K161224 · GE Medical Systems Ultrasound & Primary Care Diagnostics, LLC · IYN · Aug 5, 2016 · Radiology
Device Facts
Record ID
K161224
Device Name
LOGIQ V5 Expert, LOGIQ V5, LOGIQ V3
Applicant
GE Medical Systems Ultrasound & Primary Care Diagnostics, LLC
Product Code
IYN · Radiology
Decision Date
Aug 5, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The LOGIQ V5 Expert / LOGIQ V5 / LOGIQ V3 are general purposed ultrasound imaging and analysis systems providing digital acquisition, processing and display capability and clinical applications including: Abdominal, Obstetrical, Gynecological, Small parts, Vascular, Adult Cephalic, Pediatric, Musculoskeletal, Transcranial, Neonatal Cephalic, Transvaginal, Cardiac, Transrectal, and Tissue Biopsy/Fluid Drainage,
Device Story
The LOGIQ V5 Expert, LOGIQ V5, and LOGIQ V3 are mobile, entry-level ultrasound consoles for private clinics. They acquire, process, and display digital ultrasound images using various linear, curved, and matrix phased array transducers. Operated by clinicians, the systems support B-mode, M-mode, Doppler (PW/CW/Color/Power), and 3D/4D imaging. The device provides real-time visualization for general radiology, cardiology, vascular applications, and interventional guidance (biopsy/fluid drainage). Output is displayed on an integrated monitor, with options for digital storage, printing, and wireless connectivity. The system assists healthcare providers in diagnostic evaluation and procedural guidance, potentially improving clinical decision-making and patient outcomes through non-invasive imaging.
Clinical Evidence
No clinical studies were required to support substantial equivalence. Evidence is based on bench testing, including acoustic output, biocompatibility, cleaning/disinfection effectiveness, and thermal, electrical, electromagnetic, and mechanical safety testing in accordance with recognized standards (e.g., AAMI/ANSI ES 60601-1, IEC 60601-1-2, ISO 10993-1).
Technological Characteristics
Mobile console with keyboard control panel, color display, and wireless connectivity. Utilizes linear, curved, and matrix phased array transducers. Supports B, M, PW/CW Doppler, Color/Power Doppler, and 3D/4D imaging. Complies with NEMA UD 2/UD 3 for acoustic output and DICOM standards for connectivity. Biocompatible patient-contact materials per ISO 10993-1.
Indications for Use
Indicated for general purpose ultrasound imaging and analysis in patients requiring abdominal, obstetrical, gynecological, small parts, vascular, adult/neonatal cephalic, pediatric, musculoskeletal, transcranial, cardiac, and transrectal exams, as well as tissue biopsy and fluid drainage guidance.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the smallest and the bottom profile being the largest.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 5, 2016
GE Medical Systems Ultrasound and Primary Care Diagnostics LLC % Ms. Tracey Ortiz Regulatory Affairs Director 9900 W. Innovation Drive WAUWATOSA WI 53226
Re: K161224
Trade/Device Name: LOGIQ V5 Expert, LOGIQ V5, LOGIQ V3 Regulation Number: 21CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: June 30, 2016 Received: July 5, 2016
Dear Ms. Ortiz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K161224
Device Name LOGIQ V5 Expert / LOGIQ V5 / LOGIQ V3
#### Indications for Use (Describe)
The LOGIQ V5 Expert / LOGIQ V5 / LOGIQ V3 are general purposed ultrasound imaging and analysis systems providing digital acquisition, processing and display capability and clinical applications including: Abdominal, Obstetrical, Gynecological, Small parts, Vascular, Adult Cephalic, Pediatric, Musculoskeletal, Transcranial, Neonatal Cephalic, Transvaginal, Cardiac, Transrectal, and Tissue Biopsy/Fluid Drainage,
Type of Use (Select one or both, as applicable)
| <span style="font-size:10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|-------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size:10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a blue circle. The letters are also blue, matching the color of the circle.
