A1C EZ Glycohemoglobin Analysis System

K161533 · Biohermes Co., Ltd. · LCP · Jun 16, 2017 · Hematology

Device Facts

Record IDK161533
Device NameA1C EZ Glycohemoglobin Analysis System
ApplicantBiohermes Co., Ltd.
Product CodeLCP · Hematology
Decision DateJun 16, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.7470
Device ClassClass 2

Indications for Use

A1C EZ Glycohemoglobin Analysis System is an in vitro diagnostic test used to quantitatively measure the percent glycohemoglobin A1c or glycohemoglobin A1c fraction mmol/mol in venous whole blood samples. This system is intended for multiple patient use to monitor long term glycemic control in individuals previously diagnosed with diabetes. This test is not to be used for screening or diagnosis of diabetes or for neonatal use. The A1C EZ Glycohemoglobin Analysis System is intended for professional use in clinical laboratories only.

Device Story

The A1C EZ Glycohemoglobin Analysis System is a point-of-care diagnostic device for professional use in clinical laboratories. It accepts venous whole blood samples (2.5-4.5µL) applied to test strips. The device utilizes boronate affinity lateral chromatography; a solid-phase membrane matrix separates hemoglobin fractions based on pH-dependent charge and boronate-cis-diol binding. A built-in photometer measures reflectance to quantify the ratio of glycohemoglobin to total hemoglobin. The system provides quantitative HbA1c results in %NGSP or mmol/mol IFCC units. Results assist clinicians in monitoring long-term glycemic control in diabetic patients. The device is designed for multiple-patient use, with validated disinfection protocols using Clorox Healthcare Bleach Germicidal Wipes to ensure infection control.

Clinical Evidence

Bench testing only. Precision studies (N=360) showed total CVs of 2.3-3.1% (NGSP) and 2.8-5.3% (IFCC). Linearity verified across 3.6-14.5% range (r²=0.9989). Interference studies confirmed no significant impact from common substances, hemoglobin variants (C, D, E, S, F), or hematocrit levels (30-55%). Method comparison against Tosoh G8 and A1cNow+ showed strong correlation (r² > 0.96).

Technological Characteristics

Boronate affinity lateral chromatography; photometric reflectance measurement. Operates on venous whole blood. Dimensions/form factor: handheld analyzer. Connectivity: standalone. Sterilization: N/A (disinfection-compatible). Software: embedded firmware for signal processing and calculation. Operating conditions: 10°C-40°C, 30-75% RH.

Indications for Use

Indicated for professional use in clinical laboratories to monitor long-term glycemic control in individuals previously diagnosed with diabetes. Not for screening, diagnosis of diabetes, or neonatal use.

