← Product Code [MNS](/productcode/MNS) · K161487

# VPAP Adapt SV, VPAP Tx, S9 VPAP Tx (K161487)

_Resmed, Ltd. · MNS · Sep 9, 2016 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K161487

## Device Facts

- **Applicant:** Resmed, Ltd.
- **Product Code:** [MNS](/productcode/MNS.md)
- **Decision Date:** Sep 9, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5895
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic, Real-World Evidence, Pediatric

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K161487 · Sep 9, 2016 | VPAP Adapt SV, VPAP Tx, S9 VPAP Tx | Resmed, Ltd. | SERVE-HF clinical trial data | The SERVE-HF study is used to support a new contraindication for ASV therapy in patients with chronic, symptomatic heart failure (NYHA 2-4) with reduced left ventricular ejection fraction (LVEF ≤ 45%) and moderate to severe predominant central sleep apnea. | Labeling change; Contraindication; Heart failure; ASV therapy; Safety |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| SERVE-HF; Randomized, parallel, event-driven, international multicenter study | 1325 patients with chronic, symptomatic heart failure (NYHA 2-4) with reduced left ventricular ejection fraction (LVEF ≤ 45%) and moderate to severe predominant central sleep apnea; Sample Size: 1325; Number of Sites: International multicenter | Guideline-based medical management | Survival and cardiovascular outcomes |

## Indications for Use

The VPAP ADAPT SV is intended to provide non-invasive ventilatory support to treat adult patients with obstructive sleep apnea (OSA) and Respiratory Insufficiency caused by central and/or mixed apneas and periodic breathing. The VPAP Tx is indicated for the treatment of patients weighing more than 66 lb (> 30 kg) with obstructive sleep apnea (OSA), respiratory insufficiency, central or mixed apneas, or periodic breathing. The VPAP TX is intended to be used in a clinical environment. The S9 VPAP Tx is indicated for the treatment and titration of patients with obstructive sleep apnea (OSA), respiratory insufficiency, central or mixed apneas, or periodic breathing. CPAP, S, ST, T and PAC modes are indicated for patients weighing more than 30lb (13 kg); all other modes are indicated for patients weighing more than 66lb (30 kg). The S9 VPAP Tx is intended to be used in a clinical environment.

## Device Story

Non-invasive flow generator devices (VPAP Adapt SV, VPAP Tx, S9 VPAP Tx) provide positive airway pressure therapy. Devices use blower-based systems with flow/pressure sensors and microprocessor-controlled electronics to deliver pressurized air via mask/tubing. Systems operate in various modes (CPAP, ASV, Bi-level, VAuto, etc.) to stabilize ventilation or maintain airway patency. Devices measure instantaneous ventilation and calculate target ventilation (90% of recent average) to servo-control support. Used in clinical environments or home settings by patients under physician guidance. Output allows clinicians to treat sleep-disordered breathing; therapy stabilizes ventilation and reduces apneas. ASV therapy is specifically contraindicated for patients with chronic symptomatic heart failure (NYHA 2-4) and reduced LVEF (≤45%) due to increased cardiovascular mortality risk identified in the SERVE-HF study.

## Clinical Evidence

No clinical data generated for these submissions. Labeling change is supported by the SERVE-HF study, a randomized, parallel, event-driven, international multicenter study of 1325 patients with chronic, symptomatic heart failure (NYHA 2-4) and reduced LVEF (≤45%) with moderate to severe predominant central sleep apnea, which demonstrated increased cardiovascular mortality risk with ASV therapy.

## Technological Characteristics

Blower-based positive pressure systems; brushless low-voltage DC motors; centrifugal or multiple-impeller axial fans; servo-control loops for pressure regulation; power supply 100-240V, 50-60Hz. Software platforms include Nucleus Plus or Micrium uC/OS-II. Devices are non-invasive, non-life supporting ventilators.

## Regulatory Identification

A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.