#### Diagnostic Ultrasound Indications for Use Form LOGIQ V5 Expert / LOGIQ V5 / LOGIQ V3 Ultrasound System
| | Mode of Operation | | | | | | | | | | |
|-------------------------------------|-------------------|---|---------------|----|-------|---------|-------|--------------------|---------------------|-----------------|-------|
| Clinical Application | B | M | Doppler Modes | | | | | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse• | Other |
| Anatomy/Region of Interest | | | PW | CW | Color | Color M | Power | | | | |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics[1] | P | P | P | P | P | | P | P | P | P | [5] |
| Abdominal[1] | P | P | P | P | P | | P | P | P | P | [5] |
| Pediatric | P | P | P | P | P | P | P | P | P | P | |
| Small Organ [2] | P | P | P | P | P | | P | P | P | P | |
| Neonatal Cephalic | P | P | P | P | P | | P | P | P | P | |
| Adult Cephalic | P | P | P | P | P | | P | P | P | P | |
| Cardiac [3] | P | P | P | P | P | P | P | P | P | P | |
| Peripheral Vascular | P | P | P | P | P | | P | P | P | P | |
| Musculo-skeletal Conventional | P | P | P | P | P | | P | P | P | P | |
| Musculo-skeletal Superficial | P | P | P | P | P | | P | P | P | P | |
| Thoracic/Pleural | | | | | | | | | | | |
| Other | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transcranial | P | P | P | P | P | | P | P | P | P | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | N | N | N | | N | | N | N | N | | |
| Transvaginal | P | P | P | | P | | P | P | P | | |
| Intraoperative | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | |
| Tissue Biopsy/Fluid<br>Drainage [4] | N | N | N | N | N | | N | N | N | | [5] |
| Vascular Access (IV, PICC) | | | | | | | | | | | |
| Nerve Block | | | | | | | | | | | |
Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:
N = new indication; P= previously cleared by FDA K141768; P2= previously cleared by FDA K160277;
#### P3= previously cleared by FDA K151028
Notes: [1] Abdominal includes GYN and Urological/Prostate;
- [2] Small Organ includes breast, testes, thyroid;
[3] Cardiac includes Adult and Pediatric;
- [4] Interventional Guidance Tissue Biopsy is 2D biopsy guide;
- [5] 3D/4D imaging Mode;
- [*]Combined modes are B/M, B/ Color M, B/PWD, B/Color/PWD, B/Power/PWD
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular frame. The frame is decorated with ornamental flourishes, giving it a classic and elegant appearance. The logo is colored in blue, set against a white background.
## Diagnostic Ultrasound Indications for Use Form LOGIQ V5 Expert / LOGIQ V5 / LOGIQ V3 with 4C-RS Transducer
| | Mode of Operation | | | | | | | | | | | |
|------------------------------------|-------------------|----|---------------|----|-------|---------|-------|---------------------|---------------------|-----------------|-------|--|
| Clinical Application | B | M | Doppler Modes | | | | | Combined<br>Modes * | Harmonic<br>Imaging | Coded<br>Pulse` | Other | |
| | | | PW | CW | Color | Color M | Power | | | | | |
| Anatomy/Region of Interest | | | | | | | | | | | | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics | P | P | P | P | P | P | P | P | P | P | | |
| Abdominal[1] | P | P | P | P | P | P | P | P | P | P | | |
| Pediatric | P2 | P2 | P2 | P2 | P2 | P2 | P2 | P2 | P2 | P2 | | |
| Small Organ [2] | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | | |
| Peripheral Vascular | P | P | P | P | P | P | P | P | P | P | | |
| Musculo-skeletal Conventional | P2 | P2 | P2 | P2 | P2 | P2 | P2 | P2 | P2 | P2 | | |
| Musculo-skeletal Superficial | | | | | | | | | | | | |
| Thoracic/Pleural | | | | | | | | | | | | |
| Other | | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transcranial | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | | |
| Tissue Biopsy/Fluid<br>Drainage[4] | P2 | P2 | P2 | P2 | P2 | P2 | P2 | P2 | P2 | P2 | | |
| Vascular Access (IV, PICC) | | | | | | | | | | | | |
| Nerve Block | | | | | | | | | | | | |
| Other | | | | | | | | | | | | |
Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:
N = new indication; P= previously cleared by FDA 141768; P2= previously cleared by FDA K160277;
P3= previously cleared by FDA K151028
Notes: [1] Abdominal includes GYN and Urological/Prostate;
[2] Small Organ includes breast, testes, thyroid;
[3] Cardiac includes Adult and Pediatric;
[4] Interventional Guidance Tissue Biopsy is 2D biopsy guide;
[5] 3D/4D imaging Mode;
[*]Combined modes are B/M, B/ Color M, B/PWD, B/Color/PWD, B/Power/PWD
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a circular frame. The color of the letters and the frame is blue, and the background is white. The logo is simple and recognizable, representing the brand identity of General Electric.