Regulatory Classification

Identification

A glycosylated hemoglobin assay is a device used to measure the glycosylated hemoglobins (A1a , A1b , and A1c ) in a patient's blood by a column chromatographic procedure. Measurement of glycosylated hemoglobin is used to assess the level of control of a patient's diabetes and to determine the proper insulin dosage for a patient. Elevated levels of glycosylated hemoglobin indicate uncontrolled diabetes in a patient.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE A. 510(k) Number: k161533 B. Purpose for Submission: New Device C. Measurand: Glycosylated Hemoglobin (HbA1c) D. Type of Test: Quantitative, boronate affinity chromatography assay E. Applicant: BioHermes Co., Ltd. F. Proprietary and Established Names: A1C EZ Glycohemoglobin Analysis System G. Regulatory Information: | Classification Name | Product Code | Device Class | Regulation | Panel | | --- | --- | --- | --- | --- | | Glycosylated hemoglobin assay | LCP | II | 21 CFR 864.7470 | 81 (Hematology) | | Discrete photometric chemistry analyzer for clinical use | JJE | I | 21 CFR 862.2160 | 75 (Chemistry) | H. Intended Use: 1. Intended use(s): See Indication(s) for use below. {2} 2. Indication(s) for use: A1C EZ Glycohemoglobin Analysis System is an in vitro diagnostic test used to quantitatively measure the percent glycohemoglobin A1c or glycohemoglobin A1c fraction mmol/mol in venous whole blood samples. This system is intended for multiple patient use to monitor long term glycemic control in individuals previously diagnosed with diabetes. This test is not to be used for screening or diagnosis of diabetes or for neonatal use. The A1C EZ Glycohemoglobin Analysis System is intended for professional use in clinical laboratories only. 3. Special conditions for use statement(s): This test is not for the screening or diagnosis of diabetes or neonatal use The analyzer is used for in vitro diagnostic use only. Do not use the result of this device to change therapy without guidance from healthcare professionals. Use fresh whole blood only. Do not use plasma and serum. Patients whose hematocrit (HCT) is too high (&gt; 55%) or too low (&lt; 30%) may not use this product. It may lead to incorrect results. This test is for clinical laboratory use only. This test is for prescription use only. This test should not be used in monitoring daily glucose control and should not replace daily home testing of urine and blood glucose levels. This test should not be used for analyzing samples from patients with conditions causing shortened red blood cell survival, such as hemolytic disease, pregnancy, and significant or chronic blood loss. 4. Special instrument requirements: A1C EZ 2.0 Glycohemoglobin analyzer I. Device Description: The A1C EZ Glycohemoglobin Analysis System consists of analyzer, test strip(s), buffer A, buffer B, blood sampler (s) and a calibration chip. The analyzer has a voice function. The A1C EZ Test Strip is composed of an absorbent pad, a plastic film absorbent pad cover, a polyethylene fiber membrane, a plastic film membrane cover and a PET base plate. Buffer A contains H₂O, sodium chloride, detergent, sodium hydroxide, and Tris-Cl 2 {3} Buffer B contains H₂O, zinc chloride, detergent, sodium hydroxide, and CAPS. The sampler consists of handle grip which is made of impact polystyrene and sample loading part which is made of coiled polyester. The sampler is able to absorb the blood sample automatically when the coiled polyester touches the blood drop. The sampler is designed to be able to absorb at least 3μL of blood, which is sufficient for a single test in A1C EZ 2.0 glycohemoglobin analysis system. Quality