Reusable, self-adhering neurostimulation electrodes are indicated for use with transcutaneous electrical stimulation devices. Some common types of transcutaneous stimulation devices include, but are not limited to, transepithelial nerve stimulation (TENS) and electrical muscle stimulation (EMS) devices. Transcutaneous neurostimulation electrodes are passive devices serving as an interface between a user's skin and a neurostimulation device.
Device Story
FlexStim Neurostimulation Electrodes are passive cutaneous electrodes used to interface between a neurostimulation device and a patient's skin. The device receives electrical signals from a TENS or EMS stimulator via lead wires or snap connectors; these signals are dispersed across a conductive carbon film and transmitted through a conductive hydrogel layer to the skin. Used in hospitals, clinics, or home settings by patients or clinicians. The device facilitates electrical stimulation for therapeutic purposes, such as pain relief or muscle stimulation. It is a non-sterile, single-patient, multiple-application device available in various backing materials (non-woven cloth or PE foam).
Clinical Evidence
No clinical data. Evidence consists of bench testing: biocompatibility (cytotoxicity, sensitization, irritation per ISO 10993-1), electrical impedance testing using SEAM 10, adhesive peel testing per Afera 5001, and accelerated/real-time shelf-life stability studies (up to 24 months).
Technological Characteristics
Passive cutaneous electrode. Materials: non-woven cloth/PE foam backing, conductive carbon film, conductive hydrogel, stainless steel snap, glass-filled ABS eyelet with Ag/AgCl coating, PET release liner. Connectivity: lead wire (2.0 mm DIN) or snap connection. Complies with 21 CFR 898 and IEC 60601-1 (sub-clause 56.3(c)) for lead wire performance.
Indications for Use
Indicated for adults requiring transcutaneous electrical stimulation (e.g., TENS, EMS) via an interface between the skin and a neurostimulation device. Contraindicated for use on non-intact skin.
Regulatory Classification
Identification
A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.
K012463 — LIFECARE 5*5 CM SQUARE, LIFECARE 4*4 CM SQUARE, LIFECARE 5 CM ROUND, LIFECARE 7.5 CM ROUND, LIFECARE 4*6 CM OVAL · Everyway Medical Instruments Co.,Ltd · Feb 5, 2002
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, overlapping each other to create a sense of depth and unity. The profiles are simple and abstract, with flowing lines that suggest movement and progress.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 30, 2016
Medico Electrodes International Ltd. Amit Seth Vice President Plot 142A/11, 12, 27, 28 &29, Noida Special Economic Zone Noida, Uttar Pradesh 201305 India
#### Re: K161282
Trade/Device Name: FlexStim Neurostimulation Electrodes Model Numbers: 5050ROC1S, 5050ROC1W, 5050ROC2S, 5050ROC2W, 5050SQC1S, 5050SQC1W, 5050SQC2S, 5050SQC2W Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous Electrode Regulatory Class: Class II Product Code: GXY Dated: September 8, 2016 Received: September 28, 2016
Dear Amit Seth:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{1}------------------------------------------------
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Michael J. Hoffmann -A
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K161282
#### Device Name
FlexStim Neurostimulation Electrodes Model Numbers: 5050ROC1S, 5050ROC1W, 5050ROC2W, 5050SQC1S, 5050SQC1W, 5050SQC1W, 5050SQC2S, 5050SQC2W
#### Indications for Use (Describe)
Reusable, self-adhering neurostimulation electrodes are indicated for use with transcutaneous electrical stimulation devices. Some common types of transcutaneous stimulation devices include, but are not limited to, transepithelial nerve stimulation (TENS) and electrical muscle stimulation (EMS) devices. Transcutaneous neurostimulation electrodes are passive devices serving as an interface between a user's skin and a neurostimulation device.
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|----------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☒</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for MEDICO ELECTRODES. The logo features a stylized letter "M" intertwined with other shapes in blue and gray. The text "MEDICO ELECTRODES" is written in blue, sans-serif font to the right of the symbol.
