← Product Code [DQY](/productcode/DQY) · K161362

# ASAHI Corsair Armet (K161362)

_Asahi Intecc Co., Ltd. · DQY · Jun 14, 2016 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K161362

## Device Facts

- **Applicant:** Asahi Intecc Co., Ltd.
- **Product Code:** [DQY](/productcode/DQY.md)
- **Decision Date:** Jun 14, 2016
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1250
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The ASAHI Corsair Armet is intended to provide support to facilitate the placement of guide wires in the peripheral vasculature, and can be used to exchange one guide wire for another. The ASAHI Corsair Armet is also intended to assist in the delivery of contrast media into the peripheral vasculature. This device should not be used in coronary vasculature or neurovasculature.

## Device Story

ASAHI Corsair Armet is a percutaneous catheter used in peripheral vasculature. Device consists of distal tip, shaft tube, protector, and connector. Hydrophilic coating on distal tip and shaft; PTFE-lined inner lumen. Operated by physicians in clinical settings to facilitate guide wire placement/exchange and deliver contrast media. Provides mechanical support for navigation through peripheral vessels. Benefits patient by enabling access to peripheral vasculature for diagnostic or interventional procedures.

## Clinical Evidence

Bench testing only. Testing included appearance/dimensions, corrosion resistance, force at break, liquid/air leakage, high static pressure performance, radio-detectability, torque transmission/durability, slide durability, and kink resistance. Results demonstrate the device functions as intended and is equivalent to the predicate.

## Technological Characteristics

Percutaneous catheter; polyamide-elastomer shaft; hydrophilic coating; PTFE-lined lumen; tapered distal tip. Dimensions: 600-1500 cm length; 0.75mm distal/0.83mm proximal OD. Sterilization: Ethylene Oxide (SAL 10^-6). Single-use.

## Regulatory Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

## Predicate Devices

- ASAHI Corsair Armet ([K152249](/device/K152249.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of a human figure, with three profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the most prominent.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June14, 2016

Asahi Intecc Co., Ltd. % Candace Cederman Senior Regulatory Affairs Consultant Cardiomed Device Consultants, LLC 5523 Research Park Drive, Suite 205 Baltimore, MD 21228

Re: K161362

Trade/Device Name: Asahi Corsair Armet Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: May 13, 2016 Received: May 16, 2016

Dear Ms. Cederman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kenneth J. Cavanaugh -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K161362

Device Name ASAHI Corsair Armet

Indications for Use (Describe)

The ASAHI Corsair Armet is intended to provide support to facilitate the placement of guide wires in the peripheral vasculature, and can be used to exchange one guide wire for another. The ASAHI Corsair Armet is also intended to assist in the delivery of contrast media into the peripheral vasculature. This device should not be used in coronary vasculature or neurovasculature.

| Type of Use (Select one or both, as applicable)                                                 |                                                                                     |
|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

(as required by 21 CFR 807.92)

HI INTECC CO.LI 1703 Wakita-cho, Moriyama-ku, Nagoya, Aichi 463-0024 Japan Tel. +81-52-768-1211 Fax. +81-52-768-1221 Branch offices: Tokyo, Nagoya, Osaka, Hong Kong, Amsterdam, Singapore, Beijing Research Facilities and Factories: Osaka, Seto, Thailand, Hanoi

# ASAHI Corsair Armet

510(k) _______________________________________________________________________________________________________________________________________________________________________

| DATE PREPARED:         | May 13, 2016                                                                                                                                                                                                |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| APPLICANT              | Asahi Intecc Co., Ltd.<br>1703 Wakita-cho, Moriyama-ku<br>Nagoya, Aichi 463-0024, Japan                                                                                                                     |
| CONTACT                | Carroll Councilman, Sr. RA Manager<br>Asahi Intecc USA, Inc.<br>2500 Red Hill Avenue, Suite 210<br>Santa Ana, CA 92705<br>Tel: (949) 756-8252, FAX: (949) 756-8165<br>e-mail: ASAHI.ra-fda@ASAHI-intecc.com |
| TRADE NAME:            | ASAHI Corsair Armet                                                                                                                                                                                         |
| DEVICE CLASSIFICATION: | Class 2 per 21 CFR §870.1250                                                                                                                                                                                |
| CLASSIFICATION NAME:   | Percutaneous Catheter                                                                                                                                                                                       |
| PRODUCT CODE           | DQY – Catheter, Percutaneous                                                                                                                                                                                |
| PREDICATE DEVICES:     | ASAHI Corsair Armet (K152249)                                                                                                                                                                               |

#### Intended Use/Indications for Use

The ASAHI Corsair Armet is intended to provide support to facilitate the placement of quide wires in the peripheral vasculature, and can be used to exchange one quide wire for another.