Control material is available separately. ## J. Substantial Equivalence Information: 1. Predicate device name(s): Bayer A1CNow+ (Professional Use) 2. Predicate 510(k) number(s): k090413 3. Comparison with predicate: | Similarities/Differences Assay | | | | --- | --- | --- | | Item | Candidate Device A1C EZ Glycohemoglobin Analysis System | Predicate Device A1CNow+ k090413 | | Intended Use | Quantitative measurement of the percent of glycated hemoglobin (%HbAlc, %A1C) levels in whole blood samples to monitor long term glycemic control in individuals previously diagnosed with diabetes. | Same | | Frequency of Use | Repeated Use (analyzer can be used multiple times) | Single use (all materials are disposed at the end of the test) | | Sample Type | K2-EDTA, NaF, LiHep venous whole blood | | | Calibration | Automatic, not required by end user | Not required by end user; each unit is factory calibrated | | Testing Environment | Professional Use | Same | | Test Strip Stability | 12 months at room | 15 months at room | {4} | Similarities/Differences Assay | | | | --- | --- | --- | | Item | Candidate Device A1C EZ Glycohemoglobin Analysis System | Predicate Device A1CNow+ k090413 | | | temperature | temperature | | Methodology | Boronate Affinity Lateral Chromatography | Immunoassay | | Measuring range | 4-14% HbA1c | 4-13% HbA1c | K. Standard/Guidance Document Referenced (if applicable): IEC 62304, Medical device software – software life cycle processes IEC 61010-1, Safety requirements for electrical equipment for measurement, control, and laboratory use – Part 1: General requirements IEC 60601-1-2, Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic disturbances. CLSI EP09-A3, Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline – Third Edition CLSI EP05-A2, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline—Second Edition L. Test Principle: The A1C EZ Glycohemoglobin Analysis System utilizes the boronate affinity lateral chromatography method to quantitatively measure the percentage of glycohemoglobinA1c (HbA1c) in total hemoglobin. A solid phase separation matrix in the membrane contains both negatively charged groups and boronate groups. When the acidic buffer A flows through the matrix membrane, positively charged hemoglobin (including glycohemoglobin and non-glycohemoglobin) binds to the negatively charged groups. At this time, the device measures the amount of total hemoglobin. When buffer B is added, the pH turns basic and the hemoglobin loses its positive charge. Hemoglobin is released from the matrix, but cis-diols of glycohemoglobin can bind to the boronate groups and stay in the matrix. At this time, the device measures the amount of glycated hemoglobin. The device uses a photometer to measure the reflectance and calculate the ratio of the glycohemoglobin to the total hemoglobin. {5} M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Internal Precision Study An internal precision study was conducted according to CLSI EP05-A2. Two K2-EDTA venous whole blood samples (Level 1 = 5.8%/40 mmol/mol and Level 2 = 12.2%/110 mmol/mol HbA1c) were tested in duplicate for two runs over 20 days on one instrument using three reagent lots (N = 240 total sample measurements). Results for within-run, between-run, between-day, and total precision are as follows: NGSP units (% HbA1c): | CV % | Lot 1 | | Lot 2 | | Lot 3 | | All lots combined | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | Level 1 | Level 2 | Level 1 | Level 2 | Level 1 | Level 2 | Level 1 | Level 2 | | Within-run | 3.0% | 2.7% | 3.4% | 2.9% | 3.3% | 2.9% | 3.2% | 2.8% | | Between-run | 0.8% | 0.5% | 0.0% | 0.0% | 1.3% | 0.0% | 0.9% | 