## Diagnostic Ultrasound Indications for Use Form LOGIQ V5 Expert / LOGIQ V5 / LOGIQ V3 with 8C-RS Transducer
| | Mode of Operation | | | | | | | | | | |
|------------------------------------|-------------------|----|---------------|----|-------|--------------------|---------------------|----------------|-------|----|--|
| Clinical Application | B | M | Doppler Modes | | | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other | | |
| | | | PW | CW | Color | Color M | Power | | | | |
| Anatomy/Region of Interest | | | | | | | | | | | |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | |
| Abdominal[1] | | | | | | | | | | | |
| Pediatric | p3 | p3 | p3 | | | p3 | | p3 | p3 | p3 | |
| Small Organ [2] | | | | | | | | | | | |
| Neonatal Cephalic | p3 | p3 | p3 | | | p3 | | p3 | p3 | p3 | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac[3] | p3 | p3 | p3 | | | p3 | | p3 | p3 | p3 | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal Conventional | p3 | p3 | p3 | | | p3 | | p3 | p3 | p3 | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Thoracic/Pleural | | | | | | | | | | | |
| Other | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transcranial | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | |
| Tissue Biopsy/Fluid<br>Drainage[4] | | | | | | | | | | | |
| Vascular Access (IV, PICC) | | | | | | | | | | | |
| Nerve Block | | | | | | | | | | | |
Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:
N = new indication; P= previously cleared by FDA K141768; P2= previously cleared by FDA K160277;
P3= previously cleared by FDA K151028
Notes: [1] Abdominal includes GYN and Urological/Prostate;
[2] Small Organ includes breast, testes, thyroid;
[3] Cardiac includes Adult and Pediatric;
[4] Interventional Guidance Tissue Biopsy is 2D biopsy guide;
[5] 3D/4D imaging Mode;
[*]Combined modes are B/M, B/ Color M, B/PWD, B/Color/PWD, B/Power/PWD;
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined in a stylized script, enclosed within a blue circle. There are three small, white, curved shapes around the inside of the circle, evenly spaced.
## Diagnostic Ultrasound Indications for Use Form LOGIQ V5 Expert / LOGIQ V5 / LOGIQ V3 with 3Sc-RS Transducer
| | Mode of Operation | | | | | | | | | | |
|------------------------------------|-------------------|----|---------------|----|-------|---------|-------|--------------------|---------------------|-----------------|-------|
| Clinical Application | B | M | Doppler Modes | | | | | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse` | Other |
| | | | PW | CW | Color | Color M | Power | | | | |
| Anatomy/Region of Interest | | | | | | | | | | | |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | |
| Abdominal[1] | P | P | P | | P | | P | P | P | P | |
| Pediatric | | | | | | | | | | | |
| Small Organ [2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | P | P | P | P | P | | P | P | P | | |
| Cardiac [3] | P | P | P | P | P | P | P | P | P | | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Thoracic/Pleural | | | | | | | | | | | |
| Other | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transcranial | P | P | P | P | P | | P | P | P | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | |
| Tissue Biopsy/Fluid<br>Drainage[4] | P2 | P2 | P2 | P2 | P2 | | P2 | P2 | P2 | | |
| Vascular Access (IV, PICC) | | | | | | | | | | | |
| Nerve Block | | | | | | | | | | | |
| Other | | | | | | | | | | | |
Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:
N = new indication; P= previously cleared by FDA K141768; P2= previously cleared by FDA K160277;
P3= previously cleared by FDA K151028
Notes: [1] Abdominal includes GYN and Urological/Prostate;
[2] Small Organ includes breast, testes, thyroid;
[3] Cardiac includes Adult and Pediatric;
[4] Interventional Guidance Tissue Biopsy is 2D biopsy guide;
[5] 3D/4D imaging Mode;
[*]Combined modes are B/M, B/ Color M, B/PWD, B/Color/PWD, B/Power/PWD
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized, cursive font. The letters are enclosed within a circular border that features decorative flourishes or embellishments around the perimeter. The color of the logo is blue.
