← Product Code [LZO](/productcode/LZO) · K161192

# Wagner SL Revision Stem Lateral, Wagner Cone Prosthesis® System (K161192)

_Zimmer GmbH · LZO · Jul 8, 2016 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K161192

## Device Facts

- **Applicant:** Zimmer GmbH
- **Product Code:** [LZO](/productcode/LZO.md)
- **Decision Date:** Jul 8, 2016
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3353
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Noninflammatory degenerative joint disease (NIDJ), e.g. avascular necrosis, osteoarthritis, and inflammatory joint disease (IJD), e.g. rheumatoid arthritis. Failed previous surgery where pain, deformity, or dysfunction persists. Revision of previously failed hip arthroplasty.

## Device Story

The Wagner SL Revision Stem Lateral and Wagner Cone Prosthesis are hip arthroplasty implants. The subject of this modification is the surgical instrumentation (Awls) used to prepare the femoral medullary canal. Previously limited to manual use, the awls are now compatible with both manual and power-driven surgical tools. The implants themselves (forged titanium alloy stems) remain unchanged. The system is used by orthopedic surgeons in a clinical/OR setting to facilitate femoral canal reaming during hip replacement or revision procedures. The power-driven option aims to improve surgical efficiency. The implants provide structural support for the hip joint, benefiting patients by restoring function and reducing pain associated with degenerative or inflammatory joint disease.

## Clinical Evidence

No clinical data was required or provided. Substantial equivalence was demonstrated through non-clinical performance testing, specifically usability testing and similar device analysis.

## Technological Characteristics

Implants: Forged titanium alloy, circular cross-section, conical anchorage ribs. Awls: Modified to support both manual and power-driven operation for femoral canal reaming. Sterilization: Not specified.

## Regulatory Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

## Predicate Devices

- Wagner SL Revision Stem Lateral ([K043356](/device/K043356.md))
- Wagner Cone Prosthesis® System ([K113556](/device/K113556.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or movement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 8, 2016

Zimmer GmbH Ms. Anne-Kathrin Born Regulatory Affairs Specialist Sulzerallee 8/P.O. Box CH 8404 Winterthur Switzerland

Re: K161192

Trade/Device Name: Wagner SL Revision Stem Lateral, Wagner Cone Prosthesis® System Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: LZO, LPH, KWZ, JDI Dated: June 8, 2016 Received: June 9, 2016

Dear Ms. Anne-Kathrin Born:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

### Indications for Use

510(k) Number (if known) K161192

Device Name Wagner SL Revision Stem Lateral Wagner Cone Prosthesis® System

Indications for Use (Describe)

• Noninflammatory degenerative joint disease (NIDJ), e.g. avascular necrosis, osteoarthritis, and inflammatory joint disease (IJD), e.g. rheumatoid arthritis.

· Failed previous surgery where pain, deformity, or dysfunction persists.

· Revision of previously failed hip arthroplasty.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

EF

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for Zimmer Biomet. The logo features a stylized blue letter "Z" inside of a blue circle. Below the circle is the word "zimmer" in a lowercase, sans-serif font, also in blue.

## 510(k) Summary

| Sponsor:                                                     | Zimmer GmbH<br>Sulzerallee 8, P.O. Box<br>8404 Winterthur, Switzerland                                                                                                                                                                                                                                                                          |
|--------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:                                              | Anne-Kathrin Born<br>Specialist, Regulatory Affairs<br>Telephone: +41 58 85 48 619<br>Fax: + 41 52 244 86 58                                                                                                                                                                                                                                    |
| Date:                                                        | April 26, 2016                                                                                                                                                                                                                                                                                                                                  |
| Trade Name:                                                  | Wagner Cone Prosthesis® System<br>Wagner SL Revision Stem Lateral                                                                                                                                                                                                                                                                               |
| Classification Product Code /<br>Device Classification Name: | LZO - Prosthesis, hip, semi-constrained,<br>metal/ceramic/polymer, cemented or non-porous,<br>uncemented<br>LPH - Prosthesis, hip, semi-constrained,<br>metal/polymer, porous uncemented<br>KWZ - Prosthesis, hip, constrained, cemented or<br>uncemented, metal/polymer<br>JDI - Prosthesis, hip, semi-constrained,<br>metal/polymer, cemented |
| Regulation Number / Description:                             | 21 CFR § 888.3353 - Hip joint<br>metal/ceramic/polymer semi-constrained cemented<br>or nonporous uncemented prosthesis<br>21 CFR § 888.3358 - Hip joint metal/polymer/meta<br>semi-constrained porous-coated uncemented<br>prosthesis<br>21 CFR § 888.3310 – Hip joint metal/polymer<br>constrained cemented or uncemented prosthesis           |

