K161357 · Tearscience, Inc. · ORZ · Nov 4, 2016 · Ophthalmic
Device Facts
Record ID
K161357
Device Name
LipiFlow Thermal Pulsation System
Applicant
Tearscience, Inc.
Product Code
ORZ · Ophthalmic
Decision Date
Nov 4, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 886.5200
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LipiFlow® Thermal Pulsation System is intended for the application of localized heat and pressure therapy in adult patients with chronic cystic conditions of the eyelids, including meibomian gland dysfunction (MGD), also known as evaporative dry eye or lipid deficiency dry eye.
Device Story
LipiFlow Thermal Pulsation System provides in-office treatment for chronic cystic eyelid conditions. System consists of physician-operated Console and patient-contacting disposable Activator. Console provides electrical power, user interface, treatment monitoring, and safeguard circuitry. Activator applies controlled heat to inner eyelid surface and intermittent pressure to outer eyelid to facilitate lipid release from meibomian glands. Modification introduces Activator II (LFD-2000) with semi-permanent cable (CBL-2000) to reduce waste, plus updated Console hardware/OS. Physician monitors treatment via touchscreen. Device benefits patients by treating underlying cause of evaporative dry eye.
Clinical Evidence
Bench testing only. Performance testing validated electrical safety, EMC, temperature/pressure safeguard functions, sterility, shelf-life, and biocompatibility. Software verification and validation conducted per FDA guidance. No new clinical data provided; safety/effectiveness supported by equivalence to previously validated predicate performance.
Technological Characteristics
System comprises Console (LFTP-1000) and disposable Activator (LFD-2000). Uses localized heat and pressure. Console runs Windows 7 Embedded OS. Connectivity includes electrical/pneumatic interface between Console and Activator via semi-permanent cable. Biocompatible patient contact materials. Complies with electrical safety and EMC standards. Software is moderate level of concern.
Indications for Use
Indicated for adult patients with chronic cystic eyelid conditions, including meibomian gland dysfunction (MGD), evaporative dry eye, or lipid deficiency dry eye.
Regulatory Classification
Identification
An eyelid thermal pulsation system is an electrically-powered device intended for use in the application of localized heat and pressure therapy to the eyelids. The device is used in adult patients with chronic cystic conditions of the eyelids, including meibomian gland dysfunction (MGD), also known as evaporative dry eye or lipid deficiency dry eye. The system consists of a component that is inserted around the eyelids and a component to control the application of heat and pressure to the eyelids.
Special Controls
In addition to the general controls of the Act, the LipiFlow® Thermal Pulsation System is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Appropriate analysis/testing should validate electromagnetic compatibility (EMC) and safety of exposure to non-ionizing radiation;
(2) Design, description, and performance data should validate safeguards related to the temperature and pressure aspects of the device, including during fault conditions;
(3) Performance data should demonstrate the sterility of patient-contacting components and the shelf-life of these components;
(4) The device should be demonstrated to be biocompatible; and
(5) Performance data should demonstrate that any technological changes do not adversely effect safety and effectiveness.
