The Signia™ sterilization tray is intended to provide storage for the Signia™ reusable insertion guide and Signia™ manual retraction tool during sterilization, storage and transportation within the hospital environment. The Signia™ sterilization tray is only intended to maintain sterility of the enclosed if it is used in conjunction with an FDA cleared sterilization wrap and has only been evaluated for a non-stacked configuration. The tray can contain at a maximum: one (1) Signia™ adapter, one (1) Signia™ reusable insertion guide and one (1) Signia™ manual retraction tool. The tray is intended to allow steam sterilization of the enclosed medical devices. The validated sterilization cycle parameters are as follows: STEAM STERILIZATION Pre-vacuum (HiVac) Steam Cycles Item: Tray with components wrapped or in sterilization container system 132 °C Pre-vacuum (HiVac) Steam Cycles 134 °C Pre-vacuum (HiVac) Steam Cycles Exposure temperature 270 °F (132 °C) 273 °F (134 °C) Exposure time 4 minutes 3 minutes Vacuum dry time 20 - 40 minutes 20 - 40 minutes
Device Story
Stainless steel sterilization tray; holds Signia™ adapter, reusable insertion guide, and manual retraction tool. Perforated lid/base design allows steam penetration; includes internal inserts/strips for accessory organization. Used in hospital Central Supply Room (CSR) by qualified medical personnel. Facilitates steam sterilization of enclosed accessories; requires FDA-cleared sterilization wrap. Protects accessories during storage and transport; ensures sterility maintenance post-autoclave. Does not contact patients.
Clinical Evidence
No clinical data. Bench testing only. Cytotoxicity testing confirmed non-cytotoxic. Cleaning validation (AAMI TIR30:2011, FDA 2015 guidance) assessed carbohydrate, protein, and hemoglobin residuals; all met acceptance criteria. Sterilization validation (AAMI/ANSI ST77:2013) achieved SAL of 10^-6. Reliability evaluation confirmed material durability under reprocessing stresses.
Technological Characteristics
Stainless steel construction; perforated base and lid; includes latches and handles. Dimensions: 10.0 x 21.4 x 3.0 inches. Sterilization: Pre-vacuum steam (132°C/134°C). Non-powered, mechanical device. No software or connectivity.
Indications for Use
Indicated for storage, sterilization, and transport of Signia™ reusable insertion guide, manual retraction tool, and adapter within hospital environments. Requires use with FDA-cleared sterilization wrap in non-stacked configuration.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an emblem that resembles an abstract caduceus, with three stylized human profiles forming the staff and a serpent-like shape winding around them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 7, 2016
Covidien Katherine Choi, RAC Principal Regulatory Affairs Specialist 60 Middletown Avenue North Haven, Connecticut 06473
Re: K161347
Trade/Device Name: Signia™ Sterilization Tray Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: August 5, 2016 Received: August 8, 2016
Dear Katherine Choi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michael D. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K161347
Device Name Signia™ Sterilization Tray
#### Indications for Use (Describe)
The Signia™ sterilization tray is intended to provide storage for the Signia™ reusable insertion guide and Signia™ manual retraction tool during sterilization, storage and transportation within the hospital environment. The Signia™ sterilization tray is only intended to maintain sterility of the enclosed if it is used in coniunction with an FDA cleared sterilization wrap and has only been evaluated for a non-stacked configuration. The tray can contain at a maximum: one (1) Signia™ adapter, one (1) Signia™ reusable insertion guide and one (1) Signia™ manual retraction tool. The tray is intended to allow steam sterilization of the enclosed medical devices. The validated sterilization cycle parameters are as follows:
#### STEAM STERILIZATION
#### Pre-vacuum (HiVac) Steam Cycles
| Item: Tray with components wrapped<br>or in sterilization container system | 132 °C Pre-vacuum (HiVac) Steam<br>Cycles | 134 °C Pre-vacuum (HiVac) Steam<br>Cycles |
|----------------------------------------------------------------------------|-------------------------------------------|-------------------------------------------|
| Exposure temperature | 270 °F (132 °C) | 273 °F (134 °C) |
| Exposure time | 4 minutes | 3 minutes |
| Vacuum dry time | 20 - 40 minutes | 20 - 40 minutes |
Type of Use (Select one or both, as applicable)