{4}------------------------------------------------

21 CFR § 888.3350 - Hip joint metal/polymer semi constrained cemented prosthesis Wagner SL Revision Stem Lateral, manufactured by Predicate Device: Zimmer GmbH, K043356, cleared April 18, 2005 Wagner Cone Prosthesis® System, manufactured by Zimmer GmbH, K113556, cleared February 17, 2012 Device Description: The Wagner SL Revision Stem Lateral is a straight stem manufactured from forged titanium alloy and is available in four lengths. The neck design of the stem is provided with the standard 12/14 taper for connection with any Zimmer modular femoral head utilizing a 12/14 taper. The Wagner SL Revision Stem Lateral incorporates a circular stem crosssection and equally spaced conical anchorage ribs, which run nearly the full length of the stem. The Wagner Cone Prosthesis stem is a straight, collarless stem system designed for uncemented fixation. The surface of the prosthesis is rough blasted, and it has a tapered shape with an angle of five degrees. The stem has eight longitudinal ribs, and it is available in two different CCD angles. 125° and 135°. The stems are available in twelve diameters, ranging from 13 to 24 mm. No changes are being made to the Wagner SL Revision Stem Lateral and the Wagner Cone Prosthesis implants, but one change to the Awls is proposed in this submission: An additional energy type for driving the Wagner -SL Awls and Wagner Cone Awls is introduced: from manually powered use only to manually and power driven use. While manually reaming is still possible the proposed modification enables the use of the awls connected to a powered device to ream the femoral medullary canal.

{5}------------------------------------------------

### K161192 Page 3 of 3

| Intended Use:                                      | No changes are made to the intended use of the<br>Wagner SL Revision Stem Lateral and Wagner Cone<br>Prosthesis® implants as a result of this modification.<br>Indications for use include:<br>- Patient conditions of noninflammatory<br>degenerative joint disease (NIDJD), e.g.,<br>avascular necrosis, osteoarthritis; and<br>inflammatory joint disease (IJD), e.g.,<br>rheumatoid arthritis.<br>- Patients with failed previous surgery where pain,<br>deformity, or dysfunction persists.<br>- Revision of previously failed hip arthroplasty.<br> |
|----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparison to Predicate Device:                    | The Wagner SL Revision Lateral and Wagner Cone<br>Prosthesis® implants are not modified as compared<br>to their predicates. Instead, an additional energy type<br>for driving the Wagner SL Awls and Wagner Cone<br>Awls is introduced; from manually powered use only<br>to manually and power driven use. The Wagner SL<br>Revision Stem Lateral and Wagner Cone Prosthesis<br>® implants are identical in intended use, materials,<br>sterility, and performance characteristics to the<br>predicate devices and remain unchanged.                     |
| Performance Data (Nonclinical<br>and/or Clinical): | Non-Clinical Performance and Conclusions:<br>The results of non-clinical performance testing and<br>analyses demonstrate that the devices are safe and<br>effective and substantially equivalent to the predicate<br>devices. Performance analyses included:<br>1. Usability Testing<br>2. Similar Device Analysis<br><br>Clinical Performance and Conclusions: Clinical data<br>and conclusions were deemed not needed to<br>                                                                                                                            |

---

**Source:** [https://fda.innolitics.com/device/K161192](https://fda.innolitics.com/device/K161192)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
