← Product Code [NLH](/productcode/NLH) · K161464

# Reprocessed Dynamic Tip Steerable Diagnostic Electrophysiology Catheters (K161464)

_Innovative Health, LLC · NLH · Sep 30, 2016 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K161464

## Device Facts

- **Applicant:** Innovative Health, LLC
- **Product Code:** [NLH](/productcode/NLH.md)
- **Decision Date:** Sep 30, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1220
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Reprocessed Dynamic Tip Steerable Diagnostic Electrophysiology Catheters are intended for temporary intracardiac sensing, recording, stimulation and temporary pacing during the evaluation of cardiac arrhythmias.

## Device Story

Radiopaque, flexible, insulated polymer catheters with distal tip plunger mechanism for steerable curvature; used for temporary intracardiac sensing, recording, stimulation, and pacing during cardiac arrhythmia evaluation. Operated by physicians in clinical settings. Device is a reprocessed single-use diagnostic electrophysiology catheter; Innovative Health performs cleaning, decontamination, inspection, and functional testing. Output consists of electrical signals from heart tissue captured via electrodes for diagnostic analysis. Benefits include providing diagnostic data for arrhythmia management while utilizing reprocessed hardware to reduce waste.

## Clinical Evidence

Bench testing only. No clinical data presented. Testing included biocompatibility, cleaning validation, sterilization validation, and functional testing (visual inspection, dimensional verification, electrical continuity/resistance, simulated use, mechanical characteristics, and electrical safety/dielectric/leakage testing).

## Technological Characteristics

Radiopaque, flexible, insulated polymer shaft with plunger-actuated steerable distal tip. Features 4 to 10 electrodes with varying spacing (2mm to 10mm). 6 French size, 110cm length. Reprocessed single-use device; limited to one reprocessing cycle. Sterilization and cleaning validation performed. No software or electronic processing components.

## Regulatory Identification

An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.

## Predicate Devices

- Bard Tip Deflecting Electrode Catheter ([K891908](/device/K891908.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized caduceus symbol, which is often associated with healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 30, 2016

Innovative Health, LLC Rafal Chudzik VP, R&D and Operations 1435 North Hayden Road, Suite 100 Scottsdale, Arizona 85257

Re: K161464

Trade/Device Name: Reprocessed Dynamic Tip Steerable Diagnostic Electrophysiology Catheters (Models per Table 1 in Enclosure) Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter Or Electrode Recording Probe Regulatory Class: Class II Product Code: NLH Dated: August 26, 2016 Received: August 29, 2016

Dear Rafal Chudzik:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Mula Jellm
for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| Item Number | Description                                     | Electrodes | Spacing<br>(mm) | Curve     | French<br>Size | Usable<br>Length<br>(cm) |
|-------------|-------------------------------------------------|------------|-----------------|-----------|----------------|--------------------------|
| 200131      | Dynamic Tip Steerable<br>Diagnostic EP Catheter | 4          | 10              | Large 4.0 | 6              | 110                      |
| 200344      | Dynamic Tip Steerable<br>Diagnostic EP Catheter | 4          | 5               | Large 4.0 | 6              | 110                      |
| 6DYNTP002   | Dynamic Tip Steerable<br>Diagnostic EP Catheter | 4          | 2-5-2           | Large 4.0 | 6              | 110                      |
| 6DYNTP006   | Dynamic Tip Steerable<br>Diagnostic EP Catheter | 8          | 2               | Large 4.0 | 6              | 110                      |
| 6DYNTP001   | Dynamic Tip Steerable<br>Diagnostic EP Catheter | 10         | 2-5-2           | Large 4.0 | 6              | 110                      |

# Table 1: Reprocessed Single-Use Devices Included in K161464 Clearance

{3}------------------------------------------------

## Indications for Use

510(k) Number (if known) K161464

#### Device Name

Reprocessed Dynamic Tip Steerable Diagnostic Electrophysiology Catheter

Indications for Use (Describe)

The Reprocessed Dynamic Tip Steerable Diagnostic Electrophysiology Catheters are intended for temporary intracardiac sensing, recording, stimulation and temporary pacing during the evaluation of cardiac arrhythmias.

| Type of Use (Select one or both, as applicable)                                       |                                                                                      |
|---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| <div> <span> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span> </span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

