The Affirm Breast Biopsy Guidance System is an optional accessory for the Selenia Dimensions Mammography System. It is designed to allow the accurate localization of lesions in the breast in three dimensions. It is intended to provide guidance for interventional purposes (such as biopsy, pre-surgical localization or treatment devices).
Device Story
Affirm Lateral Arm Upright Biopsy Accessory is an optional attachment for the Hologic Selenia Dimensions Mammography System; enables lateral needle approach for breast biopsies. Device consists of lateral biopsy paddle, lateral arm stand, case, and QC phantom. Operates only when C-arm is at 0 degrees. Biopsy needle manually advanced along X-axis (Lat X). System uses X, Y, Z coordinates from the Biopsy Positioning Module to guide needle placement. Used in clinical settings by trained healthcare professionals. Provides visual guidance for interventional procedures; assists in accurate lesion targeting. Benefits include expanded surgical access options for breast biopsies.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical bench testing, including targeting accuracy tests, needle parameter validation, tensile safety factor testing (IEC 60601-1), compression plate strength testing (IEC 60601-2-45), and biopsy needle positioning accuracy testing.
Technological Characteristics
Accessory for Selenia Dimensions Mammography System. Materials: Carbon fiber (breast platform), polycarbonate (biopsy paddle). Sensing/Actuation: Manual needle advancement with coordinate-based positioning via Biopsy Positioning Module. Connectivity: Integrated with Selenia Dimensions system (software V1.8.4+). Safety: C-arm position interlock (0 degrees only). Sterilization: Not specified. Software: Moderate level of concern.
Indications for Use
Indicated for patients requiring breast lesion localization and interventional procedures, including core biopsy, vacuum-assisted biopsy, fine needle aspiration, and hook wire localization.
Regulatory Classification
Identification
A mammographic x-ray system is a device intended to be used to produce radiographs of the breast. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 10, 2016
Hologic, Inc. % Ms. Debbie Peacock Senior Manager, Regulatory Affairs 36 Apple Ridge Road DANBURY CT 06810
Re: K161575
Trade/Device Name: Affirm Lateral Arm Upright Biopsy Accessory Regulation Number: 21 CFR 892.1710 Regulation Name: Mammography x-ray system Regulatory Class: II Product Code: IZH Dated: June 8, 2016 Received: June 9, 2016
Dear Ms. Peacock:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D. O'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K161575
Device Name Affirm Lateral Arm Upright Biopsy Accessory
Indications for Use (Describe)
The Affirm Breast Biopsy Guidance System is an optional accessory for the Selenia Dimensions Mammography System. It is designed to allow the accurate localization of lesions in three dimensions. It is intended to provide guidance for interventional purposes (such as biopsy, pre-surgical localization or treatment devices).
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
{3}------------------------------------------------
# Traditional 510(k) Summary
This 510(k) Summary is submitted in accordance with the requirements of 21 CFR Part 807.92
| Date Prepared: | June 6, 2016 |
|-------------------------------|----------------------------------------------------------------------------|
| Manufacturer: | Hologic, Inc.<br>36 Apple Ridge Road<br>Danbury, CT 06810 USA |
| Est. Registration #: | 1220984 |
| Manufacturer: | Hologic, Inc.<br>37 Apple Ridge Road<br>Danbury, CT 06810 USA |
| Est. Registration #: | 1225717 |
| Contact Person: | Debbie Peacock<br>Sr. Manager, Regulatory Affairs<br>Phone: (203) 702-7794 |
| Identification of the Device: | |
### Identification of the Device:
| Proprietary/Trade Name | Affirm™ Lateral Arm Upright Biopsy Accessory |
|------------------------|----------------------------------------------|
| Classification Name: | Mammographic X-Ray System |
| Regulatory Number: | 21 CFR 892.1710 |
| Product Code: | IZH |
| Device Class: | Class II |
| Review Panel: | Radiology |
### Identification of the Legally Marketed Predicate Device:
Trade name: Affirm Breast Biopsy Guidance System General name: Mammographic X-ray system Submitter / 510(k) Holder: HOLOGIC, Inc. 510(k) #'s: K122836, cleared on 01/11/13 Product Code: IZH Regulation Number: 892.1710
### Identification of the Legally Marketed Reference Device:
Trade name: GE Senographe General name: Mammographic X-ray system Submitter / 510(k) Holder: General Electric, Inc. 510(k) #'s: K040125, cleared on 04/19/04 Product Code: IZH Regulation Number: 892.1710
{4}------------------------------------------------
## Device Description:
This submission introduces the optional Affirm Lateral Arm Upright Biopsy Accessory which attaches to the Affirm Breast Biopsy Guidance System used with the Selenia Dimensions 2D/3D Mammography System, software version 1.8.4 and higher.
