K161304 · Nephros, Inc. · FIP · Mar 2, 2017 · Gastroenterology, Urology
Device Facts
Record ID
K161304
Device Name
EndoPur Filter
Applicant
Nephros, Inc.
Product Code
FIP · Gastroenterology, Urology
Decision Date
Mar 2, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5665
Device Class
Class 2
Indications for Use
The EndoPur Endotoxin 10" Filter Flat is intended to be used to filter water used in hemodialysis devices. It assists in providing hemodialysis quality water. The device is not a complete water treatment system, but serves to remove biological contaminants. Therefore it must be used in conjunction with other water treatment equipment (Reverse Osmosis, Deionization, etc.).
Device Story
EndoPur Endotoxin 10" Filter Flat is a hollow fiber ultrafilter designed to remove biological contaminants—bacteria, viruses, endotoxin, and particulates—from water used in hemodialysis. The device functions as a component within a larger water treatment system (e.g., Reverse Osmosis or Deionization units). It is intended for use in clinical settings where hemodialysis is performed. The filter operates by passing water through hollow fibers to retain contaminants, thereby assisting in the production of hemodialysis-quality water. It is a passive, non-active device requiring no software or electrical power. Healthcare providers or technicians integrate the filter into the existing water treatment infrastructure to ensure water quality meets safety standards for patient treatment, reducing the risk of exposure to biological contaminants during hemodialysis.
Clinical Evidence
No clinical data. Bench testing only. Performance validated via flow rate vs. pressure drop, pyrogen removal, virus challenge, bacteria challenge, and disinfection compatibility testing. Biocompatibility assessed via extractable and leachable testing evaluated by a toxicologist. AAMI panel testing performed per ANSI/AAMI 13959.
Technological Characteristics
Hollow fiber ultrafilter. Materials of construction are identical to the predicate device. Passive device; no electrical energy source. Dimensions: 10-inch filter flat. Disinfection compatibility: compatible with standard disinfection processes (excluding heat sterilization above 80°C for specific housings).
Indications for Use
Indicated for filtering water used in hemodialysis devices to assist in providing hemodialysis quality water. Must be used in conjunction with other water treatment equipment (e.g., Reverse Osmosis, Deionization). Contraindicated for heat sterilization with polypropylene housings not rated for temperatures above 80°C.
Regulatory Classification
Identification
A water purification system for hemodialysis is a device that is intended for use with a hemodialysis system and that is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. This generic type of device may include a water softener, sediment filter, carbon filter, and water distillation system.
Special Controls
*Classification.* Class II (special controls). The device, when it is a water purification subsystem disinfectant, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Predicate Devices
FiberFlo™ Hollow Fiber Cartridge Water Filters (K954349)
Reference Devices
DSU/SSU
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 2, 2017
Nephros, Inc. Hollie Johnson Director Quality and Requlatory 41 Grand Ave River Edge, NJ 07661
Re: K161304
> Trade/Device Name: EndoPur Endotoxin 10" Filter Flat Regulation Number: 21 CFR§ 876.5665 Regulation Name: Water Purification System For Hemodialysis Regulatory Class: II Product Code: FIP Dated: December 23, 2016 Received: December 27, 2016
Dear Hollie Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K161304
Device Name EndoPur Endotoxin 10" Filter Flat
#### Indications for Use (Describe)
#### INDICATIONS FOR USE
Indications for Use: The EndoPur Endotoxin 10" Filter Flat is intended to filter water used in hemodialysis devices. It assists in providing hemodialysis quality water. The device is not a complete water treatment system, but serves to remove biological contaminants. Therefore it must be used in conjunction with other water treatment equipment (Reverse Osmosis, Deionization, etc.).
#### CONTRAINDICATIONS
Heat sterilization is contraindicated for use with polypropylene housings that are not rated for disinfection temperatures above 80°C
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Submitter: | Nephros Inc.<br>41 Grand Ave<br>River Edge, NJ 07661<br>Establishment Registration # 3003337893 |
|---------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Hollie Johnson ; Director of Quality and Regulatory<br>41 Grand Ave<br>River Edge, NJ 07661<br>201-343-5202 x101(p)<br>201-343-5207 (f)<br>hollie@nephros.com |
| Date Prepared | December 28, 2016 |
| Trade Name | EndoPur Endotoxin 10" Filter Flat |
| Proposed Class | Class II |
| Classification Name<br>and Number | 21 CFR Part 876.5665 Water Purification System for<br>Hemodialysis |
| Product Code | FIP |
| Predicate Devices | FiberFlo™ Hollow Fiber Cartridge Water Filters – K95439 |
| Device Description | The EndoPur Endotoxin 10" Filter Flat is a hollow fiber ultrafilter<br>that retains bacteria, viruses, endotoxin and particulate from<br>water used in hemodialysis. |
| Intended Use | The EndoPur Endotoxin 10" Filter Flat is a hollow fiber ultrafilter<br>that retains bacteria and endotoxin from fluid. |
| Indications For Use | The EndoPur Endotoxin 10" Filter Flat is intended to be used to<br>filter water used in hemodialysis devices. It assists in providing<br>hemodialysis quality water. The device is not a complete water<br>treatment system, but serves to remove biological<br>contaminants. Therefore it must be used in conjunction with |
| | other water treatment equipment (Reverse Osmosis,<br>Deionization, etc.). |
| | CONTRAINDICATIONS |
| | Heat sterilization is contraindicated for use with polypropylene<br>housings that are not rated for disinfection temperatures above<br>80°C |
| Summary of the<br>Technological<br>Characteristics | The subject device has the same technological characteristics<br>and is similar in design as compared to the primary and<br>reference devices; see executive summary located in section<br>10. |
| Assessment of Non-<br>clinical Performance<br>Data / Substantial<br>Equivalence | The EndoPur Endotoxin 10" Filter Flat has been tested for<br>performance. The tests conducted include Flow Rate versus<br>Pressure Drop, Pyrogen Removal, Virus Challenge Test,<br>Bacteria Challenge Test and Disinfection Compatibility. This<br>filter was found to be substantially equivalent to the primary<br>predicate Minntech Fiberflo Filter (K954349) and reference<br>device, DSU/SSU. |
| Biocompatibility<br>Testing | Exhaustive Extractable and Leach testing was performed,<br>evaluated by a toxicologist and found to be safe. AAMI panel<br>was executed using ANSI/AAMI 13959 Water for Hemodialysis<br>and Related Therapies to determine residual levels after each<br>disinfection process on accelerated aged product. All results<br>were found to be acceptable. |
| Electrical Safety and<br>Electromagnetic<br>Compatibility (EMC) | Not Applicable; not and Active Medical Device |
| Software Verification<br>and Validation Testing | Not Applicable; No Software contained |
| Mechanical and<br>Acoustic Study | Not Applicable |
| Animal Study | Not Applicable |
| Clinical Studies | Not Applicable |
| Conclusions | Since the predicate device is comprised of the same materials<br>of construction that have no adverse events while on the<br>market and the safety testing completed by Nephros<br>demonstrates no toxicity concerns, Nephros EndoPur<br>Endotoxin 10” filter Flat is ready for marketing for its intended<br>use, which is the same as its predicate and reference devices. |
## 510(k) Summary: EndoPur Endotoxin 10" Filter Flat
{4}------------------------------------------------
{5}------------------------------------------------
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.