← Product Code [MUD](/productcode/MUD) · K161237

# HyperView Hyperspectral Tissue Oxygenation Measurement System (K161237)

_Hypermed Imaging, Inc. · MUD · Dec 16, 2016 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K161237

## Device Facts

- **Applicant:** Hypermed Imaging, Inc.
- **Product Code:** [MUD](/productcode/MUD.md)
- **Decision Date:** Dec 16, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2700
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The HyperView™ Hyperspectral Tissue Oxygenation Measurement System is intended for use by physicians/healthcare professionals as a noninvasive tissue oxygenation measurement system that reports an approximate value of: - oxygen saturation (O2Sat), - oxyhemoglobin level (Oxy), and - deoxyhemoglobin (Deoxy) level in superficial tissue. The HyperView™ system displays two-dimensional, color-coded images of tissue oxygenation of the scanned surface and reports hyperspectral tissue oxygenation measurements for selected tissue regions. The HyperView™ system is indicated for use to determine oxygenation levels in superficial tissues for patients with potential circulatory compromise.

## Device Story

Handheld hyperspectral imaging system; captures images of superficial tissue using visible light (400-700nm) via flash illumination. Device analyzes light absorption signatures at 8 specific wavelengths to calculate oxyhemoglobin, deoxyhemoglobin, and oxygen saturation levels. Operates via resistive touchscreen and image capture button; requires no patient contact or contrast agents. Produces 2D color-coded maps and quantitative measurements. Used by clinicians to assess vascular viability and wound healing potential. Data stored in DICOM format. Benefits include noninvasive, rapid assessment of tissue perfusion in clinical settings.

## Clinical Evidence

Prospective validation study (n=20) comparing HyperView to OxyVu™-1. Primary endpoint: mean O2Sat difference between devices. Results showed mean O2Sat difference of 3.22% ± 3.38%, within the pre-specified ±7% equivalence margin. Secondary endpoints (Oxy, Deoxy) showed mean differences of 5.5% and 0.8% respectively. Stratification by skin tone, ulcer presence, and peripheral artery disease (PAD) confirmed consistent performance across subgroups. No statistically significant differences in O2Sat measurements were noted between PAD and non-PAD limbs.

## Technological Characteristics

Handheld device; visible light (400-700nm) illumination using filtered CREE LEDs. Imaging sensor, thermal sensor, and flash illumination. Connectivity via USB and docking station. Software-based spectral analysis of 8 wavelengths. Powered by internal battery. Electrical safety per ANSI/AAMI ES60601-1; EMC per IEC 60601-1-2; Laser safety per IEC 60825-1; Photobiological safety per IEC 62471.

## Regulatory Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

## Predicate Devices

- OxyVu™-1 ([K073656](/device/K073656.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{1}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three human profiles forming a wing-like shape. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 16, 2016

Hypermed Imaging, Inc. % Julie Powell Vice President, Quality Assurance Emergo Global Consulting, LLC 816 Congress Avenue, Suite 1400 Austin, Texas 78701

Re: K161237

Trade/Device Name: Hyperspectral Tissue Oxygenation Measurement System Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: MUD Dated: November 11, 2016 Received: November 14, 2016

Dear Julie Powell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{2}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Mark Jellison
for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

#### 510(k) Number (if known) K161237

#### Device Name

HyperView™ Hyperspectral Tissue Oxygenation Measurement System

#### Indications for Use (Describe)

The HyperView™ Hyperspectral Tissue Oxygenation Measurement System is intended for use by physicians/healthcare professionals as a noninvasive tissue oxygenation measurement system that reports an approximate value of.

- · oxygen saturation (O2Sat),
- · oxyhemoglobin level (Oxy), and
- deoxyhemoglobin (Deoxy) level

in superficial tissue. The HyperView™M system displays two-dimensional, color-coded images of tissue oxygenation of the scanned surface and reports hyperspectral tissue oxygenation measurements for selected tissue regions.

The HyperView™ system is indicated for use to determine oxygenation levels in superficial tissues for patients with potential circulatory compromise.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# Section 5 – 510(k) Summary

# Hyperview Hyperspectral Tissue Oxygenation Measurement System

# K161237

#### 1. Submission Sponsor

HyperMed Imaging, Inc.