# 510(k) Summary
(As required by 21 CFR 807.92)
# I. SUBMITTER
Medico Electrodes International Ltd. Plot 142A/11, 12, 27, 28 & 29, Noida Special Economic Zone, Noida - 201305, Uttar Pradesh, India
Phone: 91-120-3042984 Fax: 91-120-3042986
Contact Person: Amit Seth Date Prepared: Oct 28, 2016
## II. DEVICE
Name of Device: FlexStim Neurostimulation Electrodes Model Numbers: 5050ROC1S 5050ROC1W 5050ROC2S 5050ROC2W 5050SQC1S 5050SQC1W 5050SQC2S 5050SOC2W Common or Usual Name: Neurostimulation Electrodes Classification Name: Cutaneous electrode (21 CFR 882.1320) Regulatory Class: II Product Code: GXY
## III. PREDICATE DEVICE
PROTENS Reusable Stimulating Electrodes, K142099 (Predicate Device 1) Reusable Stimulating Electrodes, K111270 (Predicate Device 2)
These predicate devices had not been subject to a design-related recall. No reference devices were used in these submissions.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for MEDICO ELECTRODES. The logo features a stylized "M" and "J" intertwined within a circle on the left. The text "MEDICO ELECTRODES" is written in a sans-serif font to the right of the symbol.
# IV. DEVICE DESCRIPTION
- 1. Device Identification
The FlexStim Neurostimulation Electrode is used as a transcutaneous electrical nerve stimulation electrode in conjunction with an electrical stimulator for TENS or EMS and available in backing materials such as, non-woven cloth and PE foam, conductive carbon film (low impedance and standard impedance) with lead wire and snap connection configurations.
- 2. Device Characteristics
- Software: Not Applicable ●
- Biologics: Not Applicable ●
- Drugs: Not Applicable
- Any patient-contacting materials: Conductive hydrogel ●
- Coatings: Not Applicable ●
- Additives: Not Applicable ●
- The electrodes are designed for single patient/multiple application use ●
- The device is not a sterile product, therefore sterilization is not required
- 3. Environment of Use:
- Healthcare facility/hospital .
- . Home
- 4. Description of the Device
The device functions as a passive device by carrying an electrical signal from a neurostimulation device through the device cable and electrode lead wire or snap button to the user's skin.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for MEDICO ELECTRODES. The logo consists of a stylized graphic of the letters 'M' and 'J' intertwined, with the word 'MEDICO' in blue, sans-serif font to the right of the graphic. The word 'ELECTRODES' is also in blue, sans-serif font and is placed to the right of 'MEDICO'.
- 5. Materials of Use
| S. No. | Component | Description/Material of Construction |
|--------|--------------------------------------------------|-----------------------------------------------------------------------------------------|
| 1 | Backing material(s) | Non-woven cloth/PE foam |
| 2 | Lead wire having insulation on female connector | Wire length: 114.3 mm or 4.5" (104.3 mm or 4.1" - 124.3 mm or 4.9")<br>Din size: 2.0 mm |
| 3 | Stud/Snap (for electrodes without lead wire) | Stainless steel |
| 4 | Sensor/Eyelet (for electrodes without lead wire) | Glass filled ABS with Ag/AgCl coating |
| 5 | Carbon film | Conductive carbon film (low impedance and standard impedance) |
| 6 | Gel | Conductive hydrogel |
| 7 | Release liner | PET i.e., Polyethylene terephthalate |
### V. INDICATIONS FOR USE
Reusable, self-adhering Neurostimulation Electrodes are indicated for use with transcutaneous electrical stimulation devices. Some common types of transcutaneous stimulation devices include, but are not limited to, transepithelial nerve stimulation (TENS) and electrical muscle stimulation (EMS) devices. Transcutaneous Neurostimulation Electrodes are passive devices serving as an interface between a user's skin and a neurostimulation device.
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The FlexStim Neurostimulation Electrodes exhibit technological characteristics that are substantially equivalent to those of the predicate devices, as determined by both component usage and physical testing.