0.0% | | Between-day | 2.1% | 1.6% | 1.1% | 0.9% | 0.0% | 1.0% | 0.0% | 0.3% | | Total | 3.8% | 3.1% | 3.5% | 3.0% | 3.5% | 3.1% | 3.3% | 2.8% | IFCC units (mmol/mol): | CV % | Lot 1 | | Lot 2 | | Lot 3 | | All lots combined | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | Level 1 | Level 2 | Level 1 | Level 2 | Level 1 | Level 2 | Level 1 | Level 2 | | Within-run | 4.8% | 3.2% | 5.4% | 3.5% | 5.2% | 3.6% | 5.3% | 3.4% | | Between-run | 1.3% | 0.7% | 0.0% | 0.0% | 2.1% | 0.0% | 1.4% | 0.0% | | Between-day | 3.4% | 1.9% | 1.7% | 1.1% | 0.0% | 1.2% | 0.0% | 0.5% | | Total | 6.0% | 3.8% | 5.6% | 3.6% | 5.6% | 3.8% | 5.8% | 4.4% | External Precision Study An external precision study was conducted at three sites. Two operators per site evaluated three K2-EDTA venous whole blood samples. Samples were tested in two non-consecutive runs for a total of six days with three reagent lots and ten instruments. Results are as follows: {6} Site-to-site variability: NGSP units (% HbA1c): | Site | N | 5.1% HbA1c | | 7.3% HbA1c | | 11.7% HbA1c | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | Mean | %CV | Mean | %CV | Mean | %CV | | 1 | 120 | 5.04 | 3.0 | 7.29 | 2.7 | 11.49 | 2.5 | | 2 | 120 | 5.16 | 2.9 | 7.43 | 2.5 | 11.57 | 2.1 | | 3 | 120 | 5.08 | 2.8 | 7.40 | 2.5 | 11.56 | 2.2 | IFCC units (mmol/mol HbA1c): | Site | N | 32 mmol/mol HbA1c | | 56 mmol/mol HbA1c | | 104 mmol/mol HbA1c | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | Mean | %CV | Mean | %CV | Mean | %CV | | 1 | 120 | 31.6 | 5.2 | 56.2 | 3.8 | 102.0 | 3.0 | | 2 | 120 | 32.9 | 5.0 | 57.7 | 3.6 | 103.0 | 2.6 | | 3 | 120 | 32.0 | 4.9 | 57.3 | 3.5 | 102.8 | 2.7 | Lot-to-lot variability: NGSP units (% HbA1c): | Lot | N | 5.1% HbA1c | | 7.3% HbA1c | | 11.7% HbA1c | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | Mean | %CV | Mean | %CV | Mean | %CV | | 1 | 120 | 5.09 | 3.2 | 7.37 | 2.8 | 11.57 | 2.1 | | 2 | 120 | 5.10 | 3.1 | 7.35 | 2.6 | 11.47 | 2.3 | | 3 | 120 | 5.09 | 3.0 | 7.41 | 2.6 | 11.58 | 2.4 | IFCC units (mmol/mol HbA1c): | Lot | N | 32 mmol/mol HbA1c | | 56 mmol/mol HbA1c | | 104 mmol/mol HbA1c | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | Mean | %CV | Mean | %CV | Mean | %CV | | 1 | 120 | 32.1 | 5.5 | 57.0 | 4.0 | 102.9 | 2.5 | | 2 | 120 | 32.2 | 5.3 | 56.8 | 3.7 | 101.9 | 2.9 | | 3 | 120 | 32.2 | 5.2 | 57.5 | 3.6 | 103.0 | 2.9 | {7} Operator-to-operator variability: NGSP units (% HbA1c): | Operator | N | 5.1% HbA1c | | 7.3% HbA1c | | 11.7% HbA1c | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | Mean | %CV | Mean | %CV | Mean | %CV | | 1 | 60 | 5.04 | 3.0 | 7.28 | 3.0 | 11.45 | 2.4 | | 2 | 60 | 5.05 | 3.0 | 7.30 | 2.3 | 11.52 | 2.5 | | 3 | 60 | 5.18 | 2.8 | 7.47 | 2.5 | 11.56 | 2.1 | | 4 | 60 | 5.15 | 3.0 | 7.40 | 2.5 | 11.59 | 2.2 | | 5 | 60 | 5.10 | 3.1 | 7.44 | 2.7 | 11.55 | 2.2 | | 6 | 60 | 5.09 | 3.0 | 7.35 | 2.1 | 11.57 | 2.4 | IFCC units (mmol/mol HbA1c): | Lot | N | 32 mmol/mol HbA1c | | 56 mmol/mol HbA1c | | 104 mmol/mol HbA1c | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | Mean | %CV | Mean | %CV | Mean | %CV | | 1 | 60 | 31.5 | 5.2 | 56.1 | 4.2 | 101.6 | 3.0 | | 2 | 60 | 31.6 | 5.2 | 56.2 | 3.3 | 102.4 | 3.1 | | 3 | 60 | 33.0 | 4.9 | 58.1 | 3.5 | 102.8 | 2.6 | | 4 | 60 | 32.8 | 5.2 | 57.4 | 3.6 | 103.2 | 2.7 | | 5 | 60 | 31.7 | 4.8 | 57.8 | 3.8 | 102.7 | 2.5 | | 6 | 60 | 32.2 | 5.0 | 56.8 | 3.0 | 102.9 | 2.9 | Combined to include all sites, operators, instruments, and test strip lots: NGSP units (% HbA1c) | Sample | N | Total | | | --- | --- | --- | --- | | | | Mean | %CV | | 5.1% HbA1c | 360 | 5.09 | 3.1 | | 7.3% HbA1c | 360 | 7.37 | 2.7 | | 11.7 HbA1c | 360 | 11.54 | 2.3 | IFCC units (mmol/mol HbA1c) | Sample | N | Total | | | --- | --- | --- | --- | | | | Mean | %CV | | 32 mmol/mol HbA1c | 360 | 32.2 | 5.3 | | 56 mmol/mol HbA1c | 360 | 57.1 | 3.8 | | 104 mmol/mol HbA1c | 360 | 102.6 | 2.8 | {8} b. Linearity/assay reportable range: Linearity was evaluated according to CLSI EP06-A. The linearity of the A1C EZ Glycohemoglobin Analysis System was verified by mixing a high K2-EDTA venous whole blood sample (14.5%) with a low sample (3.6%) to create 11 sample levels covering the assay range (3.6, 4.7, 5.8, 6.9, 8.0, 9.1, 10.1, 11.2, 12.3, 13.4, and 14.5%). All levels were analyzed in singlicate on five instruments with one reagent lot. The mean observed %HbA1c value was determined for each level and plotted versus the % HbA1c value determined by a reference method (Tosoh G8 analyzer). The linear regression results are as follows: NGSP units (% HbA1c): $$ y = 1.0334x - 0.1628, r^2 = 0.9989 $$ IFCC units (mmol/mol HbA1c): $$ y = 1.0377x - 1.4795, r^2 = 0.9990 $$ The study supports the sponsor’s claimed measuring range of 4.0-14% HbA1c. c. Traceability, Stability, Expected values (controls, calibrators, or methods): Traceability The A1C EZ Glycohemoglobin Analysis System is certified with the National Glycohemoglobin Standardization Program (NGSP). The NGSP certification expires in one year. See NGSP website for current certification at http://www.ngsp.org. Stability: Real time stability studies are ongoing. Accelerated and real time stability protocols and acceptance criteria for the A1C EZ Glycohemoglobin Analysis System were reviewed and considered acceptable to support the sponsor’s shelf life claims: Test kit: 18 months at room temperature Test strips: 1 month after opening vial (vial is closed after every test) at 39-75°F (4-24°C) d. Detection limit: Please see linearity study above in M1b. {9} # e. Analytical specificity: An interference study was performed to assess common or known endogenous and exogenous substances that could interfere with the A1C EZ Glycohemoglobin Analysis System. The potential interferents listed below were spiked into two K2-EDTA venous whole blood samples with two different levels of $\%$ HbA1c (5.2 and $9.2\%$ ). The $\%$ HbA1c values of the spiked samples were compared to reference samples (samples containing no interferent). Samples were tested with 10 replicates and non-significant interference was defined as $\leq \pm 6\%$ difference relative to the reference sample. | Substance | Highest Concentration tested at which no interference was observed | | --- | --- | | Acetaminophen | 20 mg/dL | | Ascorbic Acid | 30 mg/dL | | Bilirubin, conjugated | 600 mg/L | | Bilirubin, unconjugated | 600 mg/L | | Triglycerides | 900 mg/dL | | Cholesterol | 1000 mg/dL | | Acetylcysteine | 1663 mg/L | | Metronidazole | 200 mg/L | | Ampicillin | 1000 mg/L | | Acetylsalicylic Acid | 1000 mg/L | | Doxycycline | 50 mg/L | | Cyclosporine A | 5 mg/L | | Theophylline | 100 mg/L | | Phenylbutazone | 400 mg/L | | Rifampicin | 64 mg/L | | Glucose | 1000 mg/dL | | Glyburide | 2 mg/L | | Ibuprofen | 500 mg/L | | Levodopa | 20 mg/L | | Salicylic acid | 599 mg/L | | Metformin | 40 mg/L | | Rheumatoid Factor | 491 IU/mL | | Hemoglobin | 500 mg/dL | # Labile A1c interference: Two K2-EDTA venous whole blood samples (5.0% and 10.5% HbA1c) containing $1200\mathrm{mg / dL}$ glucose were incubated for three hours at $37^{\circ}\mathrm{C}$ to facilitate the formation of labile A1c. The samples were tested in replicates of 10 using the A1C EZ Glycohemoglobin Analysis System. The sponsor defined non-significant interference {10} as ≤ ± 7% difference between samples containing glucose and the control samples. The results support the