## Diagnostic Ultrasound Indications for Use Form LOGIQ V5 Expert / LOGIQ V5 / LOGIQ V3 with L6-12-RS Transducer
| Clinical Application | B | M | Mode of Operation | | | | | Other | | | |
|------------------------------------|----|----|-------------------|----|-------|---------|-------|---------------------|---------------------|-----------------|--|
| | | | PW | CW | Color | Color M | Power | Combined<br>Modes * | Harmonic<br>Imaging | Coded<br>Pulse` | |
| Anatomy/Region of Interest | | | | | | | | | | | |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | |
| Abdominal[1] | | | | | | | | | | | |
| Pediatric | P | P | P | | P | | P | P | P | | |
| Small Organ [2] | P | P | P | | P | | P | P | P | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac [3] | | | | | | | | | | | |
| Peripheral Vascular | P | P | P | | P | | P | P | P | | |
| Musculo-skeletal Conventional | P | P | P | | P | | P | P | P | | |
| Musculo-skeletal Superficial | P | P | P | | P | | P | P | P | | |
| Thoracic/Pleural | | | | | | | | | | | |
| Other | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transcranial | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | |
| Tissue Biopsy/Fluid<br>Drainage[4] | p2 | p2 | p2 | | p2 | | p2 | p2 | p2 | | |
| Vascular Access (IV, PICC) | | | | | | | | | | | |
| Nerve Block | | | | | | | | | | | |
Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:
N = new indication; P= previously cleared by FDA K141768; P2= previously cleared by FDA K160277;
P3= previously cleared by FDA K151028
Notes: [1] Abdominal includes GYN and Urological/Prostate;
[2] Small Organ includes breast, testes, thyroid;
[3] Cardiac includes Adult and Pediatric;
[4] Interventional Guidance Tissue Biopsy is 2D biopsy guide;
[5] 3D/4D imaging Mode;
[*]Combined modes are B/M, B/ Color M, B/PWD, B/Color/PWD, B/Power/PWD
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a blue circle. There are three small, white, curved shapes around the inside of the circle, evenly spaced apart.
## Diagnostic Ultrasound Indications for Use Form LOGIQ V5 Expert / LOGIQ V5 / LOGIQ V3 with E8C-RS Transducer
| Clinical Application | | | Mode of Operation | | | | | | | | |
|-------------------------------------|----|----|-------------------|----|-------|---------|-------|--------------------|---------------------|-----------------|-------|
| | B | M | PW | CW | Color | Color M | Power | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse' | Other |
| <i>Anatomy/Region of Interest</i> | | | | | | | | | | | |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | P | P | P | | P | | P | P | P | | |
| Abdominal[1] | P | P | P | | P | | P | P | P | | |
| Pediatric | | | | | | | | | | | |
| Small Organ [2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac [3] | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Thoracic/Pleural | | | | | | | | | | | |
| Other | | | | | | | | | | | |
| <i>Exam Type, Means of Access</i> | | | | | | | | | | | |
| Transcranial | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | P2 | p2 | P2 | | p2 | | P2 | P2 | P2 | | |
| Transvaginal | P | P | P | | P | | P | P | P | | |
| Intraoperative | | | | | | | | | | | |
| <i>Interventional Guidance</i> | | | | | | | | | | | |
| Tissue Biopsy/Fluid<br>Drainage [4] | p2 | p2 | P2 | | p2 | | P2 | P2 | P2 | | |
| Vascular Access (IV, PICC) | | | | | | | | | | | |
| Nerve Block | | | | | | | | | | | |
Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:
N = new indication; P= previously cleared by FDA K141768; P2= previously cleared by FDA K160277;
P3= previously cleared by FDA K151028
[1] Abdominal includes GYN and Urological/Prostate; Notes:
- [2] Small Organ includes breast, testes, thyroid;
[3] Cardiac includes Adult and Pediatric;
[4] Interventional Guidance Tissue Biopsy is 2D biopsy guide;
[5] 3D/4D imaging Mode;
[*]Combined modes are B/M, B/ Color M, B/PWD, B/Color/PWD, B/Power/PWD
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. The letters are white and are surrounded by a blue background. The circle is bordered by a white line.