## Predicate Devices

- VPAP Adapt SV ([K051364](/device/K051364.md))
- VPAP Tx ([K092186](/device/K092186.md))
- S9 VPAP Tx ([K123511](/device/K123511.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The text is written in a sans-serif font.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 9, 2016

ResMed Ltd c/o Ms. Larissa D'andrea Director, Government Regulatory Affairs ReMed Corps 9001 Spectrum Center Boulevard San Diego, California 92123

Re: K161487

Trade/Device Name: VPAP Adapt SV, VPAP Tx, S9 VPAP Tx Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: Class II Product Code: MNS Dated: August 10, 2016 Received: August 11, 2016

Dear Ms. D'andrea:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

Device Name VPAP Adapt SV

Indications for Use (Describe)

The VPAP ADAPT SV is intended to provide non-invasive ventilatory support to treat adult patients with obstructive sleep apnea (OSA) and Respiratory Insufficiency caused by central and/or mixed apneas and periodic breathing.

| Type of Use (Select one or both, as applicable)            |                                                           |
|------------------------------------------------------------|-----------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
  
Food and Drug Administration
  
Office of Chief Information Officer
  
Paperwork Reduction Act (PRA) Staff
  
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## Indications for Use

Expiration Date: January 31, 2017 See PRA Statement below.

Form Approved: OMB No. 0910-0120

510(k) Number (if known)

Device Name VPAP Tx

Indications for Use (Describe)

The VPAP TX is indicated for the treatment of patients weighing more than 66 lb (> 30 kg) with obstructive sleep apnea (OSA), respiratory insufficiency, central or mixed apneas, or periodic breathing.

The VPAP TX is intended to be used in a clinical environment.

| Type of Use (Select one or both, as applicable)                                      |                                               |
|--------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
  
Food and Drug Administration
  
Office of Chief Information Officer
  
Paperwork Reduction Act (PRA) Staff
  
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known)

Device Name S9 VPAP Tx

Indications for Use (Describe)

The S9 VPAP Tx is indicated for the treatment and titration of patients with obstructive sleep apnea (OSA), respiratory insufficiency, central or mixed apneas, or periodic breathing. CPAP, S, ST, T and PAC modes are indicated for patients weighing more than 30lb (13 kg); all other modes are indicated for patients weighing more than 66lb (30 kg). The S9 VPAP Tx is intended to be used in a clinical environment.

| Type of Use (Select one or both, as applicable)                                                                                  |                                                                         |
|----------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)                 </span> </div> | <div> <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
  
Food and Drug Administration
  
Office of Chief Information Officer
  
Paperwork Reduction Act (PRA) Staff
  
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

# 510(k) Summary – VPAP Adapt SV

| Required                                                                   | By Section 807.92 (c)                                                                                                                                                                                                    |
|----------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared                                                              | 18 May, 2016                                                                                                                                                                                                             |
| Owners Name                                                                | ResMed Ltd<br>1 Elizabeth Macarthur Drive Bella<br>Vista, NSW 2153, Australia                                                                                                                                            |
| Submitter                                                                  | Jasjit Baveja<br>+ 61 2 8884 1518 (Phone)<br>+ 61 2 8884 2000 (FAX)<br>Jasjit.baveja@resmed.com.au                                                                                                                       |
| Official Contact                                                           | Larissa D'Andrea<br>Director, Government & Regulatory Affairs ResMed<br>Corp<br>9001 Spectrum Center Boulevard San<br>Diego, CA 92123<br>+1 858 836 6837 (Phone)<br>+1 858 836 5519 (Fax)<br>Larissa.D'Andrea@resmed.com |
| Product codes                                                              | 73 MNS                                                                                                                                                                                                                   |
| Class<br>Classification Reference<br>Common/Usual Name<br>Proprietary Name | II<br>21 CFR 868.5895, Product Code 73 MNS<br>Ventilator, Continuous, Non-Life Supporting<br>VPAP Adapt SV                                                                                                               |
| Predicate device(s)                                                        | VPAP Adapt SV (K051364)                                                                                                                                                                                                  |

{6}------------------------------------------------

## Reason for submission

This 510(k) is being submitted as a labelling change for the following contraindication:

ASV therapy is contraindicated in patients with chronic, symptomatic heart failure (NYHA 2-4) with reduced left ventricular ejection fraction (LVEF ≤ 45%) and moderate to severe predominant central sleep apnea.