| | Depurative Use (Part 111, 21 CFR 331, Subpart D) |
|---------------------------------------|--------------------------------------------------------|
| <div> <input type="checkbox"/> </div> | Over-The-Counter Use (Part 111, 21 CFR 331, Subpart C) |
| | Prescription Use {Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K161347 510(k) Summary
# DATE PREPARED:
September 2, 2016
#### SUBMITTER:
Covidien 60 Middletown Avenue North Haven, CT 06473 USA
# CONTACT PERSON:
Katherine Y. Choi, RAC Principal Requlatory Affairs Specialist Telephone: (203) 492-8412 Fax: (203) 492-5029
# IDENTIFICATION OF DEVICE:
| Proprietary/Trade Name: | Signia™ sterilization tray |
|-------------------------|----------------------------|
| Classification Name: | Sterilization wrap |
| Regulation Number: | 21 CFR 880.6850 |
| Product Code: | KCT |
| Device Class: | Class II |
| Review Panel: | General Hospital |
| Common Name: | Sterilization tray |
# PREDICATE DEVICE:
| Proprietary/Trade Name: | iDrive™ Ultra sterilization tray |
|-------------------------|----------------------------------------|
| 510(k) Number: | K130532 (Clearance date: May 20, 2013) |
| Classification Name: | Sterilization wrap |
| Regulation Number: | 21 CFR 880.6850 |
| Product Code: | KCT |
| Device Class: | Class II |
| Review Panel: | General Hospital |
| Common Name: | Sterilization tray |
# DEVICE DESCRIPTION:
The Signia™ sterilization tray is an optional reusable accessory to the Signia™ stapling system, and is intended to provide storage of the Signia™ accessories during sterilization, as well as, storage and transportation within the hospital environment. The Signia™ sterilization tray can simultaneously accommodate at a maximum the following Signia™ accessories:
One (1) Signia™ adapter One (1) Signia™ reusable insertion guide One (1) Signia™ manual retraction tool
The Signia™ sterilization tray is composed of a lid and base made of stainless steel. Both the lid and base are perforated to allow penetration of steam during sterilization. There are two latches on the side to secure the lid to the base, and handles for holding or transporting the tray. Inside of the base, inserts
{4}------------------------------------------------
and strips are placed to hold the respective Signia™ accessories along with the pre-printed layout of the accessory as a visual aid. Each corner has a protective cover for protection.
The Signia™ sterilization tray is designed to facilitate the sterilization process when used in conjunction within the Central Supply Room (CSR) wrapping material (an FDA cleared sterilization wrap). The tray is intended to be reprocessed, which is first to be cleaned either manually or by an automatio washerdisinfector, and then sterilization. The tray is to be handled by medical personnel qualified in the transportation, cleaning, sterilization and use of surgical devices.
# INTENDED USE/INDICATIONS FOR USE:
The Signia™ sterilization tray is intended to provide storage for the Signia™ reusable insertion guide and Signia™ manual retraction tool during sterilization, storage and transportation within the hospital environment. The Signia™ sterilization tray is only intended to maintain sterility of the enclosed devices if it is used in conjunction with an FDA cleared sterilization wrap and has only been evaluated for a non-stacked configuration. The tray can contain at a maximum: one (1) Signia™ adapter, one (1) Signia™ reusable insertion guide and one (1) Signia™ manual retraction tool. The tray is intended to allow steam sterilization of the enclosed medical devices. The validated sterilization cycle parameters are as follows:
# STEAM STERILIZATION
Pre-vacuum (HiVac) Steam Cycles
| Item: Tray with components<br>wrapped or in sterilization<br>container system | 132 °C Pre-vacuum (HiVac)<br>Steam Cycles | 134 °C Pre-vacuum (HiVac)<br>Steam Cycles |
|-------------------------------------------------------------------------------|-------------------------------------------|-------------------------------------------|
| Exposure temperature | 270 °F (132 °C) | 273 °F (134 °C) |
| Exposure time | 4 minutes | 3 minutes |
| Vacuum dry time | 20 - 40 minutes | 20 - 40 minutes |
# SUMMARY OF TECHNOLOGICAL CHARACTERISTICS:
The Signia™ sterilization tray is designed to withstand the indicated cleaning and sterilization conditions to aid in storage and sterilization of the Signia™ accessories. The tray itself is largely made of stainless steel, and does not require any special storage conditions. Repeated processing at the listed temperatures has minimal effect on the sterilization tray.