#### Diagnostic Ultrasound Indications for Use Form
#### LOGIQ V5 Expert / LOGIQ V5 / LOGIQ V3 with RAB2-6-RS Transducer
| Clinical Application | B | M | Doppler Modes | | | | | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
|------------------------------------|----|----|---------------|----|-------|---------|-------|-------------------|---------------------|----------------|-------|
| Anatomy/Region of Interest | | | PW | CW | Color | Color M | Power | | | | |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | p2 | p2 | p2 | | p2 | | p2 | p2 | p2 | | [5] |
| Abdominal[1] | p2 | p2 | p2 | | p2 | | p2 | p2 | p2 | | [5] |
| Pediatric | | | | | | | | | | | |
| Small Organ [2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Thoracic/Pleural | | | | | | | | | | | |
| Other | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transcranial | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | |
| Tissue Biopsy/Fluid<br>Drainage[4] | p2 | p2 | p2 | | p2 | p2 | p2 | p2 | p2 | | [5] |
| Vascular Access (IV, PICC) | | | | | | | | | | | |
| Nerve Block | | | | | | | | | | | |
Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:
N = new indication; P= previously cleared by FDA K141768; P2= previously cleared by FDA K160277;
P3= previously cleared by FDA K151028
[1] Abdominal includes GYN and Urological/Prostate; Notes:
[2] Small Organ includes breast, testes, thyroid;
[3] Cardiac includes Adult and Pediatric;
[4] Interventional Guidance Tissue Biopsy is 2D biopsy guide;
[5] 3D/4D imaging Mode;
[*]Combined modes are B/M, B/ Color M, B/PWD, B/Color/PWD, B/Power/PWD
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular frame. The frame is decorated with ornamental flourishes, giving it a classic and elegant appearance. The logo is colored in blue, set against a white background.
## Diagnostic Ultrasound Indications for Use Form LOGIQ V5 Expert / LOGIQ V5 / LOGIQ V3 with 6S-RS Transducer
| | | Mode of Operation | | | | | | | | | | |
|------------------------------------|--|-------------------|----|----|----|-------|---------|-------|---------------------|---------------------|----------------|-------|
| Clinical Application | | B | M | PW | CW | Color | Color M | Power | Combined<br>Modes * | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Anatomy/Region of Interest | | | | | | | | | | | | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | | |
| Abdominal[1] | | | | | | | | | | | | |
| Pediatric | | p2 | p2 | p2 | | p2 | | p2 | p2 | p2 | | |
| Small Organ [2] | | | | | | | | | | | | |
| Neonatal Cephalic | | p3 | p3 | p3 | p3 | p3 | | p3 | p3 | p3 | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac[3] | | p2 | p2 | p2 | p2 | p2 | p2 | p2 | p2 | p2 | | |
| Peripheral Vascular | | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | | |
| Thoracic/Pleural | | | | | | | | | | | | |
| Other | | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transcranial | | p2 | p2 | p2 | p2 | p2 | | p2 | p2 | p2 | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | | |
| Tissue Biopsy/Fluid<br>Drainage[4] | | | | | | | | | | | | |
| Vascular Access (IV, PICC) | | | | | | | | | | | | |
| Nerve Block | | | | | | | | | | | | |
Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:
N = new indication; P= previously cleared by FDA K141768; P2= previously cleared by FDA K160277;
P3= previously cleared by FDA K151028
Notes: [1] Abdominal includes GYN and Urological/Prostate;
[2] Small Organ includes breast, testes, thyroid;
[3] Cardiac includes Adult and Pediatric;
[4] Interventional Guidance Tissue Biopsy is 2D biopsy guide;
[5] 3D/4D imaging Mode;
[*]Combined modes are B/M, B/ Color M, B/PWD, B/Color/PWD, B/Power/PWD
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular frame. The frame is decorated with ornamental flourishes, giving it a classic and elegant appearance. The logo is colored in blue, set against a white background.
