← Product Code [LPH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH) · K161080

# Escalade Legend Acetabular Shell (K161080)

_Ortho Development Corporation · LPH · May 17, 2016 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH/K161080

## Device Facts

- **Applicant:** Ortho Development Corporation
- **Product Code:** [LPH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH.md)
- **Decision Date:** May 17, 2016
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3358
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Escalade Legend® Acetabular Shell is intended for use in total hip arthroplasty. The device is intended for uncemented, biological fixation only in cases of: Notably impaired hip joints due to osteoarthritis, rheumatoid arthritis and/or post traumatic arthritis. Previously failed hip surgery. Fractures of the femoral neck or head. Avascular necrosis of the femoral head. Congenital dysplasia or other structural abnormalities where stock exists to properly seat the prosthesis.

## Device Story

Hemispherical acetabular shell for total hip arthroplasty; used in uncemented, press-fit surgical applications. Device replaces natural articular surface of hip joint to increase mobility and decrease pain. Implanted by orthopedic surgeons in clinical settings. Subject device is a line extension of Escalade Acetabular Cup System; primary modification is transition from titanium plasma spray coating to sintered titanium beads for enhanced bone fixation. Device functions as a mechanical component of a modular hip prosthesis system; interacts with previously cleared liners, bone screws, and femoral heads. Clinical benefit derived from biological fixation to host bone, restoring joint function.

## Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing, including liner connection strength (ASTM F1820) and coating characterization (ASTM F1044, F1160, F1854, F1978).

## Technological Characteristics

Hemispherical acetabular shell; material: Ti-6Al-4V ELI (ASTM F136); coating: sintered titanium beads (CP Ti per ASTM F1580); fixation: uncemented, press-fit; modularity: snap-fit with EXLPE liner; configurations: no-hole and 3-hole options.

## Regulatory Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

## Predicate Devices

- Escalade Acetabular Cup System ([K103384](/device/K103384.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, overlapping each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 17, 2016

Ortho Development Corporation Mr. Drew Weaver Director of Quality Assurance and Regulatory Affairs 12187 South Business Park Drive Draper, Utah 84020

Re: K161080 Trade/Device Name: Escalade Legend® Acetabular Shell Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH, LZO Dated: April 15, 2016 Received: April 18, 2016

Dear Drew Weaver:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K161080

#### Device Name

Escalade Legend® Acetabular Shell

#### Indications for Use (Describe)

This device is intended for use in total hip arthroplasty. The device is intended for uncemented, biological fixation only in cases of:

- · Notably impaired hip joints due to osteoarthritis, rheumatoid arthritis and/or post traumatic arthritis.
- · Previously failed hip surgery.
- Fractures of the femoral neck or head.
- · Avascular necrosis of the femoral head.
- · Congenital dysplasia or other structural abnormalities where stock exists to properly seat the prosthesis.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

12187 So. Business Park Drive
Draper, Utah 84020
801-553-9991/fax 553-9993

orthodevelopment.com

Image /page/3/Picture/2 description: The image shows the logo for "Ortho Development". The logo consists of a purple circle with three gray arcs above it, followed by the word "ORTHO" in purple, and the word "DEVELOPMENT" in gray below it. The logo is simple and modern, and the colors are eye-catching.

# Section 5 510(k) Summary

| Name of Sponsor:     | Ortho Development Corporation<br>12187 South Business Park Drive<br>Draper, Utah 84020                                                                                                                                                        |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Contact:      | Drew Weaver<br>Director of Quality Assurance and Regulatory Affairs<br>Telephone: (801) 619-3419<br>Facsimile: (801) 553-9993<br>Email: DWeaver@orthodevelopment.com                                                                          |
| Date Prepared:       | April 15, 2016                                                                                                                                                                                                                                |
| Proprietary Name:    | Escalade Legend® Acetabular Shell                                                                                                                                                                                                             |
| Common Name:         | Acetabular Cup Prosthesis                                                                                                                                                                                                                     |
| Classification:      | 21 CFR 888.3358, Hip joint metal/polymer/metal semi-constrained<br>porous-coated uncemented prosthesis<br>21 CFR 888.3353 Hip joint metal/ceramic/polymer semi-<br>constrained cemented or nonporous uncemented prosthesis<br>Class II device |
| Device Product Code: | LPH, LZO                                                                                                                                                                                                                                      |
| Predicate Devices:   | Escalade Acetabular Cup System (K103384)<br>Ortho Development Corporation                                                                                                                                                                     |

## 5.1 Device Description

The proposed subject device Escalade Legend® Acetabular Shell is a line extension to the Escalade Acetabular Cup System (K103384). The Escalade Acetabular Cup System is a modular system intended for the replacement of the natural articular surface of the hip replacement surgery. The system consists of acetabular shells, liners, bone screws, apical screw hole cover, and femoral heads.

The Escalade Legend® Acetabular Shells are hemispherical in shape and are designed for press-fit, cementless, surgical applications. The subject device differs from the shell in K103384 device in the use 

{4}------------------------------------------------

of sintered titanium beads instead of titanium plasma spray for bone fixation. Information regarding the sintered titanium beads can be found in MAF-2172, Titanium ASYMMATRIX™ Type 1, Orchid Orthopedic Solutions, Memphis, TN.

The liners, bone screws, apical screw hole cover and femoral heads used have been cleared under K103384 and are not a part of this submission.

A summary of the Intended Use, Indications for Use, Summary of Technological Characteristics, and Testing performed on the subject device as they relate to the unmodified K103384 device can be seen in Table 5.1 below. Only information pertaining to the acetabular shell is included because the liners, bone screws, apical screw hole cover, and femoral heads are not part of this submission.

| Table 5.1 - Summary of Escalade Legend Acetabular Shell Characteristics                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                   |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|
| Escalade Acetabular Shell K103384                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Escalade Legend® Acetabular Shell |
| Intended Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                   |
| The Escalade Acetabular Cup System is intended for<br>the replacement of the natural articular surface of the<br>hip joint in a total hip replacement surgery. Total hip<br>arthroplasty is intended to provide increased patient<br>mobility and to decrease pain by replacing the<br>damaged hip joint in patients having sufficiently<br>sound bone to support the implants.                                                                                                                                                        | Same                              |
| Indications for Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                   |
| This device is intended for use in total hip<br>arthroplasty. The device is intended for uncemented,<br>biological fixation only in cases of:<br>1. Notably impaired hip joints due to osteoarthritis,<br>rheumatoid arthritis and/or post traumatic arthritis.<br>2. Previously failed hip surgery.<br>3. Fractures of the femoral neck or head.<br>4. Avascular necrosis of the femoral head.<br>5. Congenital dysplasia or other structural<br>abnormalities where sufficient bone stock exists to<br>properly seat the prosthesis. | Same                              |
| Technological Characteristics                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                   |
| Hemispherical in shape                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                              |
| Machined of Ti-6Al-4V ELI, ASTM F136                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same                              |
| Snap fit with EXLPE liner                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                              |
| Uncemented biological bone fixation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                              |
| No-hole and 3-hole shells offered                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                              |
| Coating material: CP Ti per ASTM F1580                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                              |
| Coating type: Plasma Spray                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Coating type: Sintered Bead       |

{5}------------------------------------------------

| Non-Clinical Testing                                             |      |
|------------------------------------------------------------------|------|
| Liner connection strength test per ASTM F1820                    | Same |
| Coating tests per ASTM F1044, ASTM F1160, ASTM F1854, ASTM F1978 | Same |

## 5.2 Clinical Test Summary

No clinical studies were performed.

## 5.3 Conclusions

Based on the similarities to the predicate device and the unchanged testing performance, Escalade Legend® Acetabular Shell is substantially equivalent to the predicate device that was cleared under K103384.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH/K161080](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH/K161080)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