8700 Trail Lake Drive West, Suite 200

Memphis, TN 38125

USA

Office number: (901) 453-3154

Contact: Carlyn Reynolds

Title: Vice President Regulatory Compliance & Quality Assurance

#### 2. Submission Correspondent

Emergo Global Consulting, LLC

816 Congress Avenue, Suite 1400

Austin, TX 78701

Cell Phone: (269) 720-1329

Office Phone: (512) 327-9997

Contact: Julie Powell, Vice President, QA/RA

Email: project.management@emergogroup.com

#### 3. Date Prepared

April 29, 2016

{5}------------------------------------------------

#### 4. Device Identification

| Trade/Proprietary Name: | HyperView™ Hyperspectral Tissue Oxygenation Measurement System |
|-------------------------|----------------------------------------------------------------|
| Common/Usual Name:      | Tissue Saturation Oximeter                                     |
| Classification Name:    | Oximeter                                                       |
| Regulation Number:      | 21 CFR 870.2700                                                |
| Product Code:           | MUD, Oximeter                                                  |
| Device Class:           | Class II                                                       |
| Classification Panel:   | Cardiovascular                                                 |

#### 5. Legally Marketed Predicate Device(s)

OxyVu™-1, 510(k) K073656, HyperMed, Inc.

#### 6. Device Description

HyperMed Imaging Inc.'s HyperView handheld hyperspectral imaging system consists of the HyperView Camera and the HyperView Accessories Kit that includes a Docking Station, Calibration Verification System, Power Cord and USB Cable Set.

The HyperView system provides noninvasive measurement of approximate tissue oxygenation, sensitive to local, regional, and systemic low/no-flow ischemia. HyperView uses multi-wavelength imaging technology to analyze the signature of light as it is absorbed by the patient's tissue. The HyperView system analyzes these light absorption signatures to measure oxyhemoglobin and deoxyhemoglobin levels in the tissue, and creates a color-coded image that illustrates relative concentrations of oxygenated hemoglobin and deoxygenated hemoglobin.

Quantitative data is provided for oxyhemoglobin and deoxyhemoglobin levels as well as oxygen saturation. The technology is suitable for surface assessment, as the visible spectrum light used penetrates no more than a few millimeters deep.

The HyperView is a handheld product consisting of several subassemblies including precision optical components, electro-optics, imaging sensor, a thermal sensor, flash illumination, flex circuits, printed circuit boards, a battery, plastics as well as image processing and graphics user interface software. Users will interact with the device using a built-in, resistive, touchscreen display and an image capture button similar to a digital camera. The product does not come in contact with the patient and no injectable contrast is required. The camera uses visible light to momentarily illuminate the patient's skin for photographic imaging. Such information is useful in determining viability of vascular delivery in the superficial skin and correspondingly, such information can assist the physician in determining the ability of the patient's vascular system to support normal tissue health and wound healing.

{6}------------------------------------------------

#### 7. Indication for Use Statement

The HyperView™ Hyperspectral Tissue Oxygenation Measurement System is intended for use by physicians/healthcare professionals as a noninvasive tissue oxygenation measurement system that reports an approximate value of:

- oxygen saturation (O2Sat),
- oxyhemoglobin level (Oxy), and
- deoxyhemoglobin (Deoxy) level

in superficial tissue. The HyperView™ system displays two-dimensional, color-coded images of tissue oxygenation of the scanned surface and reports hyperspectral tissue oxygenation measurements for selected tissue regions.

The HyperView™ system is indicated for use to determine oxygenation levels in superficial tissues for patients with potential circulatory compromise.