# SUBSTANTIAL EQUIVALENCE:
The Signia™ sterilization tray has the same intended use as the legally-marketed iDrive™ Ultra sterilization tray cleared via K130532 (May 20, 2013) since both are metal sterilization trays. Also the predicate device and new Signia™ sterilization tray can undergo autoclave sterilization. However, the enclosed devices are different.
{5}------------------------------------------------
The comparison between the Signia™ sterilization tray and the predicate tray K130532 is highlighted in the table below.
| | Subject Device<br>Signia™ sterilization tray | Predicate Device K130532<br>iDrive™ Ultra sterilization tray |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Enclosed<br>devices for<br>sterilization | The tray can contain at a maximum:<br>one (1) Signia™ adapter,<br>one (1) Signia™ reusable insertion guide<br>and<br>one (1) Signia™ manual retraction tool. | The tray can contain at a maximum:<br>one (1) iDrive™ Ultra powered handle,<br>two (2) Endo GIA™ adapters,<br>two (2) iDrive™ battery insertion guides and<br>one (1) iDrive™ Ultra manual adapter tool. |
| Dimensions | Approx. 10.0 x 21.4 x 3.0 (H) inches | Approx. 10 x 21 x 3 (H) inches |
| Sterilization<br>methods | Prevacuum steam | Prevacuum steam |
| Steam<br>Sterilization<br>parameters | 132 °C Pre-vacuum (Hi Vac) Steam Cycle<br>Exposure temperature: 270 °F (132 °C)<br>Exposure time: 4 minutes<br>Vacuum dry time: 20 - 40 minutes<br>134 °C Pre-vacuum (Hi Vac) Steam Cycle<br>Exposure temperature: 273 °F (134 °C)<br>Exposure time: 3 minutes<br>Vacuum dry time: 20 - 40 minutes | Prevacuum Steam Cycles<br>132°C for 4 minutes<br>134°C for 3 minutes<br>Vacuum Dry: 20 minutes |
| Base & Lid<br>Materials | Stainless steel | Aluminum |
| Stacking | Do not stack cases and trays in the<br>sterilization chamber. | Do not stack cases and trays in the<br>autoclave chamber. |
| Max. Load<br>Capacity | 10 pounds | 15 pounds |
# SUMMARY OF STUDIES:
Non-clinical performance data - Although the Signia™ sterilization tray is not intended for patient contact, all the materials that contact the enclosed Signia™ accessories during sterilization have been evaluated for cytotoxicity as an additional validation of safety to determine if there was any release of the material constituents or leachable chemicals produced from the cleaning and sterilization process that might be potentially toxic, and the testing results confirmed 'non-cytotoxic'.
The cleaning validation was designed to simulate the worst case scenario conforming to several industry standards including AAMI TIR30:2011 and the FDA's 2015 reprocessing quidance document. The cleanliness of the tray was assessed by carbohydrate, protein and hemoglobin residuals, and all residuals met predefined acceptance criteria demonstrating the adequate cleanliness of the tray using the cleaning instructions.
The sterilization validation, designed based on the FDA-recognized standard AAMI/ANSI ST77:2013, demonstrated the minimum sterility assurance level (SAL) of 10° can be achieved if the sterilization instructions in the instructions for use (IFU) were followed.
The reliability evaluation concluded the materials used in the Signia™ sterilization tray have been shown, through use in other devices, to withstand the cleaning and sterilization process without any adverse reactions and the tray will not undergo degradation due to the environmental stresses it is exposed to during reprocessing.
{6}------------------------------------------------
Clinical performance data - No clinical study has been performed. The substantial equivalence has been demonstrated by non-clinical studies.
#### CONCLUSION:
Based upon the supporting data summarized above, we concluded that Signia™ sterlization tray is as safe and as effective as the legally marketed device K130532, and does not raise different questions of safety and effectiveness than the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.