The Substantial Equivalence summary is described below:
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Medico Electrodes. The logo consists of a stylized letter "M" intertwined with other shapes, followed by the text "MEDICO ELECTRODES" in a sans-serif font. The color scheme is primarily blue, with some elements in a lighter shade of gray.
| S. No. | Areas of<br>Comparison | Subject Device<br>FlexStim<br>Neurostimulation<br>Electrodes<br>(Manufactured by Medico<br>Electrodes International<br>Ltd., India) | Predicate Device 1<br>PROTENS Reusable<br>Stimulating Electrodes<br>(K142099)<br>(Manufactured by<br>Bio Protech Inc., Korea) | Predicate Device 2<br>Reusable Stimulating<br>Electrodes (K111270)<br>(Manufactured by<br>Bio Protech Inc., Korea) |
|--------|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | Intended Use /<br>Indications for<br>Use | Reusable, self-adhering,<br>Neurostimulation Electrodes<br>are indicated for use with<br>transcutaneous electrical<br>stimulation devices. Some<br>common types of<br>transcutaneous stimulation<br>devices include, but are not<br>limited to, transepithelial<br>nerve stimulation (TENS)<br>and electrical muscle<br>stimulation (EMS) devices.<br>Transcutaneous<br>Neurostimulation Electrodes<br>are passive devices serving<br>as an interface between a<br>user's skin and a<br>neurostimulation device. | PROTENS reusable, self-<br>adhering, over-the-counter<br>Cutaneous electrodes are<br>indicated for use with<br>electrical stimulation<br>device. Some common<br>types of electrical<br>stimulation device include,<br>but are not limited to,<br>transcutaneous nerve<br>stimulation (TENS),<br>electrical muscle<br>stimulation (EMS) device,<br>Neuromuscular Electrical<br>Stimulation (NMES/FES)<br>device and Microcurrent<br>electrical nerve stimulation<br>(MENS), Interferential<br>Stimulation (IF).<br>Cutaneous electrodes are<br>passive devices serving as<br>an interface between a<br>user's skin and an<br>electrical stimulation | Bio Protech Reusable<br>Stimulating electrodes are<br>intended to be used to<br>apply electrical stimulation<br>current to the patient's skin<br>or to record physiological<br>signals.<br>Electrical stimulation<br>current applications of<br>these electrodes are:<br>a) Transcutaneous<br>Electrical Nerve Stimulation<br>(TENS) for pain relief<br>b) Electrical muscle<br>stimulation (EMS)<br>c) Functional electrical<br>stimulation (FES)<br>d) Microcurrent electrical<br>nerve stimulation (MENS)<br>e) Interferential stimulation<br>(IF)<br>f) Neuromuscular electrical<br>stimulation (NMES) |
| S. No. | Areas of<br>Comparison | Subject Device<br>FlexStim<br>Neurostimulation<br>Electrodes<br>(Manufactured by Medico<br>Electrodes International<br>Ltd., India) | Predicate Device 1<br>PROTENS Reusable<br>Stimulating Electrodes<br>(K142099)<br>(Manufactured by<br>Bio Protech Inc., Korea) | Predicate Device 2<br>Reusable Stimulating<br>Electrodes (K111270)<br>(Manufactured by<br>Bio Protech Inc., Korea) |
| 2 | Target<br>population | Adults | device. | Adults |
| 3 | Anatomical site | Intact skin (refer IFU) | Intact skin (refer IFU) | Intact skin (refer IFU) |
| 4 | Where used | Hospital, clinic and home<br>use environment | Hospital, clinic and home<br>use environment | Hospital, clinic and home<br>use environment |
| 5 | Design | Cutaneous electrode which<br>conducts an electrical signal<br>from a neurostimulation<br>device through a lead wire<br>or snap button; which is<br>dispersed from the wire<br>across a conductive<br>surface; then transmitted<br>through the conductive<br>adhesive gel to the surface<br>of the patient's skin. | Cutaneous electrode which<br>conducts an electrical<br>signal from a<br>neurostimulation device<br>through a lead wire; which<br>is dispersed from the wire<br>across a conductive<br>surface; then transmitted<br>through the conductive<br>adhesive gel to the surface<br>of the patient's skin. | Cutaneous electrode which<br>conducts an electrical signal<br>from a neurostimulation<br>device through a lead wire;<br>which is dispersed from the<br>wire across a conductive<br>surface; then transmitted<br>through the conductive<br>adhesive gel to the surface<br>of the patient's skin. |
| 6 | Materials | Basic components:<br>• Backing material<br>(Non-woven cloth/PE<br>foam)<br>• Lead wire and snap<br>connection<br>configurations | Basic components:<br>• Backing material<br>(Non-woven<br>cloth/PE<br>Foam/Tricot)<br>• Lead wire and snap<br>connection | Basic components:<br>• Backing material<br>(Non-woven cloth/PE<br>Foam/Tricot)<br>• Lead wire and snap<br>connection<br>configurations |