#### 8. Substantial Equivalence Discussion

The following table compares the HyperView system to the predicate device with respect to indications for use, principles of operation, technological characteristics, materials, and performance testing. The comparison of the devices provides more detailed information regarding the basis for the determination of substantial equivalence. The subject device does not raise any new issues of safety or effectiveness based on the similarities to the predicate device.

|                     | SUBJECT DEVICE                                                                                                                                                                                                                                                                                                                  | PREDICATE                                                                                                                                                                                                                                                                                                                       | Significant<br>Differences |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|
| Manufacturer        | HyperMed Imaging, Inc.                                                                                                                                                                                                                                                                                                          | HyperMed, Inc.                                                                                                                                                                                                                                                                                                                  |                            |
| Trade Name          | HyperView™ System                                                                                                                                                                                                                                                                                                               | OxyVu™-1                                                                                                                                                                                                                                                                                                                        |                            |
| Product Code        | MUD                                                                                                                                                                                                                                                                                                                             | MUD                                                                                                                                                                                                                                                                                                                             | Same                       |
| Regulation Number   | 21 CFR 870.2700                                                                                                                                                                                                                                                                                                                 | 21 CFR 870.2700                                                                                                                                                                                                                                                                                                                 | Same                       |
| Regulation Name     | Tissue Saturation Oximeter                                                                                                                                                                                                                                                                                                      | Tissue Saturation Oximeter                                                                                                                                                                                                                                                                                                      | Same                       |
| Indications for Use | The HyperView™<br>Hyperspectral Tissue<br>Oxygenation<br>Measurement System is<br>intended for use by<br>physicians/healthcare<br>professionals as a<br>noninvasive tissue<br>oxygenation measurement<br>system that reports an<br>approximate value of:<br>• oxygen saturation<br>(O2Sat), • oxyhemoglobin<br>level (Oxy), and | The OxyVu™-1<br>Hyperspectral Tissue<br>Oxygenation (HTO)<br>Measurement System<br>is intended for use by<br>healthcare professionals as<br>a noninvasive tissue<br>oxygenation measurement<br>system that reports an<br>approximate value of:<br>• oxygen saturation (HT-<br>Sat),<br>• oxyhemoglobin level (HT -<br>Oxy), and | Same                       |

{7}------------------------------------------------

K161237

|                   | SUBJECT DEVICE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | PREDICATE                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Significant                                                                                                                                                                                                                                                                                                                                                                                                                 |
|-------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer      | HyperMed Imaging, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | HyperMed, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Differences                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Trade Name        | HyperViewTM System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | OxyVuTM-1                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                             |
|                   | • deoxyhemoglobin<br>(Deoxy) level<br>in superficial tissue. The<br>HyperViewTM system<br>displays two-dimensional,<br>color-coded images of<br>tissue oxygenation of the<br>scanned surface and<br>reports hyperspectral<br>tissue oxygenation<br>measurements for<br>selected tissue regions.<br>The HyperViewTM system is<br>indicated for use to<br>determine oxygenation<br>levels in superficial tissues<br>for patients with potential<br>circulatory compromise.                                                                                                     | • deoxyhemoglobin (HT-<br>Deoxy) level<br>in superficial tissue. The<br>OxyVuTM-1 system displays<br>two-dimensional color<br>coded<br>images of tissue<br>oxygenation of the scanned<br>surface and reports<br>hyperspectral tissue<br>oxygenation measurements<br>for selected tissue regions.<br>The OxyVuTM-1 system is<br>indicated for use to<br>determine oxygenation<br>levels in superficial tissues<br>for patients with potential<br>circulatory compromise. |                                                                                                                                                                                                                                                                                                                                                                                                                             |
| System Components | • Hand-held, battery<br>operated device<br>including built in<br>resistive touchscreen<br>control and user<br>interface software<br>• PC software application<br>for reports<br>• Desktop docking station<br>to provide battery<br>charging and USB to<br>computer data interface<br>• USB computer interface<br>• Optionally tripod<br>mountable<br>• Calibration verification<br>cards are only required<br>to be used by user for<br>periodic verification of<br>calibration<br>• Laser verification cards<br>are used for periodic<br>verification of laser<br>alignment | • System Console: Cart,<br>System Electronics, CPU,<br>Monitor, Keyboard,<br>Pointing Device and<br>Printer (AC powered)<br>• Hyperspectral imaging<br>instrument head with<br>support arm: broadband<br>illuminator, camera and<br>spectral filter for<br>collecting hyperspectral<br>imaging cube<br>• Single use OxyVuTM Check<br>Pads for calibration                                                                                                               | Different; the update of<br>the system from the<br>cart-based unit to a<br>hand-held device<br>involved changes to the<br>product dimensions,<br>user interface, power<br>source, and packaging.<br>The change from cart-<br>based unit to hand-held<br>unit had no effect on<br>performance. There<br>were no additional<br>safety or effectiveness<br>concerns as the<br>hyperspectral<br>technology used is the<br>same. |
| Calibration       | Unit is factory calibrated                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Calibration is required                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Similar; both require                                                                                                                                                                                                                                                                                                                                                                                                       |

