← Product Code [OUR](/productcode/OUR) · K161210

# RIALTO(TM) SI Fusion System, Medtronic Reusable Instruments for Use with the IPC(TM) Powerease(TM) System, and Medtronic Navigated Manual Reusable Instruments for use with the StealthStation(TM) and IPC(TM) Powerease(TM) Systems (K161210)

_Medtronic Sofamor Danek · OUR · Aug 12, 2016 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K161210

## Device Facts

- **Applicant:** Medtronic Sofamor Danek
- **Product Code:** [OUR](/productcode/OUR.md)
- **Decision Date:** Aug 12, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The RIALTO™ SI Fusion System is intended for sacroiliac joint fusion for conditions including sacroiliac joint disruptions and degenerative sacroiliitis.

## Device Story

RIALTO™ SI Fusion System consists of cannulated titanium alloy implants (12mm diameter, 40-60mm length) for sacroiliac joint stabilization and fusion; autograft/allograft may be used. Implantation occurs via minimally invasive approach using fluoroscopy or navigated instruments. System includes reusable stainless steel taps, drivers, and drills compatible with Medtronic IPC® POWEREASE® system or manual quick-connect handles. Navigated versions integrate with StealthStation® for stereotactic guidance relative to CT/MR models or fluoroscopy. Surgeons use these tools to prepare bone and place implants. Benefits include joint stabilization and fusion for degenerative conditions.

## Clinical Evidence

Bench testing only. Mechanical verification performed per ASTM F543-13 and ASTM F2193-02 (2007). Testing included push-out, torque-to-fail, four-point bend, and shear testing. Navigation accuracy was verified for navigated instruments. Biocompatibility was established via clinical history of materials. Bacterial endotoxin testing performed per ANSI/AAMI ST72 and USP <161> confirmed compliance with the 20 EU/device limit.

## Technological Characteristics

Implants: Titanium alloy. Instruments: High-grade stainless steel. Energy: Manual or powered via IPC® POWEREASE® system. Navigation: Compatible with StealthStation® stereotactic system. Standards: ASTM F543-13, ASTM F2193-02, ANSI/AAMI ST72, USP <161>.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- MSB Sacroiliac Joint Fusion Device ([K110472](/device/K110472.md))
- iFuse Implant System ([K131405](/device/K131405.md))
- IPC® POWEREASE® System ([K111520](/device/K111520.md))
- Navigated CD Horizon Solera Screwdriver/Taps ([K140454](/device/K140454.md))
- Navigated CD HORIZON® SOLERA™ Screwdrivers, CD HORIZON® SOLERA™ Taps, CD HORIZON® SOLERA™ Iliac Taps, CD HORIZON® LEGACY™ Taps ([K124004](/device/K124004.md))
- Navigated Disc Prep Instruments and CAPSTONE Trials ([K150231](/device/K150231.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or a stylized human profile, composed of three interconnected lines or shapes.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 12, 2016

Medtronic Sofamor Danek Mr. Ankit Shah Senior Regulatory Affairs Specialist 1800 Pyramid Place Memphis, Tennessee 38132

Re: K161210

Trade/Device Name: RIALTO™ SI Fusion System, MEDTRONIC RESUSABLE INSTRUMENTS FOR USE WITH IPC™ POWEREASE™ SYSTEM, MEDTRONIC NAVIGATED MANUAK REUSABLE INSTRUMENTS FOR USE WITH THE STEALTHSTATION" AND IPC™ POWEREASE "SYSTEMS Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: OUR, OLO, HWE Dated: July 29, 2016