{1}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing to the right, stacked on top of each other, representing the department's focus on people and health.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center -WO66-G609 Silver Spring, MD 20993-0002
November 4, 2016
TearScience, Inc. Ms. Christy Coleman Vice President, Clinical, Quality and Regulatory Affairs 5151 McCrimmon Parkway Suite 250 Morrisville, NC 27560
Re: K161357
Trade/Device Name: Lipiflow® Thermal Pulsation System Regulation Number: 21 CFR 886.5200 Regulation Name: Eyelid Thermal Pulsation System Regulatory Class: Class II Product Code: ORZ Dated: October 4, 2016 Received: October 5, 2016
Dear Ms. Coleman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{2}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kesia Alexander
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name LipiFlow® Thermal Pulsation System (Component Models LFTP-1000, LFD-1000, LFD-1100, LFD-2000 and CBL-2000)
#### Indications for Use (Describe)
The LipiFlow® Thermal Pulsation System is intended for the application of localized heat and pressure therapy in adult patients with chronic cystic conditions of the eyelids, including meibomian gland dysfunction (MGD), also known as evaporative dry eye or lipid deficiency dry eye.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY
| PREPARATION DATE: | September 26, 2016 |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| APPLICANT: | TearScience, Inc.<br>5151 McCrimmon Parkway, Suite 250<br>Morrisville, NC 27560<br>Tel: (919) 467-4007<br>Fax: (919) 467-3300 |
| CONTACT PERSON: | Christy Coleman, OD, MPH<br>Vice President, Clinical, Quality & Regulatory Affairs |
| DEVICE TRADE NAME: | LipiFlow® Thermal Pulsation System<br>• Console Model LFTP-1000<br>• Activator (Disposable) Models LFD-1000, LFD-1100 and LFD-2000 with Cable Model CBL-2000 |
| CLASSIFICATION NAME: | Eyelid Thermal Pulsation System |
| DEVICE CLASSIFICATION: | Class II; 21 CFR 886.5200 |
| PRODUCT CODE: | ORZ |
| PREDICATE DEVICE: | LipiFlow® Thermal Pulsation System<br>Class II under 21 CFR 886.5200<br>Applicant: TearScience, Inc.<br>Cleared under DEN100017 on June 28, 2011<br>• Handheld Control System Model LFH-1000 and<br>• Disposable Model LFD-1000<br>Cleared under K112704 on December 19, 2011 (corrected letter received January 12, 2012)<br>• Console Model LFTP-1000 and<br>• Activator (Disposable)Model LFD-1000<br>Cleared under K133127 on December 23, 2013<br>• Console Model LFTP-1000 and<br>• Activator (Disposable)Models LFD-1000 and LFD-1100 |
#### DEVICE DESCRIPTION:
The LipiFlow® Thermal Pulsation System is used by a physician in an in-office procedure for patients with chronic cystic conditions of the eyelids to provide controlled heat to the inner eyelid surface, close to the location of the meibomian glands, and intermittent pressure to the outer eyelid to facilitate release of lipid from the cystic meibomian glands. The LipiFlow® System is comprised of a physician interface (Control component) and a patient interface (Disposable component). The Control component (labeled as Console) provides the electrical power, user interface, treatment monitoring, treatment control and safeguard circuitry used for controlling the heat and pressure applied to the patient's eyelids by the Disposable component (labeled as Activator).
{5}------------------------------------------------
This 510(k) submission is for the device modification of new model of the Disposable component (labeled Activator II, Model LFD-2000), a new semi-permanent Cable (Model CBL-2000), and a hardware and software update to the Control component (labeled as Console, Model LFTP-1000). The modified LipiFlow® System has the same intended use, indications for use and fundamental scientific technology as the legally marketed predicate device.
The new Activator II model (LFD-2000) is the same as the predicate Activator models (LFD-1000 and LFD-1100) except that the electrical and pneumatic connection point is moved closer to the patient interface. This allows the disposable portion of the device to be smaller and less wasteful. The predicate Activator models include a combined patient interface and permanently attached cable, which is connected to the Console and disposed of after each treatment. Conversely, the patient interface of the new Activator II model is used with a new semipermanent Cable, which connects the Activator II to the Console but is not disposed of after treatment. Compared to the predicate Activator models, the patient interface portion of the new Activator II model is identical in shape, size, patient contact materials, electrical safety, packaging, sterility, shelf-life, method of preventing re-use, and temperature and pressure performance and safety.
The LipiFlow® Console hardware was updated with a new equivalent computer to address computer obsolescence in the predicate device. To support the new computer, the Console was updated to a new Operating System and the software application was modified to support the hardware-specific driver differences in the Operating Systems.
## INTENDED USE:
The LipiFlow® Thermal Pulsation System is intended for the application of localized heat and pressure therapy in adult patients with chronic cystic conditions of the eyelids, including meibomian gland dysfunction (MGD), also known as evaporative dry eye or lipid deficiency dry eye. The Intended Use and Indications for Use have not changed.
### TECHNOLOGICAL CHARACTERISTICS:
The modified LipiFlow® Thermal Pulsation System has the same fundamental scientific technology as the predicate device, as described in DEN100017, K112704 and K133127. Most of the technological characteristics of the LipiFlow® System remain unchanged from the predicate device. The similarities and differences in technological characteristics are described below.