The optional Affirm Lateral Arm Upright Biopsy Accessory attaches to the Affirm Biopsy Guidance Module (BGM) to enable lateral needle approach procedures. The Lateral Arm can only be used when the Selenia Dimensions C-arm is positioned at 0 degrees. When using the Lateral Arm, the X- Y- and Z-axis movement is the same as for the Affirm standard (upright) needle approach. For the lateral approach, the biopsy needle is manually advanced into the breast along its own X-axis, referred to as "Lat X".
When performing lateral approach biopsies, the biopsy device is installed onto the lateral arm in the same manner as when performing standard (upright biopsies).
New components used with the Affirm Lateral Arm Upright Biopsy Accessory include: lateral biopsy paddle, lateral arm stand, case, and QC phantom specifically to be used for Lateral approach biopsies.
The Indications for Use is unchanged from the predicate Affirm Breast Biopsy Guidance System as shown below.
### Indications for Use:
The Affirm Breast Biopsy Guidance System is an optional accessory for the Selenia Dimensions Mammography System. It is designed to allow the accurate localization of lesions in the breast in three dimensions. It is intended to provide guidance for interventional purposes (such as biopsy, pre-surgical localization or treatment devices).
### Substantial Equivalence:
The (optional) Affirm Lateral Arm Upright Biopsy Accessory is substantially equivalent to the predicate Affirm Breast Biopsy Guidance System (K122836) and to the lateral arm attachment used as a reference device, cleared in GE's Senographe Stereo System, (K040125). Substantial equivalence is based on design, technology, Indications for Use, labeling, operation, basic constructions and materials used. The addition of the optional Lateral Arm Accessory to the cleared Affirm Biopsy Guidance System poses no additional hazards and is substantially equivalent to our predicate and reference devices listed below.
| | Affirm Lateral Arm<br>Upright Biopsy<br>Accessory<br>(Proposed) | Predicate Device<br>Affirm Breast Biopsy<br>Guidance System<br>(K122836) | Reference Device<br>GE Senographe<br>Stereo (K040125);<br>(lateral arm) |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use | The Affirm Breast Biopsy<br>Guidance System is an<br>optional accessory for the<br>Selenia Dimensions<br>Mammography System. It | The Affirm Breast Biopsy<br>Guidance System is an<br>optional accessory for the<br>Selenia Dimensions<br>Mammography System. It is | Senographe Stereo is<br>an optional accessory<br>for the Senographe DS<br>full field digital system.<br>It is designed to allow |
| | | | |
| X-ray Image<br>Device | is designed to allow the<br>accurate localization of<br>lesions in the breast in<br>two/and or three<br>dimensions. It is intended<br>to provide guidance for<br>interventional purposes<br>(such as biopsy, pre-<br>surgical localization or<br>treatment devices).<br>Hologic Selenia<br>Dimensions 2D/3D<br>Mammography System | designed to allow the<br>accurate localization of<br>lesions in the breast in three<br>dimensions. It is intended<br>to provide guidance for<br>interventional purposes<br>(such as biopsy, pre-surgical<br>localization or treatment<br>devices).<br>Same | the accurate location of<br>lesions in the breast in<br>three dimensions, using<br>information extracted<br>from stereotactic pairs<br>of two-dimensional<br>images. It is intended<br>to provide guidance for<br>interventional purposes<br>(such as biopsy, pre-<br>surgical localization, or<br>treatment devices).<br>GE Senographe DS 2D |
| Software | Selenia Dimensions V1.8.4<br>and higher) | Selenia Dimension V1.8.3<br>(current version) | Same description;<br>version unknown. |
| Method of Use | Breast lesion localization<br>and biopsy<br>core biopsy vacuum assisted<br>biopsy fine needle<br>aspiration hook wire<br>localization<br>exams. | Breast lesion localization<br>and biopsy<br>core biopsy vacuum assisted<br>biopsy fine needle<br>aspiration hook wire<br>localization exams. | Breast lesion<br>localization and biopsy,<br>such as:<br>core biopsy vacuum<br>assisted biopsy fine needle<br>aspiration hook wire<br>localization<br>exams. |
| Dedicated<br>Breast Biopsy<br>Positioner | Yes | Yes | Yes |
