Mano and Mano Personal Lubricant

K161585 · Mano and Mano Proprietary Limited · NUC · Sep 6, 2016 · Obstetrics/Gynecology

Device Facts

Record IDK161585
Device NameMano and Mano Personal Lubricant
ApplicantMano and Mano Proprietary Limited
Product CodeNUC · Obstetrics/Gynecology
Decision DateSep 6, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.5300
Device ClassClass 2

Indications for Use

Mano and Mano Personal Lubricant is a water-based personal lubricant, for penile and/or vaginal application, intended to lubricate and moisturize, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyurethane, and polyisoprene condoms.

Device Story

Mano and Mano Personal Lubricant is a clear, colorless, water-based personal lubricant. It is intended for over-the-counter use by individuals to enhance comfort during intimate sexual activity by supplementing natural lubrication. The device is applied topically to the penis or vagina. It is not compatible with natural rubber latex, polyurethane, or polyisoprene condoms. The product is formulated with purified water, propylene glycol, hydroxyethyl cellulose, sodium pyrrolidone carboxylic acid, potassium sorbate, and citric acid.

Clinical Evidence

Bench testing only. Device met physical specifications (appearance, pH, viscosity, osmolality) and microbiology specifications (USP <61>, USP <62>, USP <51>). Biocompatibility testing performed per ISO 10993-5:2009 (cytotoxicity), ISO 10993-10:2010 (sensitization and irritation), and ISO 10993-11:2006 (acute systemic toxicity). Stability testing confirmed maintenance of specifications over the shelf-life period.

Technological Characteristics

Water-based personal lubricant. Ingredients: purified water, propylene glycol, hydroxyethyl cellulose, sodium pyrrolidone carboxylic acid, potassium sorbate, citric acid. Physical properties: clear, colorless. Biocompatibility tested per ISO 10993 standards. Microbiology tested per USP <51>, <61>, <62>.

Indications for Use

Indicated for penile and/or vaginal application to lubricate and moisturize during intimate sexual activity. Not compatible with natural rubber latex, polyurethane, or polyisoprene condoms.

Regulatory Classification

Identification

A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.

Special Controls

*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of a stylized caduceus, which is a staff with two snakes entwined around it, and three human profiles facing right. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 6, 2016 Mano and Mano Proprietary Limited % Gary Mocnik Regulatory Consultant ClinReg Consulting Services, Inc. 733 Bolsana Drive Laguna Beach, CA 92651 Re: K161585 > Trade/Device Name: Mano and Mano Personal Lubricant (also branded as Nooky, Hello Sailor, or Lulu) Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: Class II Product Code: NUC Dated: June 6, 2016 Received: June 8, 2016 Dear Gary Mocnik: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Joyce M. Whang -S for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K161585 #### Device Name Mano and Mano Personal Lubricant (additionally branded as Nooky, Hello Sailor, or Lulu) Indications for Use (Describe) Mano and Mano Personal Lubricant is a water-based personal lubricant, for penile and/or vaginal application, intended to lubricate and moisturize, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyurethane, and polyisoprene condoms. | Type of Use (Select one or both, as applicable) | |------------------------------------------------------------------| | <div> <span>[Checkbox 1]</span> <span>[Checkbox 2]</span> </div> | | | Prescription Use (Part 21 CFR 801 Subpart D) |X | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary (K161585) | Applicant: | Mano and Mano Proprietary Limited<br>286 Chesterville Rd<br>Moorabbin, Victoria 3189<br>Australia | |--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------| | Contact: | Gary Mocnik<br>ClinReg Consulting Services, Inc.<br>733 Bolsana Drive<br>Laguna Beach, CA 92651<br>Phone: 949-433-0413<br>E-mail: gmocnik@cox.net) | | Date Prepared: | August 16, 2016 | | Proprietary Name: | Mano and Mano Personal Lubricant (also branded as Nooky, Hello Sailor, or<br>Lulu) | | Common Name: | Personal Lubricant | | Regulation Number: | 21 CFR 884.5300 | | Regulatory Class: | Class II | | Product Code: | NUC | | Predicate Device: | Poise Personal Lubricant (K120650) | ### Device Description: Mano and Mano Personal Lubricant (additionally branded as Nooky, Hello Sailor, or Lulu) is a clear, colorless, water-based personal lubricant composed of purified water, propylene glycol, hydroxethyl cellulose, sodium pyrrolidone carboxylic acid, potassium sorbate and citric acid. This product has an initial shelf-life of 10 months. The shelf-life will be extended pending completion of further study. #### Indications for Use: Mano and Mano Personal Lubricant is a water-based personal lubricant, for penile and/or vaginal application, intended to lubricate and moisturize, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyurethane, and polyisoprene condoms. {4}------------------------------------------------ ### Substantial Equivalence Discussion The subject and predicate devices have the same indications. They both are clear, colorless, waterbased personal lubricants. Although there are differences in pH, osmolality, viscosity and formulation of these devices, the differences do not represent a new technology as they do not raise different questions of safety and effectiveness. ### Performance Data Bench testing demonstrated that the subject device met the physical specifications (appearance, pH, viscosity and, osmolality) and the following microbiology specifications: | Criteria | Specifications | Methods | |------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------| | Total Aerobic Microbial Count | ≤100 cfu/g | USP <61> | | Total Yeast and Mold Count | ≤10 cfu/g | USP <61> | | Absence of Pathogenic<br>Organisms | Absence of pseudomonas aeruginosa, staphylococcus<br>aureus, and candida albicans | USP <62> | | Antimicrobial Effectiveness | At least 2.0 log reduction from the initial count at 14<br>days; no increase from the 14 days count at 28 days; no<br>increase from the initial calculated count at 14 and 28<br>days | USP<51> | Shelf-life (stability) testing conducted by a third-party laboratory demonstrated that the subject device maintained all physical and microbiology specifications at the end of shelf-life period. The subject device has passed the following biocompatibility testing in accordance with relevant ISO standards: - Cytotoxicity (ISO 10993-5:2009) ● - Sensitization (ISO 10993-10:2010) ● - Irritation (ISO 10993-10:2010) ● - Acute Systemic Toxicity (ISO 10993-11:2006) ● ### Conclusion: The subject and predicate devices have the same indications and fundamental technological characteristics. The differences in technological characteristics do not raise different types of questions and can be assessed by bench and biocompatibility testing. Performance data demonstrates that the subject device is substantially equivalent to the predicate device in terms of safety and effectiveness.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...