{8}------------------------------------------------

|                                                               | SUBJECT DEVICE                                          | PREDICATE                                     | Significant                                                                                                                                                                                                                                                                                      |
|---------------------------------------------------------------|---------------------------------------------------------|-----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer                                                  | HyperMed Imaging, Inc.                                  | HyperMed, Inc.                                | Differences                                                                                                                                                                                                                                                                                      |
| Trade Name                                                    | HyperView™ System                                       | OxyVu™ -1                                     |                                                                                                                                                                                                                                                                                                  |
|                                                               | and does not require<br>calibration before each<br>use. | before each use                               | calibration.<br>The change in<br>calibration frequency<br>had no effect on<br>performance. The<br>change in calibration<br>frequency does not<br>present any additional<br>safety or effectiveness<br>concerns.                                                                                  |
| Time per patient image                                        | < 1 second capture time<br>5 — 10 minutes use           | 3 minute capture time<br>10 to 15 minutes use | Similar method to<br>capture images.<br>Multiple images are<br>captured simultaneously<br>to reduce imaging time.<br>The change in image<br>time had no effect on<br>performance. The<br>patient image time<br>change does not present<br>any additional safety or<br>effectiveness concerns.    |
| Spectrum of Light Used                                        | Visible Spectrum Light<br>(400nm to 700nm)              | Visible Spectrum Light<br>(400nm to 700nm)    | Same                                                                                                                                                                                                                                                                                             |
| Wavelengths of Light<br>Used for Data Analysis<br>and Imaging | 8 Wavelengths                                           | 15 Wavelengths                                | Similar; subject device<br>uses a subset of 8 of the<br>15 wavelengths used by<br>the predicate.<br>The change in number of<br>wavelengths had no<br>effect on performance.<br>The change in number of<br>wavelengths does not<br>present any additional<br>safety or effectiveness<br>concerns. |
| LEDs                                                          | CREE LEDs (filtered)                                    | CREE LEDs (3.0 mW/cm²)                        | Similar; both use LEDs.                                                                                                                                                                                                                                                                          |

{9}------------------------------------------------

|                                  | SUBJECT DEVICE                                                                                               | PREDICATE                                                                                         | Significant<br>Differences                                                                                                                                                                                                     |
|----------------------------------|--------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer                     | HyperMed Imaging, Inc.                                                                                       | HyperMed, Inc.                                                                                    |                                                                                                                                                                                                                                |
| Trade Name                       | HyperViewTM System                                                                                           | OxyVuTM-1                                                                                         |                                                                                                                                                                                                                                |
|                                  | Illuminator 1: 11.0<br>mW/cm2<br>Illuminator 2: 6.1 mW/cm2                                                   | (unfiltered)                                                                                      | The change in LEDs and<br>the filtering of their<br>output had no effect on<br>performance. The LED<br>change does not present<br>any additional safety or<br>effectiveness concerns.                                          |
| LED/Laser<br>Classification      | Class II Laser                                                                                               | Class I LED                                                                                       | Different; LED/Laser<br>classification changed<br>from Class I to Class II.<br>The classification<br>difference had no effect<br>on performance and<br>does not present any<br>additional safety or<br>effectiveness concerns. |
| Software                         | C, C++, Java and .net (for PC<br>application)                                                                | Matlab and Java                                                                                   | Similar as both use<br>software.<br>The change in<br>programming languages<br>had no effect on<br>performance. This<br>change does not present<br>any additional safety or<br>effectiveness concerns.                          |
| Region of Interest<br>(ROI) Area | Support for freehand<br>drawing ROI selection<br>(mm2)                                                       | Support for freehand<br>drawing ROI selection<br>(mm2)                                            | Same                                                                                                                                                                                                                           |
| Image Type                       | Diagnostic                                                                                                   | Diagnostic                                                                                        | Same                                                                                                                                                                                                                           |
| Measures                         | Oxygen saturation<br>Oxyhemoglobin level<br>Deoxyhemoglobin level                                            | Oxygen saturation<br>Oxyhemoglobin level<br>Deoxyhemoglobin level                                 | Same                                                                                                                                                                                                                           |
| Method of<br>Measurement         | Spectral analysis at specific<br>wavelengths of light<br>returned from target (8<br>wavelengths, which are a | Spectral analysis at specific<br>wavelengths of light<br>returned from target (15<br>wavelengths) | Similar; as both use<br>wavelengths of light for<br>measurement.<br>The wavelengths for the                                                                                                                                    |