## Diagnostic Ultrasound Indications for Use Form LOGIQ V5 Expert / LOGIQ V5 / LOGIQ V3 with E8Cs-RS Transducer
| Clinical Application | B | M | Mode of Operation | | | | | | | | Other |
|-------------------------------------|----|----|-------------------|----|-------|---------|-------|-------------------|---------------------|----------------|-------|
| | | | PW | CW | Color | Color M | Power | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | |
| Anatomy/Region of Interest | | | | | | | | | | | |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | P2 | P2 | P2 | | P2 | | P2 | P2 | P2 | | |
| Abdominal[1] | P2 | P2 | P2 | | P2 | | P2 | P2 | P2 | | |
| Pediatric | | | | | | | | | | | |
| Small Organ [2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac [3] | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Thoracic/Pleural | | | | | | | | | | | |
| Other | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transcranial | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | P2 | P2 | P2 | | P2 | | P2 | P2 | P2 | | |
| Transvaginal | P2 | P2 | P2 | | P2 | | P2 | P2 | P2 | | |
| Intraoperative | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | |
| Tissue Biopsy/Fluid<br>Drainage [4] | P2 | P2 | P2 | | P2 | | P2 | P2 | P2 | | |
| Vascular Access (IV, PICC) | | | | | | | | | | | |
| Nerve Block | | | | | | | | | | | |
Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:
N = new indication; P= previously cleared by FDA K141768; P2= previously cleared by FDA K160277;
P3= previously cleared by FDA K151028
Notes: [1] Abdominal includes GYN and Urological/Prostate;
[2] Small Organ includes breast, testes, thyroid;
[3] Cardiac includes Adult and Pediatric;
[4] Interventional Guidance Tissue Biopsy is 2D biopsy guide;
[5] 3D/4D imaging Mode;
[*]Combined modes are B/M, B/ Color M, B/PWD, B/Color/PWD, B/Power/PWD
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a blue circle. There are three small, white, curved shapes around the inside of the circle, evenly spaced apart.
## Diagnostic Ultrasound Indications for Use Form LOGIQ V5 Expert / LOGIQ V5 / LOGIQ V3 with 12L-RS Transducer
| Clinical Application | B | M | Doppler Modes | | | | | | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
|------------------------------------|----|----|---------------|----|-------|---------|-------|----|-------------------|---------------------|----------------|-------|
| Anatomy/Region of Interest | | | PW | CW | Color | Color M | Power | | | | | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | | |
| Abdominal1 | | | | | | | | | | | | |
| Pediatric | p3 | p3 | p3 | | p3 | | p3 | p3 | p3 | p3 | | |
| Small Organ [2] | p3 | p3 | p3 | | p3 | | p3 | p3 | p3 | p3 | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac [3] | | | | | | | | | | | | |
| Peripheral Vascular | p3 | p3 | p3 | | p3 | | p3 | p3 | p3 | p3 | | |
| Musculo-skeletal Conventional | p3 | p3 | p3 | | p3 | | p3 | p3 | p3 | p3 | | |
| Musculo-skeletal Superficial | p3 | p3 | p3 | | p3 | | p3 | p3 | p3 | p3 | | |
| Thoracic/Pleural | | | | | | | | | | | | |
| Other | | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transcranial | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | | |
| Tissue Biopsy/Fluid<br>Drainage[4] | p3 | p3 | p3 | | p3 | | p3 | p3 | p3 | p3 | | |
| Vascular Access (IV, PICC) | | | | | | | | | | | | |
| Nerve Block | | | | | | | | | | | | |
Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:
N = new indication; P= previously cleared by FDA K141768; P2= previously cleared by FDA K1602777;
P3= previously cleared by FDA K151028
Notes: [1] Abdominal includes GYN and Urological/Prostate;
[2] Small Organ includes breast, testes, thyroid;
[3] Cardiac includes Adult and Pediatric;
[4] Interventional Guidance Tissue Biopsy is 2D biopsy guide;
[5] 3D/4D imaging Mode;
[*]Combined modes are B/M, B/ Color M, B/PWD, B/Color/PWD, B/Power/PWD
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized script, enclosed within a circular border. The logo is blue and white.
