S-scan

K161238 · Esaote, S.p.A. · LNH · Jul 8, 2016 · Radiology

Device Facts

Record IDK161238
Device NameS-scan
ApplicantEsaote, S.p.A.
Product CodeLNH · Radiology
Decision DateJul 8, 2016
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.1000
Device ClassClass 2

Indications for Use

S-scan is a Magnetic Resonance (MR) system that produces transversal, sagittal, coronal and oblique cross-section images of the limbs, joints and spinal column. It is intended for imaging portions of the upper limb, including the hand, wrist, forearm, elbow, arm and shoulder, imaging portions of the lower limb, including the foot, ankle, calf, knee, thigh and hip, imaging the temporomandibular joint and imaging the cervical, the thoracic and the lumbosacral sections as portions of the spinal column. S-scan images correspond to the spatial distribution of protons (hydrogen nuclei) that determine magnetic resonance properties and are dependent on the MR parameters, including spinlatice relaxation time (T1), spin-spin relaxation time (T2), nuclei density, flow velocity and "chemical shift". When interpreted by a medical expert trained in use of MR equipment, the images can provide diagnostically useful information.

Device Story

S-scan is a magnetic resonance imaging (MRI) system; produces cross-sectional images of limbs, joints, and spine based on proton spatial distribution. Input: magnetic resonance signals dependent on T1, T2, nuclei density, flow velocity, and chemical shift. Output: transversal, sagittal, coronal, and oblique images. Used in clinical settings by trained medical experts to provide diagnostic information. Modification introduces a 3-channel rigid wearable shoulder coil for improved image quality and patient accommodation. Healthcare providers interpret images to inform clinical decision-making; benefits include non-invasive diagnostic visualization of musculoskeletal structures.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

Magnetic resonance diagnostic device; 3-channel receiving shoulder coil; rigid wearable form factor with adjustable base for left/right laterality. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-33, IEC 62304, IEC 62366, ISO 10993-1, ISO 14971-2, NEMA MS-1, NEMA MS-3.

Indications for Use

Indicated for MR imaging of upper limbs (hand, wrist, forearm, elbow, arm, shoulder), lower limbs (foot, ankle, calf, knee, thigh, hip), temporomandibular joint, and spinal column (cervical, thoracic, lumbosacral sections).

Regulatory Classification

Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a silhouette effect. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the graphic. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 8, 2016 ESAOTE S.p.A. % Ms. Allison Scott Senior Regulatory Consultant Navigant Consulting, Inc. 9001 Wesleyan Road, Suite 200 INDIANAPOLIS IN 46268 Re: K161238 Trade/Device Name: S-scan Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: June 8, 2016 Received: June 9, 2016 Dear Ms. Scott: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Robert Odle Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration #### Indications for Use #### 510(k) Number (if known) #### K161238 UNKNOWN Device Name S-Scan #### Indications for Use (Describe) S-scan is a Magnetic Resonance (MR) system that produces transversal, sagittal, coronal and oblique cross-section images of the limbs, joints and spinal column. It is intended for imaging portions of the upper limb, including the hand, wrist, forearm, elbow, arm and shoulder, imaging portions of the lower limb, including the foot, ankle, calf, knee, thigh and hip, imaging the temporomandibular joint and imaging the cervical, the thoracic and the lumbosacral sections as portions of the spinal column. S-scan images correspond to the spatial distribution of protons (hydrogen nuclei) that determine magnetic resonance properties and are dependent on the MR parameters, including spinlatice relaxation time (T1), spin-spin relaxation time (T2), nuclei density, flow velocity and "chemical shift". When interpreted by a medical expert trained in use of MR equipment, the images can provide diagnostically useful information. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) {3}------------------------------------------------ 510(k) Submission S-scan Esaote, S.p.A. # Special 510(k) Summary The following 510(k) summary has been prepared pursuant to requirements specified in 21CFR 807.92. #### Submitter Information Esaote S.p.A. Via Siffredi 58 16153 Genova - Italy Contact Person: Allison Scott, RAC Senior Regulatory Consultant Tel. 317.228.8719 Fax 317.228.8701 allison.scott@navigant.com June 9, 2016 Date: Tradename: S-scan Classification Name: Magnetic Resonance Diagnostic Device Classification Panel: Radiology Classification Number: 90LNH #### Predicate Devices | Trade name | Common name | Class | Product code | Manufacturer | K number | |------------|--------------------------------------------|-------|--------------|--------------|----------| | S-scan | System, nuclear magnetic resonance imaging | II | LNH | ESAOTE S.P.A | K153039 | {4}------------------------------------------------ 510(k) Submission S-scan Esaote, S.p.A. # Device Description The change performed on the modified S-scan, with respect to the cleared version, S-scan K153039, is due to the improvement of the system performance. This modification, which does not affect the intended use or alter the fundamental scientific technology of the device, is the introduction of the Shoulder Coil 20 code 130000014. The Shoulder Coil N.20 is a three channels receiving coil. Its intended use is for shoulder examination on the S-scan. The Shoulder Coil is a wearable rigid coil equipped with an adaptor for fast placement in the patient table. The same adaptor matches for patient laterality (left /right shoulder). # Intended Use S-scan is a Magnetic Resonance (MR) system that produces transversal, sagittal, coronal and oblique cross-section images of the limbs, joints and spinal column. It is intended for imaging portions of the upper limb, including the hand, wrist, forearm, elbow, arm and shoulder, imaging portions of the lower limb, including the foot, ankle, calf, knee, thigh and hip, imaging the temporomandibular joint and imaging the cervical, the thoracic and the lumbosacral sections as portions of the spinal column. S-scan images correspond to the spatial distribution of protons (hydrogen nuclei) that determine magnetic resonance properties and are dependent on the MR parameters, including spinlattice relaxation time (T1), spin-spin relaxation time (T2), nuclei density, flow velocity and "chemical shift". When interpreted by a medical expert trained in use of MR equipment, the images can provide diagnostically useful information. {5}------------------------------------------------ 510(k) Submission S-scan Esaote, S.p.A. # Technological Characteristics The technological characteristics of the S-scan systems with the addition of the Shoulder Coil 20 option, reflected in this Special 510(k), do not alter the scientific technology of the S-scan system and are substantially equivalent to those of the predicate devices. The Shoulder Coil 20 has been designed in order to have bigger dimension (to allow an increased number of possible patient examinations and an extra receiving channel) and increased image quality than the 2Ch Shoulder Coil 7 (already cleared for the S-scan system). The 3Ch Shoulder Coil and 2Ch Shoulder Coil have a similar shape, can be tilted on their base to allow left or right shoulder examination and both are embraced by the patient during examination. # Summary of Non-Clinical Tests The devices have been evaluated for biocompatibility, electrical, electromagnetic, and mechanical safety, and have been found to conform to the following medical device safety standards. - IEC 60601-1 ● - IEC 60601-1-2 ● - IEC 60601-1-6 ● - IEC 60601-2-33 ● - IEC62304 ● - IEC62366 ● - ISO10993-1 ● - ISO 14971-2 ● - NEMA MS-1 Determination of Signal-to-Noise Ratio (SNR) in Diagnostic ● Magnetic Resonance Imaging - NEMA MS-3 Determination of Image Uniformity in Diagnostic Magnetic ● Resonance Images # Summary of Clinical Tests No clinical tests were performed. # Conclusion The changes proposed to the S-scan System are substantially equivalent to the legally marketed devices and conform to applicable medical device safety and performance standards.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...