{10}------------------------------------------------

|                  | SUBJECT DEVICE                                                                                                                                                                                                                                 | PREDICATE                                                                                                                                                                                                                                        | Significant<br>Differences                                                                                                                                                                                                                                                                                     |
|------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer     | HyperMed Imaging, Inc.                                                                                                                                                                                                                         | HyperMed, Inc.                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                |
| Trade Name       | HyperViewTM System<br>wavelengths)                                                                                                                                                                                                             | OxyVuTM-1                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                |
|                  |                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                  | hand-held subject device<br>are a subset of the<br>wavelengths used in the<br>predicate device and the<br>change from 15 to 8<br>wavelengths had no<br>effect on performance.<br>This change does not<br>present any additional<br>safety or effectiveness<br>concerns.                                        |
| Output Display   | Numeric<br>Two-dimensional color<br>map of approximate tissue<br>oxygenation                                                                                                                                                                   | Numeric<br>Two-dimensional color map<br>of approximate tissue<br>oxygenation                                                                                                                                                                     | Same                                                                                                                                                                                                                                                                                                           |
| Graphical Images | • VIS - Visible<br>• Sat – Saturation Map<br>(new map based upon<br>the oxygen saturation<br>level)<br>• Oxy - Tissue<br>oxygenation map<br>• Deoxy - Tissue<br>deoxyhemoglobin map<br>• Note: The HSI map<br>replaced with Saturation<br>Map. | • VIS - Visible<br>• HSI - Tissue oxygenation<br>map of oxyhemoglobin<br>and deoxyhemoglobin<br>• Oxy - Tissue oxygenation<br>map<br>• Deoxy - Tissue<br>deoxyhemoglobin map<br>Note: The Oxygen<br>Saturation value displayed<br>without a map. | Similar; both use same<br>graphical images for VIS,<br>Oxy, Deoxy. The HSI<br>map was replaced with<br>the Saturation Map.<br>The change from an HSI<br>to an oxygen saturation<br>map had no effect on<br>performance. This<br>change does not present<br>any additional safety or<br>effectiveness concerns. |
| Selection Tools  | Circle<br>Region<br>Perimeter with optional<br>second region                                                                                                                                                                                   | Circle<br>Region<br>Perimeter                                                                                                                                                                                                                    | Similar; both devices use<br>the same Selection<br>Tools, with added<br>feature for the<br>Perimeter Tool.<br>The additional feature in<br>the Perimeter Selection<br>Tool had no effect on<br>how the device<br>performed. This change<br>does not present any<br>additional safety or                        |

{11}------------------------------------------------

K161237 p.8/12

|                                     | SUBJECT DEVICE                                                    | PREDICATE                           | Significant<br>Differences                                                                                                                                                                                                                                    |
|-------------------------------------|-------------------------------------------------------------------|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer                        | HyperMed Imaging, Inc.                                            | HyperMed, Inc.                      |                                                                                                                                                                                                                                                               |
| Trade Name                          | HyperViewTM System                                                | OxyVuTM-1                           | effectiveness concerns.                                                                                                                                                                                                                                       |
| Data Storage                        | DICOM images stored on device                                     | Images stored in Matlab files       | Similar; as both have data storage capabilities.<br>The change in data storage type (to an industry standard format) had no effect on performance of the device. This change does not present any additional safety or effectiveness concerns.                |
| Operating Distance<br>from Patient  | Approximately 15"                                                 | Approximately 17"                   | Similar; both products have approximately the same working distance from the patient.<br>The change in operating distance from the patient had no effect on device performance. This change does not present any additional safety or effectiveness concerns. |
| Imaged Area                         | 100 mm x 100 mm                                                   | 80 mm x 80 mm                       | Similar; imaged area is increased with hand-held subject device.<br>The increase in image area had no effect on performance. This change does not present any additional safety or effectiveness concerns.                                                    |
| Electrical Safety<br>Testing Passed | Yes<br>ANSI/AAMI ES60601-<br>1:2005/(R)2012<br>IEC 60601-1-2:2007 | Yes<br>IEC 60601-1<br>IEC 60601-1-2 | Same                                                                                                                                                                                                                                                          |