| S. No. | Areas of<br>Comparison | Subject Device<br>FlexStim<br>Neurostimulation<br>Electrodes<br>(Manufactured by Medico<br>Electrodes International<br>Ltd., India) | Predicate Device 1<br>PROTENS Reusable<br>Stimulating Electrodes<br>(K142099)<br>(Manufactured by<br>Bio Protech Inc., Korea) | Predicate Device 2<br>Reusable Stimulating<br>Electrodes (K111270)<br>(Manufactured by<br>Bio Protech Inc., Korea) |
| | | Conductive carbon film Hydrogel PET liner | configurations Conductive carbon/silver film Hydrogel Transparent PET liner | Conductive carbon/silver film Hydrogel Transparent PET liner |
| 7 | Performance | Based on successful<br>biocompatibility testing of<br>the skin contacting<br>conductive hydrogel,<br>impedance testing of the<br>product, electrical<br>performance of the<br>insulated lead wire<br>components, adhesive<br>performance and stability,<br>the FlexStim<br>Neurostimulation Electrodes<br>are safe and effective when<br>used as an interface<br>between a user's skin and<br>an approved<br>neurostimulation devices. | Based on successful<br>biocompatibility testing of<br>the skin contacting<br>conductive hydrogel,<br>impedance testing of the<br>product, electrical<br>performance of the<br>insulated lead wire<br>components and adhesive<br>performance, the<br>PROTENS Reusable<br>Stimulating Electrodes are<br>safe and effective when<br>used as an interface<br>between a user's skin and<br>an approved<br>neurostimulation devices. | Based on successful<br>biocompatibility testing of<br>the skin contacting<br>conductive hydrogel,<br>impedance testing of the<br>product, electrical<br>performance of the<br>insulated lead wire<br>components and adhesive<br>performance, the Reusable<br>Stimulating Electrodes are<br>safe and effective when<br>used as an interface<br>between a user's skin and<br>an approved<br>neurostimulation devices. |
| 8 | Biocompatibility | The skin-contacting | Hydrogel | Hydrogel |
| S. No. | Areas of<br>Comparison | Subject Device<br>FlexStim<br>Neurostimulation<br>Electrodes<br>(Manufactured by Medico<br>Electrodes International<br>Ltd., India) | Predicate Device 1<br>PROTENS Reusable<br>Stimulating Electrodes<br>(K142099)<br>(Manufactured by<br>Bio Protech Inc., Korea) | Predicate Device 2<br>Reusable Stimulating<br>Electrodes (K111270)<br>(Manufactured by<br>Bio Protech Inc., Korea) |
| | | material i.e., hydrogel was<br>found to be biocompatible<br>for its intended use as per<br>ISO 10993-1 standard. | Bio-Compatible High Adhesion | Bio-Compatible High Adhesion |
| 9 | Differences | | | |
| | Intended Use | Intended to be used to<br>apply electrical stimulation<br>current to the patient's skin | Intended to be used to<br>apply electrical stimulation<br>current to the patient's<br>skin | Intended to be used to<br>apply electrical stimulation<br>current to the patient's skin<br>or to record physiological<br>signals. |
| | Materials | FlexStim<br>Neurostimulation<br>Electrodes use Non-<br>woven cloth and PE<br>foam as the backing<br>materials Conductive carbon<br>film | Reusable Stimulating Electrodes use backing<br>materials such as Non-woven cloth, PE Foam,<br>Tricot Conductive film – carbon film or silver film | Reusable Stimulating Electrodes use backing<br>materials such as Non-woven cloth, PE Foam,<br>Tricot Conductive film – carbon film or silver film |
| | Method of sale<br>and supply | We claim that these<br>electrodes can be sold as<br>OTC (21 CFR 801 Subpart<br>C) as well as Prescription<br>(21 CFR 801 Subpart D)<br>medical devices. | The devices are currently<br>being sold as OTC medical<br>devices. | The devices are currently<br>being sold as Prescription<br>medical devices. |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Medico Electrodes. The logo features the letters 'M' and 'J' in a stylized design, with the word 'MEDICO' in bold blue letters and 'ELECTRODES' in a smaller font size next to it. The color scheme is primarily blue and gray.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Medico Electrodes. The logo features a stylized letter "M" in blue, intertwined with a gray element. The text "MEDICO ELECTRODES" is written in a sans-serif font to the right of the logo, with the words in all caps and in a matching blue color.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Medico Electrodes. The logo consists of a stylized letter "M" and "J" on the left side, with the text "MEDICO ELECTRODES" on the right side. The text is in a sans-serif font and is colored in a light blue color. The logo is simple and modern, and it is likely used to represent the company's brand.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for MEDICO ELECTRODES. The logo consists of a stylized letter "M" in light blue, with a gray letter "J" behind it. To the right of the letters is the text "MEDICO ELECTRODES" in a sans-serif font, also in light blue. The logo is simple and modern, and the use of color is eye-catching.
