications and exam types include: Cardiac; Abdominal; Renal; OB/GYN; Urology; Fetal, Evaluation of Presence of Fluid; Imaging Guidance for Needle/Catheter Placement (e.g. paracentesis, pericardiocentesis, thoracentesis, amniocentesis); Peripheral Vascular Imaging (e.g. arteries and veins); Thoracic/Lung (e.g. pleural motion/sliding, line artifacts); Adult Cephalic; and Pediatrics.
> With the addition of the linear array transducer on the single dual headed probe solution, the specific clinical applications and exam types are expanded to include: Peripheral vascular imaging (e.g. lower extremity, carotid); Procedure Guidance for Arterial or Venous Vessels (e.g. central lines, upper extremity); Small Organs (e.g. thyroid); and Musculoskeletal (Long Bone; Hip, shoulder, elbow and Knee Joints); Evaluation of Presence of Fluid; Thoracic/Lung (e.g. pleural motion/sliding, line artifacts); and Pediatrics.
- Technology: The Vscan Extend employs the same fundamental scientific technology as its predicate devices.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. The letters are also blue. There are swirling designs around the letters, also in blue.
#### Determination of Comparison to Predicates Substantial Equivalence:
The Vscan Extend is substantially equivalent to the predicate devices with regards to intended use, imaging capabilities, technologicial characteristics and safety and effectiveness.
- The systems are all intended for diagnostic ultrasound ● imaging and fluid flow analysis.
- The proposed Vscan Extend and predicate Vscan with ● Dual Probe have the same clinical intended use and clinical applications except that addition of adult cephalic that was cleared on the Venue 40 predicate.
- . The proposed Vscan Extend and predicate Vscan with Dual Probe have the same imaging modes.
- . The proposed Vscan Extend and predicate Vscan with Dual Probe are manufactured with materials which have been evaluated and found to be safe for the intended use of the device.
- The proposed Vscan Extend and predicate device Vscan ● with Dual Probe have the same phased array transducer for deep scanning and linear array transducer for shallow scanning.
- The proposed Vscan Extend and predicate Vscan with Dual Probe have similar acoustic power levels. All values are below FDA applicable limits.
- . The proposed Vscan Extend and predicate Vscan with Dual Probe have the same capability in terms of performing measurements, capturing digital images and reviewing the images.
- . The proposed Vscan Extend has WiFi connectivity similar to the predicate device Venue 50.
- The proposed Vscan Extend and predicate device Venue ● 50 system have same data privacy features including internal storage encryption which allows the user to encrypt patient data.
- The proposed Vscan Extend has DICOM connectivity similar to the predicate device Venue 50.
- The proposed Vscan Extend and predicate systems have ● been designed in compliance with approved electrical and physical safety standards.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized script, enclosed within a blue circle. The circle is surrounded by swirling, wave-like shapes, also in blue, which give the logo a dynamic and recognizable appearance.
## Summary of Non-Clinical Tests:
The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection as well as thermal, electrical, electromagnetic, and mechanical safety, and has been found to conform to applicable medical device safety standards. The Vscan Extend and its applications comply with voluntary standards:
- . AAMI/ANSI ES60601-1, Medical Electrical Equipment - Part 1: General Requirements for Safety and Essential Performance, 2005
- IEC60601-1-2,Medical Electrical Equipment Part 1-. 2:General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility Requirements and Tests, 2007
- IEC60601-2-37, Medical Electrical Equipment - Part 2-37:Particular Requirements for the Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment, 2007
- NEMA UD 3, Standard for Real Time Display of . Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment, 2004
- ISO10993-1, Biological Evaluation of Medical ● Devices- Part 1: Evaluation and Testing- Third Edition, 2009
- NEMA UD 2. Acoustic Output Measurement Standard . for Diagnostic Ultrasound Equipment, 2004
- ISO14971, Application of risk management to medical ● devices: Second edition, 2007
- NEMA PS 3.1 3.20 (2011), Digital Imaging and ● Communications in Medicine (DICOM) Set. (Radiology)
The following quality assurance measures were applied to the development of the system:
- Risk Analysis
- Requirements Reviews ●
- Design Reviews ●
- Testing on unit level (Module verification) ●
- Integration testing (System verification)
- Performance testing (Verification) ●
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized script, enclosed within a blue circle. The circle has a wavy, decorative border, also in blue, giving the logo a classic and recognizable appearance.
- . Safety testing (Verification)
- . Verification of Privacy & Security Design
Transducer materials are biocompatible.
## Summary of Clinical Tests:
The subject of this premarket submission, Vscan Extend, did not require clinical studies to support substantial equivalence.
- Conclusion: GE Healthcare considers the Vscan Extend to be as safe, as effective, and performance is substantially equivalent to the predicate device(s).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.