Received: August 1, 2016

Dear Mr. Shah:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing

{1}------------------------------------------------

(21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| DEPARTMENT OF HEALTH AND HUMAN SERVICES         |                                                                                                                                                                                                                                                                                                                                                          |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Food and Drug Administration                    |                                                                                                                                                                                                                                                                                                                                                          |
| <b>Indications for Use</b>                      | Form Approved: OMB No. 0910-0120                                                                                                                                                                                                                                                                                                                         |
|                                                 | Expiration Date: January 31, 2017                                                                                                                                                                                                                                                                                                                        |
|                                                 | See PRA Statement below.                                                                                                                                                                                                                                                                                                                                 |
| 510(k) Number (if known)                        | K161210                                                                                                                                                                                                                                                                                                                                                  |
| Device Name                                     | RIALTOTM SI Fusion System                                                                                                                                                                                                                                                                                                                                |
| Indications for Use (Describe)                  | The RIALTOTM SI Fusion System is intended for sacroiliac joint fusion for conditions including sacroiliac joint disruptions and degenerative sacroiliitis.                                                                                                                                                                                               |
| Type of Use (Select one or both, as applicable) |                                                                                                                                                                                                                                                                                                                                                          |
|                                                 | <span> <svg height="12" width="12"> <rect height="12" style="fill:none;stroke:#000000;stroke-width:1" width="12"></rect> <line style="stroke:#000000;stroke-width:1" x1="0" x2="12" y1="0" y2="12"></line> <line style="stroke:#000000;stroke-width:1" x1="0" x2="12" y1="12" y2="0"></line> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) |
|                                                 | <span> <svg height="12" width="12"> <rect height="12" style="fill:none;stroke:#000000;stroke-width:1" width="12"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C)                                                                                                                                                                      |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services  
Food and Drug Administration  
Office of Chief Information Officer  
Paperwork Reduction Act (PRA) Staff  
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

әбед

ー

go

ധ

{3}------------------------------------------------

| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration<br><b>Indications for Use</b> | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below.                                                                                                                                                                                                                                                                                                                                                                                                                        |
|-------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number (if known)                                                                              | K161210                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Device Name                                                                                           | MEDTRONIC REUSABLE INSTRUMENTS FOR USE WITH THE IPCTM POWEREASEM SYSTEM                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Indications for Use (Describe)                                                                        | IPCT™ System is indicated for the incision/cutting, removal, drilling and sawing of soft and hard tissue and bone, and biomaterials in Neurosurgical (Cranial, Craniofacial), Orthopedic, Arthroscopic, Spinal, Sternotomy, and General surgical procedures.  The IPCTM POWEREASET™ System is indicated for drilling, tapping and driving screws and working end attachments during spinal surgery, including open and minimally invasive procedures. It is also used in placement or cutting of screws, posts and rods. |
| Type of Use (Select one or both, as applicable)                                                       | <div> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)             </div> <div> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)             </div>                                                                                                                                                                                                                                                                                                              |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services  

Food and Drug Administration  

Office of Chief Information Officer  

Paperwork Reduction Act (PRA) Staff  

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

ə bey

2

구이

3

{4}------------------------------------------------

| DEPARTMENT OF HEALTH AND HUMAN SERVICES |
|-----------------------------------------|
| Food and Drug Administration            |
| <b>Indications for Use</b>              |

| Form Approved: OMB No. | 0910-0120                |
|------------------------|--------------------------|
| Expiration Date:       | January 31, 2017         |
|                        | See PRA Statement below. |

| 510(k) Number (if known)       | K161210                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name                    | MEDTRONIC NAVIGATED MANUAL REUSABLE INSTRUMENTS FOR USE WITH THE STEALTHSTATION <sup>TM</sup> AND IPC <sup>TM</sup> POWEREASETM SYSTEMS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Indications for Use (Describe) | Medtronic Navigated Reusable Instruments are intended to be used during the preparation and placement of Medtronic             screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open, or minimally             invasive, procedures. Medtronic Navigated Reusable Instruments are specifically designed for use with the             StealthStation™ System, which is indicated for any medical condition in which the use of stereotactic surgery may be             appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra can be identified             relative to a CT or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy. Medtronic Navigated             Reusable Instruments are also compatible with the IPCTM POWEREASET™ System. |

| Type of Use (Select one or both, as applicable)                                                       |                                                                                    |
|-------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <label><input checked="checked" type="checkbox"/>Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/>Over-The-Counter Use (21 CFR 801 Subpart C)</label> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

 Department of Health and Human Services  

 Food and Drug Administration  

 Office of Chief Information Officer  

 Paperwork Reduction Act (PRA) Staff  

[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
 information unless it displays a currently valid OMB number."

| FORM FDA 3881 (8/14) | Page 1 of 1 |
|----------------------|-------------|
|----------------------|-------------|

PSC Publishing Services (301) 443-6740 EF

ə bəya

3

go

3

{5}------------------------------------------------

## RIALTO™ SI Fusion System 510(k) SUMMARY July 2016

- I. Medtronic Sofamor Danek USA, Inc. Submitter 1800 Pyramid Place Memphis, TN 38132 (901)396-3133
Contact Ankit K. Shah Senior Regulatory Affairs Specialist

