K161470 · KLS Martin L.P. · MQN · Dec 1, 2016 · Dental
Device Facts
Record ID
K161470
Device Name
Internal Distraction - Sterile
Applicant
KLS Martin L.P.
Product Code
MQN · Dental
Decision Date
Dec 1, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4760
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Internal Distraction - Sterile includes devices intended as bone stabilizers and lengthening (and or transport) devices for correction of congenital deficiencies or post traumatic defects of the mandible (including ramus, body, alveolar ridge, palate, symphysis) and mid-face bones that require gradual distraction.
Device Story
Internal distraction devices for bone stabilization and lengthening via distraction osteogenesis; used for mandible and mid-face bone correction. Device consists of titanium footplates and threaded drive screws; fixed to bone using bone screws. Activation occurs via patient-operated driver, either directly or through an activation arm extending through soft tissue. Mechanical rotation of drive screws separates footplates, inducing bone growth and soft tissue expansion. Used in clinical settings by surgeons; patient performs daily activation. Device provides temporary stabilization; explanted after bone consolidation. Provided sterile via gamma irradiation.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including axial load, bending torsion, axial-torsion, and transverse shear testing, confirming mechanical properties meet design requirements. Biocompatibility and endotoxin testing (ANSI/AAMI ST72:2011, USP <85>) confirm safety.
Technological Characteristics
Materials: Titanium Alloy (Ti-6Al-4V) and CP Titanium. Principle: Mechanical distraction via threaded drive screws. Form factor: Footplates with fixed/detachable activation arms. Sterilization: Gamma irradiation. Standards: Biocompatibility per ISO 10993; Endotoxin per ANSI/AAMI ST72:2011, USP <85>, EP 2.6.14.
Indications for Use
Indicated for patients requiring gradual distraction for correction of congenital deficiencies or post-traumatic defects of the mandible (ramus, body, alveolar ridge, palate, symphysis) and mid-face bones.
Regulatory Classification
Identification
A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 1, 2016
KLS Martin LP Gary Moore Quality Mgt & Regulatory Affairs Manager 11201 Saint Johns Industrial Pkwy S Jacksonville, Florida 32246
Re: K161470
Trade/Device Name: Internal Distraction - Sterile Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: Class II Product Code: MQN Dated: May 25, 2016 Received: May 27, 2016
Dear Gary Moore:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161470/S002
Device Name Internal Distraction - Sterile
#### Indications for Use (Describe)
Internal Distraction - Sterile includes devices intended as bone stabilizers and lengthening (and or transport) devices for correction of congenital deficiencies or post traumatic defects of the mandible (including ramus, body, alveolar ridge, palate, symphysis) and mid-face bones that require gradual distraction.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for KLS martin L.P. The letters "KLS" are in red, while the word "martin" is in gray. The letters "L.P." are in black.
P.O. Box 16369 · Jacksonville, FL 32245-6369 904-641-7746 or 800-625-1557 ● Fax 904-641-7378 www.klsmartinnorthamerica.com
#### 5. 510(K) SUMMARY
| Submitter: | KLS Martin LP |
|----------------------|---------------------------------------------------------|
| | 11201 Saint Johns Industrial Pkwy S |
| | Jacksonville, FL 32246 |
| Contact Person: | Jennifer Damato |
| | Director of Quality MGT and Regulatory Affairs |
| | Phone: 1-800-625-1557 |
| | Fax: 904-641-7378 |
| Date Prepared: | November 18, 2016 |
| Trade Name: | Internal Distraction - Sterile |
| Common Name: | Bone plate |
| Classification Name: | External Mandibular Fixator And/or Distractor |
| | Class II, 21 CFR 872.4760, Product Code MQN |
| Predicate Devices: | Zurich Distraction System (K010139) (Primary Predicate) |
| | KLS Martin Intraoral Distractor (K973275) |
| | KLS Intra-oral Vertical Distractor (K983515) |
| | KLS Martin Intraoral Zurich Ramus Distractor (K983809) |
| | K-L-W Intraoral Distractor (K991875) |
| | Zurich Pediatric Maxillary Distractor (K000580) |
| | Track 1.0 & 1.5 mm Systems - Tissue Regeneration by |
| | Alveolar Callus Distraction (K002152) |
| | Molina Orbital Malar Distractor (K003883) |
## Device Description
Internal Distraction - Sterile consists of sterile internal distraction devices intended for the correction of the mandible and mid-face bones that are comprised of several different designs and components intended for bone stabilization and elongation through distraction osteogenesis. It is composed of multiple sizes and shapes of distractor footplates and either fixed or detachable activator arms. The devices are positioned internally with a connected activation arm extending through the soft tissue for external activation. Some devices, due to their anatomical positioning, are directly activated using a patient activation driver, eliminating the need for attaching an activation arm to the device. The distractor footplates are fixated to the bone on either side of the osteotomy using previously cleared bone screws (K943347, K944561, K944565, K971297, K060177). Distraction is achieved by rotating the distractor threaded drive screws with the patient driver, often with an activation arm, which causes a separation of the distractor footplates and induces the body to grow bone and expand soft tissue as a response. Various lengths of distractor drive screws are available to achieve the desired distraction length. Upon completion of
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for KLS Martin L.P. The letters "KLS" are in red, while "martin" is in gray. The letters "L.P." are in black. The logo is simple and modern.
distraction and consolidation of the bone, the screws are removed from the footplates and the distractor is explanted.