The ASAHI Corsair Armet is also intended to assist in the delivery of contrast media into the peripheral vasculature.

This device should not be used in coronary vasculature or neurovasculature.

## Device Description:

The ASAHI Corsair Armet consists of a distal tip, a shaft tube that is inserted into vasculature, a protector, and a connector for catheter control and infusion of contrast media. The device has a hydrophilic coating on the

{4}------------------------------------------------

outer surface of distal tip and the shaft tube to provide a smooth transition in blood vessels. The distal tip of the Corsair Armet has a tapered shape. PTFE is applied to the inner lumen of the catheter for the purposes of a smooth transition and exchange of guidewires.

The primary change in this Special 510(k) involves a minor change in the shape of the distal tip.

#### Comparison with Predicate Devices:

Comparisons of the modified ASAHI Corsair Armet to its predicate device show that the technological characteristics of the Subject device such as the product performance, intended use/indications, components, materials, sterilization method, shelf life, manufacturing process, and operating principle are identical to the currently marketed predicate devices. There are only minor dimensional variations between the Subject and predicate device.

| Name of Device               | ASAHI Corsair Armet                                                                                                                                                                                                                                                                                                                                                                              | ASAHI Corsair Armet |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|
| 510(k)                       | Current Application                                                                                                                                                                                                                                                                                                                                                                              | K152249             |
| Intended Use and Indications | The ASAHI Corsair Armet is intended to provide support to facilitate the placement of guide wires in the peripheral vasculature, and can be used to exchange one guide wire for another.<br>The ASAHI Corsair Armet is also intended to assist in the delivery of contrast media into the peripheral vasculature.<br>This device should not be used in coronary vasculature or neurovasculature. |                     |
| Target Body Location         | Peripheral                                                                                                                                                                                                                                                                                                                                                                                       |                     |
| Hydrophilic coating          | Yes                                                                                                                                                                                                                                                                                                                                                                                              |                     |
| Effective Length             | 600-1500 cm                                                                                                                                                                                                                                                                                                                                                                                      |                     |
| Nominal Outer Diameter       | Distal : 0.75mm<br>Proximal : 0.83 mm                                                                                                                                                                                                                                                                                                                                                            |                     |
| Catheter Shaft Material      | Polyamide-elastomer                                                                                                                                                                                                                                                                                                                                                                              |                     |
| Distal Tip Length            | 1.2mm                                                                                                                                                                                                                                                                                                                                                                                            | 2.0mm               |
| Single Use                   | Yes                                                                                                                                                                                                                                                                                                                                                                                              |                     |
| Sterilization                | Provided sterile via Ethylene Oxide to SAL10-6                                                                                                                                                                                                                                                                                                                                                   |                     |
| Shelf Life                   | 3 Years                                                                                                                                                                                                                                                                                                                                                                                          |                     |

{5}------------------------------------------------

## Non Clinical testing / Performance Data:

Confirmatory non clinical laboratory testing was performed on the modified ASAHI Corsair Armet to determine substantial equivalence.

The following testing/assessments were performed:

- Appearance/Dimensions
- Corrosion Resistance ●
- Force at Break
- Liquid Leakage under Pressure ●
- Air Leakage ●
- Leak and Damage under High Static Pressure ●
- Radio-Detectability ●
- Torque Transmission ●
- Slide Durability ●
- Kink Resistance ●
- . Torque Durability

The in vitro bench tests demonstrated that the modified ASAHI Corsair Armet met all acceptance criteria and performed similarly to the predicate devices. Performance data demonstrate that the device functions as intended, and is substantially equivalent to the predicate devices.

## BIOCOMPATIBILITY:

The modified ASAHI Corsair Armet was compared to the predicate device. Based on the identical materials and manufacturing process used in the subject device to its predicate, the biocompatibility of the modified ASAHI Corsair Armet was verified to be the same as those of the predicate.

## CONCLUSION:

The modified ASAHI Corsair Armet has identical intended use and the same or similar technological characteristics such as components, design, materials, sterilization method, shelf life and operating principles as the predicate devices. Performance data demonstrates that the device functions as intended.

Therefore, the modified ASAHI Corsair Armet is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K161362](https://fda.innolitics.com/device/K161362)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