### VII. PERFORMANCE DATA
The following performance data is being provided in support of the substantial equivalence determination.
#### 1. Biocompatibility
The biocompatibility evaluation for the FlexStim Neurostimulation Electrodes was conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process", as recognized by FDA. The battery of testing included the following tests:
- Cytotoxicity
- · Sensitization
- · Irritation
The skin-contacting material i.e., conductive hydrogel was found to be biocompatible for its intended use.
#### 2. Electrical Performance
Impedance testing of the product was performed using the Surface Electrode Analysis Meter (SEAM 10). The equipment has been designed by CALM Technologies Inc., Canada to test the impedance of Neurostimulation Electrodes under a wide range of current pulse inputs, thereby assuring that the electrode has conductive properties appropriate to the device's intended use.
As per the Certificate of Compliance (CoC) received from the vendor of lead wire, the lead wire assembly is in compliance with the requirements of FDA performance standard 21 CFR Part 898 by testing under IEC 60601-1, sub-clause 56.3 (c).
#### 3. Adhesive Performance
In order to assure that the electrode's adhesive performance is substantially equivalent to the predicate devices, the Peel Test was performed as per the International Standard Afera 5001. Based on the results, it was concluded that the design of the electrode ensures that it will adhere to the patient's skin for a duration of use compatible with the intended use of the device.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for MEDICO ELECTRODES. The logo consists of a stylized letter "M" in light blue, intertwined with a gray shape. To the right of the symbol, the words "MEDICO ELECTRODES" are written in a bold, sans-serif font, also in light blue.
# 4. Stability
- FlexStim electrode samples had undergone an accelerated shelf-life study when kept at 40°C ± 2°C and relative humidity of 75 % ± 5 % in a Calibrated Stability Chamber (SC-01) and tested at 0, 3 & 6 Months to monitor the critical parameters of finished device such as impedance and peel adhesion.
- Apart from samples kept at accelerated ageing temperature, in parallel, samples were ● also kept at ambient temperature (25°C ± 2°C and relative humidity of 60 % ± 5 %) and tested for impedance & peel adhesion to demonstrate performance of the hydrogel for up to 11 months.
- As per the accelerated and real time stability study results, it is concluded that the . product passes electrical (impedance testing) and adhesive performance test. The product is suitable for use up to 24 months when stored at 5°C - 30°C. This further implies that the hydrogel is stable and resist physical and chemical breakdown as a result of conducting electrical current and extended periods of storage over a range of environmental conditions.
## VIII. CONCLUSIONS
Medico Electrodes International Ltd., India considers the FlexStim Neurostimulation Electrodes to be as safe and effective as the predicate devices PROTENS Reusable Stimulating Electrodes (K142099) and Reusable Stimulating Electrodes (K111270) as the subject device has identical indications for use, technological characteristics (including design and materials) and performance specifications as compared to the predicate devices already being legally marketed in the United States. Thus, the device is substantially equivalent to the predicates and any difference between the devices do not pose new questions of safety and effectiveness.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.