July 28, 2016 Date Prepared

- Device II.
	- RIALTO™ SI Fusion System, Medtronic Reusable Name of Device Instruments for Use with the IPC™ Powerease™ System, and Medtronic Navigated Manual Reusable Instruments for use with the StealthStation™ and IPC™M Powerease™ Systems
	- Smooth or threaded metallic bone fixation fastener Common Name (Implant) Taps, Drivers and Drills (Instrument)
	- Classification Name Sacroiliac Joint Fixation - OUR (For Implant)

Stereotaxic Instrument - OLO (For Navigated Instruments)

Surgical Instrument Motors and Accessories/Attachments -HWE

{6}------------------------------------------------

|                | (For Instruments Compatible with the IPC®<br>POWEREASE® System)                                                                                                  |
|----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification | Class II (Implant and Instruments)                                                                                                                               |
| Product Codes  | OUR (Implants)<br>21 CFR 888.3040<br>OLO (Navigated Instruments)<br>21 CFR 882.4560                                                                              |
|                | HWE (IPC® POWEREASE® Compatible Instruments)<br>21 CFR 878.4820                                                                                                  |
| Predicates     | There are 6 Predicates.                                                                                                                                          |
|                | MSB Sacroiliac Joint Fusion Device<br>K110472, S.E. 05/29/2012 (Primary Predicate)                                                                               |
|                | iFuse Implant System<br>K131405, S.E. 10/16/2013                                                                                                                 |
|                | IPC® POWEREASE® System<br>K111520, S.E. 10/26/2011                                                                                                               |
|                | Navigated CD Horizon Solera Screwdriver/Taps<br>K140454, S.E. 05/22/2014                                                                                         |
|                | Navigated CD HORIZON® SOLERA™ Screwdrivers, CD<br>HORIZON® SOLERA™ Taps, CD HORIZON®<br>SOLERA™ Iliac Taps, CD HORIZON® LEGACY™ Taps<br>K124004, S.E. 03/22/2013 |
|                | Navigated Disc Prep Instruments and CAPSTONE Trials<br>K150231, S.E. 06/16/2015                                                                                  |
|                | The predicates have not been subject to a design related<br>recall.                                                                                              |

{7}------------------------------------------------

#### III.Product Description

#### RIALTO™ SI Fusion System

The subject RIALTOTM SI Fusion System consists of cannulated devices available in various lengths, used to provide stabilization when fusion of the sacroiliac joint is desired. Autograft and/or allograft may be placed in conjunction with the RIALTO™ SI Fusion System. The RIALTO™ implant are made using Titanium Alloy and are 40mm-60mm in length with a diameter of 12mm. This device may be implanted via a minimally invasive approach using fluoroscopy or navigated instruments compatible with Medtronic StealthStation® and IPC® POWEREASE®.

### Medtronic Reusable Instruments Only Compatible with the IPC® POWEREASE® System

The subject Medtronic Reusable Taps, Drivers and Drills are spine preparation instruments made of high grade stainless steel and are designed specifically for subject RIALTO™ SI Fusion System. The subject Taps, Drivers and Drills are compatible with Medtronic's IPC® POWEREASE® System may be connected to the IPC® POWEREASE® handpiece. The subject Taps, Drivers and Drills can be used manually using existing Medtronic Class I Exempt quick connect handles in place of the IPC® POWEREASE® handpiece.

### Medtronic Reusable Instruments Compatible with the STEALTHSTATION® System and IPC® POWEREASE® System

The subject Medtronic Navigated Reusable Taps, Drivers and Drills are spine preparation instruments made of high grade stainless steel. These instruments were specifically designed for use with subject RIALTO™ SI Fusion System in procedures where the use of stereotactic surgery may be appropriate. The subject Taps, Drivers and Drills are also compatible with Medtronic's IPC® POWEREASE™ System when connected to the POWEREASE™ handpiece or may be used manually with existing Medtronic Class I Exempt quick connect handles in place of the IPC® POWEREASE® handpiece or the NavLock™ tracker.

{8}------------------------------------------------

#### IV. Indications for Use:

#### RIALTO™ SI Fusion System

The RIALTO™ SI Fusion System is intended for sacroiliac joint fusion for conditions including sacroiliac joint disruptions and degenerative sacroiliitis.