#### 510(k) Summarv
In accordance with 21 CFR 807.92 the following summary of information is provided: Date: April 28, 2016 GE Medical Systems Ultrasound and Primary Care Diagnostics, Submitter: 9900 Innovation Drive Wauwatosa, WI 53226 Tracey Ortiz Primary Contact Person: Regulatory Affairs Director GE Healthcare T:(262)676-6120 F:(414)918-8275 Secondary Contact Person: Jian Xie Regulatory Affairs GE Medical Systems (China) Co, Ltd. T: +86 510 8527 8651 F: +86 510 8522 7347 LOGIO V5 Expert / LOGIO V5 / LOGIO V3 Device: Trade Name: Common/Usual Name: Ultrasound system Classification Names: Class II Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550 90-IYN Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO Product Code: Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX Primary Predicate LOGIQ V3 / LOGIQ V5 - K141768 Device(s): Reference Predicate LOGIQ F8 Expert / LOGIQ F8 / LOGIQ F6 - K160277 LOGIO e - K151028 Device(s): The LOGIQ V5 Expert / LOGIQ V5 / LOGIQ V3 are entry level Device Description: ultrasound scanners from the LOGIQ family for private clinics focusing on OB/GYN. The systems are for general purpose imaging and analysis providing real-time digital acquisition, processing and display capability intended for general radiology imaging, evaluation with some cardiology and vascular applications, and providing guidance during tissue biopsy and fluid drainage procedures.
> The track 3 systems consist of a mobile console with keyboard control panel; color display and optional image storage, wireless
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a circular border. The logo is colored in blue, with a white outline around the letters and the circular border.
capability, and printing devices. They utilize a variety of linear, curved linear and matrix phased array transducers. The systems have the same hardware and software but may have different options available.
- Intended Use: The LOGIO V5 Expert / LOGIO V5 / LOGIO V3 are general purposed ultrasound imaging and analysis systems providing digital acquisition, processing and display capability and clinical applications including: Abdominal, Obstetrical, Gynecological, Small parts, Vascular/Peripheral Vascular, Adult Cephalic, Pediatric, Musculoskeletal, Transcranial, Neonatal Cephalic, Transvaginal, Urological, Cardiac, Transrectal, and Tissue Biopsy/Fluid Drainage
- Technology: The LOGIO V5 Expert / LOGIO V5 / LOGIO V3 employ the same fundamental scientific technology as its predicate devices.
#### Comparison to Predicates Determination of Substantial Equivalence:
The following is an overview of the difference between the proposed LOGIQ V5 Expert / LOGIQ V5 / LOGIQ V3 and the predicate LOGIQ V5 / LOGIQ V3 (K141768). Most of feartures and probes are currently cleared on other GE ultrousund systems.
- The systems are all intended for diagnostic ultrasound imaging and fluid flow analysis.
- The proposed LOGIQ V5 Expert / LOGIQ V5 / LOGIQ V3 and the predicate LOGIQ V5 / LOGIQ V3 have the same intended use and clincial applications except the addition of the transrectal and biopsy applications that were cleared on the LOGIQ F8 Expert / LOGIQ F8 / LOGIQ F6 predicate (K160277).