### Similarities :
The modified device and the predicate device have the following same design features:
- Both devices have the same principle of operation.
- . Both devices have a single-use, hermetically sealed, biocompatible Disposable component, which has the same design features for temperature safety, pressure safety, mechanical safety, biocompatibility, sterility and method to prevent re-use.
- Both devices use a Cable to join the patient interface (Disposable) to the physician interface ● (Console) that has the same electrical and pneumatic design requirements.
{6}------------------------------------------------
- . Both devices have a Control component with the same treatment hardware, treatment control, temperature and pressure specifications, safety features, device self-tests, software user interface and touchscreen display.
- . Both devices comply with the same performance standards for electrical safety and electromagnetic compatibility (EMC).
## Differences:
The differences between the new Activator II (Model LFD-2000) with the new semi-permanent Cable (Model CBL-2000) and the predicate Activator models (LFD-1000 and LFD-1100) include:
- . Activator II does not have a permanently attached cable for air tubing and electrical wiring. Activator II has a new electrical and pneumatic connector located behind the patient interface area that attaches to the new semi-permanent Cable. The single-use fuse that disables the Disposable component after treatment to prevent re-use was moved from the cable connector in the predicate Activator model to the Activator II.
- The new Cable is not disposable; it is semi-permanent and designed and tested for over 2,500 ● uses. A locking clip was added to the Cable connector to the Console to prevent accidental disposal of the Cable and to keep the connection with the Console secure. An electrical switch was added to the Cable connector to the Console, which when pressed will electrically disconnect the Cable from the Console facilitating self-test and troubleshooting system error messages. The Cable has a new electrical and pneumatic connector to attach to the Activator II. The Cable has three additional electrical conductors needed to detect when the Activator II is connected and to deactivate the Activator II when treatment is complete to prevent reuse. A new Cable holder was designed to store the Cable on the Console when not in use.
The differences between the modified Console and the predicate Console include:
- . The modified Console hardware has a new equivalent computer because the computer in the predicate device became obsolete.
- . The modified Console utilizes off-the-shelf Microsoft Windows 7 Embedded Operating System with drivers to support a new computer; whereas, the predicate device utilizes off-the-shelf Microsoft Windows XP Embedded Operating System.
- . The modified LipiFlow® software application has been updated from the predicate software to support the hardware-specific driver differences in the Operating Systems.
### PERFORMANCE TESTING:
Performance testing was conducted to demonstrate substantial equivalence of the modified LipiFlow® System to the predicate LipiFlow® System, and to validate continued conformance to the following special controls for an eyelid thermal pulsation system per 21 CFR 886.5200.
- 1) Performance testing demonstrates that the modified LipiFlow® System conforms to the same electrical safety and EMC performance standards as for the predicate device.
{7}------------------------------------------------
- 2) Performance testing shows that the modified LipiFlow® System has equivalent temperature and pressure performance and safeguard functions, including during fault conditions, as the predicate device. The modified device meets the same design requirements as the predicate device based on known safe and effective temperature and pressure specifications, previously validated in bench, animal and clinical studies of the LipiFlow® System.
- 3) The packaging, sterility and shelf-life for the new Activator II model are the same as for the predicate Activator models. The new Activator II model meets the same performance standards for sterility and shelf-life as the predicate Activator models.
- 4) The patient contact materials for the new Activator II model are the same as for the predicate Activator models. Therefore, the new Activator II model meets the same biocompatibility performance standards as the predicate Activator models.
- 5) Performance testing shows that the device modifications do not raise new questions of safety and effectiveness, and do not adversely affect safety and effectiveness of the device.
Furthermore, software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern because the software is an accessory to a medical device that has a moderate level of concern. Prior to the mitigation of hazards, a failure of the device could result in minor injury to the patient or user.
#### CONCLUSIONS:
The modified LipiFlow® System has the same intended use and the same fundamental scientific technology as the predicate device. Performance testing demonstrates the modified LipiFlow® System is at least as safe and effective as the legally marketed LipiFlow® System. Therefore, the modified LipiFlow® System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.