| Mechanism of<br>Action | Guidance for breast biopsy<br>-Standard (vertical)<br>approach, and<br>-Right or left lateral<br>approach. | Guidance for breast biopsy<br>-Standard (vertical) approach | Guidance for breast<br>biopsy<br>-Standard (vertical),<br>and right or left lateral<br>approach. |
| Technology | | | |
| Coordinate<br>Determination | Yes | Yes | Yes |
| Needle<br>Positioning | Biopsy Positioning<br>Module gets X, Y, Z<br>coordinates of target area<br>from AWS | Biopsy Positioning Module<br>gets X, Y, Z coordinates of<br>target area from AWS | Biopsy Positioning<br>Module gets X, Y, Z<br>coordinates of target<br>area from AWS |
{5}------------------------------------------------
{6}------------------------------------------------
| Stereotactic/<br>Tomographic<br>Angle | Stereo:<br>+/- 15 °<br>Tomo:<br>15 degrees total (+/- 7.5<br>degrees) | Stereo:<br>Same<br>Tomo:<br>Same | Stereo:<br>+/- 15 °<br>Tomo:<br>N/A |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| Stated<br>Accuracy | +/- 1 mm in X, Y and Z<br>Axis<br>[overall targeting<br>accuracy is equal to<br>combined targeting<br>accuracy of the Biopsy<br>arm controller and the<br>biopsy device (maximum<br>deviation from target<br>coordinate will not be<br>more than 2 mm from<br>either side)]. | Same | +/- 1 mm in X, Y and Z<br>Axis |
| Compression<br>Method | Manual, automatic<br>compression release is<br>disabled automatically<br>when the BGM is<br>installed. | Same | Manual, compression<br>range is 10-100 mm,<br>with compression<br>controlled manually. |
| Compression<br>Paddle | Vertical approach<br>w/window,<br>Lateral approach, no<br>window | Vertical approach w/window | Vertical approach<br>w/window,<br>Lateral approach, no<br>window: |
| Safety<br>Features | Same<br>Plus:<br>When the system is in<br>lateral approach biopsy<br>mode, the system shall<br>only allow exposures at C-<br>arm position of 0 degrees. | Automatic detection of<br>mounting, latching and<br>connection of biopsy module<br>C-arm motion disabled if<br>biopsy module is not locked<br>in place.<br>Automatic compression<br>release disabled when biopsy<br>module installed<br>C-Arm Motion disabled<br>when breast is under<br>compression<br>Motorized movement of<br>biopsy device only under<br>user control.<br>Audible alert when biopsy<br>device motion may result in | Unknown |
| | | mechanical interference. | |
| Materials (Short-term Skin Contracting) | | | |
| Breast<br>Platform<br>Material | Carbon Fiber | Same | Same |
| Biopsy Paddle<br>Material | Polycarbonate plastics | Same | Same |
{7}------------------------------------------------
# Summary of Testing:
The (optional) Affirm Lateral Arm Upright Biopsy Accessfully performed system design control verification and validation tests, which are summarized in accordance with FDA's Guidance for the Content of premarket Submissions for Software Contained in Medical Devices (issued on May 11, 2005) based on a moderate level of concern.
In addition, the following test reports are included:
- Lateral Arm Targeting Accuracy Test ●
- Default Needle Parameter Validation Test Report .
The Affirm CB test report was updated to include the optional Lateral Arm Accessory and the new lateral biopsy paddle. Third party testing was conducted on the tests listed below which were considered necessary to evaluate the noted changes:
- Tensile Safety Factor (IEC 60601-1, Clause 9.8.2)
- Strength of Compression Plates (IEC 60601-2-45, Subclause 203.8.5.4.102.5)
- Biopsy Needle Positioning Accuracy of Mammographic Stereotactic Devices *IEC ● 60601-2-45, Subclause 201.9.2.1013 a,b,c)
No clinical studies were performed. Substantial equivalence has been demonstrated by nonclinical testing.
# Conclusion:
The Affirm Lateral Arm Upright Biopsy Accessory is a modification to the predicate Affirm Breast Biopsy Guidance System (K122836) and is also substantially equivalent to the lateral arm component used in the reference GE Senographe Stereo System. (K040125). The design, operation, basic construction and materials used are substantially equivalent to the above predicate devices. Selenia Dimensions software V1.8.4 and higher enables use of either 2D or tomosynthesis-guided biopsy in a standard (vertical) or lateral approach. The addition of the (optional) Affirm Lateral Arm Upright Biopsy Accessory to the cleared Affirm Biopsy Guidance System is substantially equivalent to our predicate and reference devices listed above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.