{12}------------------------------------------------

K161237 p.9/12

#### 9. Non-Clinical Performance Data

As part of demonstrating safety and effectiveness of the HyperView system, and in showing substantial equivalence to the predicate device of this 510(k) submission, HyperMed a number of tests. The HyperView system meets all requirements for design characteristics, non-clinical performance testing, EMC/EMI testing, electrical safety, laser safety and battery safety testing to confirm that the output meets the design inputs and specifications for the device.

The HyperView system passed all testing in accordance with internal requirements, national standards, and international standards shown below to support substantial equivalence of the subject device:

- . Software verification and validation testing per FDA Guidance and IEC 62304: conformance of software development life cycle for the HyperView System and compliance to the requirements of the FDA guidance document for software contained in a medical device.
- Electrical safety testing per ANSI/AAMI ES60601-1: PASS
- . Electromagnetic compatibility testing per IEC 60601-1-2: PASS
- Laser safety testing per IEC 60825-1:2007: PASS
- . Photobiological safety of lamps and lamp systems per IEC 62471:2006: PASS
- . Secondary cells and batteries safety testing per IEC 62133:2012 : PASS
- . Transportation safety testing of Lithium Batteries per UN 38.3:2009: PASS
- Transit testing per ASTM D4169:2014: PASS

The HyperView passed all the testing in accordance with internal requirements, national standards, and international standards shown above to support substantial equivalence of the subject device.

#### 10. Clinical Performance Data

A validation study was performed with the HyperView system for assessing the hyperspectral imaging in persons with vascular disease and / or diabetes mellitus and potentially compromised tissue oxygenation. The study was conducted to evaluate the HyperView handheld system with the predicate OxyVu™-1 console system device. The study was performed in compliance with Good Clinical Practices (GCP) with subjects enrolled in an IRB approved study that were consented for participation according to the intended use of the device, defined inclusion criteria, and defined exclusion criteria; with the purpose of meeting the study objectives.

The study objectives were defined as the following:

{13}------------------------------------------------

#### p.10/12

- Comparison of hyperspectral imaging tissue oxygenation as the difference in the mean tissue . oxygen saturation (O2Sat) measurements of the OxyVu™-1 compared to the HyperView handheld system.
- . Secondary endpoints for this study include testing of mean oxyhemoglobin (Oxy) and deoxyhemoglobin (Deoxy) levels with stratification by the following:
	- Skin pigmentation and melanin concentration levels O
	- Non-ulcerated and ulcerated areas O
	- Degree of peripheral artery disease in the study limb o

The clinical study was planned for a minimum of 20 (twenty) subjects to be enrolled, which was met. The subjects had images taken of target areas with both the HyperView handheld system and the predicate device, during a single session. No follow up visits were required.

The outcomes measurements for the clinical study are summarized as follows:

#### Primary Efficacy Endpoint Results

The pre-specified primary efficacy endpoint for the validation study was to compare the O2Sat measurement difference between the investigational HyperView and predicate OxyVu™-1 device ((HyperView O2Sat minus OxyVuTM-1 O2Sat).

A Student's t-tests was performed using SAS version 9.4 (SAS Institute Inc., Cary, NC) to compare the mean differences with non-adjusted results. A difference that is not greater than 7% nor less than -7% between the mean O2Sat values for each device was deemed substantially equivalent.

A mean O2Sat difference of 3.22 +/- 3.38% was demonstrated between the HyperView and the OxyVu™-1 devices in this study. Lower confidence intervals and upper confidence intervals were 2.58% and 3.87% respectively.