#### Medtronic Reusable Instruments For Use with the IPC® POWEREASE® System

IPC® System is indicated for the incision/cutting, removal, drilling and sawing of soft and hard tissue and bone, and biomaterials in Neurosurgical (Cranial, Craniofacial), Orthopedic, Arthroscopic, Spinal, Sternotomy, and General surgical procedures.

The IPC® POWEREASE® System is indicated for drilling, tapping and driving screws and working end attachments during spinal surgery, including open and minimally invasive procedures. It is also used in placement or cutting of screws, posts and rods.

### Medtronic Navigated Reusable Instruments For Use with STEALTHSTATION® and IPC® POWEREASE™ Systems

Medtronic Navigated Reusable Instruments are intended to be used during the preparation and placement of Medtronic screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open, or minimally invasive, procedures. Medtronic Navigated Reusable Instruments are specifically designed for use with the StealthStation® System, which is indicated for any medical condition in which stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra, can be identified relative to a CT or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy. Medtronic Navigated Reusable Instruments are also compatible with the IPC® POWEREASE™ System.

### V. Comparison of Technological Characteristics

The primary predicate for the RIALTO™ SI Fusion System is the predicate SI-FIX Sacroiliac Joint Fusion System K110472, S.E. 05/29/2012 (Primary Predicate).

{9}------------------------------------------------

The subject RIALTO™ SI Fusion System implants have the same or similar intended use, fundamental scientific technology, material, and indications as the following FDA cleared predicates K110472 (S.E. 05/29/2012) and K131405 (S.E. 10/16/2013).

The subject RIALTO™ SI Fusion System instruments have the same or similar intended use, fundamental scientific technology, material, and indications as the following FDA cleared predicates K111520 (S.E. 10/26/2011), K110472 (S.E. 05/29/2012), K140454 (S.E. 05/22/2014), K124004 (S.E. 03/22/2013) and K150231 ( S.E. 06/16/2015)

#### VI. Performance Data

The following performance data are provided in support of the substantial equivalence determination.

#### Biocompatibility

Identical to the primary predicate devices the implants in the subject RIALTO™ SI Fusion System are made using Titanium alloy.

The non-sterile instruments are manufactured using stainless steel and are identical to the materials used for the instruments cleared under predicate submissions.

The titanium alloy and stainless steel material used for the subject RIALTO™ SI Fusion System implants and instruments have a long clinical history of safe and effective use in similar commercially available medical devices. Therefore, no additional biocompatibility testing is required.

#### Mechanical Testing

In accordance with the Guidance for Industry and FDA Staff - Spinal System 510(k)'s, Medtronic has evaluated the subject devices to demonstrate substantial equivalence to the predicate devices.

{10}------------------------------------------------

Design verification testing for the subjectimplants was completed in accordance with

- ASTM F543-13 Standard Specification and Test Methods for Metallic Medical . Bone Screws
- ASTM F2193-02 (2007) Standard Specifications and Test Methods for . Components Used in the Surgical Fixation of the Spinal Skeletal System

The tests completed were:

- Push-Out -
- Torque—to-Fail -
- Four Point Bend -
- Shear Testing -

The verification testing on the instruments included Navigation accuracy on the instruments compatible with StealthStation™ and IPC™ Powerease™.

The subject devices met the pre-determined acceptance criteria for all tests. Therefore, design verification testing determined that the subject screws and instruments are substantially equivalent to the predicate devices.

Design validation testing was performed that demonstrated that the subject instruments performed as intended and to validate the surgical approach.

#### Non-Pyrogenicity Endotoxin Testing

The bacterial endotoxin test, also known as Limulus Amebocyte Lysate (LAL) test, was performed utilizing worst case subject RIALTO™ SI Fusion implants to verify that the subject implants meet the 20 endotoxin units (EU)/device pyrogen limit specification. Testing was successfully performed and it was confirmed that the subject implants meet the 20 EU/device testing limit for general medical devices that are implanted as outlined in ANSI/AAMI ST72, Bacterial endotoxins - Test methods, routine monitoring, and 

{11}------------------------------------------------

alternatives to batch testing and USP <161>, Transfusion and Infusion Assemblies and Similar Medical Devices.

#### VII. Conclusions

Based on the test results and additional supporting information provided in this premarket notification, the subject devices demonstrated substantial equivalence to the legally marketed predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K161210](https://fda.innolitics.com/device/K161210)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
