K161513 · Euroimmun US · LSR · Aug 25, 2016 · Microbiology
Device Facts
Record ID
K161513
Device Name
Anti-Borrelia burgdorferi US Westernblot (IgG)
Applicant
Euroimmun US
Product Code
LSR · Microbiology
Decision Date
Aug 25, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3830
Device Class
Class 2
Attributes
Software as a Medical Device
Intended Use
The EUROIMMUN Anti-Borrelia burgdorferi US Westernblot (IgG) kit is a Western blot assay intended for the qualitative determination of IgG class antibodies against Borrelia burgdorferi in human serum and plasma (K⁺-EDTA, Li⁺-heparin, Na⁺-citrate) samples that have been found positive or equivocal/borderline using an enzyme immunoassay (EIA) or immunofluorescence assay (IFA) test procedure for Borrelia burgdorferi antibodies. Results can be read manually or automated utilizing EUROLineScan. This test is used as an aid in the diagnosis of infections with B. burgdorferi and the associated diseases, in conjunction with other laboratory and clinical findings.
Device Story
Western blot assay for qualitative detection of IgG antibodies against Borrelia burgdorferi. Input: human serum or plasma samples (K+-EDTA, Li+-heparin, Na+-citrate) previously screened as positive/equivocal by EIA/IFA. Process: patient samples incubated with test strips containing electrophoretically separated B. burgdorferi antigens; specific IgG antibodies bind to antigens; enzyme-labeled anti-human IgG conjugate added; color reaction catalyzed by substrate. Output: visual or automated (EUROLineScan) identification of specific protein bands on nitrocellulose membrane. Used in clinical laboratories; interpreted by technicians or software according to CDC criteria. Results aid diagnosis of Lyme disease in conjunction with clinical findings.
Clinical Evidence
No clinical data provided in the document.
Technological Characteristics
Western blot assay using B. burgdorferi (strain B31) antigens separated by SDS-PAGE and transferred to nitrocellulose membranes. Detection uses alkaline phosphatase anti-human IgG conjugate and NBT/BCIP substrate. Qualitative format. Manual or automated (EUROLineScan) reading. Compatible with serum and plasma (K+-EDTA, Li+-heparin, Na+-citrate).
Indications for Use
Indicated for the qualitative detection of IgG antibodies against Borrelia burgdorferi in human serum and plasma samples previously tested as positive or equivocal/borderline by EIA or IFA. Used as an aid in the diagnosis of B. burgdorferi infections and associated diseases.
Regulatory Classification
Identification
Treponema pallidum treponemal test reagents are devices that consist of the antigens, antisera and all control reagents (standardized reagents with which test results are compared) which are derived from treponemal sources and that are used in the fluorescent treponemal antibody absorption test (FTA-ABS), the Treponema pallidum immobilization test (T.P.I.), and other treponemal tests used to identify antibodies to Treponema pallidum directly from infecting treponemal organisms in serum. The identification aids in the diagnosis of syphilis caused by bacteria belonging to the genus Treponema and provides epidemiological information on syphilis.
Related Devices
K991180 — BORRELIA BURGDORFERI IGG WESTERN BLOT TEST SYSTEM · Zeus Scientific, Inc. · Sep 3, 1999
K051071 — VIRAMED BIOTECH BORRELIA B31 IGG VIRABLOT · Viramed Biotech AG · Aug 11, 2005
K991185 — BORRELIA BURGDORFERI IGM WESTERN BLOT TEST SYSTEM · Zeus Scientific, Inc. · Sep 3, 1999
K970761 — LIMITED HUMAN LYME IGG WESTERN BLOT KIT (CAMBRIDGE DIAGNOSTICS IRELAND) · Cambridge Diagnostics Ireland, Ltd. · Apr 9, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.
Public Health Service
August 25, 2016
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
EUROIMMUN US Michael Locke Director of Regulatory Affairs 1 Bloomfield Avenue Mountain Lakes, NJ 07046
Re: K161513
Trade/Device Name: EUROIMMUN Anti-Borrelia burgdorferi US Westernblot (IgG) Regulation Number: 21 CFR 866.3830 Regulation Name: Treponema pallidum treponemal test reagents Regulatory Class: II Product Code: LSR Dated: May 26, 2016 Received: June 2, 2016
Dear Mr. Locke:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Steven R. Gitterman -S
for Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161513
#### Device Name
EUROIMMUN Anti-Borrelia burgdorferi US Westernblot (IgG)
Indications for Use (Describe)
The EUROIMMUN Anti-Borrelia burgdorferi US Westernblot (IgG) kit is a Western blot assay intended for the qualitative determination of IgG class antibodies against Borrelia burgdorferi in human serum and plasma (K+-EDTA, Li+-heparin, Na+-citrate) samples that have been found positive or equivocal/borderline using an enzyme immunoassay (EIA) or immunofluorescence assay (IFA) test procedure for Borrelia burgdorferi antibodies. Results can be read manually or automated utilizing EUROLineScan. This test is used as an aid in the diagnosis of infections with B. burgdorferi and the associated diseases, in conjunction with other laboratory and clinical findings.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| <div> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span style="font-size: 20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches