Next Science Irrigation Solution

K161165 · Next Science, LLC · FQH · Sep 7, 2016 · General Hospital

Device Facts

Record IDK161165
Device NameNext Science Irrigation Solution
ApplicantNext Science, LLC
Product CodeFQH · General Hospital
Decision DateSep 7, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5475
Device ClassClass 2

Indications for Use

Next Science™ Irrigation Solution is to be used with a jet lavage system and is indicated for use in cleansing and removal of debris, including microorganisms, from wounds.

Device Story

Next Science Irrigation Solution is an aqueous solution for wound irrigation and debridement; used to remove debris and microorganisms from wounds. Device functions via mechanical action of fluid moving across wound surface (hydrodynamic shear). Provided in 1000 mL polypropylene containers; labeled 'not for IV use'. Used in clinical settings; applied via jet lavage system. Healthcare providers use the solution to irrigate wounds; mechanical action aids in cleaning and removal of foreign material. Benefits include effective wound cleansing and debris removal, supporting wound healing.

Clinical Evidence

Bench testing and animal study provided. Biocompatibility testing included ISO Intracutaneous Reactivity Test and ISO Guinea Pig Maximization Sensitization Test. Animal study: Porcine Dermal Full-Thickness Wound Wash Study using 8 Yorkshire pigs (64 total wounds). Results showed normal wound healing comparable to control and untreated wounds over 24 days.

Technological Characteristics

Aqueous solution containing ethanol (100g/L), acetic acid (50g/L), sodium acetate (30g/L), benzalkonium chloride (1.3g/L), and purified water. Packaged in 1000 mL polypropylene containers. Classified as a surface device with limited contact per ISO 10993-1:2007. Mechanical sensing/actuation via hydrodynamic shear.

Indications for Use

Indicated for use with a jet lavage system for cleansing and removing debris, including microorganisms, from wounds.

Regulatory Classification

Identification

A jet lavage is a device used to clean a wound by a pulsatile jet of sterile fluid. The device consists of the pulsing head, tubing to connect to a container of sterile fluid, and a means of propelling the fluid through the tubing, such as an electric roller pump.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 880.9.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol featuring three stylized human profiles facing to the right, layered on top of each other to create a sense of depth and unity. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 7, 2016 Next Science, LLC Matthew Myntti, Ph.D. President 8130 Baymeadows Way West, Suite 200 Jacksonville, FL 33256 Re: K161165 Trade/Device Name: Next Science Irrigation Solution Regulation Number: 21 CFR 880.5475 Regulation Name: Jet lavage Regulatory Class: Class II Product Code: FQH Dated: August 9, 2016 Received: August 9, 2016 Dear Dr. Myntti: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, ## David Krause -S Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices > Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ | ̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣<br>리를 | of the looking to the first to the first to the forme the forme the forme the forme to the forme of the forme of the forme of the formation of the formation of the formation | 00 000 000 000 000 000 000 000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000 | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | the production in a stamilies nebud zich guident commont to notismoni to noticello and weiver bris<br>a mil<br>ent gribulari şenodser ned strong et belemires si noticemoni to noireline sint not emit nebud edit<br>(inisi in introlution, in surposition, in resultant that in is in isti<br>o uoticello e 'o' puodsar of painted tou si uossed e pue 'sosuods 10 louploo you yew your of the prove a<br>NOJE SEBROCA JIAMS FRATS ARA JHT OT MAO3 COTTLISMOS AUCY COMBE TOM OD<br>ວຍອາ ໄດ້ ເວີລ ການປ່ານປອກ ສາວຫາອຸດສາ ອຸດັງ ໄດ້ ໄດ້ ຢາກອາຫອງແນງອາ ດຳ ປຸ່ມາ ຂອງແຜນລາວລາວໄທ<br>้าอย่านมา BMO bilev ปุ่นตอนนว ธ รับธุรกิจ วัน รรอบม ของเธนนองก่<br>Department of Health and Human Services<br>apperwork Reduction Art (PRA) Stat<br>vog.zdyd.leb]@MelSAAAq<br>Office of Chief Information Office<br>For a ministration Drug Asministration | | | | ค้าที่ 1 คือ ค.ศ. 1994 ค.ศ. 1994 ค.ศ. 1995 น. 1995 บทเป็นที่เกิด 2 | | | O | (D heading 108 973 12 heat 21 CFR 800 Support D | Type of Use (Select one or both, as applicable) | | | | | | | ໄດ ໂຮບດຕອງ ປ່າ ຂອງຕ່າງປີ ອອນ ແລະ ອອນ ອອກສາ ລາວ ຂອງຣາຍໄລວ໌ ຄ.ສ. ຄ.ສ. 2012 ເດີຍນາ ປະຊາທິນາປຸລິຍາມາລາວ | spunow mont , sustime , sustims , from wound<br>Next<br>Indications for Use (Describe) | | | | Nort Secience Irrigation Solution<br>Device Name | | | | 510(k) Number (if known<br>K1616165 | | sing the statement below<br>iration Date: January 31, 2017 | Food animbA Quog Aministration<br>dications for Us | | | Form Approved: OMB No. 0910-0120 | гээгдэг иамин ома нттаан то тиэмтяачээ | | {3}------------------------------------------------ ## 510(k) Summary: K161165 ## Next Science™ Irrigation Solution | Submitter: | Next Science TM, LLC<br>8130 Baymeadows Way West<br>Suite 200<br>Jacksonville, Florida 32256 | |----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Matthew Myntti, Ph.D.<br>President, Next Science TM, LLC<br>855-564-2762 ext. 1004 | | Date Prepared: | September 2, 2016 | | Device Trade Name: | Next Science TM Irrigation Solution | | Device Common Name: | Irrigation Solution | | Classification Name: | Dressing, Wound, Drug | | Product Code: | FRO and FQH | | Classification: | Unclassified | | Predicate Device: | Prontosan Wound Irrigation Solution (K072876) | | Reference Device: | Atteris Antimicrobial Skin & Wound Cleanser<br>(K160192). This device is referenced because, like<br>the proposed device, it is intended for cleansing and<br>removal of debris, including microorganisms, from<br>wounds. | | Device Description: | Next Science TM Irrigation Solution is an aqueous<br>solution for irrigation and debridement of wounds.<br>The solution is a clear, colorless, low-odor aqueous<br>solution that is used to remove debris, including<br>microorganisms, from wounds.<br>The mechanical action of fluid moving across the<br>wound provides the mechanism of action and aids in<br>the removal of debris, including microorganisms, from<br>wounds.<br>Next Science TM Irrigation Solution will be provided in<br>a 1000 mL polypropylene plastic container with a | {4}------------------------------------------------ single port and will be labeled "not for IV use". The container will be provided in a polyethylene overwrap and packaged with six (6) containers per case. The formulation for Next Science™ Irrigation Solution is composed of ethanol, acetic acid, sodium acetate, benzalkonium chloride and water. - Indications For Use: Next Science™ Irrigation Solution is to be used with a jet lavage system and is indicated for use in cleansing and removal of debris, including microorganisms, from wounds. - The proposed device and the predicate device are Substantial Equivalence: intended for use in the irrigation of wounds. Although the proposed device and the predicate device consist of different ingredients, these differences do not raise different questions of safety or effectiveness. The safety and effectiveness of the Next Science™ Irrigation Solution is supported by biocompatibility testing, functional performance testing, animal testing, and shelf-life testing conducted on the proposed device. | Product | Next Science™ Irrigation<br>Solution | Prontosan Wound Irrigation<br>Solution (K072876) | |----------------|--------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------| | Company | Next Science, LLC | B. Braun | | 510(k) Number | K161165 | K072876 | | Classification | FQH, Jet Lavage<br>FRO, Dressing, Wound,<br>Drug | FRO, Dressing, Wound, Drug | | Composition | Ethanol, 100g/L<br>Acetic acid, 50g/L<br>Sodium acetate, 30g/L<br>Benzalkonium chloride,<br>1.3g/L<br>Purified water | 0.1% Undecylenamidopropyl<br>betaine<br>0.1% Polyaminopropyl<br>biguanide<br>Sodium hydroxide<br>Purified water | | Description | Next Science™ Irrigation<br>Solution is a clear, colorless<br>liquid containing ethanol,<br>acetic acid, sodium acetate, | Prontosan Wound Irrigation<br>Solution is a clear, colorless<br>liquid containing<br>undecylenamidopropyl betaine, | {5}------------------------------------------------ | Product | Next Science™ Irrigation<br>Solution | Prontosan Wound Irrigation<br>Solution (K072876) | |---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | and benzalkonium chloride. | polyaminopropyl biguanide,<br>sodium hydroxide and purified<br>water. | | Indications | Next Science Irrigation<br>Solution is to be used with a<br>jet lavage system and is<br>indicated for use in<br>cleansing and removal of<br>debris, including<br>microorganisms, from<br>wounds. | Prontosan Wound Irrigation<br>Solution is intended for cleaning<br>wounds and for moistening and<br>lubricating absorbent wound<br>dressings for ulcers, burns,<br>post-surgical wounds and<br>abrasions. | | Mechanism(s)<br>of Action | Mechanical removal of<br>debris via hydrodynamic<br>shear.<br>The mechanical action of<br>moving across the wound<br>aids in the removal of<br>foreign material such as dirt<br>and debris. | Mechanical removal of debris<br>via hydrodynamic shear.<br>The mechanical action of<br>moving across the wound aids<br>in the removal of foreign<br>material such as dirt and debris. | {6}------------------------------------------------ ## Performance testing: Next Science™ Irrigation Solution is categorized as a surface device with limited contact per ISO10993-1:2007, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" and FDA's Draft Guidance for Industry and Food and Drug Administration Staff "Use of International Standard ISO 10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" (April 23, 2013). Biocompatibility testing was conducted in accordance with International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA, and included the following: - > ISO Intracutaneous Reactivity Test to determine if the solution was capable of causing local dermal irritation. - > ISO Guinea Pig Maximization Sensitization Test (GLP Method for Liquid Test Article) to evaluate the solution's allergenic potential or sensitizing capacity. Animal testing: A Porcine Dermal Full-Thickness Wound Wash Study was conducted to evaluate the effect of the Irrigation Solution components on the wound healing process. In the animal study conducted, eight Yorkshire pigs were used each receiving eight wounds, four on either side of the dorsal midline. The four wounds were treated with the same control or test solution, or an empty control (no treatment). The empty control article was included to serve as a baseline. Under the conditions of the porcine wound healing study, Next Science Irrigation Solution resulted in normal wound healing and was comparable to wounds that received the control article and wounds without treatment over 24 days in this porcine full-thickness dermal wound model. Substantial Equivalence Conclusions: The Indications for Use statement for Next Science™ Irrigation Solution is similar to that for the predicate device, and the differences in ingredients do not raise different questions of safety or effectiveness. The performance testing and animal testing demonstrate that Next Science™ Irrigation Solution is at least as safe and effective as the predicate device. Therefore, the information in this submission demonstrates that Next Science™ Irrigation Solution is substantially equivalent to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...