This contraindication is supported by SERVE-HF, which was a randomized, parallel, event-driven, international multicenter study in 1325 patients with chronic, symptomatic heart failure (NYHA 2-4) with reduced left ventricular ejection fraction (LVEF≤45%) and moderate to severe predominant central sleep apnea. The study investigated the effects of adding ASV to guideline-based medical management on survival and cardiovascular outcomes. The addition of ASV did not improve outcomes and showed an increased risk of cardiovascular mortality despite effective control of central sleep apnea.

#### Intended Use

The VPAP ADAPT SV is intended to provide non-invasive ventilatory support to treat adult patients with obstructive sleep apnea (OSA) and Respiratory caused by central and/or mixed apneas and periodic breathing.

### Substantial Equivalence

The new device has the following similarities to the previously cleared predicate devices.

- Same intended use
- Same operating principle
- Similar technologies
- . Same manufacturing process

ResMed has determined that the new device has not altered the safety and effectiveness of providing non- invasive ventilatory support for treatment for patients with Obstructive Sleep Apnoea (OSA) and respiratory insufficiency caused by central and/or mixed apneas, and periodic breathing. The new device complies with the applicable requirements referenced in the FDA guidance documents:

- FDA Draft Reviewer Guidance for Ventilators (July 1995)
- FDA Reviewer Guidance for Premarket Notification Submissions (November 1993)
- FDA Reviewer's and Industry, Guidance for the content of premarket submissions for software contained in medical devices, May 1998

As this was a labelling change only, no further clinical testing was required to show substantial equivalence to the predicate device VPAP Adapt SV (K051364)

The new device is as safe and effective as the predicate device.

#### Device description

The VPAP Adapt SV is identical to the predicate device VPAP Adapt SV (K051364), using a blower based positive pressure system. The device platform is identical to the VPAP Adapt SV (K051364) and contains a blower (motorfan assembly), flow and pressure sensors, and processing electronics. The blower supplies pressurized air to the patient via a mask and air tubing.

The VPAP Adapt SV is a non-invasive flow generator device designed to provide adaptive servo-ventilation therapy to stabilize a patient's ventilation. The device continually measures instantaneous ventilation, and calculates a target ventilation equal to 90% of the patient's recent average ventilation (time constant 100 seconds). It then adjusts the degree of support to servo-control the patient's ventilation to at least equal the target ventilation.

{7}------------------------------------------------

Therapy modes contained in the VPAP Adapt SV are

- CPAP mode;
- Auto Servo Ventilation (ASV) ●

Therapy modes come from the predicate VPAP Adapt SV (K051364).

| Characteristic                     | VPAP Adapt SV (K051364)                                                                                                                                                                                                                         | New Device (VPAP Adapt SV)                                                                                                                                                                                                                      | Comments                                                          |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
| Indication for<br>use              | The VPAP ADAPT is intended to<br>provide non-invasive ventilatory<br>support to treat adult patients with<br>obstructive sleep apnea (OSA)<br>and Respiratory Insufficiency<br>caused by central and/or mixed<br>apneas and periodic breathing. | The VPAP ADAPT is intended to<br>provide non-invasive ventilatory<br>support to treat adult patients with<br>obstructive sleep apnea (OSA) and<br>Respiratory Insufficiency caused by<br>central and/or mixed apneas and<br>periodic breathing. | Equivalent<br>Only labeling change to<br>include contraindication |
| Location of use                    | Hospital/Home                                                                                                                                                                                                                                   | Hospital/Home                                                                                                                                                                                                                                   | Equivalent                                                        |
| Pressure Range and Treatment Modes | 4-13 cm H2O (CPAP)<br>4-25 cm H2O (ASV)                                                                                                                                                                                                         | 4-13 cm H2O (CPAP)<br>4-25 cm H2O (ASV)                                                                                                                                                                                                         | Equivalent                                                        |
| RAMP<br>Settings                   | Feature enables a slow<br>increase in the patient airflow<br>to allow the patient to fall<br>asleep before the device<br>delivers the full pressure; the<br>timer may be set to 0 – 45<br>minute delay periods, in 5-<br>minutes increments     | Feature enables a slow<br>increase in the patient airflow<br>to allow the patient to fall<br>asleep before the device<br>delivers the full pressure; the<br>timer may be set to 0 - 45<br>minute delay periods, in 5-<br>minutes increments     | Equivalent                                                        |
| Flow generator<br>operating system | Nucleus Plus                                                                                                                                                                                                                                    | Nucleus Plus                                                                                                                                                                                                                                    | Equivalent                                                        |
| Motor Type                         | Brush-less low voltage DC                                                                                                                                                                                                                       | Brush-less low voltage DC                                                                                                                                                                                                                       | Equivalent                                                        |
| Fan Type                           | Centrifugal                                                                                                                                                                                                                                     | Centrifugal                                                                                                                                                                                                                                     | Equivalent                                                        |
| Electronic<br>Controller           | Servo-control loop                                                                                                                                                                                                                              | Servo-control loop                                                                                                                                                                                                                              | Equivalent                                                        |
| Power supply                       | 100-240V, 50-60Hz                                                                                                                                                                                                                               | 100-240V, 50-60Hz                                                                                                                                                                                                                               | Equivalent                                                        |

# Conclusion

The VPAP Adapt SV is substantially equivalent to the predicate device, VPAP Adapt SV (K051364).

{8}------------------------------------------------

| 510(k) Summary -VPAP Tx                                                    |                                                                                                                                                                                                                          |
|----------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Required                                                                   | By Section 807.92 (c)                                                                                                                                                                                                    |
| Date Prepared                                                              | 18 May, 2016                                                                                                                                                                                                             |
| Owners Name                                                                | ResMed Ltd<br>1 Elizabeth Macarthur Drive<br>Bella Vista, NSW 2153, Australia                                                                                                                                            |
| Submitter                                                                  | Jasjit Baveja<br>+ 61 2 8884 1518 (Phone)<br>+ 61 2 8884 2000 (FAX)<br>Jasjit.baveja@resmed.com.au                                                                                                                       |
| Official Contact                                                           | Larissa D'Andrea<br>Director, Government & Regulatory Affairs<br>ResMed Corp<br>9001 Spectrum Center Boulevard<br>San Diego, CA 92123<br>+1 858 836 6837 (Phone)<br>+1 858 836 5519 (Fax)<br>Larissa.D'Andrea@resmed.com |
| Product codes                                                              | 73 MNS                                                                                                                                                                                                                   |
| Class<br>Classification Reference<br>Common/Usual Name<br>Proprietary Name | II<br>21 CFR 868.5895, Product Code 73 MNS<br>Ventilator, Continuous, Non-Life Supporting<br>VPAP Tx                                                                                                                     |
| Predicate device(s)                                                        | VPAP Tx (K092186)                                                                                                                                                                                                        |

{9}------------------------------------------------

#### Reason for submission

This 510(k) is being submitted as a labelling chanqe for the following contraindication:

ASV therapy is contraindicated in patients with chronic, symptomatic heart failure (NYHA 2-4) with reduced left ventricular ejection fraction (LVEF ≤ 45%) and moderate to severe predominant central sleep apnea.

This contraindication is supported by SERVE-HF, which was a randomized, parallel, event-driven, international multicenter study in 1325 patients with chronic, symptomatic heart failure (NYHA 2-4) with reduced left ventricular ejection fraction (LVEF≤45%) and moderate to severe predominant central sleep apnea. The study investigated the effects of adding ASV to quideline-based medical management on survival and cardiovascular outcomes. The addition of ASV did not improve outcomes and showed an increased risk of cardiovascular mortality despite effective control of central sleep apnea.

#### Indication for Use

The VPAP Tx is indicated for the treatment of patients weighing more than 66 lb (> 30 kq) with obstructive sleep apnea (OSA), respiratory insufficiency, central or mixed apneas, or periodic breathing. The VPAP Tx is intended to be used in a clinical environment.

## Substantial Equivalence

The new device has the following similarities to the previously cleared predicate devices.

- Same intended use
- Same operating principle
- Similar technologies
- . Same manufacturing process

ResMed has determined that the new device has not altered the safety and effectiveness of treatment for patients with obstructive Sleep Apnoea (OSA), respiratory insufficiency central and/or mixed apneas, or periodic breathing who weigh more than 66 lb (>30 kg). The new device complies with the applicable requirements referenced in the FDA quidance documents:

- FDA Draft Reviewer Guidance for Ventilators (July 1995)
- FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005)
- FDA Off-the-Shelf Software Use in Medical Devices (September 9,1999)

As this was a labelling change only, no further clinical testing was required to show substantial equivalence to the predicate device VPAP Tx (K092186).

The new device is as safe and effective as the predicate device.

## Device Description

The VPAP Tx is identical to the predicate device VPAP Tx (K092186), using a blower based positive pressure system. The device platform is identical to the VPAP Tx (K092186) and contains a Micro- processor controlled blower system that generates airway pressures as required to maintain an "air splint" for effective treatment of OSA and/or respiratory insufficiency.

The VPAP Tx system comprises the flow generator, patient interface) and optional humidifier.

{10}------------------------------------------------

Therapy modes contained in the VPAP Tx are:

- СРАР
- · Auto-titrating
- Bi-level
- VAuto
- . ASV

These modes come from the VPAP Tx (K092186).

| Characteristic                     | VPAP Tx (K092186)                                                                                                                                                                                                                                                                              | New Device (VPAP Tx)                                                                                                                                                                                                                                                                        | Comments                                                          |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
| Indication for<br>use              | The VPAP Tx is indicated for the<br>treatment of patients weighing<br>more than 66 lb (> 30 kg) with<br>obstructive sleep apnea (OSA),<br>respiratory insufficiency, central or<br>mixed apneas, or periodic<br>breathing.<br>The VPAP Tx is intended to be<br>used in a clinical environment. | The VPAP Tx is indicated for the<br>treatment of patients weighing more<br>than 66 lb (> 30 kg) with obstructive<br>sleep apnea (OSA), respiratory<br>insufficiency, central or mixed<br>apneas, or periodic breathing.<br>The VPAP Tx is intended to be used<br>in a clinical environment. | Equivalent<br>Only labeling change to<br>include contraindication |
| Location of use                    | Hospital/Home                                                                                                                                                                                                                                                                                  | Hospital/Home                                                                                                                                                                                                                                                                               | Equivalent                                                        |
| Pressure Range and Treatment Modes |                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                             |                                                                   |
|                                    | 4-20 cm H2O (CPAP)                                                                                                                                                                                                                                                                             | 4-20 cm H2O (CPAP)                                                                                                                                                                                                                                                                          | Equivalent:                                                       |
|                                    | 4-20 cm H2O (Autoset)                                                                                                                                                                                                                                                                          | 4-20 cm H2O (Autoset)                                                                                                                                                                                                                                                                       |                                                                   |
|                                    | 3-30 cm H2O (Bilevel)                                                                                                                                                                                                                                                                          | 3-30 cm H2O (Bilevel)                                                                                                                                                                                                                                                                       |                                                                   |
|                                    | 4-25 cm H2O (VAuto)<br>4-25 cm H2O (ASV)                                                                                                                                                                                                                                                       | 4-25 cm H2O (VAuto)<br>4-25 cm H2O (ASV)                                                                                                                                                                                                                                                    |                                                                   |
| Flow generator<br>operating system | ResMed proprietary 'FG Engine'<br>using Diamond architecture                                                                                                                                                                                                                                   | ResMed proprietary 'FG Engine'<br>using Diamond architecture                                                                                                                                                                                                                                | Equivalent                                                        |
| Motor Type                         | Brush-less low voltage DC                                                                                                                                                                                                                                                                      | Brush-less low voltage DC                                                                                                                                                                                                                                                                   | Equivalent                                                        |
| Fan Type                           | Centrifugal                                                                                                                                                                                                                                                                                    | Centrifugal                                                                                                                                                                                                                                                                                 | Equivalent                                                        |
| Electronic<br>Controller           | Servo-control loop                                                                                                                                                                                                                                                                             | Servo-control loop                                                                                                                                                                                                                                                                          | Equivalent                                                        |
| Power supply                       | 100-240V, 50-60Hz                                                                                                                                                                                                                                                                              | 100-240V, 50-60Hz                                                                                                                                                                                                                                                                           | Equivalent                                                        |