sponsor's claim that &gt; 5% labile HbA1c does not interfere with the A1C EZ Glycohemoglobin Analysis System. ## Carbamylated hemoglobin interference: Two K2-EDTA venous whole blood samples (5.2% and 7.4% HbA1c) were treated with 4mM potassium cyanate and incubated for two hours at 37°C to facilitate the formation of carbamylated hemoglobin. The samples were tested in replicates of 10 using the A1C EZ Glycohemoglobin Analysis System. The sponsor defined non-significant interference as ≤ ± 7% difference between samples containing potassium cyanate and the control samples. The results support the sponsor's claim that &gt; 5% carbamylated hemoglobin does not interfere with the A1C EZ Glycohemoglobin Analysis System. ## Acetylated hemoglobin interference: Two K2-EDTA venous whole blood samples (5.4% and 9.8% HbA1c) were treated with 20 mM acetylsalicylate and incubated for 16 hours at 37°C to facilitate the formation of acetylated hemoglobin. The samples were tested in replicates of 10 using the A1C EZ Glycohemoglobin Analysis System. The sponsor defined non-significant interference as ≤ ± 7% difference between samples containing acetylsalicylate and the control samples. The results support the sponsor's claim that acetylated hemoglobin does not interfere with the A1C EZ Glycohemoglobin Analysis System. ## Total hemoglobin: The effect of different levels of total hemoglobin was evaluated using K2-EDTA venous whole blood samples (5.7% and 8.7% HbA1c). Erythrocytes and plasma of the same sample were mixed to obtain 10 levels containing total hemoglobin concentrations of 69 – 226 g/L (21.2% - 66.4% hematocrit). The samples were tested in replicates of 10 using the A1C EZ Glycohemoglobin Analysis System. The sponsor defined non-significant interference as &lt; ± 7% difference between samples containing hemoglobin and the known HbA1c concentration. The data supports the claimed total hematocrit range of 30 – 55% g/L. ## Hemoglobin variants: A hemoglobin variant study was performed using 117 K2-EDTA venous whole blood samples (ranging from 3.9 to 10.6% HbA1c) containing known levels of hemoglobin variants C, D, E, S, and F. The samples were tested for % HbA1c in replicates of 10 using the A1C EZ Glycohemoglobin Analysis System and results were compared to results obtained with a reference method (Sebia Capillarys 2 Flex Piercing, k122101). Non-significant interference was defined as ≤ ± 8% difference between the candidate and reference method. 10 {11} The testing results indicate that there is no significant interference for Hemoglobin C (≤ 44.7%), Hemoglobin D (≤ 41.7%), Hemoglobin E (≤ X32.6%), Hemoglobin S (≤ 37.8%), and Hemoglobin F (≤ 14.7%). The labeling contains the following statement: "The results from the A1C EZ 2.0 Glycohemoglobin Analysis System show that there is no significant interference for Hemoglobin C (≤ 44.7%), Hemoglobin D (≤ 41.7%), Hemoglobin E (≤ X32.6%), Hemoglobin S (≤ 37.8%) and Hemoglobin F (≤ 14.7%)." ## Hemolysis The effect of hemolysis was evaluated using three K2-EDTA venous whole blood samples (5.7%, 7.9% and 9.9% HbA1c) containing up to 7700 mg/dL hemolysate generated from the same samples. Samples were each tested with 10 A1C EZ 2.0 Glycohemoglobin Analysis instruments and one test strip lot, and results were compared to results from control samples without added hemolysate. The sponsor defined non-significant interference as ≤ ± 6% difference between hemolyzed samples and control samples. The results support the sponsor's claim that hemolysis of up to 5000 mg/dL has no significant effect on HbA1c measurements obtained with the A1C EZ Glycohemoglobin Analysis System. f. Assay cut-off: Not applicable. 2. Comparison studies: a. Method comparison with predicate device: ### Internal Method Comparison Study An internal method comparison study was conducted according to CLSI EP09-A2. Up to 120 K2-EDTA venous whole blood samples ranging from 4.2% to 13.3% HbA1c were analyzed in singlicate with three reagent lots over five days on the A1C EZ Glycohemoglobin Analysis System and compared to measurements obtained for the same samples on the Tosoh G8 HPLC analyzer and the A1cNow+ device. The linear regression results are as follows: {12} NGSP units (% HbA1c): | Comparative Method | N | % HbA1c | Lot | Slope | Intercept | R² | | --- | --- | --- | --- | --- | --- | --- | | Tosoh G8 | 120 | 4.2 to 13.3 | 1 | 0.971 | 0.201 | 0.993 | | | | | 2 | 0.977 | 0.176 | 0.992 | | | | | 3 | 0.987 | 0.124 | 0.993 | | A1cNow+ | 118 | 4.1 to 12.7 | 1 | 0.966 | 0.203 | 0.963 | | | | | 2 | 0.976 | 0.152 | 0.967 | | | | | 3 | 0.989 | 0.079 | 0.971 | IFCC units (mmol/mol HbA1c): | Comparative Method | N | % HbA1c | Lot | Slope | Intercept | R² | | --- | --- | --- | --- | --- | --- | --- | | Tosoh G8 | 120 | 23.5 to 121.9 | 1 | 0.975 | 1.326 | 0.985 | | | | | 2 | 0.979 | 1.266 | 0.984 | | | | | 3 | 0.989 | 0.957 | 0.985 | | A1cNow+ | 118 | 21.3 to 115.3 | 1 | 0.966 | 1.431 | 0.963 | | | | | 2 | 0.975 | 1.127 | 0.967 | | | | | 3 | 0.988 | 0.652 | 0.971 | # b. Matrix comparison: 46 matched K2-EDTA, lithium heparin, and sodium fluoride venous whole blood samples ranging from 4.4 to $13.4\%$ HbA1c were tested in singlicate with one reagent lot on the A1C EZ Glycohemoglobin Analysis System. Results were compared to measurements obtained for the same samples with the Tosoh HLC-723 G8 (k131580). Linear regression analysis: NGSP units (% HbA1c): | | K2-EDTA | Lithium Heparin | Sodium Fluoride | | --- | --- | --- | --- | | Slope | 1.0048 | 0.9786 | 0.9976 | | Intercept | 0.0074 | 0.1037 | -0.0430 | | R² | 0.9940 | 0.9942 | 0.9947 | IFCC units (mmol/mol HbA1c) : | | K2-EDTA | Lithium Heparin | Sodium Fluoride | | --- | --- | --- | --- | | Slope | 1.0036 | 0.9790 | 0.9986 | | Intercept | 0.3156 | 0.6082 | -0.5984 | | R² | 0.9878 | 0.9886 | 0.9886 | {13} 13 3. Clinical studies: a. Clinical Sensitivity: Not applicable. b. Clinical specificity: Not applicable. c. Other clinical supportive data (when a. and b. are not applicable): Not applicable. 4. Clinical cut-off: Not applicable. 5. Expected values/Reference range: The sponsor provides the following in the labeling: The American Diabetes Association (ADA) recommendations are summarized in the following table: | % HbA1c (mmol/mol) | Glycemic Control | | --- | --- | | <8% (64 mmol/mol) | Less stringent | | < 7% (53 mmol/mol) | General (non-pregnant adult) | | < 6% (48 mmol/mol) | More stringent | HbA1c values above 6.5% HbA1c (48mmol/mol) are an indication of hyperglycemia during the preceding 2 to 3 months or longer. Source: American Diabetes Association. Position Statement: Standards of medical care in diabetes - 2012. Diabetes Care 2012;35 (Suppl 1):S11–S63. N. Instrument Name: A1C EZ Glycohemoglobin analyzer O. System Descriptions: 1. Modes of Operation: Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device? {14} Yes _______ or No ☐ X Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission? Yes _______ or No ☐ X 2. Software: FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types: Yes ☐ X or No _______ 3. Specimen Identification: There is no sample identification function with this device. Samples are applied directly to the device as they are collected. 4. Specimen Sampling and Handling: The user collects venous blood samples by pipetting 4-5μL of whole blood onto the blood sampling device that is supplied as part of the kit. 5. Calibration: Calibration is automatic and obtained by inserting a calibration code chip into the device. Each kit contains a lot-specific calibration code chip. The calibration is traceable to the International Federation of Clinical Chemistry (IFCC) and the device is certified with the NGSP standardization program. 6. Quality Control: Low and high levels of quality control material intended for use with the A1c EZ 2.0 Glycohemoglobin Test System are available for purchase. P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above: Vibration Study In order to validate the effect of vibration of device during the test, the devices were put on the surface of a shaker during testing. The shaker rotated in a manner where the buffer used in the test would not be shaken out of the device (~15rpm). Three K2-EDTA venous whole blood samples (5.0%, 6.4% and 10.3% HbA1c) were tested with ten devices in replicates of 20 for each sample (two measurements for each sample per device). The average value, 14 {15} coefficient of variation, and % bias were calculated and compared with the results obtained from tests under normal conditions. The results support the sponsor’s claim that vibration during testing does not affect test results. ## Sample Volume Study The blood sampling device is designed to absorb 2.5-4.5 μL of capillary fingerstick or venous whole blood. The sponsor provided a study to show that the total volume of blood absorbed by the blood sampling device ranges from 2.9 to 4.0 μL with a mean volume of 3.5 μL. The sponsor then investigated the effect of low or high sample volume on the test. 0.5 μL, 1.5 μL, 3 μL, 4 μL, or 5.5 μL of K2-EDTA venous whole blood (5.9%, 8.4% and 10.4% HbA1c) was pipetted onto the coiled polyester of the blood sampler. Each sample was tested once on a total of ten devices. The bias and % bias compared to the known HbA1c concentration in the samples were calculated. For 0.5 μL, the device reported error code E7 for all data points. The results support the sponsor’s claim that low or high blood volume does not affect the test results. ## Operating Conditions Study In order to verify the effectiveness of the A1C EZ Glycohemoglobin Analysis System at extreme operating conditions, the sponsor tested three K2-EDTA venous whole blood samples (5.2%, 7.4%, and 10.2% HbA1c) on 10 EZ 2.0 instruments at the following operating condition combinations: 11°C/11% RH, 10°C/90%RH, 38°C/10%RH, and 40°C/90%RH. Results were compared to %HbA1c observed for the same samples at 22°C/40%RH and %bias was calculated. The results support the sponsor’s claim for operation of the device at temperatures of 50 to 104°F (10 to 40°C) and 30 to 75% relative humidity. ## Infection Control Studies Disinfection efficacy studies were performed on the materials comprising the meter to demonstrate complete inactivation of Hepatitis B virus with Clorox Healthcare Bleach Germicidal Wipes (EPA #67619-12). Robustness studies were performed to demonstrate that there was no change in performance or in external materials of the meter after 18250 disinfection cycles with Clorox Healthcare Bleach Germicidal Wipes. The robustness studies were designed to simulate 5 years of multiple-patient use with 10 disinfections per day. ## Readability assessment A Flesch-Kinkaid reading level assessment was conducted demonstrating that the user manuals and test strip package inserts were written at an 7th grade reading level. ## Q. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. 15 {16} R. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 16
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...