The purpose of this bundled submission is to offer the previously cleared KLS Martin LP non-sterile predicate devices intended for mandible and mid-face bones sterile via gamma irradiation and encompasses the following devices:
Zurich Distraction System - K010139: The Zurich Distraction System includes devices intended as a bone stabilizer and lengthening (and or transport) device when correction of congenital deficiencies or post traumatic defects of the mandible (including ramus, body, alveolar ridge, palate, symphysis), mid-face, and cranial bones require gradual distraction.
KLS Intraoral Distractor - K973275: The KLS Martin Intraoral Distractor is intended to be used in the mandible for conditions such as mandibular deficiency or post-traumatic effects of the mandible, where gradual bone distraction is required. (Micrognathia, Hemifacial Microsomia, Congenital Craniofacial Syndromes).
KLS Intra-oral Vertical Distractor - K983515: The KLS Martin Intra-oral Distractor is intended to be used in the conditions of mandibular deficiency or post-traumatic effects of the mandible.
KLS Martin Intraoral Zurich Ramus Distractor - K983809: A uni-directional intraoral device for the distraction of the ascending ramus and mandibular body.
K-L-W Intraoral Distractor - K991875: The indications for use are a uni-directional, intraoral device for the distraction of facial bones.
Zurich Pediatric Maxillary Distractor - K000580: The Zurich Pediatric Maxillary Distractor is intended for use in the maxilla as a bone stabilizer and lengthening device when correction of congenital mid facial deficiencies or posttraumatic defects require gradual bone distraction.
TRACK Distractors - K002152: The KLS Martin TRACK distractors are intended for use in patients who have totally or partially edentulous mandible or small craniofacial bone deficiencies and need an increase in bone height and mass by means of distraction osteogenesis. These patients include those who have traumatic defects, periodontal disease or birth abnormality. The TRACK distractors are designed to provide temporary stabilization and gradual expansion across an osteotomy in the mandibular bone or maxillary bone and thereby increasing the height of the adjacent Alveolar ridge. This device is intended to be removed after consolidation of the callus and prior to final prosthetic reconstruction. The mechanism to achieve distraction is a threaded titanium rod with attached titanium plates, which are fixed to the mandible, or maxilla with titanium bone screws.
Molina Orbital Malar Distractor - K003883: The KLS Martin Molina Orbital Malar Distractor is intended for use in the treatment of cranial and midface conditions such as syndromic craniosynostosis and congenital midface deficiencies in which osteotomies and
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for KLS Martin L.P. The letters "KLS" are in red, while "martin" is in gray. The letters "L.P." are in black and are larger than the other letters in the logo.
gradual bone distraction are indicated. This device is intended to provide temporary stabilization and gradual lengthening of the cranial and midface bones.
## Indications for Use
Internal Distraction - Sterile includes devices intended as bone stabilizers and lengthening (and or transport) devices for correction of congenital deficiencies or post traumatic defects of the mandible (including ramus, body, alveolar ridge, palate, symphysis) and mid-face bones that require gradual distraction.
## Technological Characteristics/Substantial Equivalence Discussion
#### Similarities to Predicate Devices
Internal Distraction - Sterile is very similar to the predicate devices with respect to intended use, but identical in materials, manufacturing process, principles of operation, design mechanisms, and performance. Placement and fixation of the Internal Distraction - Sterile devices is identical to the methods used for the predicate devices.
#### Differences to Predicate Devices
Internal Distraction - Sterile differs from the predicate devices in that it will be provided in sterile packaging and will have different stock numbers from the originally cleared stock numbers to identify the product as sterile. In addition, the intended use differs slightly in that Internal Distraction - Sterile will be intended for correction of bones in the mandible and mid-face only, and will not include cranial bones.
## Non-Clinical Performance Data
#### Biocompatibility Testing
Biocompatibility requirements were assessed in accordance with FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO 10993, "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process." as recognized by FDA. The patient-contacting materials Titanium Alloy (Ti-6Al-4V) and CP Titanium used for the Internal Distraction - Sterile devices and their components were tested for biocompatibility and are previously cleared in all predicate devices. The Internal Distraction - Sterile devices and their components are comprised of the same materials, have the same chemical composition, undergo the same manufacturing processes, and have the same body contact duration as these devices; therefore, biocompatibility testing for Internal Distraction - Sterile is not needed.
## Endotoxin Testing
Bacterial Endotoxins testing was conducted in accordance to ANSI/AAMI ST72:2011, USP <85>, and EP 2.6.14 for detection and quantitation of bacterial endotoxins. The results of the testing demonstrate that the subject devices conform to the required endotoxin units per device for medical devices and meet pyrogen limit specifications.
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for KLS martin L.P. The letters "KLS" are in red, while "martin" is in gray. The letters "L.P." are in black and are smaller than the other letters in the logo.
# Performance Testing - Bench
Axial load testing, bending torsion testing, axial-torsion testing, and transverse shear testing were performed to evaluate the mechanical properties of the subject devices. All devices passed design requirements for material properties, manufacturing tolerances, axial resistance, and torsional loading. The results of the testing demonstrate the devices are sufficiently capable of withstanding the anatomical loads placed upon them and allow for effective and safe boney movement during their intended use.
# Clinical Performance Data
Clinical testing was not necessary for the determination of substantial equivalence.
## Conclusion
Internal Distraction - Sterile is very similar in its intended use, is identical in design, function, manufacturing process, and is composed of the same materials as the predicate devices. The similarities and differences in technological characteristics do not raise new questions of safety or effectiveness. Internal Distraction - Sterile will encompass all the previously cleared predicate devices intended for correction of mandible and mid-face bones and provide them sterile. Therefore, the information presented supports substantial equivalence of Internal Distraction - Sterile to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.