# Conclusion

The VPAP Tx is substantially equivalent to the predicate device, VPAP Tx (K092186).

{11}------------------------------------------------

# 510(k) Summary – S9 VPAP Tx

| Required                                                                   | By Section 807.92 (c)                                                                                                                                                                                                    |
|----------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared                                                              | 18 May, 2016                                                                                                                                                                                                             |
| Owners Name                                                                | ResMed Ltd<br>1 Elizabeth Macarthur Drive<br>Bella Vista, NSW 2153, Australia                                                                                                                                            |
| Submitter                                                                  | Jasjit Baveja<br>+ 61 2 8884 1518 (Phone)<br>+ 61 2 8884 2000 (FAX)<br>Jasjit.baveja@resmed.com.au                                                                                                                       |
| Official Contact                                                           | Larissa D'Andrea<br>Director, Government & Requlatory Affairs<br>ResMed Corp<br>9001 Spectrum Center Boulevard<br>San Diego, CA 92123<br>+1 858 836 6837 (Phone)<br>+1 858 836 5519 (Fax)<br>Larissa.D'Andrea@resmed.com |
| Product codes                                                              | 73 MNS                                                                                                                                                                                                                   |
| Class<br>Classification Reference<br>Common/Usual Name<br>Proprietary Name | <br>21 CFR 868.5895, Product Code 73 MNS<br>Ventilator, Continuous, Non-Life Supporting<br>S9 VPAP Tx                                                                                                                    |
| Predicate device(s)                                                        | S9 VPAP Tx (K123511)                                                                                                                                                                                                     |

{12}------------------------------------------------

### Reason for submission

This 510(k) is being submitted as a labelling change for the following contraindication:

ASV therapy is contraindicated in patients with chronic, symptomatic heart failure (NYHA 2-4) with reduced left ventricular ejection fraction (LVEF ≤ 45%) and moderate to severe predominant central sleep apnea.

This contraindication is supported by SERVE-HF, which was a randomized, parallel, event-driven, international multicenter study in 1325 patients with chronic, symptomatic heart failure (NYHA 2-4) with reduced left ventricular ejection fraction (LVEF≤45%) and moderate to severe predominant central sleep apnea. The study investigated the effects of adding ASV to quideline-based medical management on survival and cardiovascular outcomes. The addition of ASV did not improve outcomes and showed an increased risk of cardiovascular mortality despite effective control of central sleep apnea.

#### Indication for Use

The S9 VPAP Tx is indicated for the treatment and titration of patients with obstructive sleep apnea (OSA), respiratory insufficiency, central or mixed apneas, or periodic breathing. CPAP, S, ST, T and PAC modes are indicated for patients weighing more than 30lb (13 kg); all other modes are indicated for patients weighing more than 66lb (30 kg).

The S9 VPAP Tx is intended to be used in a clinical environment.

## Substantial Equivalence

The new device has the following similarities to the previously cleared predicate devices.

- Same intended use ●
- Same operating principle
- . Similar technologies
- Same manufacturing process

ResMed has determined that the new device has not altered the safety and effectiveness of treatment for patients with Obstructive Sleep Apnoea (OSA), central and/or mixed apneas, or periodic breathing who weigh more than 66 lb (>30 kq).

The new device complies with the applicable requirements referenced in the FDA guidance documents:

- FDA Draft Reviewer Guidance for Ventilators (July 1995)
- FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005)

As this was a labelling change only, no further clinical testing was required to show substantial equivalence to the predicate device S9 VPAP Tx (K123511).

The new device is as safe and effective as the predicate device.

## Device Description

The S9 VPAP Tx is identical to the predicate device S9 VPAP Tx (K123511), using a blower based positive pressure system. The device platform is identical to the S9 VPAP Tx (K092186) and contains a Micro- processor controlled blower system that generates airway pressures as required to maintain an "air splint" for effective treatment of OSA and/or respiratory insufficiency.

The S9 VPAP Tx comprises the flow generator, patient interface) and optional humidifier.

Therapy modes contained in the VPAP Tx are

- . СРАР
- . Auto-titrating
- o Bi-level
- o VAuto
- ASV

These modes come from the S9 VPAP Tx (K123511).

{13}------------------------------------------------

| Characteristic                     | S9 VPAP Tx (K123511)                                                                                                                                                                                                                                                                                                                                                                                                                                                      | New Device (S9 VPAP Tx)                                                                                                                                                                                                                                                                                                                                                                                                                                                | Comments                                                             |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| Indication for<br>use              | The S9 VPAP Tx is indicated<br>for the treatment and titration of<br>patients with obstructive sleep<br>apnea (OSA), respiratory<br>insufficiency, central or mixed<br>apneas, or periodic breathing.<br>CPAP, S, ST, T and PAC<br>modes are indicated for<br>patients weighing more than<br>30lb (13 kg); all other modes<br>are indicated for patients<br>weighing more than 66lb (30<br>kg).<br>The S9 VPAP Tx is intended to<br>be used in a clinical<br>environment. | The S9 VPAP Tx is indicated for<br>the treatment and titration of<br>patients with obstructive sleep<br>apnea (OSA), respiratory<br>insufficiency, central or mixed<br>apneas, or periodic breathing.<br>CPAP, S, ST, T and PAC modes<br>are indicated for patients<br>weighing more than 30lb (13 kg);<br>all other modes are indicated for<br>patients weighing more than<br>66lb (30 kg).<br>The S9 VPAP Tx is intended to<br>be used in a clinical<br>environment. | Equivalent<br>Only labeling change<br>to include<br>contraindication |
| Location of use                    | Hospital/Home                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Hospital/Home                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Equivalent                                                           |
| Pressure Range and Treatment Modes | 4-20 cm H2O (CPAP)<br>3-30 cm H2O (Bi-level)<br>3-25 cm H2O (VAuto)<br>3-30 cm H2O (ASV)<br>3-30 cm H2O (ASVAuto)                                                                                                                                                                                                                                                                                                                                                         | 4-20 cm H2O (CPAP)<br>3-30 cm H2O (Bi-level)<br>3-25 cm H2O (VAuto)<br>3-30 cm H2O (ASV)<br>3-30 cm H2O (ASVAuto)                                                                                                                                                                                                                                                                                                                                                      | Equivalent:                                                          |
| Flow generator<br>operating system | Micrium uC/OS-II Software/digital                                                                                                                                                                                                                                                                                                                                                                                                                                         | Micrium uC/OS-II Software/digital                                                                                                                                                                                                                                                                                                                                                                                                                                      | Equivalent                                                           |
| Motor Type                         | Brush-less low voltage DC                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Brush-less low voltage DC                                                                                                                                                                                                                                                                                                                                                                                                                                              | Equivalent                                                           |
| Fan Type                           | Multiple impeller axial motor                                                                                                                                                                                                                                                                                                                                                                                                                                             | Multiple impeller axial motor                                                                                                                                                                                                                                                                                                                                                                                                                                          | Equivalent                                                           |
| Electronic Controller              | Servo-control loop                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Servo-control loop                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Equivalent                                                           |
| Power supply                       | 100-240V, 50-60Hz                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 100-240V, 50-60Hz                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Equivalent                                                           |

# Conclusion

The S9 VPAP Tx is substantially equivalent to the predicate device, S9 VPAP Tx (K123511).

---

**Source:** [https://fda.innolitics.com/device/K161487](https://fda.innolitics.com/device/K161487)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