## SECTION 5: 510(k) SUMMARY

As required by 21 CFR 807.92(c)

### Submitter's Name and Address:

Innovative Health, LLC. 1435 N. Hayden Road, Suite 100 Scottsdale, AZ 85257

### Contact Name and Information:

Rafal Chudzik VP, R&D and Operations (480) 525-6006 (office) (844) 965-9359 (fax) rchudzik@innovative-health.com

### Date prepared:

May 25, 2016

### Device Information:

| Trade/Proprietary Name: | Reprocessed Dynamic Tip Steerable Diagnostic |
|-------------------------|----------------------------------------------|
|                         | Electrophysiology Catheters                  |
| Common Name:            | Diagnostic Electrophysiology Catheter        |
| Classification Name:    | Catheter, Recording, Electrode, Reprocessed  |
| Classification Number:  | Class II, 21 CFR 870.1220                    |
| Product Code:           | NLH                                          |

#### Predicate Device:

| 510(k) Number | 510(k) Title                           | Manufacturer    |
|---------------|----------------------------------------|-----------------|
| K891908       | Bard Tip Deflecting Electrode Catheter | C.R. Bard, Inc. |

### Device Description:

The Reprocessed Dynamic Tip Steerable Diagnostic Electrophysiology (EP) Catheters are radiopaque, flexible, insulated catheters with a polymer shaft. The catheters have a plunger mechanism, which, when moved forward or back, results in curvature of the distal tip.

The item numbers included in the scope of this submission are as follows:

| Description                                           | Item Number | Number of<br>Electrodes | French<br>Size | Electrode<br>Spacing<br>(mm) | Curve     | Length (cm) |
|-------------------------------------------------------|-------------|-------------------------|----------------|------------------------------|-----------|-------------|
| Dynamic Tip<br>Steerable<br>Diagnostic EP<br>Catheter | 200131      | 4                       | 6              | 10                           | Large 4.0 | 110         |
|                                                       | 200344      | 4                       | 6              | 5                            | Large 4.0 | 110         |
|                                                       | 6DYNTP002   | 4                       | 6              | 2,5,2                        | Large 4.0 | 110         |
|                                                       | 6DYNTP006   | 8                       | 6              | 2                            | Large 4.0 | 110         |
|                                                       | 6DYNTP001   | 10                      | 6              | 2,5,2                        | Large 4.0 | 110         |

Table 5.1: Device Scope

{5}------------------------------------------------

## Indications for Use:

The Reprocessed Dynamic Tip Steerable Diagnostic EP Catheters are intended for temporary intracardiac sensing, recording, stimulation and temporary pacing during the evaluation of cardiac arrhythmias.

## Technoloqical Characteristics:

The purpose, design, materials, function, and intended use of the Reprocessed Dynamic Tip Steerable Diagnostic Electrophysiology (EP) Catheters are identical to the predicate devices. There are no changes to the claims, clinical applications, patient population, performance specifications, or method of operation. In addition, Innovative Health's reprocessing of these devices includes removal of visible soil and decontamination. Each device is inspected and function tested prior to packaging and labeling.

## Functional and Safety Testing:

Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of the Reprocessed Dynamic Tip Steerable Diagnostic EP Catheter. This included the following:

- Biocompatibility ●
- Cleaning Validation ●
- Sterilization Validation ●
- Functional Testing ●
	- Visual Inspection ●
	- . Dimensional Verification
	- Electrical Continuity and Resistance .
	- . Simulated Use
	- Mechanical Characteristics ●
- . Electrical Safety Testing
	- . Dielectric and Current Leakage
- Packaging Validation ●

The Reprocessed Dynamic Tip Steerable Diagnostic EP Catheters are reprocessed no more than one (1) time. Each device is marked and tracked. After the device has reached the maximum number of reprocessing cycles, the device is rejected from further reprocessing. Reprocessing is performed only by Innovative Health. Innovative Health restricts its reprocessing to exclude devices previously reprocessed by other reprocessors.

## Conclusion:

Innovative Health concludes that the Reprocessed Dynamic Tip Steerable Diagnostic EP Catheters are as safe and effective as the predicate devices described herein.

---

**Source:** [https://fda.innolitics.com/device/K161464](https://fda.innolitics.com/device/K161464)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
