# SIGNA Voyager (K161567)

_GE Healthcare · LNH · Sep 1, 2016 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K161567

## Device Facts

- **Applicant:** GE Healthcare
- **Product Code:** [LNH](/productcode/LNH.md)
- **Decision Date:** Sep 1, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The SIGNA Voyager is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan time imaging. The SIGNA Voyager is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images and/or spectra, dynamic images, and parametric maps of the internal structures and organs of the entire body. Body structures for evaluation include, but are not limited to: head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced by the SIGNA Voyager reflect the spatial distribution and/or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

## Device Story

SIGNA Voyager is a 1.5T superconducting whole-body MRI scanner with a 70cm bore. It utilizes time-varying magnetic fields (gradients) and RF transmissions to acquire data on the density and position of nuclei exhibiting magnetic resonance. The system generates axial, sagittal, coronal, and oblique images, spectra, dynamic images, and parametric maps. Operated by trained clinicians in clinical settings, the device supports high-resolution, high signal-to-noise ratio imaging with reduced scan times. Output images are interpreted by physicians to assist in clinical diagnosis. The system is available in fixed or mobile configurations.

## Clinical Evidence

No clinical trials were conducted. Evidence consists of sample clinical images demonstrating acceptable diagnostic imaging performance and image quality equivalent to the predicate devices. Bench testing included module verification, system integration, performance, safety, and simulated use testing, alongside compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-33, ISO 10993-1, and NEMA standards.

## Technological Characteristics

1.5T superconducting magnet, 70cm bore. Employs time-varying magnetic fields and RF transmissions. Features updated RF receive chain, modified magnet and RF body coil, updated TDI coil suite, and new computers. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-33, ISO 10993-1, NEMA MS standards, and NEMA PS3 (DICOM).

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- SIGNA Pioneer ([K160621](/device/K160621.md))
- Optima MR450w 1.5T (DV25) ([K142085](/device/K142085.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of the department's name arranged in a circular fashion around a symbol. The symbol is a stylized representation of three human profiles facing to the right, stacked on top of each other. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 1, 2016

Ge Healthcare Glen Sabin Regulatory Affairs Director 3200 N. Grandview Blvd Waukesha. Wisconsin 53188

Re: k161567

Trade/Device Name: Signa Voyager Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH Dated: June 3, 2016 Received: June 7, 2016

Dear Glen Sabin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Michael D.'Hara

For

Robert Ochs Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized, cursive font, enclosed within a circular frame. The frame has decorative swirls or flourishes around the perimeter, giving it a classic and recognizable appearance. The color of the logo is blue.

## Section 4: Indications for Use Statement

SIGNA Voyager

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K161567

Device Name SIGNA Voyager

Indications for Use (Describe)

The SIGNA Voyager is a whole body magnetic resonance scanner designed to support high resolution, high signal-tonoise ratio, and short scan time imaging. The SIGNA Voyager is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images and/or spectra, dynamic images, and parametric maps of the internal structures and organs of the entire body. Body structures for evaluation include, but are not limited to: head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used.

The images produced by the SIGNA Voyager reflect the spatial distribution and/or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

| Type of Use (Select one or both, as applicable)                                   |                                                                                  |
|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| <div> <span>     ☒ Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span>     ☐ Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in white inside. There are also some white swirls around the letters. The logo is simple and recognizable.

## Section 5: 510(k) Summary

SIGNA Voyager

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in white inside. There are white swirls around the letters. The logo is simple and recognizable.