- The proposed LOGIO V5 Expert / LOGIO V5 / LOGIO ● V3 and the predicate LOGIQ V5 / LOGIQ V3 have the same probes except for 5 being added: 6S-RS, E8Cs-RS, and RAB2-6-RS were cleared in predicate the LOGIQ F8 Expert / LOGIQ F8 / LOGIQ F6 predicate (K160277); 12L-RS and 8C-RS were cleared in predicate LOGIQ e system (K151028).
- The 6S-RS probe application is being expanded to include ● Neonatal Cephalic which was cleared in the LOGIO e predicate system (K151028).
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script. The letters are surrounded by a circular border with wave-like designs around the perimeter. The logo is colored in a light blue hue.
510(k) Premarket Notification Submission
- Adding a 3D/4D mode with the RAB2-6-RS probe that was cleared on the predicate LOGIQ F8 Expert / LOGIQ F8 / LOGIQ F6 (K160277).
- . A USB wireless adaptor is being added to allow for wireless connectivity like that in predicate LOGIQ e (K151028).
- Adding optional LCD monitor display sizes and a VGA/S-video/composite output.
- Expanding the Scan Coach software feature
- Adding 3D/4D imaging which was cleared in predicate ● device LOGIQ F8 Expert/LOGIQ F8/LOGIQ F6 (K160277).
- The LOGIQ V5 Expert / LOGIQ V5 / LOGIQ V3 and the predicate LOGIQ V5 / LOGIQ V3 have been designed in compliance with electrical and physical safety standards.
- . The LOGIO V5 Expert / LOGIO V5 / LOGIO V3 and the predicate LOGIQ V5 / LOGIQ V3 have the same capability in terms of performing measurements, capturing digital images, reviewing and reporting studies.
- The systems are manufactured with materials which have been evaluated and found to be safe for the intended use of the device.
- The system has acoustic power levels which are below the ● applicable FDA limits.
### Summary of Non-Clinical Tests:
The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic, and mechanical safety, and has been found to conform to applicable medical device safety standards. The LOGIQ V5 Expert / LOGIQ V5 / LOGIQ V3 and its applications comply with voluntary standards:
- AAMI/ANSI ES 60601-1, Medical Electrical Equipment ● - Part 1: General Requirements for Safety, 2005
- IEC 60601-1-2, Medical Electrical Equipment Part 1-● 2:General Requirements for Safety – Collateral Standard: Electromagnetic Compatibility Requirements and Tests, 2007
- IEC 60601-2-37, Medical Electrical Equipment Part 2-● 37:Particular Requirements for the Safety of Ultrasonic
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized, cursive font. The letters are enclosed within a circular border, and there are decorative swirls or flourishes around the circle, giving it a classic and recognizable appearance. The logo is presented in a blue color.
510(k) Premarket Notification Submission
Medical Diagnostic and Monitoring Equipment, 2007
- NEMA UD 3, Standard for Real Time Display of Thermal ● and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment, 2004
- . ISO 10993-1, Biological Evaluation of Medical Devices-Part 1: Evaluation and Testing, 2009
- NEMA UD 2, Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment, 2004
- ISO 14971, Application of risk management to medical ● devices, 2007
- NEMA, PS 3.1 3.20 (2011), Digital Imaging and ● Communications in Medicine (DICOM) Set. (Radiology)
The following quality assurance measures were applied to the development of the system:
- Risk Analysis
- Requirements Reviews
- Design Reviews
- Testing on unit level (Module verification) ●
- Integration testing (System verification)
- Performance testing (Verification)
- Safety testing (Verification) ●
- Simulated use Testing (Validation) ●
Transducer materials and other patient contact materials are biocompatible.
#### Summary of Clinical Tests:
The subject of this premarket submission, LOGIQ V5 Expert / LOGIO V5 / LOGIO V3. did not require clinical studies to support substantial equivalence.
GE Healthcare considers the proposed LOGIQ V5 Expert / Conclusion: LOGIO V5 / LOGIO V3 to be as safe, as effective, and performance is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.