A second analysis was performed comparing the study limbs by image sequence. The conclusions remained the same. The mean differences in O2Sat between the devices were 3.17%, 3.0% and 3.5% respectively.

#### Secondary Efficacy Endpoint Results

Secondary objectives entailed assessing O2Sat, Oxy, and Deoxy differences stratified by skin tone, presence of ulcers, and an ABI indicative of peripheral artery disease.

#### Stratification by Skin Tone:

The mean difference was 2.42% for the light skin tone and 4.64% for medium skin tone. While the means of these two groups were statistically different (P>.0002), the mean values and confidence intervals were all within a 5% difference between the two cameras.

An analysis on the index forearm region was also performed, as all three skin tone groups were represented for this location (light, medium, and dark). In this location, there were no

{14}------------------------------------------------

statistically significant differences noted between the skin tone groups (P>0.1968). The mean differences were 4% or less for all three groups.

#### Stratification by Non-ulcerated and Ulcerate Study Limbs

Eleven study limbs (31%) had a foot ulcer present. Each study limb had 3 sets of paired images for a total of 33 paired data sets. The mean difference in O2Sat between the devices was 1.91% with lower and upper confidence intervals of 0.55% and 3.26% in study limbs with a foot ulcer and 3.8% when ulcers were not present. The mean O2Sat difference was significantly smaller between the devices when ulcers were present on the foot (P>0.0068).

#### Stratification by Peripheral Artery Disease (PAD) of the Study Limb

There were 15 study limbs that met this definition. The mean difference in O2Sat for this group was 2.6% with lower and upper confidence intervals of 1.14% and 4.06% respectively.

There were no statistically significant differences noted between the device measurements of O2Sat between study limbs with PAD compared to study limbs without PAD (P>0.4449).

### Oxy Differences

An assessment was made of the difference in Oxy for paired images. Oxy values are on a numeric scale of 0 to 255. A mean Oxy difference of 14.07 (5.5%) was observed between the two devices.

Oxy differences by skin tone on the plantar surface of the foot demonstrated mean difference of 11.94 (4.7%) in light skin toned feet and 17.85 (7.0%) in medium skin toned feet.

Oxy differences by skin tone on the forearm demonstrated a mean difference of 12.62 (4.9%) in light toned skin, 13.64 (5.3%) in medium toned skin, and 17.5 (6.8%) in dark toned skin.

Differences in Oxy between the two devices with an ulcer present was 11.3 (4.4%).

Difference in Oxy between for subjects with peripheral artery disease was 12.27 (4.8%).

## Deoxy Differences

An assessment was made of the difference in Deoxy for paired images. Deoxy values are on a numeric scale of 0 to 255. A mean difference of 1.95 (0.8%) was observed.

Deoxy differences by skin tone on the plantar surface of the foot demonstrated mean difference of 2.9 (1.1%) in light skin toned feet and 0.28 (0.1%) in medium skin toned feet.

Deoxy differences by skin tone of the forearm demonstrated a mean difference of 2.47 (1.0%) in light toned skin, 1.1 (0.4%) in medium toned skin, and 2.38 (0.9%) in dark toned skin.

Deoxy values between the two devices for study limbs with ulcerated tissue (challenged tissue) demonstrated a mean difference of 3.09 (1.2%).

The difference in Deoxy between the two devices in study limbs with PAD was 3.27 (1.3%).

The validation study concluded that the HyperView system is safe and effective for its intended use and the outcomes of the study met the stated objectives. The validation study supports the indication for use as a noninvasive tissue oxygenation measurement system that reports approximate values of

{15}------------------------------------------------

#### K161237

## p.12/12

oxygen saturation (O2Sat), oxyhemoglobin level (Oxy), and deoxyhemoglobin (Deoxy) level in superficial tissue.

## 11. Statement of Substantial Equivalence

By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device or the device has the same intended use and different technological characteristics provided it can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise additional questions regarding its safety and effectiveness as compared to the predicate device.

The HyperView™ Hyperspectral Tissue Oxygenation Measurement System, as designed and manufactured, is determined to be substantially equivalent to the referenced predicate device.

---

**Source:** [https://fda.innolitics.com/device/K161237](https://fda.innolitics.com/device/K161237)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