## Section 5: 510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

| Date:                     | June 03, 2016                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter:                | GE Healthcare<br>3200 N. Grandview Blvd.<br>Waukesha, WI 53188. USA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Primary Contact Person:   | Glen Sabin<br>Regulatory Affairs Director<br>GE Healthcare, (GE Medical Systems, LLC)<br>Phone: 262-521-6848                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Secondary Contact Person: | Steve Kachelmeyer<br>Regulatory Affairs Director - MR<br>GE Healthcare, (GE Medical Systems, LLC)<br>Phone: (262) 548-2432                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Device Trade Name:        | SIGNA Voyager                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Common/Usual Name:        | Magnetic Resonance Diagnostic Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Classification Names:     | Magnetic Resonance Diagnostic Device per 21 CFR 892.1000                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Product Code:             | LNH, MOS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Predicate Device:         | SIGNA Pioneer (K160621)<br>Optima MR450w 1.5T (DV25) (K142085)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Device Description:       | The SIGNA Voyager system is a whole body magnetic<br>resonance scanner designed to support high resolution, high<br>signal-to-noise ratio, and short scan times. The system uses a<br>combination of time-varying magnetic fields (gradients) and RF<br>transmissions to obtain information regarding the density and<br>position of elements exhibiting magnetic resonance. The system<br>can image in the sagittal, coronal, axial and oblique planes, using<br>various pulse sequences and reconstruction algorithms. The<br>system is offered as a new system installation, in either a fixed or<br>a mobile configuration. The system features A 1.5T<br>superconducting magnet with a 70cm bore size. |

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a circular frame. The logo is colored in a light blue hue, and the background is white.

510(k) Premarket Notification Submission

| Intended Use:                     | The SIGNA Voyager is a whole body magnetic resonance<br>scanner designed to support high resolution, high signal-to-noise<br>ratio, and short scan time imaging. The SIGNA Voyager is<br>indicated for use as a diagnostic imaging device to produce axial,<br>sagittal, coronal, and oblique images, spectroscopic images<br>and/or spectra, dynamic images, and parametric maps of the<br>internal structures and organs of the entire body. Body structures<br>for evaluation include, but are not limited to: head, neck, TMJ,<br>spine, breast, heart, abdomen, pelvis, joints, prostate, blood<br>vessels, and musculoskeletal regions of the body. Depending on<br>the region of interest being imaged, contrast agents may be used.<br>The images produced by the SIGNA Voyager reflect the spatial<br>distribution and/or molecular environment of nuclei exhibiting<br>magnetic resonance. These images and/or spectra when<br>interpreted by a trained physician yield information that may<br>assist in diagnosis. |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological<br>Characteristics: | The SIGNA Voyager employs the same fundamental scientific<br>technology as its predicate devices.<br><br>The following is a summary of the different technology<br>characteristics from the predicate devices:<br>Updated RF Receive Chain New version of software Modified magnet and RF body coil Updated TDI Coil Suite New Computers                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are three water droplet shapes around the letters. The logo is simple and recognizable.

| Performance Data: | Summary of Non-Clinical Tests:<br>Like the predicate device, the SIGNA Voyager complies with the<br>following voluntary standards:<br>• IEC 60601-1<br>• IEC 60601-1-2<br>• IEC 60601-2-33<br>• ISO 10993-1<br><br>In addition, the SIGNA Voyager complies with the applicable<br>NEMA MS standards for MRI and NEMA PS3 standard for<br>DICOM, as does the predicate device.<br><br>The following quality assurance measures were applied to the<br>development of the system, as they were for the predicates:<br>• Risk Analysis<br>• Requirements Reviews<br>• Design Reviews<br>• Testing on unit level (Module verification)<br>• Integration testing (System verification)<br>• Performance testing (Verification)<br>• Safety testing (Verification)<br>• Simulated use testing (Validation) |
|-------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                   | Summary of Clinical Tests:<br>The sample clinical images demonstrate the acceptable<br>diagnostic imaging performance of the SIGNA Voyager<br>including the additional enhanced software features and all coils.<br>The image quality of SIGNA Voyager is substantially equivalent<br>to that of the predicate device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Conclusion:       | The SIGNA Voyager has the same intended use as the predicate<br>devices. The differences in technological characteristics between<br>the SIGNA Voyager and the predicate devices do not raise any<br>different questions of safety or effectiveness. Performance data<br>provided in this submission demonstrate that the SIGNA<br>Voyager is as safe, as effective, and performs as well as or better<br>than the predicate devices.<br><br>GE Healthcare considers the SIGNA Voyager to be substantially<br>equivalent to the SIGNA Pioneer and Optima MR450w devices.                                                                                                                                                                                                                             |

---

**Source:** [https://fda.innolitics.com/device/K161567](https://fda.innolitics.com/device/K161567)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).